4-20mA ULT Connection

The ULT product line come with built in 4-20mA support. To use this the following will need to be done.
Part of what comes with the ULT is 315672H01 terminal block, 7 position.

Thermo Scientific 4 20mA ULT Connection - FIGURE 1

This terminal block will plug into the relay enclosure as shown below.

Thermo Scientific 4 20mA ULT Connection - FIGURE 2

Once the Terminal block is connected you will use Pin #1, and Pin #2 to connect the unit to the facility BMS.
The Parameters to be set are: 4mA = -115c, 20mA = +50c

Thermo Scientific 4 20mA ULT Connection - FIGURE 3