TSI 800320 Fume Hood Sidewall Velocity Sensor Installation Guide

September 25, 2024
tsi

800320 Fume Hood Sidewall Velocity Sensor

“`html

Specifications:

  • Product Name: Fume Hood Sidewall Velocity Sensor
  • Part Number: 800320

Product Usage Instructions:

1. Determine Vent Kit Requirement:

Check if a vent kit is required for installation as shown in
Figure 1.

2. Sidewall Velocity Sensor Location and Mounting:

  1. Install the Mounting Plates to the fume hood inner wall lining
    using two of the supplied screws (Figure 3).

  2. The Mounting Plates must be installed exactly as shown in
    Figure 3.

  3. Plug the Sensor Cable into the Sidewall Velocity Sensor.

  4. Thread the unconnected end of the Sensor Cable through the
    mounting hole location towards the Controller or Monitor. Leave at
    least 8 inches of slack at the Controller or Monitor end for
    connection and mounting adjustment. Do not cut the sensor cable
    length.

  5. Mount the Sidewall Velocity Sensor to the Mounting Plate using
    the remaining two supplied screws (Figure 4).

  6. The mounted Sidewall Velocity Sensor must be installed exactly
    as shown in Figure 4.

  7. Plug the Sensor Cable into the Controller or Monitor when
    wiring.

FAQ:

Q: Can I cut the sensor cable length during installation?

A: No, do not cut the sensor cable length as it may affect the
performance and functionality of the Sidewall Velocity Sensor.

“`

Fume Hood Sidewall Velocity Sensor
P/N 800320
Installation Instructions

NOTICE This document serves as a guide for the installation of the 800320 Fume Hood Sidewall Velocity Sensor. Please read these instructions thoroughly before beginning installation.
For wiring or other Fume Hood Controller or Monitor setup information, refer to the installation instructions included with that equipment.

Included Parts

Description

Qty

Sidewall Velocity Sensor Assembly

1

Sidewall Velocity Sensor Cable ­ 6 ft

1

Mounting Plates

2

Mounting Screws (8-32 x 5/8″)

4

Installation Instructions

1

www.tsi.com

Sidewall Velocity Sensor Installation
Determine if Vent Kit is Required
The 800320 Sidewall Velocity Sensor must be referenced to room air in order to function properly. If the fume hood panels extend to the ceiling, a TSI® Sensor Vent Kit must be installed (Figure 1). Contact TSI® for Sensor Vent Kit information. If a vent kit is used follow the installation instructions included with that equipment.
Figure 1: Check for Vent Kit Requirement
Page 2 of 4

Sidewall Velocity Sensor Location and Mounting
NOTICE The Sidewall Velocity Sensor and the Controller or Monitor must be mounted on the same side of the fume hood. The Sidewall Velocity Sensor is mounted to the inner wall lining of the fume hood. 1. Determine the Sidewall Velocity Sensor Location (Figure 2). 2. Drill a 2¼” hole through the fume hood inner wall lining at this location (Figure 2).
Figure 2: Sidewall Velocity Sensor Location 3. Install the Mounting Plates to the fume hood inner wall lining using two of the supplied screws (Figure 3). 4. The Mounting Plates must be installed exactly as shown (Figure 3).
Figure 3: Install Mounting Plates Page 3 of 4

5. Plug the Sensor Cable into the Sidewall Velocity Sensor. 6. Thread the unconnected end of the Sensor Cable through the mounting hole location and over to the
Controller or Monitor. Leave at least 8″ of slack at the Controller or Monitor end to allow for connection to the Controller or Monitor and mounting adjustment. DO NOT CUT SENSOR CABLE LENGTH. 7. Mount the Sidewall Velocity Sensor to the Mounting Plate using the remaining two supplied screws (Figure 4). 8. The mounted Sidewall Velocity Sensor must be installed exactly as shown (Figure 4). 9. Plug the Sensor Cable into the Controller or Monitor when wiring the Controller or Monitor.
Figure 4: Final Sidewall Velocity Sensor Mounting


TSI and TSI logo are registered trademarks of TSI Incorporated in the United States and may be protected under other country’s trademark registrations.

TSI Incorporated ­ Visit our website www.tsi.com for more information.

USA UK France Germany

Tel: +1 800 680 1220 Tel: +44 149 4 459200 Tel: +33 1 41 19 21 99 Tel: +49 241 523030

India

Tel: +91 80 67877200

China

Tel: +86 10 8219 7688

Singapore Tel: +65 6595 6388

P/N 6004050 Rev D

©2024 TSI Incorporated

Printed in U.S.A.