LUTRON HT-315 Humidity Meter User Manual
- June 9, 2024
- Lutron
Table of Contents
LUTRON HT-315 Humidity Meter
Model : HT-315
FEATURES
- Humidity + Temperature + Dew point are combined into one meter, intelligent and professional.
- 0.01 %RH resolution for the humidity reading, 0.01 degree resolution for the Temp. reading.
- Pocket size with Separate humidity & temp. probe, easy operation. & remote measurement.
- Fast humidity measuring response time.
- High accuracy and high precision.
- Dew point measurement.
- Manual and auto manual data logger.
- Just few panel buttons, easy operation.
- Microprocessor circuit assures maximum possible accuracy, provides special functions and features.
- LCD with two display, easy readout.
- Heavy duty & compact housing case, designed for easy carry out & operation.
- Records Maximum and Minimum readings with Recall.
- Auto shut off saves battery life.
- Data hold function for freezing the desired value on display.
- RS 232 PC serial interface.
- Show the humidity & temperature values on the LCD display at same time.
- Built-in low battery indicator. Wide humidity & temp. measuring range.
- DC 9V power adapter input socket.
SPECIFICATIONS
General Specifications
Circuit | Custom one-chip of microprocessor LSI circuit. |
---|---|
Display | LCD size : 44 mm x 29 mm dual function LCD display. |
Measurement Unit | Humidity : %RH ( Relative Humidity ) |
Temperature : ℃ or ℉.
Dew point : ℃ or ℉.
Response Time| 5 to 30 seconds typically.
@ Reach the 85% reading value
Temperature Compensation| Automatic temp. compensation for the humidity
function.
Data Hold| Freeze the display reading.
Memory Recall| Maximum & Minimum value.
Sampling Time| Approx. 0.8 second.
Power off| Auto shut off saves battery life or manual off by push button.
Data Output| RS 232 PC serial interface.
Operating Temperature| 0 to 50 ℃.
Operating
Humidity
| Main instrument : Less than 85% R.H.
Probe : 0 to 95 %RH.
Power Supply| 006P DC 9V battery
( Alkaline or Heavy duty type ).
Power Current | Approx. DC 4.6 mA. |
---|---|
Weight | 264 g/0.67 LB. |
@ Battery is included.
Dimension| Main instrument : 135 x 60 x 33 mm,
( 5.3 x 2.4 x 1.3 inch ).
Humidity Sensor Probe :
197 mm ( 7.8 inch ) in length.
Accessories Included| Instruction manual………………….1 PC
Humidity probe………………………1 PC
Optional Accessories| RS232 cable, UPCB-02
Data Acquisition software, SW-801-WIN Case wall holder
Power adapter ( ACV to DC 9V )
Electrical Specifications 23± 5 ℃
Humidity/ Temperature
__
Humidity
| Range| 0 % to 95 % R.H.
---|---|---
Resolution| 0.01 % R.H.
Accuracy| ≧70% RH
± (3% reading + 1% RH).
< 70% RH – 3% RH.
± 3% RH.
__
Temperature
| Range| 0 ℃ to 50 ℃,32 ℉ to 122 ℉.
Resolution| 0.01 degree
Accuracy| ℃ – 0.8 ℃.
℉ – 1.5 ℉.
Dew Point
℃ | Range | -25.3 ℃ to 48.9 ℃ |
---|---|---|
Resolution | 0.01 ℃ | |
℉ | Range | -13.5 ℉ to 120.1 ℉. |
Resolution | 0.01 ℉. |
Remark :
* Dew Point display value is calculated from the Humidity/Temp. measurement automatically.
* The Dew Point accuracy is sum accuracy value of Humidity & Temperature measurement..
@ Above specification tests under the environment RF Field Strength less than 3 V/M & frequency less than 30 MHz only.
FRONT PANEL DESCRIPTION
- Display
- Power Button
- Hold Button ( Esc Button )
- REC Button ( Enter Button )
- Function Button ( Send Button )
- Set Button ( Logger Button )
- Battery Compartment/Cover
- RS-232 Output Terminal
- Probe Input Socket
- Probe Plug
- Probe handle
- Probe head ( Humidity & Temperature )
- DC 9V Power Adapter Input Socket
GENERAL MEASURING PROCEDURE
The meter default value are :
- The temperature reading unit is ℃.
- The auto power off.
Humidity and Temperature measurement
-
Plug the ” Probe Plug ” ( 3-10, Fig. 1 ) into the ” Probe Input Socket ” ( 3-9, Fig. 1 ).
-
Power on the meter by pressing the ” Power Button ” ( 3-2, Fig. 1 ), the LCD shows the unit ” %RH ” & ” ℃ ” at the same time and measured value will show on
the display ( upper display is Humidity value, the
lower display is the temperature value ) . -
The meter Temp. display unit is defaulted to ” ℃ “. If intend to let the meter’s temperature unit default to ” ℉ ” , then please refer section 5-1 ( page 9 ).
Dew point measurement
The procedures of ” Dew point measurement ” are same as the above ” 4-1
Humidity and Temperature measurement ” except select the ” Dew point ”
function by pressing the ” Function Button ” ( 3-5, Fig. 1 ) once, the LCD
will show the unit ” DEW ” & ” ( or ℃℉ ). The upper display show the Dew point
value, the lower display show the temperature value
Data Hold
During the measurement, press the ” Hold Button ” ( 3-3, Fig. 1 ) once will
hold the measured value & the LCD will display a ” HOLD ” symbol. * Press the
” Hold Button ” once again will release the data hold function.
Data Record ( Max., Min. reading )
- The data record function records the maximum and minimum readings. Press the ” REC Button ” ( 3-4, Fig. 1 ) once to start the Data Record function and there will be a ” REC. ” symbol on the display.
- With the ” REC. ” symbol on the display :
- a ) Press the ” REC Button ” ( 3-4, Fig. 1 ) once, the ” REC. MAX. ” symbol along with the maximum value will appear on the display. If intend to delete the maximum value, just press the ” Hold Button ” ( 3-3, Fig. 1 ) once, then the display will show the ” REC. ” symbol only & execute the memory function continuously.
- b ) Press the ” REC Button ” ( 3-4, Fig. 1 ) again, the ” REC. MIN. ” symbol along with the minimum value will appear on the display. If intend to delete the minimum value, just press the ” Hold Button ” ( 3-3, Fig. 1 ) once, then the display will show the ” REC. ” symbol only & execute the memory function continuously.
- c ) To exit the memory record function, just press the ” REC ” button for 2 seconds at least. The display will revert to the current reading.
ADVANCED MEASURING PROCEDURE
Before executing advanced adjustment procedures, exit the ” Hold function ” and the ” Record ” function.
-
a. Hold the ” Set Button ” ( 3-6, Fig. 1 ) at least two seconds until the lower display show ” COde “, then release the ” Set Button “, the upper display will show ” 1000 “.
-
1000 is the password code that allow to execute the Advanced Measuring Procedure following.
After display show ” COde 1000 “, push the ” Enter Button ” ( 3-4, Fig. 1 ) once will go to the following b. procedures. -
If push the ” ESC Button ” ( 3-3, Fig. 1 ) will escape the selecting function and return to the normal measuring display.
-
b. One by one to press the ” Set Button ” ( 3-6, Fig. 1 ) once a while to select the two main function that show on the lower display as
-
℉ ……. Change the Temp ℃, ℉ unit
-
OFF…….Auto power ON/OFF management
Change the Temp ℃, ℉ unit
Use the ” Set Button ” to select the main function to ” ℉ “, then one by one
to press the ” Function Button ” ( 3-5, Fig. 1 ) a while will determine the
default Temp. unit to ℃ or ℉
- Press the ” Function Button “, if the upper display value show ” 0 “, the default Temp. unit is ℃
- Press the ” Function Button “, if the upper display value show ” 1 “, the default Temp. unit is ℉.
After the function is determined, press the ” Enter Button ” ( 3-4, Fig. 1 ) to confirm and save the selection data into memory IC permanently. Press the ” Esc Button ” ( 3-3. Fig. 1 ) will revert to normal display screen.
Auto power ON/OFF
Use the ” Set Button ” to select the main function to ” OFF “, then one by one
to press the ” Function Button ” ( 3-5, Fig. 1 ) a while will determine the
default the power management system is Auto Power Off enable or disable.
- Press the ” Function Button ” once, if the upper display value show ” 0 “, it is not Auto Power Off management ( disable ).
- Press the ” Function Button ” once, if the upper display value show ” 1 “, it is the Auto Power Off management ( enable ).
After the function is determined, press the ” Enter Button ” ( 3-4, Fig. 1 ) to save the selection function into memory IC permanently. Press the ” Esc Button ” ( 3-3. Fig. 1 ) will revert to normal display screen.
RS232 PC SERIAL INTERFACE
The instrument has RS232 PC serial interface via a 3.5 mm terminal ( 3-8, Fig. 1 ). The data output is a 16 digit stream which can be utilized for user’s specific application. A RS232 lead with the following connection will be required to link the instrument with the PC serial port.
The 16 digits data stream will be displayed in the following format : D15 D14 D13 D12 D11 D10 D9 D8 D7 D6 D5 D4 D3 D2 D1 D0
Each digit indicates the following status
D0 | End Word |
---|---|
D1 & D8 | Display reading, D1 = LSD, D8 = MSD For example : |
If the display reading is 1234, then D8 to D1 is : 00001234
D9| Decimal Point(DP), position from right to the left
0 = No DP, 1= 1 DP, 2 = 2 DP, 3 = 3 DP
D10| Polarity
0 = Positive 1 = Negative
D11 & D12| Annunciator for Display
℃ = 01| ℉ = 02| % RH = 04
D13| When send the upper display data = 1 When send the lower display data = 2
D14| 4
D15| Start Word
RS232 FORMAT : 9600, N, 8, 1
BATTERY REPLACEMENT
- When the left corner of LCD display show “ “, it is necessary to replace the battery. However, in-spec. measurement may still be made for several hours after low battery indicator appears before the instrument become inaccurate.
- Slide the ” Battery Cover ” ( 3-7, Fig. 1 ) away from the instrument and remove the battery.
- Replace with 9V battery ( Alkaline or Heavy duty type ) and reinstate the cover.
- Make sure the battery cover is secured after changing the battery.
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>