ROCKER Mars 320 Graphite Hotplate Instruction Manual
Important Notice
This instrument is designed for laboratory usage only. Please read this manual
carefully before
installing and operating. The instrument shall not be modified in any way. Any
modification
will void the warranty and may result in potential hazard. We are not
responsible for any injury
or damage caused by any non-intended purposes and modifying the instrument
without
authorization.
- Check the voltage specified on the name plate and ensure it matches the line voltage in your location.
- Install the instrument in a clean, dust-less and ventilated area under 40°C. Please keep the instrument away from flammable or explosive material.
- Mars hotplate should be operated in a functional fume hood, whereas the controller must be placed outside the fume hood.
- Mars hotplate is a high-temperature digestion system with exposed hot surface. Be careful around the instrument during or after operation and do NOT touch it to avoid serious burns.
- Please cool the instrument to room temperature before changing blocks to avoid burns.
- Hotplate and blocks are coated with PTFE, please do NOT contact the sharp objects to prevent from PTFE coating damage.
- Do NOT bundle the cable and power cord to prevent any danger from overcurrent.
- This product is not warranted and tested for using with perchloric acid. User should take full responsibility of use with perchloric acid.
- This instrument must be grounded to prevent the risk of electrical short circuit. If there’s any doubt about grounding installation, please consult a qualified electrician.
- Please discard packing material according to local related regulations.
- Operating condition
(a) Ambient temperature: 5~ 40°C
(b) Relative humidity: 80% RH Max.
(c) Power supply: 200-240V~, 50/60Hz
(d) Fuse of Mars 320: T6A, 250V~
Fuse of Mars 430: Tl 0A, 250V~
(e) Altitude: up to 2000 m
(f) Pollution degree: II
(g) Overvoltages category: CAT II
(h) Indoor use
Caution: Hot surface
The equipment must be disconnected from the mains supply before replacing the
fuse.
Unpacking
Please check if the package is complete without any damage before unpacking.
When unpacking, please make sure you have all accessories that indicated on
the list. If there is any
problem, please keep the serial number along with packing case and contact
your local distributor immediately for assistance.
Main Part Diagram
(1 ). Graphite Hotplate
(2). Controller
Display/ Button / Symbol Explication
(1 ). Operation Panel
(2). Button
(3). Symbols
(4). Display
Installation
(1 ). Instruction of Wall-Mounted Accessary
Loose the front foot pads of controller, pull out wall-mounted accessory then
screw the front
foot pads tightly. Hang Mars controller outside of fume hood to prevent
corrosion.
Controller can be placed upright as well as hung outside of the fume hood. 3M
17004 adhesive hook is recommended.
(2). Instruction of Heating Cable Connection
Curmec.t the hotplote arid c.ontruller tug ether by pl ug-i 11 heating ubl2.
De ~JCT buidle lre ca;ile. Plea•e aligr lhe” rO•N, er, bolh ·;ea lir;, cc1c)ie
anc sockeL anj imer. iL.
Operation
(1 ). Program Menu
(2). Program Operation
(a). Continuous Heating Mode (Timer sets as 00:00)
Note 2. (a).
(1) Controller will memorize the selected program and maintain when it
switches on next time.
(2) TE represents the reading of external temperature sensor and shows” –” if
no sensor is connected.
(3) How to stop or select other programs: Press / the screen shows”STOP
HEATING7″ message. Press O to stop heating and return to program list, press /
to continue heating.
(4) 00:00 stands for continuous heating. The counter will stop until 99:59
,whereas hotplate maintains heating continuously with heating indication, stop
heating by pressing / .
(b). Countdown Mode (Timer should not be set as 00:00)
Note 2. (b).
(1) Controller will memorize the selected program and maintain when it
switches on next time.
(2) TE represents the reading of external temperature sensor and shows” –” if
no sensor is connected.
(3) How to stop or select other programs: Press / the screen shows”STOP
HEATING1″message. Press O to stop heating and return to
program list, press / to continue heating.
(3). Parameter Setting
(4). Temperature Calibration
Insert electronic thermometer into hole for reference sensor. Calibrate
temperature as following procedure.
Please make sure thermometer and temp. setpoint is calibrated regularly or
certified by standard organization to ensure the accuracy is ±1 “C.
Note 4.
- Please set the calibrated setpoints before calibration, refer to (3) Parameter Setting.
- TE represents the reading of external temperature sensor and shows “—” if no sensor is connected. Press 0 to calibrate the reading of sensor when sensor is connected.
- If the actual temperature is higher than setting temperature, a cooling progress appears.
- To stop calibration, press / and screen shows “STOP HEATING7” message. Press O to stop heating and return to program list, press / to continue calibration.
- When connecting an external PTFE temperature sensor, press O to switch the calibration parameters of hotplate and external sensor. Press or to adjust the temperature while the figure is flashing, and hold O for 1 second to save the parameters and go for next step.
- Error message (“Cal-# Error”) appears if calibration is error. (eg. T2 temperature is lower than Tl). Press any button to reset to default calibration settings automatically.
- With an unsatisfied calibration result, you can re-calibrate or select the “Rst” program and press / to reset to default calibration settings.
Maintenance
- Please unplug the instrument and cool it down before cleaning.
- Please operate the instrument in a well-ventilated area and keep it clean. If there’s any solution drop or splash to hotplate, block, heating cable and controller, please clean the surface with damp cloth or mild soap after cooling to room temperature.
- If fuse blows, please troubleshoot and solve problems first. When replace fuse on controller, get the spare fuse from the fuse holder by a flathead screwdriver.
Troubleshooting
Ordering Information
178320-22 Mars 320, Graphite Hotplate AC220V, 50/60Hz
178430-22 Mars 430, Graphite Hotplate AC220V, 50/60Hz
178200-45 PTFE-Coated Temperature Sensor (1.SM calbe)
178200-36 Rack for 50 ml Digestion Tube (for 1 Block)
178200-36-1 Rack for 50 ml Digestion Tube (for 1 Block)
Digestion Tube
*Max.Operating Temperature PP: 130 C Glass, PTFE: 200 C
Graphite Block (L x W): 295 x 197 mm