MACHEREY-NAGEL 931219 VISOCOLOR ECO Free Chlorine 6 Test Kit Instructions

June 3, 2024
MACHEREY-NAGEL

MACHEREY-NAGEL 931219 VISOCOLOR ECO Free Chlorine 6 Test Kit

MACHEREY-NAGEL-931219-VISOCOLOR-ECO-Free-Chlorine-6-Test-Kit-
PRO

Method

Photometric determination of free and total chlorine. At a pH value of 6.2 to 6.5 in a phosphate buffered system, free chlorine reacts with N,N-diethyl-1,4-phenylene diamine (DPD) and forms a red-vi- olet dye. In the presence of iodide ions, the content of total chlorine (free and combined chlorine together) can be determined.

Measurement range: 0.05–6.00 mg/L Cl2

Contents

  • REF 931217 (free and total) sufficient for 200 tests
    • 28 g Cl2-1
    • 30 mL Cl2-2
    • 1 measuring spoon 85 mm
    • 1 plastic syringe 5 mL
    • 1 instructions for use
  • REF 931219 (free) sufficient for 400 tests
    • 2 x 28 g Cl2-1
    • 1 measuring spoon 85 mm
    • 1 plastic syringe 5 mL 1 instructions for use

Hazard warning: Information regarding safety can be found on the box’ label and in the safe- ty data sheet. You can download the SDS from www.mn- net.com/SDS.

Procedure

Requisite accessories: test tubes 16 mm OD (REF 91680)

  • a) Free chlorine
    • Rinse test tube 16 mm OD several times with the sample (the pH value of the sample must be between pH 4 and 8) and fill with 5 mL sample.
    • Place test tube in photometer as blank value and adjust for zero.
    • Add 1 level measuring spoon of Cl2-1, close and shake well for 20 s.
    • Clean outside of test tube and measure after 1 min.
  • b) Total chlorine (only 931217)
    • Open test tube again, add 3 drops of Cl2-2, close and mix.
    • Clean outside of test tube and measure after 2 min.
  • c) Combined chlorine
    The content of combined chlorine can be calculated as difference of total and free chlorine.

Measurement:
The reagents are also suitable for photometric evaluation. See manual for MN photometer. After use, rinse out test tubes thoroughly and seal them. The method can be applied also for the analysis of sea water.

Interferences

The temperature of the water sample should be between 10 and 50 °C.
The determination of free chlorine measures bromine, bromamine, chloramine, iodine and, in part, chlorine dioxide as well. Higher manganese compounds simulate free chlorine. Chlorine concentrations above 10 mg/L can bleach the red reaction color (low results).
Rinse test tubes several times thoroughly. Residues of Cl2-2 can cause higher values for free chlorine.

Conversion

1.0 mg/L Cl2 ≜ 1.9 mg/L ClO2 ≜ 1.5 mg/L OCl– ≜ 2.1 mg/L NaOCl ≜ 2.3 mg/L Br2 ≜ 3.6 mg/L I2
Disposing of the samples: Information regarding disposal can be found in the safety data sheet. You can download the SDS from www.mn- net.com/SDS.
Storage: Store the test kit in a cool (< 25 °C) and dry place.