ERMENRICH TC40 Digital Multimeter User Manual

September 13, 2024
ERMENRICH

ERMENRICH TC40 Digital Multimeter

PRODUCT OVERVIEW

  1. Power button

  2. Charge jack

  3. NCV sensor

  4. Warning indicator

  5. LCD display

  6. SEL (Selection)/
    Flashlight button

  7. LED indicator

  8. A jack

  9. mA jack

  10. COM jack

  11. INPUT jack

  12. Flashlight

  13. RANGE (Range) button

  14. HOLD/REL
    (Data hold/
    Relative value)
    button

  15. FUNC
    (Functions) button

  16. MAX/MIN (Max.min. values)button

Ermenrich Zing TC40 Digital Multimeter

Please carefully read the safety instructions and the user manual before using this product. Keep away from children. Use the device only as specified in the user manual.

The kit includes: digital multimeter, test leads (red and black), charger, carry bag, user manual, and warranty.

Charging mode
Whenever the charger is connected, the multimeter enters the charging mode.
The charge symbol is displayed on screen and an analog bar with the battery charge level in percent appears. It takes about 3 hours to be fully charged. The warning symbol is displayed.

Do not take measurements while charging.

Getting started

ATTENTION! This device must only be operated by a qualified specialist.
(A qualified specialist is a person whose competence is officially confirmed by relevant documents, who has the necessary knowledge and experience to work safely with this device and is authorized to work with circuits with the appropriate voltage class.)

Press and hold the Power button (1) for 2 seconds to turn the device on/off. This multimeter has two modes – Smart (Auto) and Professional (Manual). When turned on, it displays Auto (Automatic mode) and enters the Smart (Auto) measurement mode.

Range selection
Press the RANGE button (13) to enter the manual range selection mode, and then press again to switch the range. Only voltage, resistance, and current (in mA) measurements are available in this mode. Press and hold the RANGE button (13) for 2 seconds to return to the automatic range selection mode.

This function can’t be used in the Smart measurement mode.

Function selection
Press the FUNC button (15) to activate the manual mode. Then press the SEL/Flashlight button (6) to select the required function.
To return to Smart (Auto) mode, press and hold the FUNC button (15) for 2 seconds.

Data hold
During a measurement, press the HOLD/REL button (14) to turn on data hold function and display the current reading. Press again to turn the function off and return to measurement display.

This function can’t be used in NCV/Live mode.

Flashlight
Press and hold the SEL/Flashlight button (6) for 2 seconds to turn the flashlight on/off.

MAX/MIN mode
Press the MAX/MIN button (16) to switch between maximum and minimum values. Press and hold the FUNC button (15) for 2 seconds to exit this function.

  • This function can’t be used in capacitance, frequency/duty cycle, temperature, NCV/Live modes.
  • During measurement of maximum/minimum values, the multimeter will automatically enter the manual range mode.

Relative value measurement
Press the HOLD/REL button (14) for at least 2 seconds to turn relative value measurement on or off.

This function can’t be used in capacitance, frequency/duty cycle, temperature, NCV/Live modes.
During measurement of relative values, the multimeter will automatically enter the manual range mode.

Relative value measurement
Press the HOLD/REL button (14) for at least 2 seconds to turn relative value measurement on or off.

  • This function can’t be used in capacitance, frequency/duty cycle, temperature, NCV/Live modes.
  • During measurement of relative values, the multimeter will automatically enter the manual range mode.

Fuse Broken indication
When the fuse is blown, the symbol tis displayed.
When the current measurement function is selected, the symbol FUSt will be displayed at the same time and no further measurements will be taken. Please replace the fuse in time.

Input jack indicator
When the function is changed, the corresponding input jack light will flash 5 times to indicate that the test lead should be inserted into the corresponding jack.

Automatic recognition of current measurement
When the test lead is inserted into the A jack, the multimeter automatically switches to the automatic recognition of current measurement function ( ); switching functions will not be allowed at this time.
When the multimeter is switched manually to the current measurement function and no lead is inserted in the A jack, the display shows LEAd (Lead needs to be inserted) and no further measurement is taken.

Auto power turn-off
When you press the Power button ( ) to power on, the automatic turn-off function is activated by default and the display shows the symbol. After 15 minutes without any key operation, the multimeter will turn off automatically to save battery energy.
To cancel the automatic turn-off function, press and hold the SEL/Flashlight button ( ). The symbol is not displayed when the automatic turn-off function is cancelled.

Display information

Smart (Auto) measurement

The multimeter is turned on in Smart (Auto) measurement mode by default. In this mode, DC voltage, AC voltage, resistance, and continuity can be measured, and the multimeter can automatically identify the measurement signal.
Plug the black test lead into the COM jack (10) and the red test lead into the INPUT jack (11). Connect the probes to the measuring points or in parallel to the circuit to be measured. The multimeter will automatically recognize the measured signal. The measurement results will appear on the display (5).

  • Do not measure the voltage higher than 1000V DC or 750V AC; otherwise, the device may be damaged.
  • Pay special attention to safety when measuring high voltage to avoid electric shock or personal injury.
  • Before use, test the known voltage with the multimeter to confirm that the multimeter is in good condition.

Professional measurement

AC/DC voltage measurement (in V)
Press the FUNC button (15) to select the function. Press the SEL/Flashlight button (6) to select the AC voltage or DC voltage. The or DC symbol is displayed accordingly. Plug the black test lead into the COM jack (10) and the red test lead into the INPUT jack (11). Connect the probes to the measuring points. The measurement results will appear on the display (5).

Resistance measurement
Press the FUNC button (15) to select the  function. Plug the black test lead into the COM jack (10) and the red test lead into the INPUT jack (11). Connect the probes to the resistance measuring points of circuit or resistor. The measurement results will appear on the display (5).

Continuity test
Press the FUNC button (15) to select the function. Plug the black test lead into the COM jack (10) and the red test lead into the INPUT jack (11). Connect the probes to the circuit or component to be tested. The measurement results will appear on the display (5). If the resistance is <50Ω, an acoustic signal will be emitted continuously. The measurement results will appear on the display (5).

AC/DC voltage measurement (in mV)
Press the FUNC button (15) to select the function. Press the SEL/Flashlight button (6) to select AC voltage or DC voltage. The or symbol is displayed accordingly. Plug the black test lead into the COM jack (10) and the red test lead into the INPUT jack (11). Connect the probes (in parallel) to the measuring points. The measurement results will appear on the display (5).

Frequency and duty cycle measurement
Press the FUNC button (15) to select the function. Plug the black test lead into the COM jack (10) and the red test lead into the INPUT jack (11). Connect the probes to the measuring points or in parallel to the circuit to be measured. The measurement results will appear on the display (5).

Capacitance measurement
Press FUNC button (15) to select the function. Plug the black test lead into the COM jack (10) and the red test lead into the INPUT jack (11). Connect the probes (in parallel) to the capacitance to be measured. The measurement results will appear on the display (5).

Diode test
Press the FUNC button (15) to select the function. Plug the black test lead into the COM jack (10) and the red test lead into the INPUT jack (11). Connect the black test probe to the cathode side and the red test probe to the anode side of the diode to be measured. If the polarity of the test probes is reverse to the diode polarity, the OL (Overload) will appear on the screen. This can be used for distinguishing the anode and cathode side of a diode. The measurement results will appear on the display (5).

Temperature measurement
Press the FUNC button (15) to select the function. Plug the negative test lead of K-type thermocouple into the COM jack (10) and the positive test lead into the INPUT jack (11). Connect the thermocouple probe to the object to be measured. The measurement results will appear on the display (5).

When using the included K-type thermocouple, temperature measurement range is –40… +260°C / –40… +500°F.

Non-contact voltage detection (NCV)
Press the FUNC button (15) to select the NCV/Live function. NCV (Non-contact voltage detection) is displayed. Slowly bring the NCV sensor (3) close to the point to be detected. When the signal of a weak electromagnetic field is detected, the “– – – L” will appear on the screen, the built-in buzzer will produce a slow beep sound, and the LED indicator (4) will glow green. When the signal of a strong electromagnetic field is detected, the “– – – H” will appear on the screen, the built-in buzzer will produce a quick beep sound, and the LED indicator (4) will glow red.

When using this function, remove the test leads from the jacks.

Live detection
Press the FUNC button (15) to select the NCV/Live function. LIVE (Live detection) is displayed. Plug only the red test lead into the INPUT jack (11). Connect the red test probe to the conductor to be measured. When the signal of a weak electromagnetic field is detected,the “– – – L” will appear on the screen, the built-in buzzer will produce a slow beep sound, and the LED indicator (4) will glow green. When the signal of a strong electromagnetic field is detected, the “– – – H” will appear on the screen, the built-in buzzer will produce a quick beep sound, and the LED indicator (4) will glow red.

AC/DC current measurement (in mA)
Press the FUNC button (15) to select the A function, or insert the red test lead into the mA jack (9) to automatically select A the function.
Press the SEL/Flashlight button (6) to select the AC current or DC current. The symbol or symbol is displayed accordingly. Plug the black test lead into the COM jack (10) and the red test lead into the mA jack (9). Connect the test leads in series to the measuring points.
The measurement results will appear on the display (5).

AC/DC current measurement (in A)
Press the FUNC button (15) to select the function, or insert the red probe into the 10A jack (8) to automatically select the function.
Press the SEL/Flashlight button (6) to select the AC current or DC current. The symbol or symbol is displayed accordingly. Plug the black test lead into the COM jack (10) and the red test lead into the 10A jack (8). Connect the test leads in series to the measuring points.
The measurement results will appear on the display (5).

Do not measure current >10A in this function, otherwise the fuse will be burnt out.

Fuse replacement

  • Turn off the multimeter power and remove the probes.
  • Remove the screws fixing the back cover and remove the back cover.
  • Remove the burnt out fuse, replace it with a new one of the same specification, and ensure that the fuse is installed in the safety clip and clamped tightly.
  • Install the back cover and fix it with screws.

After opening the back cover of the multimeter, do not use the instrument for measurement to prevent electric shock or damage to the instrument.

Specifications

DC voltage, range 99.99mV / 999.9mV / 9.999V / 99.99V / 999.9V ±(0.5%+3)
AC voltage, range 99.99mV / 999.9mV / 9.999V / 99.99V / 750V ±(0.8%+3)
DC current, range 1 9.999mA / 99.99mA / 600.0mA ±(0.8%+3)
DC current, range 2 9.999A ±(1.2%+3)
AC current, range 1 9.999mA / 99.99mA / 600.0mA ±(0.8%+3)
AC current, range 2 9.999A ±(1.2%+3)
Resistance, range 1 999.9Ω / 9.999kΩ / 99.99kΩ / 999.9kΩ / 9.999MΩ ±(1.0%+5)
Resistance, range 2 99.99MΩ ±(2.0%+10)
Capacitance, range 1 9.999nF / 99.99nF / 999.9nF / 9.999μF / 99.99μF /

999.9μF ±(4.0%+3)
Capacitance, range 2| 9.999mF / 99.99mF ±(5.0%+5)
Frequency, range| 9.999Hz / 99.99Hz / 999.9Hz / 9.999kHz / 99.99kHz / 999.9kHz / 9.999MHz ±(1.0%+3)
Duty, range| 1–99% ±(3.0%+3)
Temperature measurement range| –40… +1000°C / –40… +1832°F
Accuracy| –40… 0°C / –40… +32°F: ±5%

0… +400°C / +32… +752°F: ±1%

+400… +1000°C / +752… +1832°F: ±2%

Counts| 10 000
Auto-off| 15 min.
Operating temperature range| 0… +40°C (32… 104°F)
Operating humidity range| 0–80% RH
Sorage temperature range| –10… +60°C (14… 140°F)
Storage humidity range| 0–70% RH
Power supply| 3.7V, 1400mA·h rechargeable lithium battery
Safety rating| EN61010-1, EN61010-2-030, EN61010-2-033, EN61326-1; Cat. III, 1000V; Cat. IV, 600V

The manufacturer reserves the right to make changes to the product range and specifications without prior notice.

Care and maintenance

Use the device only within the permitted range. Failure to follow these instructions may result in electric shock, fire and/or personal injury.
When measuring voltage, do not exceed the operating conditions (see “Specifications”). When changing functions and ranges, always remove the test probes from the test points. Do not use the device if it is not working properly. Please note that the parameters of the powersupply must comply with the technical characteristics of the device. Do not try to disassemble the device on your own for any reason. For repairs and cleaning of any kind, please contact your local specialized service center. Protect the device from sudden impact and excessive mechanical force. Do not use the device in aggressive atmosphere. Store the device in a dry cool place. Only use accessories and spare parts for this device that comply with the technical specifications. Never attempt to operate a damaged device or a device with damaged electrical parts! If a part of the device or battery is swallowed, seek medical attention immediately.

Battery safety instructions
The device is equipped with a built-in rechargeable lithium-ion battery. This avoids frequent battery replacement. Always switch the device off when not in use. If the battery charge is low, please recharge the device in time to ensure that the measurements are accurate. Do not overheat the battery. Do not discharge the battery completely. Only use the original charger.

Ermenrich Warranty
Ermenrich products, except for their accessories, carry a 5-year warranty against defects in materials and workmanship. All Ermenrich accessories are warranted to be free of defects in materials and workmanship for six months from the purchase date. The warranty entitles you to the free repair or replacement of the Ermenrich product in any country where a Levenhuk office is located if all the warranty conditions are met.

For further details, please visit: levenhuk.com/warranty
If warranty problems arise, or if you need assistance in using your product, contact the local Levenhuk branch.

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

Related Manuals