RENISHAW RFP1 Fringe Probe Installation Guide

June 3, 2024
RENISHAW

RENISHAW - logo

RFP1 Fringe Probe
Installation Guide

RFP1 Fringe Probe

RENISHAW RFP1 Fringe Probe Highly accurate structured light measurement probe for REVO-2
The REVO® fringe probe (RFP1) is a non-contact structured light sensor for the REVO system offering highly accurate measurement of surface topology on an infinite positioning, 5-axis CMM platform.
RFP1 increases the multi-sensor capability of the REVO system, which now offers interchangeable tactile scanning, touch-trigger, surface finish, non- contact structured light and vision probes.
The RFP1 probe is designed for inspection of free-form surfaces and complex geometry, rapidly delivering patches of surface data with a high point density. The probe projects a fringe pattern onto the part surface, capturing variations in the fringe pattern with the camera to build a point cloud of 3D surface data.
Automatic exposure controls compensate for different surface materials, colours and textures, ensuring optimal data results without the need for matt coatings.
Unlike other non-contact structured light systems, the RFP1 fringe probe does not require reference markers to stitch together data from different patches – this is done automatically by the REVO system. Two new software tools provide easy-to-use inspection planning and digitising. The RFP inspection planner is a tool for path planning and generating DMIS part programs from CAD, whilst the RFP digitiser collects data from parts without CAD models as part of a reverse engineering process.

Key benefits

Multi-sensor system with unrivalled feature access
RFP1 data is automatically stitched together in one co-ordinate frame with no need for reference markers, and combined with data from all other REVO-2 sensors maximises part inspection capability on one CMM.
Automatic exposure compensation
Ensures optimal data results with automatic compensation for different materials, surface colours and textures without the need for matt coatings.
Easy-to-use inspection planning and digitising software
RFP inspection planner generates DMIS part programs from CAD quickly and easily, whilst the RFP digitiser collects data from parts without CAD models as part of a reverse engineering process.

Specification

Field of view 50 mm x 50 mm
Depth of field 20 mm
Stand-off 160 mm
Rotational capability A-axis (from REVO-2)
B-axis (from REVO-2) Infinite positioning
XY point density (resolution) 65 pm
Probe repeatability (plane) 4 pm (2a)
Probe accuracy (position error of a plane) 9 pm
Measurement form error (plane) 15 pm RMS
Equivalent scanning speed 50,000 pts/s
Mounting Magnetised coupling
Sensor type 1/1.8″ CMOS sensor
Projector type Fixed sinusoidal filtering grating
Illumination A wavelength of 450 nm blue LED
Operating modes 3D surface point capture 2D vision measurement
Probe head REVO-2 only
Rack port compatibility VPCP
Change rack MRS2 recommended
Software compatibility UCCsuite 5.6 onwards

MODUS 1.11 onwards
Integration| I++ DME protocol
Weight| 530 g
Operating temperature range| +10 °C to +40 °C
Metrology specification range| 20 °C ± 2 °C
Storage temperature range| -25 °C to +70 °C
Operating humidity| 0% to 80% (non-condensing)
Calibration artefact| FA10
Output| Point cloud data file (.xyz)

© 2019–2022 Renishaw plc. All rights reserved. RENISHAW® and the probe symbol are registered trade marks of Renishaw plc.
Renishaw product names, designations and the mark ‘apply innovation’ are trade marks of Renishaw plc or its subsidiaries.
Other brand, product or company names are trade marks of their respective owners.
Renishaw plc. Registered in England and Wales. Company no: 1106260. Registered office: New Mills, Wotton-under-Edge, Glos, GL12 8JR, UK.
WHILE CONSIDERABLE EFFORT WAS MADE TO VERIFY THE ACCURACY OF THIS DOCUMENT AT PUBLICATION, ALL WARRANTIES, CONDITIONS, REPRESENTATIONS AND LIABILITY, HOWSOEVER ARISING, ARE EXCLUDED TO THE EXTENT PERMITTED BY LAW. RENISHAW -
logo

+44 (0) 1453 524524
[
uk@renishaw.com](mailto:uk@renishaw.com)
Part no.: H-1000-2272-02-A
Issued: 09.2022
**** #renishaw
www.renishaw.com/RFP1

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals