ibx H30SQC Digital Magnetic Stirrer with Heating User Manual

June 13, 2024
ibx

H30SQC Digital Magnetic Stirrer with
Heating
user manual

H30SQC Digital Magnetic Stirrer with Heating

ibx H30SQC Digital Magnetic Stirrer with Heating

Please read the User Manual carefully before use, and follow all operating and safety instructions!
Preface
Users should read this Manual carefully, follow the instructions and procedures, and beware of all the cautions when using this instrument.
Service
In order to guarantee this equipment works safely and efficiently, it must receive regular maintenance. In case of any faults, do not try to repair it yourself. If help is needed, you can always contact your dealer or Labbox via www.labbox.com
Please provide the customer care representative with the following information:

  • Serial number
  • Description of problem
  • Your contact information

Warranty
This instrument is warranted to be free from defects in materials and workmanship under normal use and service, for a period of 24 months from the date of invoice. The warranty is extended only to the original purchaser. It shall not apply to any product or parts which have been damaged on account of improper installation, improper connections, misuse, accident or abnormal conditions of operation.
For claim under the warranty please contact your supplier.

Safety Instructions

| Warning!
• Read the operating instructions carefully before use.
• Ensure that only trained staff works with the instrument.
---|---
| Risk of burn!
• Caution when touching the housing parts and the hotplate, which can reach a temperature of 500 ℃.
• Pay attention to the residual heat after switching off.
| Protective ground contact
• Make sure that the socket is grounded (has a protective ground contact) before use.

  • When working, wear personal safety guards to avoid the risk from:
    – Evaporation of liquids
    – Release of toxic or combustible gases

  • Set up the instrument on a stable, clean, non-slip, dry, and fireproof surface. Do not operate the instrument in an explosive atmosphere, with hazardous substances, or underwater.
    Gradually increase the speed, but reduce it if:
    – The stirring bar breaks away due to high speed.
    – The instrument is not running smoothly, or the container moves on the base plate.

  • Temperature must always be set to at least 50 ℃ lower than the fire point of the media used.

  • Be aware of hazards due to:
    – Flammable materials or media with a low boiling temperature
    – Overfilling of media
    – Unsafe container

  • Process pathogenic materials only in closed vessels.

  • Check the instrument and accessories prior to each use. Do not use damaged components. Accessories must
    be securely attached to the device and cannot come off by themselves. Always disconnect the plug before
    fitting accessories.

  • When using an external temperature sensor, ensure that the tip of the measuring sensor is at least 5-10mm
    away from the vessel bottom and wall.

  • The instrument can only be disconnected from the main power supply by pulling out the main or the
    connector plug.

  • Ensure that the voltage stated on the label matches the main power supply.

  • Make sure that the main power supply cable does not touch the hotplate, and do not cover the device during use.

  • Do not apply pressure or overheat the media on the surface of the glass ceramic as this may cause it to break.

  • The instrument must only be opened by experts.

  • Keep the instrument away from high magnetic fields.

Proper Use

The instrument is designed for mixing and/or heating liquids in schools, laboratories, or factories.

  • Ensure that there is a minimum distance of 100 mm between the devices, between the device and the wall, and above the assembly (min. 100 mm).ibx H30SQC Digital Magnetic Stirrer with Heating - Ensure

This device is not suitable for use in residential areas or other constraints mentioned in Chapter 1.

Inspection

3.1. Receiving Inspection
Carefully unpack the equipment and check for any damages that may have occurred during transport. If you need technical support, please contact the supplier.
Note:
If there is any visible damage to the system, please do not plug it into the power line.
3.2. List of Items Included
The package includes the following items:

Items Qty
Main unit 1
User manual 1

Table 1

Control

4.1. Receiving Inspectionibx H30SQC Digital Magnetic Stirrer with
Heating - Receiving Inspection

Items Descriptions
Temperature control knob Controls the heating function.
Speed Control Knob Sets the rated rotary speed. Turns the stirring function

ON or OFF.
LED Heat| Lights up when the heating function is turned ON.
LED Stir| Lights up when the stirring function is turned ON.
Timer| Press the knob, and then press “-” or “+” to set the heating time value.
LCD Display| Displays the temperature setting value (on the right side) and the real value (on the left side) when the heating knob is rotated.
SD card| Used to save the data of the test.
Power Switch| Turns the instrument ON or OFF.

Table 2
4.2. Display
ibx H30SQC Digital Magnetic Stirrer with Heating -
Display

Temp and °C Display the temperature when the heating function is switched ON.
Stir Display the stirring state when the stirring function is switched ON.
Timer Display the timer parameter when the timer function is switched ON.
Err Display an error message if an error occurs.
Hot Display a warning if the heating plate temperature is above 50 °C after

switching OFF the heating function.
Probe| Display when an external probe is being used.
SD| Display when an SD card is being used.
Set value / Real value| Display values when the heating or timer functions are switched ON.

Note:
If both heating and stirring functions have been started simultaneously, the heat display always takes higher priority. If in this case, the speed is changed using the stirring knob, it displays stirring speed and reverts to temperature after 5 seconds. The timer can only control the heating time, and the button will not work for the stirring function as it has already been started.

Trial Run

  • Make sure the required operating voltage and power supply voltage match.
  • Ensure that the socket is properly grounded.
  • Plug in the power cable and ensure that the power is on to begin initializing.
  • Place the vessel on the work plate.
  • Set the target heating time.
  • Set the target temperature and start heating.
  • Stop the heating functions.

If these operations above are normal, the device is ready to operate. If these operations are not normal, the device may be damaged during transportation, please contact your local dealer /supplier.
Warning!
Forbid to transfer the vessel when the instrument working.

Working with External Temperature Sensor

ibx H30SQC Digital Magnetic Stirrer with Heating -
ExternalThe external temperature sensor PT1000 is a manufacturer’s standard accessory. If the sensor is plugged in, “Probe” will be shown on the LCD digital display to indicate that the sensor is operating. The setting value of the external temperature sensor and the actual temperature are displayed. The safe circuit controls the hotplate temperature. Compared to the temperature control of the hotplate, the external temperature sensor can control the medium’s temperature more precisely. The heating function will be stopped automatically under abnormal conditions. Please follow the instructions below to operate:

  • Switch OFF the instrument.
  • Ensure the external temperature sensor is inserted into the media being heated.
  • Switch ON the instrument and run the heating function.

If you experience any issues with the heating function, please contact you dealer or supplier for assistance.

Residual Heat Warning (HOT)

To prevent the risk of burns from the hotplate, the unit is equipped with a residual heat warning function. When the heating function is switched off and the heating plate temperature is still above 50°C, “HOT” will flash to warn of the hazard of burns from the hotplate. When the hotplate temperature drops below 50°C, the unit will automatically switch off. If users want to turn off the LCD screen immediately, they can simply pull out the plug. Note that the residual heat warning function cannot be activated when the unit is powered off.

Logging Function

With the power switch turned off, insert the SD card into the slot on the left side of the panel.
After inserting it, turn on the power switch and operate the heating as described above, setting the temperature at 1-minute intervals.
The current temperature will be automatically logged.
The logged data can be checked in Excel format by connecting the SD card to a personal computer.”
Note:
The logging function starts automatically when you press the temperature control knob and stops when you press it again. Be sure to turn off the power switch before removing the SD card to avoid malfunctions.

Faults

  • The instrument cannot be powered ON:
    – Check whether the power line is unplugged.
    – Check whether the fuse is broken or loose.

  • Fault in power ON self-test:
    – Switch OFF the unit, then switch ON and reset the instruments to the factory default setting.

  • Unit cannot be powered OFF when switched OFF:
    – Check if the residual heat warning function is still ON and the hotplate temperature is above 50℃ (the LCD screen still works and “Hot” flashes).

If these faults are not resolved, please contact dealer/suppler.

Maintenance and Cleaning

  • Proper maintenance can keep instruments working properly and lengthen their lifetime.
  • Do not spray cleanser into the instrument when cleaning.
  • Unplug the power line when cleaning.
  • Only use recommended cleansers:
Dyes Isopropyl alcohol
Construction materials Water-containing tenside / Isopropyl alcohol
Cosmetics Water-containing tenside / Isopropyl alcohol
Foodstuffs Water-containing tenside
Fuels Water-containing tenside

Wear proper protective gloves during cleaning of the instrument.
Before using any other method for cleaning or decontamination, the user must ascertain with the manufacturer that this method will not harm the instrument.
When sending the instrument in for service, send it back in the packaging carton. Storage packing is not sufficient for the return shipment. Additionally, use a suitable transportation packing.
The enamel makes the hotplate easier to care for and more resistant to acids and bases. However, due to the enamel coating, the hotplate is more susceptible to extreme fluctuations in temperature and the force of impact. This can result in cracks forming or the coating flaking off.

Associated Standards and Regulations

Construction in accordance with the following safety standards:
EN 61010-1
UL 3101-1
CAN/CSA C22.2(1010-1)
EN 61010-2-10
Construction in accordance with the following EMC standards:
EN 61326-1
Associated EU guidelines: EMC-
guidelines: 89/336/EWG Instrument
guidelines: 73/023/EWG
Changes or modifications not expressly approved by the party responsible for compliance could void the user’s authority to operate the equipment.
NOTE: This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense.

Specifications

Items Specifications

LCD digital model
Voltage [VAC], Frequency [Hz]| 110V,60Hz / 220V,50Hz
Power [W]| 1400
Stirring point position quantity| 1
Max. stirring quantity (H2O) [L)| 30
Max. magnetic bar [Lx0, mm)| 80x 10
Motor type| DC brushless motor
Speed range [rpm]| 100-1500
Rotary speed display| LCD
Plate material| Glass ceramic
Dimensions of the work plate (mm)| 254×254
Temperature range [°C)| RT-500, increment: 1
Temperature display [°C]| LCD
Temperature display accuracy [°C)| ±0.1
Control accuracy of heating temperature [°C)| ±1°C (<100°C), ±1% (>100°C)
The safety temperature of the hotplate [°C]| 540
Temperature sensor in medium| PT1000
Control accuracy of heating temperature with temperature sensor [°C]| +0.2
*Residual heat warning| 50°C
Dimensions (mm)| 328x424x138
Permitted ambient temperature [°C]| 5-40
Permitted relative humidity| 80%
Protection class acc. to DIN 60529| IP21
SD card| Yes

www.labbox.com

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>