ORB-B2 Orbital Shaker with Digital Screen and Timing Function
Preface
Users should read this Manual carefully, follow the instructions and
procedures, and beware of all the cautions when using this instrument.
Service
In order to guarantee this equipment works safely and efficiently, it must
receive regular maintenance. In case of any faults, do not try to repair it
yourself. If help is needed, you can always contact your dealer or Labbox via
www.labbox.com
Please provide the customer care representative with the following information:
- Serial number
- Description of problem
- Your contact information
Warranty
This instrument is warranted to be free from defects in materials and
workmanship under normal use and service, for a period of 24 months from the
date of invoice. The warranty is extended only to the original purchaser. It
shall not apply to any product or parts which have been damaged on account of
improper installation, improper connections, misuse, accident, or abnormal
conditions of operation. For claim under the warranty please contact your
supplier.
Installation Instructions
- Carefully unpack the instrument and place it on a stable, clean, non-slip, and fireproof surface.
- Ensure a good working environment free of any combustible media or flammable materials to avoid additional hazards.
- Place the shaker at a distance of more than 10cm from the wall and other units if using multiple units together.
- Ensure that there is no accidental spillage during speed setting.
- Do not use any combustible media or flammable materials to avoid any additional hazards.
- Ensure that the label indicates the correct voltage before connecting the unit to the power supply.
- Do not operate the unit with a damaged power cord.
- Disable the power supply while fitting the accessories.
- Ensure that the unit and its accessories are in good working condition before each operation
Operating Procedure
- Switch ON the power and the instrument will start self-checking.
- Press the “Mode” key and the unit will enter the adjustment mode.
- Press the “+/-” keys and confirm the set speed and time. Adjust the speed and time to the desired settings.
- Press the “Start” key and the instrument will begin shaking.
- Press the “Stop” key to stop the operation and return the unit to adjustment mode.
Troubleshooting and Error Code Information
Error Code | Problem | Reason | Solution |
---|---|---|---|
E01 | No operating |
response found (LED
OFF)| Power switch is OFF| Check and conned the power supply before
restarting. Check the adapter.
Switch position is OFF| Turn on the instrument and check the
speed set on the LED display.
Internal cables disconnected| Open the instrument and check
allconnections before reconnecting them.
E02| The shaking board
operation failure| Target speed is not set| Set the target speed and make sure
the indicator lamp is ON.
Driver board damage| Replace the driver board.
Motor damage| Replace the motor.
E03| Inaccurate shaking
speed| Driver board damage| Replace the driver board.
Photoelectric switch damage| Replace the photoelectric switch.
E04| Shaking speed is
jammed| Driver board damage| Replace the driver board.
Motor damage| Replace the motor.