- Method: Chloride reacts with mercury(II) thiocyanate to undissociated mercury(II) chloride. The free thiocyanate forms a blood-red complex with iron(III) ions.
- Reagents: (REF 91667)
- Standards: REF: 985019
- Sampling and Preparation: Quality control involves measuring a blank value and a standard before each series.
Product Usage Instructions
Procedure
- Test a sample of distilled water (REF 918932) to generate a blank value for the reagent.
- Ensure measurements are possible in test tubes (16 mm OD) when using other photometers. Calibrate the method accordingly.
- If using a standard, ensure the measured value remains constant for at least 30 minutes.
Notes
Safety information can be found on the box label and in the safety data
sheet available for download from www.mn-net.com/SDS.
FAQ
-
Q: How often should quality control measures be performed?
A: It is recommended to perform quality control measures by measuring a blank value and a standard before every measuring series. -
Q: Where can I find safety information for the product?
A: Safety information can be found on the box label and in the safety data sheet available for download from www.mn-net.com/SDS.
Overview
The test is suitable for the photometric determination of the chloride
content. The principle of the reaction is in accordance with DIN EN ISO
15682‑D31.
The test is suitable for surface water, ground and drinking water.
- Measuring range:
- 5–200 mg/L Cl (method 0191)
- 0.10–1.00 g/L Cl (method 0192)
- Number of tests: 20
- Wavelength for photometric determination: 470 nm
- Shelf life: 12 months
- Reaction time: 3 minutes
- Storage temperature: 15–25 °C
- Storage conditions: upright
Method
Chloride reacts with mercury(II) thiocyanate to undissociated mer‑ cury(II)
chloride. The free thiocyanate forms a blood-red complex with iron(III) ions.
The colour is evaluated photometrically.
Interferences
The following contaminants do not interfere with the test up to the
indicated concentrations. The cumulative effect of different interfe‑ ring
ions has not been tested.
- Data in mg/L:
- Br, I, SCN, S2O3 2, S2: 0
- F: 20
Turbidities cause higher measurement values. The method is suitable for the analysis of seawater after 1+199 dilution.
Reagents and accessories
- Contents of reagents set:
- 20 test tubes R0
- 2 reagent R2
- Required devices:
- MACHEREY-NAGEL photometer
- Digital piston pipette 1–5 mL (REF 916909) with pipette tips (REF 916916)
- Digital piston pipette 200–1000 µL (REF 91671) with pipette tips (REF 91667)
Standards
- NANOCONTROL Multistandard Metals 1 (REF 925015)
Sampling and preparation
- See DIN EN ISO 5667-3-A 21.
- Adjust to pH 1–13 prior to analysis.
Quality control
The measurement of a blank value and a standard is recommended before every
measuring series as quality control measure.
Quality data:
The following data were determined during production according to ISO 8466-1
and DIN 38402-A51:
- Number of LOTs: 7
- Standard deviation of the method: ± 5 mg/L Cl
- Coefficient of variation of the process: ± 5 %
- Confidence interval: ± 15 mg/L ClSpecified data for procedure:
- Sensitivity (absorbance of 0.010 A corresponds to): ± 1 mg/L Cl
- Accuracy of a measurement value: ± 2 mg/L ClLOT-specific certificates are available at www.mn-net.com.
Procedure
5–200 mg/L Cl-(method 0191)
- Open test tube. Pipette 1 mL of sample into test tube
- Add 1 mL R2
- Mix
- Wait 3 min
- Clean outside of test tube
- Measure
0.10–1.00 g/L Cl-(method 0192)
- Open test tube. Pipette 0.2 mL of sample into test tube
- Add 1 mL R2
- Mix
- Wait 3 min
- Clean outside of test tube
- Measure
Notes
Test a sample of distilled water (REF 918932) to generate a blank value for
the reagent. When using other photometers, make sure measurements are possible
in test tubes (16 mm OD) and calibrate the method. When using a standard, the
measured value is constant over a period of min. 30 min.
Information regarding safety can be found on the box’ label and in the safety
data sheet. You can download the SDS from www.mn-net.com/SDS.
MACHEREY‑NAGEL GmbH & Co. KG
Valencienner Str. 11
52355 Düren · Germany
DE Tel.: +49 24 21 969‑0
info@mn‑net.com
CH Tel.: +41 62 388 55 00
sales‑ch@mn‑net.com
FR Tel.: +33 388 68 22 68
sales‑fr@mn‑net.com
US Tel.: +1 888 321 62 24
sales‑us@mn‑net.com
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>