MACHEREY-NAGEL TNb 220 Tube test NANOCOLOR Robot Total Nitrogen Instructions
- June 3, 2024
- MACHEREY-NAGEL
Table of Contents
MACHEREY-NAGEL TNb 220 Tube test NANOCOLOR Robot Total Nitrogen
Instructions
Overview
The test is suitable for the photometric determination of total nitrogen.
The test is suitable for surface water, ground and drinking water and
wastewater.
- Measuring range: 5–220 mg/L N (method 0881)
- Number of tests: 20
- Wavelength for photometric determination: 345 / 350 / 365 nm
- Shelf life: 12 months
- Reaction time: 10 minutes
- Storage temperature: 15–25 °C
- Storage conditions: upright
Method
Oxidative decomposition with subsequent interference compensation and photometric determination with 2,6-dimethylphenol in a sulfuric acid/phosphoric acid mixture. Decomposition: Analogous to DIN EN ISO 11905-1 H36. Determination: Analogous to ISO 7890-1; DIN 38405-D9.
Interferences
The foreign materials shown here do not interfere with the test up to the indicated concentrations (in mg/L). The cumulative effect of different interfering ions has not been tested. Data in mg/L:
- Cl- , COD: 10000
This method is not suitable for analyzing seawater.
Turbidities cause higher measurement values.
Reagents and accessories
Contents of reagents set:
-
2 x 20 test tubes R0
-
1 reagent R2
-
1 decomposition reagent
-
1 NANOFIX compensation reagent
-
1 measuring spoon orange
Required devices: -
MACHEREY-NAGEL photometer
-
MACHEREY-NAGEL heating block
-
Digital piston pipette 200–1000 µL (REF 91671) with pipette tips (REF 91667)
Standards
- NANOCONTROL Multi standard Sewage inflow (REF 925012)
Sampling and preparation
See DIN EN ISO 5667-3-A 21.
Adjust to pH 5–9 prior to analysis
Quality control
The measurement of a blank value and a standard is recommended before every measuring series as quality control measure.
Quality data:
The following data were determined during production according to ISO 8466-1 and DIN 38402-A51:
- Number of LOTs: 22
- Standard deviation of the method: ± 2.79 mg/L N
- Coefficient of variation of the process: ± 2.54 %
- Confidence interval: ± 6.44 mg/L N
Specified data for procedure:
- Sensitivity (absorbance of 0.010 A corresponds to): 1.55 mg/L N
- Accuracy of a measurement value: ± 4.79 mg/L N
LOT-specific certificates are available at www.mn-net.com.
Procedure
- Open test tube A
- Pipette 0.5 mL of sample into test tube
- Add 1 level measuring spoon of decomposition reagent
- Seal test tube and shake vigorously
- Heat for 30 min at 120°C
- Take the tube from the heating block
- Swirl
- Cool to room temperature
- Open test tube
- Add 1 NANOFIX compensation reagent
- Seal test tube and shake vigorously
- Open test tube TNb 220
- Pipette 0.5 mL decomposition solution into the tube
- Add 0.5 mL R2
- Seal test tube and turn upside down 3×
- Wait 10 min
- Clean outside of test tube
- Measure
Notes
When using other photometers, make sure measurements are possible in test
tubes (16 mm OD) and calibrate the method.
Correction value e. g. for colored or turbid samples possible (see photometer
manual).
The pH value of the sample to be decomposed must be between pH 5 and 9; if
necessary, adjust with sodium hydroxide solution or sulfuric acid. Nitrogen
concentrations above the double measuring range can simulate results within
the single measuring range and thus cause a wrong evaluation. Dilute the
sample until the measured value is within the measuring range previously
indicated by the test. For waters of unknown concentrations we recommend that
you perform the test with very different dilutions (e.g. 1+9, 1+99) until the
last dilution confirms the previous value. For samples which consume large
amounts of oxidizing substances (e.g. for COD values above 10,000 mg/L O2),
decomposition can be incomplete.
In such cases repeat the decomposition with the previously diluted original
sample.
Information regarding safety can be found on the box’ label and in the safety
data sheet. You can download the SDS from www.mn-net.com/SDS.