ibx instruments WB-T Transparent Water Bath User Manual
- June 9, 2024
- IbX instruments
Table of Contents
user manual
WB-T Transparent Water Bath
WB-T Transparent Water Bath
Please read the User Manual carefully before use, and follow all operating and
safety instructions!
Preface
Thank you for purchasing our Heating Mantle. Users should read this manual carefully, follow the instructions and procedures, and beware of all the preventive measures when using this instrument.
Service
If help is needed, you can always contact your dealer or Labbox via
www.labbox.com
Please, provide the customer service representative with the following
information:
- Serial number (on the back side)
- Description of the problem
- Your contact information
Warranty
This instrument is guaranteed to be free from defects in materials and workmanship under normal use and service, for a period of 12 months from the date of invoice. The warranty is extended only to the original purchaser. It shall not apply to any product or parts that have been damaged due to improper installation, improper connections, misuse, accident, or abnormal conditions of operation. For claim under the warranty, please contact your supplier.
Safety Instructions
| Connect the device to an earthed socket to ensure the safety of the machine
and the experiment. Make sure the voltage is the same as required by the
device.
---|---
| The use of this instrument in flammable, explosive, poisonous, or highly
corrosive environments or with hazardous substances is forbidden.
| Place the incubator on a horizontal, flat, clean, non-slip and stable
surface. Adjust the feet so that the device is leveled.
| This item must only be used by qualified staff that has read the
instructions manual and knows how to operate it.
| Do not place the device near any heat source. Keep away from high magnetic
fields. Do not put volatile, flammable and explosive materials in the machine,
otherwise it could cause an explosion or a fire.
| Non-professionals must not disassemble or repair this machine.
| Read the instructions manual before using this device.
-
When working, wear the necessary personal protective equipment to avoid the risk of:
– Burns caused by contact with hot surfaces or materials
– Burns caused by splashing and evaporation of liquids
– Intoxication caused by release of toxic or flammable gases. -
Set up the instrument on a spacious, stable, clean, non-slip, dry, and fireproof surface that can support the weight of the equipment. Do not operate the instrument in explosive atmospheres or with hazardous substances.
-
During use, do not touch the heating pipe with your hands as it could cause burns.
-
The temperature must be set at least 25ºC below the flash point of the substances used.
-
Process pathogenic materials only in closed vessels.
-
Beware of hazards due to:
– Flammable materials or media with a low boiling temperature
– Overfilling of the vessel
– Unsafe vessels -
The instrument can only be disconnected from the main power supply by pulling from the plug, not the cable.
-
Check the instrument and accessories before handling for damage prior to every use. Do not use damaged components. Safe operation is only guaranteed with the accessories described in the “Accessories” chapter. Accessories must be securely attached to the device and cannot come off by themselves. Always disconnect the plug before the assembly or disassembly of accessories.
-
The voltage stated on the label must correspond to the main power supply.
-
Ensure that the mains cable does not touch the heating base.
-
Do not cover the device.
-
Keep away from high magnetic fields.
-
Pay attention to the temperature setting. Never leave the equipment unattended while the heating function is ON.
Product description
2.1. Summary
This product is suitable for distillation, concentration, drying and
thermostatic heating in medical units, universities and schools, scientific
research units, and laboratories of industrial and mining enterprises like
chemical painting and dyeing enterprises or pharmaceutical enterprises.
2.2. Structure features
– The innovative design structure that effectively prevents steam leakage.
– Intelligent temperature control instrument with high-brightness digital
display and touch key that allows simple operation and high-precision
temperature control.
– The shell, interior and cover of the product are all made of transparent
engineering plastics which allow to clearly see the object that is being
heated.
2.3. Cleaning and maintenance
– Before use, add water until it reaches the clapboard, then connect to the
power supply and heat up. Samples cannot be heated without sufficient water.
– The equipment must be used only when the resistor is covered by water; doing
otherwise will harm the resistor.
– Avoid violent vibration, collisions, and keep away from corrosive gas.
– After use, empty the tank, dry it, and clean it. Proper maintenance could
lengthen the device’s working lifetime.
– Do not spray cleanser directly onto the device.
– If sent in for repair, clean, disinfect and pack it in its original
packaging before shipping.
2.4. Main technical parameters
Function
| Voltage (V)| 200-240 /100-120
---|---|---
Power (W)| 800
Sensor| Pt100
Temperature Motion (℃)| ±0.5
Temperature Range (℃)| 0~100
Temperature Sensitivity (℃)| ≤±1
Time Range| 0~9999 min or hours
Measurement error| < 0.3%
Structure
| Shelves Height (mm)| Distance from bottom 100mm
Shelves Holes Diameter| 5 X104mm
Inner Chamber Size (LWH)| 280 x 220 x150 mm
Exterior Size (LWH)| 400 x 260x 190 mm
Packing Size (LWH)| 450 x 350 x270 mm
N.W. (KG)| 3.5
G.W. (KG)| 5
Working conditions| Environmental temperature (°C)| 5~40
Relative humidity| RH < 85%
2.5. Control monitor Indicators
RUN/AT: lights up when the functions are running. Flashes during auto-
tuning.
OUT: lights up when the heating output is working.
ALM: lights up when there is an over-temperature alarm.
SW: lights up when thee is not enough water.
Keys
◄: Press to shift the set value in the setting state.
▼: Press to reduce the set value in the setting state. Hold to reduce
continuously.
▲: Press to increase the set value in the setting state. Hold to increase
continuously.
SET: Press to enter the setting state and to enter a value.
Screen
PV: Displays the measurement value.
SV: Displays the parameter value.
Operation
- Place the water bath on a stable surface that can support the weight of the device and leave 30 cm clear on each side.
- Fill the tank to 2/3 of the total volume.
- Connect to a power supply that is consistent with the voltage requirement of this instrument. Make sure it is properly earthed.
- Switch the device on. The power supply indicator will light up.
- Set the desired temperature (see instructions below). The temperature controller will display the temperature inside the working chamber.
- If desired, set the time (see instructions below).
Note: The thermostatic bath cannot be used with deionized, demineralized or distilled water as it will not conduct electricity properly
Temperature and time setting
4.1. No timer function
- Press the SET button to enter the setting state → the indicator SP and the temperature set value will show.
- Use ◄ ▼▲ to modify the value.
- Press SET to enter → the controller will return to the normal display and the value will be saved
4.2. Timer function
- Press SET to enter the setting state → the indicator SP and the temperature set value will show.
- Press SET again → the indicator SP and the time set value will show.
- Use ◄ ▼▲ to modify the value.
- Press SET to return to the initial screen. The value will be saved automatically.
When the time is set to 0, the controller will run continuously until manually
stopped.
Otherwise, when the timer is set, the lower window will display the
temperature set value or the running time depending on the ntd parameter
configuration (see later). When the time is over, the lower window will
display End and the alarm will go off and ring for 60s. Stop it by pressing
any button. Press the RST button for 3s to restart the controller.
Self-tuning of PID
When the temperature control effect is not ideal, a system self-tuning can be
caried out where the equipment will automatically configurate its PID
parameters. To do that, follow the next instructions:r
– Press AT for 6s → the controller will run the auto-tuning function. The AT
indicator will flash.
– To stop the process, press AT for 6s while it is running.
– During the process, the SET button does not work, and the lower window
always displays the temperature set value.
Internal parameters setting
6.1. How to enter
To adjust the internal parameters, you must enter the required password. To do that, follow these instructions:
- Press SET for 3s → the controller will display the password prompt Lc.
- Use ◄ ▼▲ to choose the required value (see the password values below).
- Press SET to enter → the controller will enter the internal parameters setting state.
To return to the regular state at any time, press SET for 3s.
6.2. Parameter tables
INDICATOR| NAME| INSTRUCTIONS OF THE FUNCTION| (RANGE)
DEFAULT VALUE
---|---|---|---
Lc| Password| The password to modify the parameters in this table is
Lc = 3| 0
ALH| Over-temp alarm| If SV > (SP + ALH), the ALM light turns
on. The buzzer sounds and the heating output turns off.| (0~100.0℃) 20.0
ALL| Under-temp alarm| If SV < (SP – ALL), the ALM light
flashes, the buzzer sounds.| (0~100.0℃) 0
T| Control cycle| The temperature control cycle.| (1~60S) 5
P| Proportional band| Adjustment of proportional function.| (1~400.0℃)
35.0
I| Integration time| Adjustment of integration function.| (1~2000S) 200
d| Differential time| Adjustment of differential function.| (0~1000S)
200
| Pb| Zero point adjustment| When the error is comparatively larger,
this value might need adjustment.
Pb = actual value – measure value| (-12.0~12.0℃) 0
| PK| Large value adjustment| Correction of errors caused by sensor high
temperature values. actual value – measured value
| |
(-999~999)0
INDICATOR| NAME| INSTRUCTIONS OF THE FUNCTION| (RANGE)
DEFAULT VALUE
---|---|---|---
Lc| Password key| The password to modify the parameters in this table is
Lc = 9| 0
ndA| Temp alarm mode| 0 : With over-temp alarm only.
1 : With over-temp alarm and under-temp alarm at the same time.| (0~1) 0
|
ndt
|
Timer mode
| 0 : No timer function.
1 : With timer function, the lower window displays the running time after
the measured temperature reaches to the set value.
2 : With timer function, the lower window always displays the setting
value.| |
(0~2) 1
Hn| Timer unit| 0 : Minutes 1: Hours| (0~1) 0
|
EH
|
Timer end mode
| 0 : Continue to maintain the constant temperature when the running time
is over.
1 : Stop the temperature control when the running time is over.| |
(0~1) 0
oPn| Auxiliary parameter| 0 : ON
1 : OFF| (0~1) 0
nP| Maximum power| Percentage of maximum power heating output.| (0~100%)
100
Co| Off point| If SV > (SP + Co), stop the heating output forcibly.|
(0~100.0℃) 50.0
SPL| Minimum set point| Adjustment of the minimum temperature.| (0~0) 0
SPH| Maximum set point| Adjustment of the maximum temperature.|
(0~400.0) 300.0
Fault analysis
FAILURE | CAUSE | HANDLING METHOD |
---|
No power supply
| Bad contact between plug and socket| Replace the plug or socket tube
The fuse is burnt| Replace the fuse with one with the same specifications
No temperature rise
| The temperature controller is broken| Replace the instrument
The sensor is broken| Replace the sensor
The set temperature is lower than water temperature| Reset the temperature
The heating pipe is burnt| Replace the heating pipe
There is a big difference between display temp. and actual temp.| The
temperature controller| Replace the temperature controller
The temperature sensor is broken.| Replace the temperature sensor
Wiring diagrams