APPA sFLEX 18D Flexible Current Probe Meter
Product Information
The sFLEX 18D is a flexible current probe meter that is designed to measure AC
current. It has a 3000-count display and over-range display. The dimensions of
the device are approximately 130mm x 350mm x 25mm and weighs 200g (without
battery). It requires two AAA size batteries and has a battery life of 200
hours. The device is designed to conform to EU directives and safety standards
such as EN61010-1, EN61010-2-032, and EN61326-1 CAT IV 600V, CAT III 1000V.
The conversion rate is 4 times/second, and the frequency response is 45 –
500Hz (Sine Wave).
Product Usage Instructions
Safety Information: To ensure safe operation and service of the meter, follow the instructions provided in the user manual. Failure to observe warnings can result in severe injury. Identify hazardous conditions and actions that could cause bodily harm or death – individual protective equipment should be used if necessary.
Maintenance: Do not attempt to repair this clamp meter. It contains no user-serviceable parts. Repair or servicing should only be performed by qualified personnel. Periodically wipe the case with a dry cloth and detergent. Do not use abrasives or solvents.
Power on/off: To turn off the device, press the power button for more than 2 seconds.
Flexible AC current probe: To unlock the probe, slide the unlock button to the right. Do not touch the area above the LCD during measurement.
Auto Power off Function: The device has an auto power off function that turns off the device after 20 minutes of inactivity. To disable this function, press the power button and during the LCD shows battery capacity, press the backlight button until the LCD shows “AoFF.”
Backlight: To turn on the backlight, press the
backlight button.
Data Hold: To freeze the reading, press the
data hold button.
Replacing Batteries: Before replacing
batteries, remove the meter from the conductor and turn off power. Unlock the
battery cover by sliding the unlock button to the right and then remove the
old batteries. Insert new batteries into the battery compartment and lock the
battery cover by sliding the lock button to the left.
Read First
Safety Information
To ensure safe operation and service of the meter, follow these instructions. Failure to observe warnings can result in severe Injury.
Warning
Identify hazardous conditions and actions that could cause BODILY HARM or
DEATH .
- Individual protective equipment should be used if HAZARDOUS LIVE parts in the installation where measurement is to be carried out could Be ACCESSIBLE.
- If the equipment is used in a manner not specified by the manufacturer, the protection provided by the equipment may be impaired.
- To reduce the risk of fire or electric shock, do not expose this product to rain or moisture.
- Verify the Meter operation by measuring a known Current. If in doubt, have the Meter serviced.
- Do not apply more than the rated voltage/Current, as marked on Meter.
- To avoid false readings that can lead to electric shock and injury, replace battery as soon as low battery indicator.
- Do not use Meter around explosive gas or vapor.
- Do not use a flexible current sensor if the inner copper wire of the flexible cord is visible.
- De-energize the installation under test or wear suitable protective clothing during fitting a removal of the Flexible Current Probe.
- Do not apply around or remove from UNINSU LATED HAZARDOUS LIVE conductors, which may render electric shock, electric burn, or arc flash.
Caution
Identify conditions and actions that could cause DAMAGE the meter or
equipment under test. Do not expose Meter to extremes in temperature of high
humidity.
Symbols as marked on the Meter and Instruction Card
| See instruction manual
---|---
| Do not apply to or remove from HAZARDOUS LIVE conductors.
| Equipment protected by double or reinforced insulation
| Battery
| Earth
| Conforms to EU directives
| Do not discard this product or throw away
****| AC measurement
Maintenance
Do not attempt to repair this Clamp Meter. It contains no user-serviceable parts. Repair or servicing should only be performed by qualified personnel.
Cleaning
Periodically wipe the case with a dry cloth and detergent. Do not use
abrasives or solvents.
Power on/off
Flexible AC current probe
Do not touch the area above the LCD during measurement
Auto Power off Function
The meter automatically turns the power off after powering up and no operation for 20 minutes ± 30 seconds.
Disable Auto Power off Function
Press “Power” button, and during the LCD shows battery capacity, press “backlight” button until the LCD shows “AoFF” to disable Auto Power Off Function.
Backlight
Data Hold
Replacing Batteries
Caution
Remove Meter from conductor and turn off power before opening the battery
cover.
General Specifications
Display : 3000 counts.
Over range Display : “OL”
Dimensions (W x H x D) : 18″ coil : Approx. 130mm x 350mm x 25mm
Weight(without Battery) : 18″ coil : 200g
Power Requirements : AAA Size Battery x 2
Battery Life : 200 hours ALKALINE Battery Low Batteries Indication :
Voltage will flash. drops below operating voltage,
Maximum Conductor Size : coil size Safety Standard : EN61010-1, EN61010-2-032,
EN61326-1 CAT IV 600V, CAT III 1000V
Conversion Rate : 4 times/second
Environmental Specifications
Indoor Use.
Pollution Degree : 2
- Maximum operating altitude : 2000m (6562 ft)
Operating temperature & Relative Humidity :
-
0°C – 30°C, ≦80%RH
-
30°C – 40°C, ≦75%RH
-
40°C – 50°C, ≦45%RH
-
Storage temperature :
-20°C – 60°C, 0 to 80% RH (no batteries). -
Temperature coefficient :
0.2 x (Specified accuracy) / °C, < 18°C, > 28°C -
Shock vibration : Sinusoidal vibration
MIL-PRF-28800F (5-55 Hz, 3g max.) -
Drop Protection : 4 feet drop to hardwood on concrete Floor
Electrical Specifications
Accuracy is given as ±(% of reading + counts of least significant digit) at 23°C ± 5°C, with relative humidity Less than 80% R.H. Accuracy is specified for a period of one year after calibration.
Position Error of Clamp
Accuracy and position error assumes centralized primary conductor at optimum
position, no external electrical or magnetic field, and within operating
temperature range.
| sFlex-18| Error
---|---|---
Distance from Optimum (mm)
| 35| ±1.0%
50| ±1.5%
60| ±2.0%
AC Current
Range
| OL
Reading
| ****
Resolution
| Accuracy
(45Hz~500Hz)
---|---|---|---
30.00A| 33.00A| 0.01A| ±(3.0% + 5D)
300.0A| 330.0A| 0.1A| ±(3.0% + 5D)
3000A| 3300A| 1A| ±(3.0% + 5D)
LCD displays 0 counts when the reading < 10 counts Frequency Response : 45 – 500Hz (Sine Wave)
AC Function
ACA specifications are ac coupled, true R.M.S. For non-sinusoidal waveforms,
Additional
- Accuracy by Crest Factor (C.F.):
- Add 3.0% for C.F. 1.0 – 2.0.
- Add 5.0% for C.F. 2.0 – 2.5.
- Add 7.0% for C.F. 2.5 – 3.0
Limited Warranty
This Meter is warranted to the original purchaser against defects in material and workmanship for 3 years from the date of purchase. During this warranty period, manufacturer will, at its option, replace or repair the defective unit, subject to verification of the defect or malfunction. This warranty does not cover fuses, disposable batteries, or damage from abuse, neglect, accident, unauthorized repair, alteration, contamination, or abnormal conditions of operation or handling. Any implied warranties arising out of the sale of this product, including but not limited to implied warranties of merchantability and fitness for a particular purpose , are limited to the above. The manufacturer shall not be liable for loss of use of the instrument or other incidental or consequential damages, expenses, or economic loss, or for any claim or claims for such damage, expense or economic loss. Some states or countries laws vary, so the above limitations or exclusions may not apply to you.