MACHEREY-NAGEL Test 0-98 NANOCOLOR Aluminum 07 Test Tube User Manual

June 3, 2024
MACHEREY-NAGEL

REF 985 098
Test 0-98
08.16
Aluminum 07

Test 0-98 NANOCOLOR Aluminum 07 Test Tube

Method:
Photometric determination with eriochrome cyanine R

Contents of reagent set:
19 test tubes Aluminum 07 with NANOFIX
1 test tube with 11 mL Aluminum 07 R2
1 test tube with 11 mL Aluminum 07 R3
1 test tube with 11 mL Aluminum 07 R4
1 test tube with blank value ”NULL”

Hazard warning:

This test does not contain any harmful substances which must be specially labelled as hazardous.
Preliminary tests:
If the order of magnitude of the concentration in a sample is not known, a preliminary test with QUANTOFIX® Aluminum (5–500 mg/L Al3+, REF 913 07) rapidly gives this information. From the order of magnitude the required dilution can be calculated and prepared directly.
Interferences:
Fluoride interferes.
Turbid samples are to be filtered (membrane filter 0.45 μm, REF 916 50). The total Aluminum can be determined with NANOCOLOR® NanOx Metal (REF 918 978) and microwave decompo-sition.

The following ions will not interfere:
≤ 100 mg/L SiO2
≤ 10 mg/L Cu2+, Fe2/3+, Mn2+, Ni2+, Zn2+
≤ 5 mg/L Cr2+, Cd2+
≤ 1 mg/L Co2+
The method can be applied also for the analysis of sea water.

Procedure:

Requisite accessories: piston pipette with tips

Open test tube, add
500   μL (= 0.5 mL) R2,
500   μL (= 0.5 mL) R3 and
4.0   mL sample solution (the pH value of the sample must be between pH 3 and 6), close and mix.
Open test tube again, add
500   μL (= 0.5 mL) R4, close and mix.
Clean outside of test tube and measure after 5 min.
Adjust photometer to zero by using blank value “NULL”.

Measurement:

For NANOCOLOR® photometers and PF-12 see manual, test 0-98.
Measurement when samples are colored or turbid:
For all NANOCOLOR® photometers see manual, use key for correction value.
Photometers of other manufacturers:
For other photometers check whether measurement of round glass tubes is possible. Verify factor for each type of instrument by measuring standard solutions. The factor depends strongly on the wavelength.
Analytical quality control:
NANOCONTROL Multistandard Drinking water (REF 925 018)

MACHEREY-NAGEL GmbH & Co. KG · Neumann-Neander-Str. 6–8 · 52355 Düren · Germany
Tel.: +49 24 21 969-0 · Fax: +49 24 21 969-199 · info@mn-net.com · www.mn-net.com
PD 14122 / A025134 / 985 098 / 0861

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

MACHEREY-NAGEL User Manuals

Related Manuals