MACHEREY-NAGEL 985 098 Tube test NANOCOLOR Aluminum 07 Instructions
- June 3, 2024
- MACHEREY-NAGEL
Table of Contents
MACHEREY-NAGEL 985 098 Tube test NANOCOLOR Aluminum 07
Method
Photometric determination with eriochrome cyanine R
Contents of reagent set:
19 test tubes Aluminum 07 with NANOFIX
- 1 test tube with 11 mL Aluminum 07 R2
- 1 test tube with 11 mL Aluminum 07 R3
- 1 test tube with 11 mL Aluminum 07 R4
- 1 test tube with blank value ”NULL”
Hazard warning:
This test does not contain any harmful substances which must be specially
labelled as hazardous. Preliminary tests:
If the order of magnitude of the concentration in a sample is not known, a
preliminary test with QUANTOFIX® Aluminum (5–500 mg/L Al3+, REF 913 07)
rapidly gives this information. From the order of magnitude the required
dilution can be calculated and prepared directly.
Interferences
Fluoride interferes.
Turbid samples are to be filtered (membrane filter 0.45 μm, REF 916 50). The
total Aluminum can be determined with NANOCOLOR® NanOx Metal (REF 918 978) and
microwave decompo-sition.
The following ions will not interfere:
The method can be applied also for the analysis of sea water.
Procedure
Requisite accessories: piston pipette with tips
Measurement:
For NANOCOLOR® photometers and PF-12 see manual, test 0-98.
Measurement when samples are colored or turbid:
For all NANOCOLOR® photometers see manual, use key for correction value.
Photometers of other manufacturers:
For other photometers check whether measurement of round glass tubes is
possible. Verify factor for each type of instrument by measuring standard
solutions. The factor depends strongly on the wavelength.
Analytical quality control:
NANOCONTROL Multistandard Drinking water (REF 925 018)