Ibx instruments ORB-B2 Orbital Shaker with Digital Screen and Timing Function User Manual
- June 12, 2024
- IbX instruments
Table of Contents
ORB-B2 Orbital Shaker with Digital Screen and Timing Function
Product Information
The ORB-B2 Orbital Shaker with Digital Screen and Timing Function is a
high-quality instrument designed for laboratory use. It features a digital
screen and timing function, allowing for precise and controlled shaking of
samples. The instrument comes with a 24-month warranty, covering defects in
materials and workmanship under normal use and service. The warranty is valid
for the original purchaser and does not apply to products or parts damaged due
to improper installation, connections, misuse, accidents, or abnormal
conditions of operation. For warranty claims, please contact your supplier.
Installation Instructions
- Carefully unpack the instrument and place it on a stable, clean, non-slip, and fireproof surface.
- Ensure a good working environment free of any combustible media or flammable materials to avoid additional hazards.
- Place the shaker at a distance of more than 10cm from the wall and other units if using multiple units together.
- Ensure that there is no accidental spillage during speed setting.
- Do not use any combustible media or flammable materials to avoid any additional hazards.
- Ensure that the label indicates the correct voltage before connecting the unit to the power supply.
- Do not operate the unit with a damaged power cord.
- Disable the power supply while fitting the accessories.
- Ensure that the unit and its accessories are in good working condition before each operation.
Operating Procedure
- Switch ON the power and the instrument will start self-checking.
- Press the Mode key and the unit will enter the adjustment mode.
- Press the +/- keys and confirm the set speed and time. Adjust the speed and time to the desired settings.
- Press the Start key and the instrument will begin shaking.
- Press the Stop key to stop the operation and return the unit to adjustment mode.
Troubleshooting and Error Code Information
Error Code | Problem | Reason | Solution |
---|---|---|---|
E01 | No operating response found (LED OFF) | Power switch is OFF | Check and |
connect the power supply before restarting.
E02| Internal cables disconnected| Switch position is OFF| Check the adapter.
Turn on the instrument and check the speed
set on the LED display.
E03| Inaccurate shaking speed| Driver board damage| Set the target speed and
make sure the indicator lamp is ON.
Replace the driver board if necessary.
E04| Shaking speed is jammed| Motor damage| Replace the motor. Replace the
driver board if necessary.
E05| Target speed is not set| Driver board damage| Set the target speed and
replace the driver board if
necessary.
E06| Photoelectric switch damage| Motor damage| Replace the photoelectric
switch. Replace the driver board if
necessary. Replace the motor if necessary.
For more information and support, please visit our website at www.labbox.com.
ORB-B2 Orbital Shaker with Digital Screen and Timing Function
Please read the User Manual carefully before use, and follow all operating and
safety instructions!
Preface
Users should read this Manual carefully, follow the instructions and procedures, and beware of all the cautions when using this instrument.
Service
In order to guarantee this equipment works safely and efficiently, it must
receive regular maintenance. In case of any faults, do not try to repair it
yourself. If help is needed, you can always contact your dealer or Labbox via
www.labbox.com
Please provide the customer care representative with the following
information:
- Serial number
- Description of the problem
- Your contact information
Warranty
This instrument is warranted to be free from defects in materials and workmanship under normal use and service, for a period of 24 months from the date of invoice. The warranty is extended only to the original purchaser. It shall not apply to any product or parts which have been damaged on account of improper installation, improper connections, misuse, accident, or abnormal conditions of operation. For a claim under the warranty please contact your supplier.