MACHEREY-NAGEL Test 0-74 NANOCOLOR Phenolic index 5 Tube Test Instructions
- June 3, 2024
- MACHEREY-NAGEL
Table of Contents
MACHEREY-NAGEL Test 0-74 NANOCOLOR Phenolic index 5 Tube Test
Method
Photometric determination of phenols and other compounds capable of oxidative coupling, which forms antipyrine dyes with 4-amino antipyrine, and, if wanted, subsequent extraction with isobutyl methyl ketone (MIBK).
Tube test | 10 mm cuvette after extraction | |
---|---|---|
Range: | 0.2–5.0 mg/L phenolic index | 0.2–5.0 mg/L phenolic |
index
Wavelength| |
(HW = 5–12 nm):| 520 nm| 445 nm 470 nm
Reaction time:| 5 min (300 s)| 15 min (900 s)
Reaction temperature:| 20–25 °C| 20–25 °C
Contents of reagent set
- 20 test tubes Phenolic index 5
- 2 test tubes each with 11 mL Phenolic index 5 R2
- 1 tube NANOFIX Phenolic index 5 R3
Note:
The content of the 20 test tubes with Phenolic index 5 may be colored light
yellow, which has no influence on the color reaction.
Hazard warning
Test tubes contain 4-amino antipyrine 25–100 %, reagent R2 contains ammonia 1–5 %, and reagent R3 contains potassium peroxodisulfate 20–100 %. H317 and H334 May cause an allergic skin reaction. May cause allergy or asthma symptoms or breathing difficulties if inhaled. P261, P272, P280, P302+352, P304+340, P333+313, P342+311, P363 Avoid breathing dust. Contaminated work clothing should not be allowed out of the workplace. Wear protective gloves/eye protection. IF ON SKIN: Wash with plenty of water / … IF INHALED: Remove person to fresh air and keep comfortable for breathing. If skin irritation or rash occurs: Get medical advice/attention. If experiencing respiratory symptoms: Call a POISON CENTER/doctor/ … Wash contaminated clothing before reuse. For further information ask for a safety data sheet.
Interferences
- oxidizing substances
- reducing substances
- cyanides
- applicable for turbid water after extraction
- applicable for seawater after extraction
To ensure the measuring results, extraction with isobutyl methyl ketone (MIBK, REF 918 929) is recommended, analogous to DIN 38 409 H16.
Procedure
Requisite accessories: piston pipette with tips
Open the test tube, add
- 1.0 mL R2 and
- 4.0 mL test sample (the pH value of the sample must be between pH 1 and 13), close, and mix. Add
1 NANOFIX R3, close and mix.
- (Close the NANOFIX tube immediately after use.) Clean the outside of the test tube and measure after 5 min.
Procedure with extraction
Requisite accessories: piston pipette with tips, glass cuvettes 10 mm, isobutyl methyl ketone R4 (MIBK, REF 918 929).
Test sample | Blank value |
---|
Open the test tube, add
1.0 mL R2 and
4.0 mL test sample (the pH value of the sample must be between pH 1 and 13), close, and mix. Add
1 NANOFIX R3, close and mix.
(Close the NANOFIX tube immediately after use.)
Wait 5 min. Add
4.0 mL R4 (MIBK), close and shake vigorously for 30 s.
| Open test tube, add
1.0 mL R2 and
4.0 mL distilled water, close and mix.
Add
1 NANOFIX R3, close and mix.
(Close the NANOFIX tube immediately after use.)
Wait 5 min. Add
4.0 mL R4 (MIBK), close and shake vigorously for 30 s.
After 15 min open test tubes with the test sample and blank value and pipet each about 2 mL of the upper organic phase into 10 mm cuvettes and measure [method 1742].
Measurement
For NANOCOLOR® photometers and PF-12 see manual, test 0-74.
Measurement when samples are colored or turbid:
For all NANOCOLOR® photometers see the manual and use the key for correction
value.
Photometers of other manufacturers:
For other photometers check whether the measurement of round glass tubes is
possible. Verify the factor for each type of instrument by measuring standard
solutions.
MACHEREY-NAGEL GmbH & Co. KG · Neumann-Neander-Str. 6–8 · 52355 Düren · Germany Tel.: +49 24 21 969-0 · Fax: +49 24 21 969-199 · info@mn- net.com. www.mn-net.com.
PD 14122 / A011703 / 985 074 / 0861.