ROCKER CD 200 Chemical Oxygen Demand Detector Instruction Manual

June 1, 2024
ROCKER

ROCKER CD 200 Chemical Oxygen Demand Detector Instruction Manual
ROCKER CD 200 Chemical Oxygen Demand Detector

Please read this instruction manual before using this product.

Important Notice

This instrument is designed for laboratory usage only. Please read this manual carefully before installing and operating. The instrument shall not be modified in any way. Any modification will void the warranty and may result in potential hazard. We are not responsible for any injury or damage caused by any non-intended purposes and modifying the instrument without authorization.

  1. Check the voltage specified on the name plate and ensure it matches the line voltage in your location.

  2. Install the instrument in a clean, dust-less and well-ventilated area under 40°C.

  3. The instrument can be powered either by a power adaptor or a DC9V battery. For safety reasons, it’s recommended to use the power adaptor supplied by the manufacturer.
    If using a DC9V battery, please remember to remove it when instrument will be unused for an extended period to prevent leakage.

  4. Review Safety Data Sheets (SDS) before handling reagents, as they may be hazardous.

  5. Please discard packing material according to local related regulations.

  6. For more product details, please refer to the latest catalog and website.

  7. Operating condition

(a) Ambient temperature: 5~ 40 °C
(b) Relative humidity: 80% RH Max.
(c) Power supply: 100~120V, 50/60Hz or 200~240V, 50/60Hz
(d) Altitude: up to 2000 m
(e) Pollution degree: II
(f) Indoor use

* Before starting COD measurement, allow the vials to cool down to room temperature and ensure the reagent is uniformly mixed and completely precipitate.

Unpacking

Please check if the package is complete without any damage before unpacking. When unpacking, please make sure you have all accessories that indicated on the list. If there is any problem, please keep the serial number along with packing case and contact your local distributor immediately for assistance.

Unpacking

Item Content
1 CD 200 Detector
2 Power Adaptor
3 Instruction Manual

Main Part Diagram

Main Part Diagram

Item Designation
1 Vial Dust Cap
2 Vial Holder
3 Display
4 Buttons
5 DC Socket
6 Battery Cover
7 Power Adaptor
8 DC9V Battery (Optional)

Button / Symbol Explication

Buttons

Button Designation Description
Data Data Button Display information and data on the screen.
Menu Menu Button Return to the previous menu.
Up Button Scroll upward through menus or stored data.
Down Button Scroll downward through menus or stored data
Right Button Scroll to right through menus or stored data
Enter Button (1) Select the current menu option. (2) Confirm and accept the

input value.
| Power Button| Turns the power on or off

Symbols

Symbol Description
Indication of power adaptor connection.
100% full power.
66% power
33% power
5% low battery. Please replate the battery promptly. (Backlight and sound

disabled)
| Selected option
| Selected option

Operation

Installation

  • Power Supply
    The instrument can be powered either by a power adaptor or a DC9V battery.

    • Power Adaptor:
      Ensure the power supply voltage matches the voltage rating for CD 200 (100-240V~), with an output voltage of DC12V.

    • DC9V Battery:
      Users can install a DC9V battery by removing / installing the battery cover at the back of CD200.

  • Please remove the battery if CD 200 will be unused for an extended period to prevent leakage.

  • Power On / Off

    • Power On: Press the power button , and you can start using CD 200.
    • Power Off: Press the power button , and select the “Yes” option to shut it down

Menu

Menu Item Description Remark
Main Menu 1: COD TEST Select the Curve Brand.
2: Time Date and time setting.
3: Backlight Backlight setting.
4: Sound Sound setting.
5: Auto Shut Auto power setting.
6: Memory Memory setting.
Curve Menu 1: HACH HACH COD standard curve.
2: Lovibond Lovibond COD standard curve.
3: CHEMetrics CHEMetrics COD standard curve
4: HANNA HANNA COD standard curve.
5: Merck Merck COD standard curve
6: User-1 User Curve Self-Calibration.
7: User-2 User Curve Self-Calibration.
8: User-3 User Curve Self-Calibration.
9: User-4 User Curve Self-Calibration.
Curve Range 150 Measuring range 10 ~ 150 mg/L
1500 Measuring range 100 ~ 1500 mg/L
15000 Measuring range 1000 ~ 15000 mg/L
Operating Menu Info (1) Standard Curve: display curve measuring range. (2)
User Curve: display calibration data
CAL Calibration Parameters

Curve Operation

Built-In Standard Curves – Hach, Lovibond, CHEMetrics, HANNA, Merck

  • Built-in standard curves are preloaded with saved calibration values for direct COD testing.
  • Insert sample vials in the same direction to keep COD values consistent.
Step Operation Screen Display
1 Switch on the instrument and LOGO shows on the screen. Switch on the
instrument and LOGO shows on the screen. ![ROCKER Logo](https://manuals.plus

/wp-content/uploads/2024/05/Screenshot_1-82.png)
2| The curve menu shows on the screen in 2 seconds.|

  1. HACH<<
  2. Lovibond

3| Select desired curve based on the used reagent brand by using 「▲」 or 「▼」. Press 8 button to enter the curve range|

  1. HACH<<
  2. Lovibond

4| Press 「 ▶」and select curve range based on reagent concentration: 150, 1500, or 15000. Press 8 button to enter the operating menu.| HACH-1500 150 1500< 15000
5| While Start is selected, press 「 ▶」 to start a COD test..| HACH-1500 Start< Info
6| Insert Blank vial into the vial holder, and “Blank OK” message appears in 2 seconds.| HACH-1500 Insert Blank
7| Remove the Blank vial and insert sample vial when “Insert Sample” message appears| HACH-1500 Insert Sample
8| The COD value will appear in 2 seconds and the data will be saved automatically.| HACH-1500 1001 mg/L
9| Remove the sample vial from the vial holder and insert another vial to test COD value.| HACH-1500 487 ﹒ 1 mg/L

  • “<<” and “<” symbols indicate the selected option.
  • Press the Data button to display stored data on the screen.
  • The Info on the operating menu displays the range of reading value.
  • “Over Range” or “Under Range” errors occur when the testing concentration is outside the range of the current COD curve.

User Curves Calibration – User-1, User-2, User-3, User-4

  • Ensure to calibrate using standard solution and reagent vials during initial use.

  • Set the concentration values equally to prevent errors from close absorbance readings.

  • Complete the Blank calibration and use more than 1 standard to finish the calibration.
    You can save up to 6 standard setpoints in one curve.

  • Avoid calibrating the same concentration twice to prevent errors in calibration parameters.

  • Insert sample vials in the same direction to keep COD values consistent.

Step Operation Screen Display
1 Switch on the instrument and LOGO shows on the screen. ![ROCKER

Logo](https://manuals.plus/wp-content/uploads/2024/05/Screenshot_1-82.png)
2| The curve menu shows on the screen in 2 seconds.|

  1. HACH<<
  2. Lovibond

3| Select user curve by using 「▲」 or 「▼」. Press button to enter the curve range.|

  1. Merck
  2. User-1<<

4| Press 「 ▶」and select curve range based on reagent concentration: 150, 1500, or 15000. Press button to enter the operating menu.| User-1-150 150< 1500 15000
5| Press 「 ▶」and select CAL. Press button to proceed calibration.| User-1-150 Start Info CAL<
6| Insert Blank vial when “Insert Blank” message appears.| User-1-150 Insert Blank
7| Insert Blank vial into the vial holder, and “Blank OK” message appears in 2 seconds.| User-1-150 Blank OK
8| Remove the Blank vial and press button to confirm.| User-1-150 Blank OK
9| Insert 1st standard vial into vial holder when “Insert Std 1” message appears.| User-1-150 Insert Std 1
10| Input the standard concentration value by using「▲」, 「▼」 and 「 ▶」.| User-1-150 00050 mg/L
11| Remove the sample vial from the vial holder and press button to save the 1st setpoint.| User-1-150 S1:50.0 mg/LOK<
12| Repeat step 9~11 until standard vials are finished. Press Menu and select Save to complete the calibration. Save: Save the calibration parameters and exit. Exit: Exit without saving CAL parameters. CAL: Back to the calibration.| User-1-150 Save< Exit CAL

  • After completing the calibration, follow steps in P.10 3.3 Calibration Verification to verify its success.

Calibration Verification – User-1, User-2, User-3, User-4

  • After completing the calibration, follow these steps to verify its success.
  • If the calibration fails, please refer to P.8 3.2 User Curves Calibration for instructions on redoing the calibration.
Step Operation Screen Display
1 Switch on the instrument and LOGO shows on the screen ![ROCKER

Logo](https://manuals.plus/wp-content/uploads/2024/05/Screenshot_1-82.png)
2| The curve menu shows on the screen in 2 seconds.| 1: HACH<< 2: Lovibond
3| Select user curve by using 「▲」 or 「▼」. Press button to enter the curve range.| 5: Merck 6: User-1<<
4| Press 「 ▶」and select curve range based on reagent concentration: 150, 1500, or 15000. Press button to enter the operating menu.| User-1-150 150< 1500 15000
5| Press 「 ▶」and select Info. Press 8 button to verify the calibration result| User-1-150 Start Info< CAL
6| CAL point: Number of calibration setpoints. OK: Calibration parameters are correct. NG: Calibration parameters are incorrect, please redo the calibration.| User-1-150 CAL point=2OK<
7| Check the concentration of setpoints using 「▲」 or 「▼」. Press  or Menu button to return to the previous menu.| User-1-150 S1=1500.0mg/L<

Change Name of User Curve – User-1, User-2, User-3, User-4

Step Operation Screen Display
1 Switch on the instrument and LOGO shows on the screen. ![ROCKER

Logo](https://manuals.plus/wp-content/uploads/2024/05/Screenshot_1-82.png)
2| The curve menu shows on the screen in 2 seconds.| 1: HACH<< 2: Lovibond
3| Select user curve by using 「▲」 or 「▼」. Press button to enter the curve range.| 5: Merck 6: User-1<<
4| Press and hold 「 ▲」for 2 seconds during the curve range menu to change the curve name.| User-1-150 150< 1500 15000
5| Use「▲」, 「▼」 and 「 ▶」to change the characters (up to 10 characters).| User-1 =>User-1
6| Press button to save the name. Press Menu button to exit without saving.| User-1-150 =>ROCKER

Check Testing Data

CD 200 can store up to 100 records and automatically delete the oldest record when the limit is exceeded.

Step Operation Screen Display
1 Press the Data button to display stored information under every menu page.

1: HACH<< 2: Lovibond
2| The displayed data includes the date, time, curve name, curve range, data number and COD value.| 2 0 1 6 / 0 2 / 0 6 0 8 : 3 4 022 487 1 mg/L
3| The data, time, curve name and curve range will alternate in the first row every 2 seconds. Press「 ▶」to stop the automatic switching and switch it manually.| HACH-1500 022 4871 mg/L
4| Use 「▲」 or 「▼」 to display different data. Press Menu button to exit.| 2016/02/0608:33 021 999 ﹒ 9 mg/L

Setup of Basic Settings – Time, Backlight, Sound, Auto Shut, Memory

Step Operation Screen Display
1 Switch on the instrument and press the Menu button to return to Main Menu.

1: HACH<< 2: Lovibond
2| 1: CODTest 2: Time<<| 1: CODTest 2: Time<<
3| Use「▲」, 「▼」 and 「 ▶」to change the settings. Press 8 button to save and exit.| Data:2016/02/06 Time:09:34

List of Settings and Their Meanings

Settings Options Description
Time N/A The time setting is automatically saved upon modification.
Backlight On Backlight remains continuously on
Auto Backlight turns on but automatically turns off after 3-min idle time.
Off Backlight remains off.
Sound On Sound is enabled
Off Sound is disabled
Auto Shut On The instrument will automatically switch off after 10-min idle

time
Off| Auto Shut feature is disabled
Memory| On| After switching on, the last menu displayed before shutdown.
Off| After switching on, the curve menu will be displayed.

Maintenance

  1. Please clean the surface with a damp cloth, mild soap, or 75% ethanol and dry it.
  2. Use the dust cap to prevent from dust accumulation for reliable performance.
  3. Remove the battery if the instrument will be unused for an extended period to prevent leakage.
  4. Clean the vial holder with ethanol to avoid reading interference due to dirt accumulation.

Troubleshooting

Problem Reason and Solution
Fail to start or Abnormal Display
  • Wrong power supply Connect to power specified on mani unit.
  • Loose plug Reconnect plug to power supply and instrument.
  • Battery error Replace with a new 9V battery.
  • Display or components failure Contact distributor for assistance

Dark Display|

  • Incorrect backlight setting Turn on the backlight.
  • Low battery Replace with a new 9V battery.
  • Display or components failure Contact distributor for assistance.

Error Messages|

  • Incorrect range of reagent Ensure range of curve and reagent is corresponding.
  • Over / under range Testing value exceeds than used curve range.
  • Blank NG Used wrong range of reagent vials.

Other problems|

  • Please contact distributor for assistance

Ordering Information

198200-01(02) CD 200, COD Detector, AC100~240V adaptor, US plug (EU plug)
179250-11(22) CR 25, COD Reactor, AC110V,60Hz (AC220V,50Hz)
179250-43 Protective Lid
191100-01(02) WD 100, Multiparameter Colorimeter, AC100~240V adaptor, US plug (EU plug)
AWE770110 1000~5000 ul Pipette (Adjustable)
SSI 5000-50 5 ml Pipette Tip (50/PK)