TSI 800320 Fume Hood Sidewall Velocity Sensor Installation Guide
- September 25, 2024
- tsi
Table of Contents
800320 Fume Hood Sidewall Velocity Sensor
“`html
Specifications:
- Product Name: Fume Hood Sidewall Velocity Sensor
- Part Number: 800320
Product Usage Instructions:
1. Determine Vent Kit Requirement:
Check if a vent kit is required for installation as shown in
Figure 1.
2. Sidewall Velocity Sensor Location and Mounting:
-
Install the Mounting Plates to the fume hood inner wall lining
using two of the supplied screws (Figure 3). -
The Mounting Plates must be installed exactly as shown in
Figure 3. -
Plug the Sensor Cable into the Sidewall Velocity Sensor.
-
Thread the unconnected end of the Sensor Cable through the
mounting hole location towards the Controller or Monitor. Leave at
least 8 inches of slack at the Controller or Monitor end for
connection and mounting adjustment. Do not cut the sensor cable
length. -
Mount the Sidewall Velocity Sensor to the Mounting Plate using
the remaining two supplied screws (Figure 4). -
The mounted Sidewall Velocity Sensor must be installed exactly
as shown in Figure 4. -
Plug the Sensor Cable into the Controller or Monitor when
wiring.
FAQ:
Q: Can I cut the sensor cable length during installation?
A: No, do not cut the sensor cable length as it may affect the
performance and functionality of the Sidewall Velocity Sensor.
“`
Fume Hood Sidewall Velocity Sensor
P/N 800320
Installation Instructions
NOTICE This document serves as a guide for the installation of the 800320 Fume
Hood Sidewall Velocity Sensor. Please read these instructions thoroughly
before beginning installation.
For wiring or other Fume Hood Controller or Monitor setup information, refer
to the installation instructions included with that equipment.
Included Parts
Description
Qty
Sidewall Velocity Sensor Assembly
1
Sidewall Velocity Sensor Cable 6 ft
1
Mounting Plates
2
Mounting Screws (8-32 x 5/8″)
4
Installation Instructions
1
www.tsi.com
Sidewall Velocity Sensor Installation
Determine if Vent Kit is Required
The 800320 Sidewall Velocity Sensor must be referenced to room air in order to
function properly. If the fume hood panels extend to the ceiling, a TSI®
Sensor Vent Kit must be installed (Figure 1). Contact TSI® for Sensor Vent Kit
information. If a vent kit is used follow the installation instructions
included with that equipment.
Figure 1: Check for Vent Kit Requirement
Page 2 of 4
Sidewall Velocity Sensor Location and Mounting
NOTICE The Sidewall Velocity Sensor and the Controller or Monitor must be
mounted on the same side of the fume hood. The Sidewall Velocity Sensor is
mounted to the inner wall lining of the fume hood. 1. Determine the Sidewall
Velocity Sensor Location (Figure 2). 2. Drill a 2¼” hole through the fume hood
inner wall lining at this location (Figure 2).
Figure 2: Sidewall Velocity Sensor Location 3. Install the Mounting Plates to
the fume hood inner wall lining using two of the supplied screws (Figure 3).
4. The Mounting Plates must be installed exactly as shown (Figure 3).
Figure 3: Install Mounting Plates Page 3 of 4
5. Plug the Sensor Cable into the Sidewall Velocity Sensor. 6. Thread the
unconnected end of the Sensor Cable through the mounting hole location and
over to the
Controller or Monitor. Leave at least 8″ of slack at the Controller or Monitor
end to allow for connection to the Controller or Monitor and mounting
adjustment. DO NOT CUT SENSOR CABLE LENGTH. 7. Mount the Sidewall Velocity
Sensor to the Mounting Plate using the remaining two supplied screws (Figure
4). 8. The mounted Sidewall Velocity Sensor must be installed exactly as shown
(Figure 4). 9. Plug the Sensor Cable into the Controller or Monitor when
wiring the Controller or Monitor.
Figure 4: Final Sidewall Velocity Sensor Mounting
TSI and TSI logo are registered trademarks of TSI Incorporated in the United States and may be protected under other country’s trademark registrations.
TSI Incorporated Visit our website www.tsi.com for more information.
USA UK France Germany
Tel: +1 800 680 1220 Tel: +44 149 4 459200 Tel: +33 1 41 19 21 99 Tel: +49 241 523030
India
Tel: +91 80 67877200
China
Tel: +86 10 8219 7688
Singapore Tel: +65 6595 6388
P/N 6004050 Rev D
©2024 TSI Incorporated
Printed in U.S.A.