PPI 48X48 Automatic Autoclave Controllers Instruction Manual
- June 11, 2024
- PPI
Table of Contents
PPI 48X48 Automatic Autoclave Controllers Instruction Manual
This brief manual is primarily meant for quick reference to wiring connections and parameter searching. For more details on operation and application; please log on to www.ppiindia.net
OPERATOR PARAMETERS : PAGE 0
Parameters| Settings (Default Value)
Control Setpoint
| Setpoint Low Limit to Setpoint High Limit (Default : 121.0)
Air Outlet Setpoint
| Setpoint Low Limit to Setpoint High Limit (Default : 100.0)
Soak Time
| 1 to 999 Minutes (Default : 20)
High Alarm Deviation Setpoint
| 1 to 10 or 0.1 to 10.0 (Default : 5.0)
Fail-Safe Deviation Setpoint
| 3 to 20 or 0.3 to 20.0 (Default : 10.0)
Cycle Abort Command
| No Yes
Cycle Abort Pass Code
| 0 to 250 (Default : 0)
I / O CONFIGURATION PARAMETERS : PAGE 10
Parameters | Settings (Default Value) |
---|
Hysteresis
| 1 to 999°C or 0.1 to 99.9°C (Default : 0.2)
Setpoint Low Limit
| Min. Range for the selected Input Type to Setpoint High (Default : 90.0)
Setpoint HighLimit
| Setpoint Low to Max. Range for the selected Input Type (Default : 150.0)
Units for Pressure
| Kg/cm² PSI
(Default : Kg/cm²)
SOAK TIMER PARAMETERS : PAGE 15\
Parameters | Settings (Default Value) |
---|
Holdback Strategy
| None Up Down Both (Default : Down)
Hold Band
| 1 to 5 or 0.1 to 5.0 (Default : 1.0)
End-of-Soak Alarm Time
| 5 to 250 Sec. (Default : 10 Sec.)
Power-fail Recovery Method
| Abort (Re)Start Continue (Default : Continue)
SUPERVISORY PARAMETERS : PAGE 11
Parameters | Settings (Default Value) |
---|
Offset for Temperature Value
| -1999 to 9999 or -199.9 to 999.9 (Default : 0.0)
Digital Filter for Temperature Value
| 0.5 to 25.0 Seconds in steps of 0.5 Seconds (Default : 1.0)
Operator Parameter Locking
| No Yes (Default : No)
Pass Code for Cycle Abort
| 1 to 250 (Default : 22)
Utility Option
| None Serial Comm. Cycle Start Water Level (Default : None)
Slave ID
| 1 to 127 (Default : 1)
Baud Rate
| 1200 2400 4800 9600 (Default : 9600) 1200 2400 4800 9600
Parameters | Settings (Default Value) |
---|
Communication Write Enable
| No Ye (Default : Yes)
Water Level Switch Logic
| Normal Reverse (Default : Norma
Pressure View
| Disable Enable (Default : Enable
Applicable only if Utility Option is Selected as Water Level Input
ELECTRICAL CONNECTIONS
FRONT PANEL LAYOUT
48X48
S ymbol | Key | Function |
---|---|---|
PAGE | Press to enter or exit set-up mode. | |
DOWN | Press to decrease the parameter value. Pressing once decreases the |
value by one count; keeping pressed speeds up the change.
| UP| Press to increase the parameter value. Pressing once increases the value
by one count; holding pressed speeds up the change.
| ENTER| Press to store the set parameter value and to scroll to the next
parameter on the PAGE.
PV Error Indications
Message | PV Error Type |
---|---|
Over-range (PV above Max. Range) | |
Under-range (PV below Min. Range | |
Open (Sensor open / broke |
101, Diamond Industrial Estate, Navghar,
Vasai Road (E), Dist. Palghar – 401 210.
Sales : 8208199048 / 8208141446
Support : 07498799226 / 08767395333
E: sales@ppiindia.net,
support@ppiindia.net