MACHEREY-NAGEL 985 049 Test 0-49 NANOCOLOR Silver 3 Instructions
- June 3, 2024
- MACHEREY-NAGEL
Table of Contents
MACHEREY-NAGEL 985 049 Test 0-49 NANOCOLOR Silver 3
Method:
Silver ions react with an indicator to form a blue dye.
Contents of reagent set
20 test tubes Silver 3
- 1 test tube with 11 mL Silver 3 R2
- 1 test tube with 11 mL Silver 3 R3
Hazard warning:
This test does not contain any harmful substances which must be specially
labelled as hazardous.
Interferences
Silver compounds like silver bromide, silver chloride, silver iodide, silver
cyanide or silver thiocyanate are not detected by the determination. These
compounds can be determined after pretreatment
with NANOCOLOR® NanOx Metal (REF 918 978).
The following ions will not interfere:
- < 1000 mg/L Pb2+, F–, NO3–, SO42–
- < 500 mg/L PO4 3–
- < 200 mg/L Mn2+, Ni2+
- < 100 mg/L Al3+, Cr(III)
- < 50 mg/L Cd2+
- < 20 mg/L Ca2+, Cu2+, Fe3+, Hg2+, Mg2+, Zn2+
- < 10 mg/L Cr(VI), Mo(VI)
The method can not be applied for the analysis of sea water.
Procedure
Requisite accessories: piston pipette with tips
Lower silver concentrations (0.08–0.50 mg/L Ag+) can be determined by using 50 mm semi-micro cuvettes (REF 919 50):
Pour the contents of test tubes into 50 mm semi-micro cuvettes and measure after 10 min [method 1491].
Measurement
For NANOCOLOR® photometers see manual, test 0-49.
Measurement when samples are colored or turbid:
For all NANOCOLOR® photometers see manual, use key for correction value.
Photometers of other manufacturers:
For other photometers check whether measurement of round glass tubes is
possible. Verify factor for each type of instrument by measuring standard
solutions.
MACHEREY-NAGEL GmbH & Co. KG · Neumann-Neander-Str. 6–8 · 52355 Düren ·
Germany
Tel.: +49 24 21 969-0 · Fax: +49 24 21 969-199 · info@mn-net.com · www.mn-net.com
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>