Ibx instruments ORB-B2 Orbital Shaker Instruction Manual

June 16, 2024
IbX instruments

Ibx instruments ORB-B2 Orbital Shaker Instruction Manual
Kenmore 251.55530510 Dehumidifier User Guide Ibx instruments ORB-B2 Orbital
Shaker

ORB-B2 Orbital Shaker with Digital Screen and Timing Function

Preface
Users should read this Manual carefully, follow the instructions and procedures, and beware of all the cautions when using this instrument.

Service
In order to guarantee this equipment works safely and efficiently, it must receive regular maintenance. In case of any faults, do not try to repair it yourself. If help is needed, you can always contact your dealer or Labbox via www.labbox.com

Please provide the customer care representative with the following information:

  • Serial number
  • Description of problem
  • Your contact information

Warranty
This instrument is warranted to be free from defects in materials and workmanship under normal use and service, for a period of 24 months from the date of invoice. The warranty is extended only to the original purchaser. It shall not apply to any product or parts which have been damaged on account of improper installation, improper connections, misuse, accident, or abnormal conditions of operation. For claim under the warranty please contact your supplier.

Installation Instructions

  1. Carefully unpack the instrument and place it on a stable, clean, non-slip, and fireproof surface.
  2. Ensure a good working environment free of any combustible media or flammable materials to avoid additional hazards.
  3. Place the shaker at a distance of more than 10cm from the wall and other units if using multiple units together.
  4. Ensure that there is no accidental spillage during speed setting.
  5. Do not use any combustible media or flammable materials to avoid any additional hazards.
  6. Ensure that the label indicates the correct voltage before connecting the unit to the power supply.
  7. Do not operate the unit with a damaged power cord.
  8. Disable the power supply while fitting the accessories.
  9. Ensure that the unit and its accessories are in good working condition before each operation

Operating Procedure

  1. Switch ON the power and the instrument will start self-checking.
  2. Press the “Mode” key and the unit will enter the adjustment mode.
  3. Press the “+/-” keys and confirm the set speed and time. Adjust the speed and time to the desired settings.
  4. Press the “Start” key and the instrument will begin shaking.
  5. Press the “Stop” key to stop the operation and return the unit to adjustment mode.

Troubleshooting and Error Code Information

Error Code Problem Reason Solution
E01 No operating

response found (LED
OFF)| Power switch is OFF| Check and conned the power supply before restarting. Check the adapter.
Switch position is OFF| Turn on the instrument and check the
speed set on the LED display.
Internal cables disconnected| Open the    instrument   and   check allconnections before reconnecting them.
E02| The shaking board
operation failure| Target speed is not set| Set the target speed and make sure the indicator lamp is ON.
Driver board damage| Replace the driver board.
Motor damage| Replace the motor.
E03| Inaccurate shaking
speed| Driver board damage| Replace the driver board.
Photoelectric switch damage| Replace the photoelectric switch.
E04| Shaking speed is
jammed| Driver board damage| Replace the driver board.
Motor damage| Replace the motor.