MACHEREY-NAGEL 931288 VISOCOLOR ECO Oxygen Test Kit Instructions
- June 3, 2024
- MACHEREY-NAGEL
Table of Contents
MACHEREY-NAGEL 931288 VISOCOLOR ECO Oxygen Test Kit
Oxygen
Test kit for performing colorimetric tests on dissolved oxygen in surface
water and sewage
Method:
modified Winkler method
In the alkaline range, dissolved oxygen oxidizes manganese(II) ions to higher
manganese hydroxides. Acidification leads to the release of manganese(III)
ions which react with a special reagent forming a dark red dye.
Measurement range:
- 1–10 mg/L O2
Contents of test kit (*refill pack)
sufficient for 50 tests
- 15 mL O2-1*
- 15 mL O2-2*
- 30 mL O2-3*
- 2 screw-plug measuring glasses
- 1 slide comparator
- 1 colour chart
- 1 plastic syringe 1 mL
- 1 instructions for use*
additionally required: oxygen reaction bottle (REF 915498)
Hazard warning
Information regarding safety can be found on the box’ label and in the safety data sheet. You can download the SDS from www.mn‑net.com/SDS.
Instructions for use
also refer to the pictogram on the back of the color chart
Cover the working surface with a polyethylene-coated filter paper.
-
Pour a 1 mL water sample into one of the measuring glasses and place it on position A in the comparator.
-
Rinse the oxygen reaction bottle several times with the water to be tested and fill until it overflows without air bubbles.
-
Add 5 drops of O2‑1.
-
Add 5 drops of O2‑2, close the bottle with the stopper (avoid air bubbles) and mix by shaking.
-
After 1 min add 12 drops of O2‑3, close the bottle and shake well until the deposit is dissolved.
-
Pour 1 mL of the resultant reaction solution into the second measuring glass and place it on position B in the comparator.
-
Slide the comparator until the colors match in the inspection hole on top.
Check the measurement reading in the recess on the comparator reed. Midvalues can be estimated. -
After use, rinse out the oxygen reaction bottle and both measuring glasses thoroughly and seal them.
The reagents can be used for the photometric evaluation with photometer
PF-12 / PF-12Plus.
The method can be applied also for the analysis of sea water.
Disposing of the samples:
Information regarding disposal can be found in the safety data sheet. You can
download the SDS from www.mn‑net.com/SDS.
Interferences:
Most oxidizing and reducing substances interfere, e. g. active chlorine,
higher manganese compounds, ascorbic acid, iodide, nitrite, sulfide and
sulfite.
Organic compounds interfere, if the potassium permanganate consumption exceeds
60 mg/L.
Conversion table
mg/L O2 mmol/m3
1 31
2 63
3 94
4 **** 125
6 190
8 250
10 310
Storage:
Store the test kit in a cool (< 25 °C) and dry place.
MACHEREY‑NAGEL GmbH & Co. KG · Neumann-Neander-Str. 6–8 · 52355 Düren ·
Germany
Tel.: +49 24 21 969-0 · Fax: +49 24 21 969-199 · info@mn-net.com · www.mn‑net.com