MACHEREY-NAGEL 963029 Chemical Oxygen Demand Instructions

June 3, 2024
MACHEREY-NAGEL

MACHEREY-NAGEL 963029 Chemical Oxygen Demand

MACHEREY-NAGEL-963029-Chemical-Oxygen-Demand-
PRODUCT

Method

Photometric determination of chromium(III) concentration after oxidation with potassium dichromate / sulfuric acidMACHEREY-NAGEL-963029-Chemical-Oxygen-
Demand-FIG-1

Contents of reagent set:
20 test tubes COD 1500 Hg-free
Hazard warning:
Information regarding safety can be found on the box’ label and in the safety data sheet. You can download the SDS from www.mn-net.com/SDS.

Interferences

Chloride interferes: 100 mg/L Cl– ≜ approx. 22 mg/L COD. Samples containing more than 1000 mg/L Cl– should be diluted prior to determining the COD. Moreover, chloride contents up to 2000 mg/L can be eliminated using NANOCOLOR® cartridges for chloride elimination (REF 963911). For determination of the concentration of chlorides we recommend a preliminary test with QUANTOFIX® Chloride (REF 91321).
The method cannot be applied for the analysis of sea water.

Procedure

Requisite accessories: NANOCOLOR® heating block, piston pipette with tips

Decomposition at 148 °C

Open test tube, hold it diagonally and slowly add 2.0 mL test sample to contents without mixing so that two sepa ra te lay ers are formed; screw cap securely on to test tu be, hold tube by the cap, place tube into the safety bottle and shake (Cau tion, test tube becomes hot), then pla ce tube into the heating block.
After 2 h remove test tube from heating block, after about 10 min (test tube is still warm) shake once and allow to cool to room temperature.
Clean outside of test tube and measure.

Short time COD at 160 °C

Open test tube, hold it diagonally and slowly add 2.0 mL test sample to contents without mixing so that two sepa ra te lay ers are formed;
screw cap securely on to test tu be, hold tube by the cap, place tube into the safety bottle and shake
(Cau tion, test tube becomes hot), then pla ce tube into the heating block.
After 30 min remove test tube from heating block, after about 10 min (test tube is still warm) shake once and allow to cool to room temperature.
Clean outside of test tube and measure.

  • In contrast to the digestion at 148 °C, the short time COD is characterized by a higher digestion temperature and reduced reaction time. Therefore we recommend to compare the results of the short time COD from time to time (150 ± 5 °C / 2 h ± 10 min).

Measurement

For MACHEREY-NAGEL photometers see manual, test 0-29.

Photometers of other manufacturers:
For other photometers check whether measurement of round glass tubes is pos sible. Verify factor for each type of instrument by measuring standard solutions.

Analytical quality control:
NANOCONTROL COD 1500 (REF 92529)

Disposal

Information regarding disposal can be found in the safety data sheet. You can download the SDS from www.mn-net.com/SDS.

Storage
Store the test kit in a cool and dry place. Avoid exposing the test kit to sunlight.