ibx LBX MF12 Muffle Furnace For Laboratory User Manual

June 9, 2024
ibx

ibx LBX MF12 Muffle Furnace For Laboratory

Preface

Users should read this Manual carefully, follow the instructions and procedures, and beware of all the cautions when using this instrument.

Service

If help is needed, you can always contact your dealer or Labbox via www.labbox.com (declare an incidence)
Please, provide the customer service representative with the following information:

  • Serial number
  • Description of the problem
  • Your contact information

Warranty

This instrument is guaranteed to be free from defects in materials and workmanship under normal use and service, for a period of 24 months from the date of invoice. The warranty is extended only to the original purchaser. It shall not apply to any product or parts which have been damaged on account of improper installation, improper connections, misuse, accident or abnormal conditions of operation.
For claim under the warranty, please contact your supplier.

Main Technical Parameters

Model MUFU-020-001 MUFU-072-001 MUFU-120-001
Heating mode Alloy wire heating in three sides left; right; top side.
Function Temp. Range 100-1200℃
Temp. Resolution Ratio 1℃
Temp. motion ±1℃
Temp. Rising time to max temp ≤30min
Structure Chamber material Ceramic fiber
Outer shell cold rolling steel electrostatic spraying exterior
Insulation layer Ceramic fiber
Heater Alloy heating wire
Power rating 1.5kW 3.0kW 4.5kW
Exhaust hole φ30mm(chimney size 80*60mm)
Controller Temp. control mode single controller
Temp. setting mode Touch button setting
Temp. display mode Measuring temperature: LED upper row, setting temperature:

lower row
Timer| 0-9999 min (with timing wait function)
Operation function| Fixed temperature operation, timing function, auto stop.
Additional function| Sensor deviation correction, Temperature overshoot self- tuning, Internal parameter locked, Power-off parameter memory
Sensor| Platinum-rhodium sensor
Safety device| Manual door security lock, over temperature sound-light alarm, door opening electrical outage, over-temperature protection thermocouple failure
Specification| Inner Chamber size| 12020080| 200300120| 200300200
(WLH)(mm)
Exterior size (WLH)(mm)| 450685600| 530785640| 530785720
Packing size
(WLH)(mm)| 580775730| 660875770| 660875850
Volume| 2L| 7L| 12L
Current rating (50/60HZ)| AC220V/6.6A| AC220V/13.6A| AC220V/20.4A
NW/GW (kg)| 33/37| 45/50| 62/68

Characteristics

  1. Double insulation ceramic inner chamber with stable performance.
  2. Efficient insulation and three side heating of special alloy heating wire to provide optimum temperature uniformity.
  3. High precision micro-computer controller and accurate sensor.
  4. Double-shell hollow thermal insulation and double ventilation duct with excellent ventilation, quick internal cooling.
  5. Multiple safety protection measures, such as thermocouple failure, over door opening electrical outage, audio-visual alarm, etc.

Installation And Operation

  1. Open the package, check the furnace and make sure that no part is damaged. Place the furnace on smooth ground or on a table. The equipment should avoid contact with vibrating surfaces.
  2. Install the power switch at the original power line. To make sure the equipment operates safely, furnace and controller must be grounded reliably.
  3. Connect the controller direction to the power line. Turn on the power and set the temperature on the meter. It starts heating when the indicator light of the meter turns green. Adjust the power to reach the target temperature as it is needed, but make sure the voltage and electric current of the product will not surpass the rated power.

Attentions

Install the outer ground protection to ensure safety of machine and experiment; ensure power as the machine requires.

It is forbidden to use this equipment in inflammable, explosive, poisonous or strong corrosive experiments.
Make sure the equipment is horizontally installed.
Non-professionals are not allowed to disassemble and repair this machine.
Pay attention to the setting temperature when dealing with inflammable matters.
Make sure to dry the resin container, if the temperature is set too high by accident, the container could be dissolved and then fall on the heater, which could cause fire.
Overfilling of sample will lead to overheating of working room, which could dissolve the inflammable material and cause fire.
While the machine is working, don’t touch the device top, as well as observation window and exhaust port to keep away from high-temperature burns.
Do not open the door when the temperature more than 500℃
Set the temp.50℃ under the Max. temp. for long experiments
Read the instruction book before operation.

Meter Operation Instruction

Basic display status

When powered on, the screen shows the basic display status, upper window displays real temperature value (PV) in red while lower window displays the set temperature value (SV) in green. If the real value overruns measure range (thermocouple breaks for example), upper window will display “orAL” and the highest and lowest values, at this time, the controller will automatically the output.
In the controller, there are various LED lights: OP1, AU1, AU2, RUN respectively stand for output, first alarm, second alarm, and working condition

Temperature and time program setting:

In basic display status, if the parameter lock “Loc” isn’t locked, we can set the temperature point (SV) by pressing or . When any of those keys is pressed, the upper window will display “SP” in red while a dot (.) will appear in the lower-right of the digit to be modified in the lower window (in green). Press key to decrease the value of the digit, key to increase the value of the digit, and key to move to the digit expected to modify. The temperature value must be input in °C.
Once the expected temperature is set, press until the upper window displays “t-1”, then you can set time.
The time value must be input in minutes and the last digit is decimal parts of a minute, for example: 60.0 = 60 minutes, 60.5 = 60 minutes and 30 seconds.

Working control

When the power is on, the controller is in stop status, you can press for 2 seconds until the below window shows “run” to start controller’s working. Press for 2 seconds until it shows “stop” to stop the controller’s working.

Auto tuning:

When auto tuning control method is chosen, the PID parameters can be obtained by running auto-tuning. In basal display status, press for 2 seconds, the “At” parameter will appear. Press to change the value of “At” from “oFF” to “on”, then press to active the auto-tuning process. During auto tuning, the instrument executes on-off control. After 2-3 times of on-off action, the instrument will obtain the optimal control parameter value.
If you want to escape from auto tuning status, press and hold the key for about 2 seconds until the “At” parameter appear again. Change “At” from “on” to “oFF”, press to confirm, then the auto tuning process will be cancelled.

Note1 : If the setpoint is different, the parameters obtained from auto- tuning are possibly different. So you’d better set setpoint to an often-used value or middle value first, and then start auto-tuning.
Depending on the system, the auto-tuning time can be from several seconds to several hours.
Note2: Setting of CHYS may have influence on AT. The lower value of CHYS the higher accuracy. But too low CHYS isn’t advice, CHYS=2.0 is adviced.
Note3: At the beginning after AT, the result maybe not table but it will get best result after a time.

Parameter Setting

In basic display status, press and hold for about 2 seconds to access the Field Parameter Table. Pressing will move to the next parameter; pressing or can modify a parameter. Press and hold can return to the preceding parameter. Press (don’t release) and then press key simultaneously to escape from the parameter table. The instrument will escape automatically from the parameter table if no key is pressed within 30 seconds. Setting Loc=808 and then press can access System Parameter Table.

Field parameter table (Press and hold for 2 seconds to access)

Code Name Description Factory value
HIAL High limit alarm Alarm on when PV (Process Value) >HIA; alarm off when
PV<HIA-AHY 3000
LoAL Lower limit alarm Alarm on when PV (Process Value) <LoA; alarm off when
PV>LoA-AHY -999
HdAL Deviation high alarm Alarm on when PV-SV>HdA; alarm off when PV-SV<HdA-
AHY 50
LdAL Deviation low alar m Alarm on when PV-SV<LdA; alarm off when PV-SV>HdA-
AHY -999
Loc Parameter Lock Loc=0: Allowed to modify parameters and do AT Loc=1:

Allowed to modify parameters but cannot AT Loc=2: Allowed to modify parameters and AT
Loc=4-255: NOT allowed to modify parameters ecept Loc.
Loc=808, Set to 808 and press        , allowed modify all parameters.| 0
AHYS| Hysteresis| Avoid wrong frequent alarm caused by wrong value setting| 2
AoP| Alarm Output assignment| AoP is to define the place of HIAL,LoAL,HdAL alarms like :
Value 0~4.0 means never alarm, 3and4 mean alarm from AU1,AU2.
E.g. AOP=403 means HIAL alarm from AU1, HdAL from AU2.LoAL no alarm| 0~4444
CrL| Control mode| onoF : On-off control, used in normal case
APld: high precision control PID mode nPld: standard PID control| APId
Srun| Running status| Run, normal running status, PRG light on Stop, stop status, below windo shows “stop”,PRG light out
Hold, keep current status, program stop counting time at this moment| stop
Act| Direct/reverse acting function| rE: Reverse acting.Increase in measured variable causes a decrease in the output,such as heating control.
dr: Direct acting.Increase in measured variable causes an increase in the output,such as refrigerating control.
rEbA: Reverse acting with low limit alarm and deviation low alarm blocking at the beginning of power on.
drbA: Direct acting with high limit alarm and deviation high alarm blocking at the beginning of power on.| rE
P| Proportional band| Proportional band in PID with unit℃/F.
Notes: normally use At to confirm P,I,D and Ctl .But can set known correct value.| 30
I| Time of Integral| Time of integral in PID.No integral effect when I=0 unit is 1 second| 100
d| Time of Derivative| Time of derivative in PID.No derivative effect when d=0. Display unit is 0.1second| 50
Ctl| Control period| Small value can improve control accuracy. For SSR output, generally 0.5 to 3.0 seconds.
For Relay output, generally 15~40 seconds, because small value will cause the frequent On-Off of mechanical switch and shorten its service life. CtI is recommended to be 1/4 ~ 1/10 of derivative time. When control under on-off control,Ctl use as restart delay time affer off,for protect compressor application.| 2.0 or 20
CHYS| Control Hysteresis| CHY is used for on-off control, if PV > SV, output turns off; PV<SV-CHYS, output turns on.| 2
InP| Input specification| | InP| Input spec| InP| Input spec
---|---|---|---
0| K| 1| S
2| R| 3| Spare
4| E| 5| J
6| Spare| 7| N
8-20| Spare| 21| Pt100
1
dPt| Resolution| “0” means 1℃or 1F, “0.0” means 0.1℃/F| 0
Scb| Input Shift| Scb is used to make input shift to compensate the error produced by sensor or input signal.
PV after compensation= PV before compensation + Scb.
Note: normally set it as 0| 0
FILt| PV input filter| The value of FIL will determine the ability of filtering noise. If great interference exists, then you can increase parameter “FIL” gradually to make momentary fluctuation of measured value less than 2 to 5.
When the instrument is being metrological verified, “FIL” s can be set to 0 or 1 to shorten the response time.| 1
Fru| Power frequency and display unit| 50C –frequency 50Hz,display unit ℃. 50F– frequency 50Hz,display unit F.
60C –frequency 50Hz,display unit ℃. 60F– frequency
50Hz,display unit F.| 50C
OPH| Output highest limit| When PV<OEF, max limit 100%| 100
OEF| OPH valid range| When PV<OEF,OUTP output highest limit is OPH, When PV>OEF, no limit ,100% output
Note: if you want to avoid too quick temperature raising, and temperature is lower than 150C, only 30% is allowed for heating power, then you can set:
OEF=150.0 ℃, OPH=30%| 3200
AF| Senior function code| AF=A1+B2+E16+G64
A=0,HdAL and LdAL is for Hysteresis alarm;A=1,they are absolute alarm
B=0,single lop alarm, B=1,Dual loops alarm
E=0,HIAL ,LOAL are absolute highest alarm and howest alarm E=1,HIAL,LOAL are hysteresis highest alarm and hysteresis lowest alarm
G=0,sensor allow ,. G=1,not allow higher than highest alarm value
Set it as 0 unless an expert tells you to change.| 0
SPL| Lower limit of SV| Mininmum value of SV| 0
SPH| Upper limit of SV| Maximum value that SV allowed to be. When SPH=400,the SV range will 0~400℃.| 1200
SPr| Limit of temperature rising speed| If SPr is set as valid, when controller works, RUN light will shine if wrong or unormal speed occurs.| 0
PonP| Auto running when power on| Cont, controller stop work
StoP, when power on, it’s in stop status.
Run1, continue working
dASt, if now alarm, then work, if alarm, stop working
HoLd,(only for AI-518P),if accident short of electricity, stop work.| cont
EP1~E P8| Senior function code| Can set 1~8 field parameter. If not need or less than 8, can set it as nonE.|

Fault analysis

Phenomena Causation Treatment Method
No power
  1. No Power Supply
  2. Switch broken
  3. Wire short or fuse broken

|

  1. Check the power or change the plug socket
  2. Change the switch
  3. Check the wire or change the fuse

SX3| SV display OraL| Sensor broken| Change the sensor
Alarm or Over temperature light on| Machine body higher than the limited temperature, protect now!| The Temperature down to safety temperature and recover by itself (Inspect the reason or change the limited temperature)
Not work| Specification wrong changed| Change the specification
Temperature no rising|

  1. Controller broken
  2. Heating Element broken

| Contact with the repair

Customer Support

www.Iabbox.com

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

ibx User Manuals

Related Manuals