ERMENRICH TC40 Digital Multimeter User Manual
- September 13, 2024
- ERMENRICH
Table of Contents
ERMENRICH TC40 Digital Multimeter
PRODUCT OVERVIEW
-
Power button
-
Charge jack
-
NCV sensor
-
Warning indicator
-
LCD display
-
SEL (Selection)/
Flashlight button -
LED indicator
-
A jack
-
mA jack
-
COM jack
-
INPUT jack
-
Flashlight
-
RANGE (Range) button
-
HOLD/REL
(Data hold/
Relative value)
button -
FUNC
(Functions) button -
MAX/MIN (Max.min. values)button
Ermenrich Zing TC40 Digital Multimeter
Please carefully read the safety instructions and the user manual before using this product. Keep away from children. Use the device only as specified in the user manual.
The kit includes: digital multimeter, test leads (red and black), charger, carry bag, user manual, and warranty.
Charging mode
Whenever the charger is connected, the multimeter enters the charging mode.
The charge symbol is displayed on screen and an analog bar with the battery
charge level in percent appears. It takes about 3 hours to be fully charged.
The warning symbol is displayed.
Do not take measurements while charging.
Getting started
ATTENTION! This device must only be operated by a qualified specialist.
(A qualified specialist is a person whose competence is officially confirmed
by relevant documents, who has the necessary knowledge and experience to work
safely with this device and is authorized to work with circuits with the
appropriate voltage class.)
Press and hold the Power button (1) for 2 seconds to turn the device on/off. This multimeter has two modes – Smart (Auto) and Professional (Manual). When turned on, it displays Auto (Automatic mode) and enters the Smart (Auto) measurement mode.
Range selection
Press the RANGE button (13) to enter the manual range selection mode, and then
press again to switch the range. Only voltage, resistance, and current (in mA)
measurements are available in this mode. Press and hold the RANGE button (13)
for 2 seconds to return to the automatic range selection mode.
This function can’t be used in the Smart measurement mode.
Function selection
Press the FUNC button (15) to activate the manual mode. Then press the
SEL/Flashlight button (6) to select the required function.
To return to Smart (Auto) mode, press and hold the FUNC button (15) for 2
seconds.
Data hold
During a measurement, press the HOLD/REL button (14) to turn on data hold
function and display the current reading. Press again to turn the function off
and return to measurement display.
This function can’t be used in NCV/Live mode.
Flashlight
Press and hold the SEL/Flashlight button (6) for 2 seconds to turn the
flashlight on/off.
MAX/MIN mode
Press the MAX/MIN button (16) to switch between maximum and minimum values.
Press and hold the FUNC button (15) for 2 seconds to exit this function.
- This function can’t be used in capacitance, frequency/duty cycle, temperature, NCV/Live modes.
- During measurement of maximum/minimum values, the multimeter will automatically enter the manual range mode.
Relative value measurement
Press the HOLD/REL button (14) for at least 2 seconds to turn relative value
measurement on or off.
This function can’t be used in capacitance, frequency/duty cycle, temperature,
NCV/Live modes.
During measurement of relative values, the multimeter will automatically enter
the manual range mode.
Relative value measurement
Press the HOLD/REL button (14) for at least 2 seconds to turn relative value
measurement on or off.
- This function can’t be used in capacitance, frequency/duty cycle, temperature, NCV/Live modes.
- During measurement of relative values, the multimeter will automatically enter the manual range mode.
Fuse Broken indication
When the fuse is blown, the symbol tis displayed.
When the current measurement function is selected, the symbol FUSt will be
displayed at the same time and no further measurements will be taken. Please
replace the fuse in time.
Input jack indicator
When the function is changed, the corresponding input jack light will flash 5
times to indicate that the test lead should be inserted into the corresponding
jack.
Automatic recognition of current measurement
When the test lead is inserted into the A jack, the multimeter automatically
switches to the automatic recognition of current measurement function ( );
switching functions will not be allowed at this time.
When the multimeter is switched manually to the current measurement function
and no lead is inserted in the A jack, the display shows LEAd (Lead needs to
be inserted) and no further measurement is taken.
Auto power turn-off
When you press the Power button ( ) to power on, the automatic turn-off
function is activated by default and the display shows the symbol. After 15
minutes without any key operation, the multimeter will turn off automatically
to save battery energy.
To cancel the automatic turn-off function, press and hold the SEL/Flashlight
button ( ). The symbol is not displayed when the automatic turn-off function
is cancelled.
Display information
Smart (Auto) measurement
The multimeter is turned on in Smart (Auto) measurement mode by default. In
this mode, DC voltage, AC voltage, resistance, and continuity can be measured,
and the multimeter can automatically identify the measurement signal.
Plug the black test lead into the COM jack (10) and the red test lead into the
INPUT jack (11). Connect the probes to the measuring points or in parallel to
the circuit to be measured. The multimeter will automatically recognize the
measured signal. The measurement results will appear on the display (5).
- Do not measure the voltage higher than 1000V DC or 750V AC; otherwise, the device may be damaged.
- Pay special attention to safety when measuring high voltage to avoid electric shock or personal injury.
- Before use, test the known voltage with the multimeter to confirm that the multimeter is in good condition.
Professional measurement
AC/DC voltage measurement (in V)
Press the FUNC button (15) to select the function. Press the SEL/Flashlight
button (6) to select the AC voltage or DC voltage. The or DC symbol is
displayed accordingly. Plug the black test lead into the COM jack (10) and the
red test lead into the INPUT jack (11). Connect the probes to the measuring
points. The measurement results will appear on the display (5).
Resistance measurement
Press the FUNC button (15) to select the function. Plug the black test lead
into the COM jack (10) and the red test lead into the INPUT jack (11). Connect
the probes to the resistance measuring points of circuit or resistor. The
measurement results will appear on the display (5).
Continuity test
Press the FUNC button (15) to select the function. Plug the black test lead
into the COM jack (10) and the red test lead into the INPUT jack (11). Connect
the probes to the circuit or component to be tested. The measurement results
will appear on the display (5). If the resistance is <50Ω, an acoustic signal
will be emitted continuously. The measurement results will appear on the
display (5).
AC/DC voltage measurement (in mV)
Press the FUNC button (15) to select the function. Press the SEL/Flashlight
button (6) to select AC voltage or DC voltage. The or symbol is displayed
accordingly. Plug the black test lead into the COM jack (10) and the red test
lead into the INPUT jack (11). Connect the probes (in parallel) to the
measuring points. The measurement results will appear on the display (5).
Frequency and duty cycle measurement
Press the FUNC button (15) to select the function. Plug the black test lead
into the COM jack (10) and the red test lead into the INPUT jack (11). Connect
the probes to the measuring points or in parallel to the circuit to be
measured. The measurement results will appear on the display (5).
Capacitance measurement
Press FUNC button (15) to select the function. Plug the black test lead into
the COM jack (10) and the red test lead into the INPUT jack (11). Connect the
probes (in parallel) to the capacitance to be measured. The measurement
results will appear on the display (5).
Diode test
Press the FUNC button (15) to select the function. Plug the black test lead
into the COM jack (10) and the red test lead into the INPUT jack (11). Connect
the black test probe to the cathode side and the red test probe to the anode
side of the diode to be measured. If the polarity of the test probes is
reverse to the diode polarity, the OL (Overload) will appear on the screen.
This can be used for distinguishing the anode and cathode side of a diode. The
measurement results will appear on the display (5).
Temperature measurement
Press the FUNC button (15) to select the function. Plug the negative test lead
of K-type thermocouple into the COM jack (10) and the positive test lead into
the INPUT jack (11). Connect the thermocouple probe to the object to be
measured. The measurement results will appear on the display (5).
When using the included K-type thermocouple, temperature measurement range is –40… +260°C / –40… +500°F.
Non-contact voltage detection (NCV)
Press the FUNC button (15) to select the NCV/Live function. NCV (Non-contact
voltage detection) is displayed. Slowly bring the NCV sensor (3) close to the
point to be detected. When the signal of a weak electromagnetic field is
detected, the “– – – L” will appear on the screen, the built-in buzzer will
produce a slow beep sound, and the LED indicator (4) will glow green. When the
signal of a strong electromagnetic field is detected, the “– – – H” will
appear on the screen, the built-in buzzer will produce a quick beep sound, and
the LED indicator (4) will glow red.
When using this function, remove the test leads from the jacks.
Live detection
Press the FUNC button (15) to select the NCV/Live function. LIVE (Live
detection) is displayed. Plug only the red test lead into the INPUT jack (11).
Connect the red test probe to the conductor to be measured. When the signal of
a weak electromagnetic field is detected,the “– – – L” will appear on the
screen, the built-in buzzer will produce a slow beep sound, and the LED
indicator (4) will glow green. When the signal of a strong electromagnetic
field is detected, the “– – – H” will appear on the screen, the built-in
buzzer will produce a quick beep sound, and the LED indicator (4) will glow
red.
AC/DC current measurement (in mA)
Press the FUNC button (15) to select the A function, or insert the red
test lead into the mA jack (9) to automatically select A the function.
Press the SEL/Flashlight button (6) to select the AC current or DC current.
The symbol or symbol is displayed accordingly. Plug the black test lead into
the COM jack (10) and the red test lead into the mA jack (9). Connect the test
leads in series to the measuring points.
The measurement results will appear on the display (5).
AC/DC current measurement (in A)
Press the FUNC button (15) to select the function, or insert the red probe
into the 10A jack (8) to automatically select the function.
Press the SEL/Flashlight button (6) to select the AC current or DC current.
The symbol or symbol is displayed accordingly. Plug the black test lead into
the COM jack (10) and the red test lead into the 10A jack (8). Connect the
test leads in series to the measuring points.
The measurement results will appear on the display (5).
Do not measure current >10A in this function, otherwise the fuse will be burnt out.
Fuse replacement
- Turn off the multimeter power and remove the probes.
- Remove the screws fixing the back cover and remove the back cover.
- Remove the burnt out fuse, replace it with a new one of the same specification, and ensure that the fuse is installed in the safety clip and clamped tightly.
- Install the back cover and fix it with screws.
After opening the back cover of the multimeter, do not use the instrument for measurement to prevent electric shock or damage to the instrument.
Specifications
DC voltage, range | 99.99mV / 999.9mV / 9.999V / 99.99V / 999.9V ±(0.5%+3) |
---|---|
AC voltage, range | 99.99mV / 999.9mV / 9.999V / 99.99V / 750V ±(0.8%+3) |
DC current, range 1 | 9.999mA / 99.99mA / 600.0mA ±(0.8%+3) |
DC current, range 2 | 9.999A ±(1.2%+3) |
AC current, range 1 | 9.999mA / 99.99mA / 600.0mA ±(0.8%+3) |
AC current, range 2 | 9.999A ±(1.2%+3) |
Resistance, range 1 | 999.9Ω / 9.999kΩ / 99.99kΩ / 999.9kΩ / 9.999MΩ ±(1.0%+5) |
Resistance, range 2 | 99.99MΩ ±(2.0%+10) |
Capacitance, range 1 | 9.999nF / 99.99nF / 999.9nF / 9.999μF / 99.99μF / |
999.9μF ±(4.0%+3)
Capacitance, range 2| 9.999mF / 99.99mF ±(5.0%+5)
Frequency, range| 9.999Hz / 99.99Hz / 999.9Hz / 9.999kHz / 99.99kHz / 999.9kHz
/ 9.999MHz ±(1.0%+3)
Duty, range| 1–99% ±(3.0%+3)
Temperature measurement range| –40… +1000°C / –40… +1832°F
Accuracy| –40… 0°C / –40… +32°F: ±5%
0… +400°C / +32… +752°F: ±1%
+400… +1000°C / +752… +1832°F: ±2%
Counts| 10 000
Auto-off| 15 min.
Operating temperature range| 0… +40°C (32… 104°F)
Operating humidity range| 0–80% RH
Sorage temperature range| –10… +60°C (14… 140°F)
Storage humidity range| 0–70% RH
Power supply| 3.7V, 1400mA·h rechargeable lithium battery
Safety rating| EN61010-1, EN61010-2-030, EN61010-2-033, EN61326-1; Cat. III,
1000V; Cat. IV, 600V
The manufacturer reserves the right to make changes to the product range and specifications without prior notice.
Care and maintenance
Use the device only within the permitted range. Failure to follow these
instructions may result in electric shock, fire and/or personal injury.
When measuring voltage, do not exceed the operating conditions (see
“Specifications”). When changing functions and ranges, always remove the test
probes from the test points. Do not use the device if it is not working
properly. Please note that the parameters of the powersupply must comply with
the technical characteristics of the device. Do not try to disassemble the
device on your own for any reason. For repairs and cleaning of any kind,
please contact your local specialized service center. Protect the device from
sudden impact and excessive mechanical force. Do not use the device in
aggressive atmosphere. Store the device in a dry cool place. Only use
accessories and spare parts for this device that comply with the technical
specifications. Never attempt to operate a damaged device or a device with
damaged electrical parts! If a part of the device or battery is swallowed,
seek medical attention immediately.
Battery safety instructions
The device is equipped with a built-in rechargeable lithium-ion battery. This
avoids frequent battery replacement. Always switch the device off when not in
use. If the battery charge is low, please recharge the device in time to
ensure that the measurements are accurate. Do not overheat the battery. Do not
discharge the battery completely. Only use the original charger.
Ermenrich Warranty
Ermenrich products, except for their accessories, carry a 5-year warranty
against defects in materials and workmanship. All Ermenrich accessories are
warranted to be free of defects in materials and workmanship for six months
from the purchase date. The warranty entitles you to the free repair or
replacement of the Ermenrich product in any country where a Levenhuk office is
located if all the warranty conditions are met.
For further details, please visit:
levenhuk.com/warranty
If warranty problems arise, or if you need assistance in using your product,
contact the local Levenhuk branch.
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>