IBX Orb-B2 Orb-B2 Orbital Shaker with Digital Screen User Manual

June 14, 2024
ibx

IBX Orb-B2 Orb-B2 Orbital Shaker with Digital Screen

Orb-B2 Orbital Shaker with Digital Screen and Timing Function

Product Information

The Orb-B2 Orbital Shaker is a high-quality laboratory instrument designed for shaking and mixing applications. It features a digital screen and timing function, allowing precise control over speed and duration. The shaker is equipped with a reliable motor and a sturdy shaking board to ensure efficient and consistent performance. The instrument comes with a 24-month warranty from the date of invoice, covering defects in materials and workmanship under normal use and service.

Usage Instructions

Installation Instructions

  1. Carefully unpack the instrument and place it on a stable, clean, non-slip, and fireproof surface.
  2. Ensure a good working environment free of any combustible media or flammable materials to avoid accidents.

Operating Procedure

  1. Switch ON the power and the instrument will start self-checking.
  2. Press the Mode key and the unit will enter the adjustment mode.
  3. Press the +/- keys and confirm the set speed and time. Adjust the speed and time to the desired settings.
  4. Press the Start key and the instrument will begin shaking.
  5. Press the Stop key to stop the operation and return the unit to adjustment mode.

Troubleshooting and Error Code Information

  • Error Code E01: No operating response found (LED OFF) Reason: Power switch is OFF Solution: Check and connect the power supply before restarting. Check the adapter.
  • Error Code E02: The shaking board operation failure Reason: Inaccurate shaking speed Solution: Turn on the instrument and check the speed set on the LED display. Set the target speed and make sure the indicator lamp is ON.
  • Error Code E03: Shaking speed is jammed Reason: Driver board damage Solution: Open the instrument and check all connections before reconnecting them. Replace the driver board.
  • Error Code E04: Target speed is not set Reason: Motor damage Solution: Replace the motor.

For more information and support, please visit www.labbox.com.

User Manual

Orb-B2 Orbital Shaker with Digital Screen and Timing Function

Preface
Users should read this Manual carefully, follow the instructions and procedures, and beware of all the cautions when using this instrument.

Service
In order to guarantee this equipment works safely and efficiently, it must receive regular maintenance. In case of any faults, do not try to repair it yourself. If help is needed, you can always contact your dealer or Labbox via www.labbox.com.

Please provide the customer care representative with the following information

  • Serial number
  • Description of problem
  • Your contact information

Warranty

This instrument is warranted to be free from defects in materials and workmanship under normal use and service, for a period of 24 months from the date of invoice. The warranty is extended only to the original purchaser. It shall not apply to any product or parts which have been damaged on account of improper installation, improper connections, misuse, accident, or abnormal conditions of operation. For claim under the warranty please contact your supplier.

Installation Instructions

  1. Carefully unpack the instrument and place it on a stable, clean, non-slip, and fireproof surface.
  2. Ensure a good working environment free of any combustible media or flammable materials to avoid additional hazards.
  3. Place the shaker at more than 10 cm from the wall and other units if using multiple units together.
  4. Ensure that there is no accidental spillage during speed setting.
  5. Do not use any combustible media or flammable materials to avoid any additional hazards.
  6. Ensure that the label indicates the correct voltage before connecting the unit to the power supply.
  7. Do not operate the unit with a damaged power cord.
  8. Disable the power supply while fitting the accessories.
  9. Ensure that the unit and its accessories are in good working condition before each operation.

Operating Procedure

  1. Switch ON the power and the instrument will start self-checking.
  2. Press the “Mode” key and the unit will enter the adjustment mode.
  3. Press the “+/-” keys and confirm the set speed and time. Adjust the speed and time to the desired settings.
  4. Press the “Start” key and the instrument will begin shaking.
  5. Press the “Stop” key to stop the operation and return the unit to adjustment mode.

Troubleshooting and Error Code Information

Error Code Problem Reason Solution
**** E01 **** No operating response found (LED OFF) **** Power switch is
OFF Check and connect the power supply before restarting. Check the adapter.
**** Switch position is OFF Turn on the instrument and check the speed set

on the LED display.
Internal cables disconnected| Open the instrument and check all connections before reconnecting them.
E02| The shaking board operation failure| Target speed is not set| Set the target speed and make sure the indicator lamp is ON.
Driver board damage| Replace the driver board.
Motor damage| Replace the motor.
**** E03| Inaccurate shaking speed| Driver board damage| Replace the driver board.
Photoelectric switch damage| Replace the photoelectric switch.
E04| Shaking speed is jammed| Driver board damage| Replace the driver board.
Motor damage| Replace the motor.

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

ibx User Manuals

Related Manuals