MACHEREY-NAGEL Visocolor ECO Chloride Colorimetric Test Kit Range Instructions

June 3, 2024
MACHEREY-NAGEL

MACHEREY-NAGEL Visocolor ECO Chloride Colorimetric Test Kit Range

Test kit for performing colorimetric tests on chloride ions in surface water and sewage

Method:
Chloride ions react with mercuric thiocyanate to produce undissociated mercuric chloride and to liberate thiocyanate ions. In the presence of ferric salts this thio- cyanate ions produce a characteristic orange color

Measurement range:

1–60 mg/L Cl–

Contents ot test kit (*refill pack)

sufficient for 90 tests

  • 2 x 20 mL Cl-1*
  • 24 mL Cl-2*
  • 2 screw-plug measuring glasses
  • 1 slide comparator
  • 1 color chart
  • 1 plastic syringe 5 mL
  • 1 instructions for use*

Hazard warning:
Information regarding safety can be found on the box’ label and in the safety data sheet. You can download the SDS from www.mn-net.com/SDS.

Instructions for use

also refer to the pictogram on the back of the color chart

  1. Pour a 5 mL water sample into each of the measuring glasses using the plastic syringe. Place one measuring glass on position A in the comparator. Only add the reagents to measuring glass B.
  2. Add 10 drops of Cl-1. Seal the glass and mix.
  3. Add 10 drops of Cl-2. Seal the glass and mix.
  4. Open the glass after 1 min and place it on position B in the comparator.
  5. Slide the comparator until the colors match in the inspection hole on top. Check the measurement reading in the recess on the comparator reed. Mid-values can be estimated.
  6. After use, rinse out both measuring glasses thoroughly (see „Disposal“) and seal them.

The method can not be applied for the analysis of sea water.

Measurement up to 300 mg/L chloride

  1. Pour a 1 mL water sample and 4 mL of distilled water into each of the mea- suring glasses.
  2. Same procedure as described above. Multiply the read-off value by 5 (see „Conversion table“).

Conversion table for measurements up to 300 mg/L chlorides:

  • Read-off value in mg/L Cl– Chloride concentration in
  • mg/L Cl– (read-off value x 5)
  • 1 5
  • 2 10
  • 4 20
  • 7 35
  • 12 60
  • 20 100
  • 40 200
  • 60 300

The reagents can be used for the photometric evaluation with photometer PF-12 / PF-12Plus.
Disposal:
Information regarding disposal can be found in the safety data sheet. You can download the SDS from www.mn-net.com/SDS.
Interferences:
Bromide, cyanide, iodide, sulfide, thiocyanate and thiosulfate all interfere since they react in the same way as chloride. The following ions will not interfere: ≤ 2000 mg/L NO2 ; ≤ 20 mg/L F–.
Note:
For the determination of chloride in concrete, please contact MACHEREY-NAGEL for special working instructions.
Storage:
Store the test kit in a cool (< 25 °C) and dry place.

MACHEREY-NAGEL GmbH & Co. KG · Neumann-Neander-Str. 6–8 · 52355 Düren · Germany
Tel.: +49 24 21 969-0 · Fax: +49 24 21 969-199 · info@mn-net.com · www.mn-net.com

References

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

MACHEREY-NAGEL User Manuals

Related Manuals