MACHEREY-NAGEL Iron 1 Performing Colorimetric Tests Kit Instructions

June 3, 2024
MACHEREY-NAGEL

MACHEREY-NAGEL Iron 1 Performing Colorimetric Tests Kit

MACHEREY-NAGEL-Iron-1-Performing-Colorimetric-Tests-Kit-
featured

Test kit for performing colorimetric tests on iron ions in surface water and sewage

Method

Colorimetric determination of Fe(II) and Fe(III) ions with a triazine derivate

Measurement range

0.04–1.0 mg/L Fe

Contents of test kit (refill pack)

sufficient for 200 tests

  • 30 mL Fe-1*
  • 2 screw-plug measuring glasses
  • 1 slide comparator
  • 1 color chart
  • 1 plastic syringe 5 mL
  • 1 instruction for use

Hazard warning:
Information regarding safety can be found on the box label and in the safety data sheet. You can download the SDS from www.mn-net.com/SDS.

Instructions for use

Also, refer to the pictogram on the back of the color chart

  1. Pour a 5 mL water sample into each of the measuring glasses using the plastic syringe. Place a measuring glass on position A in the comparator.
    Only add the reagent to measuring glass B.

  2. Add 5 drops of Fe-1, seal the glass, and mix.

  3. Open the glass after 3 min and place it on position B in the comparator.

  4. Slide the comparator until the colors match in the inspection hole on top. Check the measurement reading in the recess on the comparator reed. Mid-values can be estimated.

  5. After use, rinse out both measuring glasses thoroughly and seal them.

The reagents can also be used for the photometric evaluation with photometers PF-12 and PF-12Plus.

This technique can be used also for analyzing seawater.

Disposing of the samples:
Information regarding disposal can be found in the safety data sheet. You can download the SDS from www.mn-net.com/SDS.

Interferences

The following quantities of ions will not interfere: ≤ 5 mg/L Co2+, Ni2+, MoO42–; ≤ 3 mg/L Cu2+.

Conversion table

mg/L Fe

|

mmol/m 3

---|---
0.04| 0.7
0.07| 1.3
0.10| 1.8
0.15| 2.7
0.20| 3.6
0.30| 5.4
0.50| 9.0
1.0| 18

Storage:
Store the test kit in a cool (< 25 °C) and dry place.

MACHEREY-NAGEL GmbH & Co. KG·
Neumann-Neander-Str. 6–8 · 52355 Düren · Germany Tel.: +49 24 21 969-0 · info @mn-net.com· www.mn-net.com.

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

MACHEREY-NAGEL User Manuals

Related Manuals