Druck DPI104, DPI104-IS Digital Pressure Indicator User Guide

July 25, 2024
Druck

Druck DPI104, DPI104-IS Digital Pressure Indicator

Product Specifications

  • Model: DPI104, DPI104-IS
  • Color: Blue (DPI104), Yellow (DPI104-IS)
  • Static Dissipative: Yes
  • Battery Specification: Refer to Batteries on page 2 and Replacing the Batteries on page 2
  • Conformances: CE, UKCA, ETL, IECEx, INMETRO, ATEX, UKEX

Product Usage Instructions

Safety Instructions

The DPI104 / DPI104-IS is designed to be safe when operated following the procedures outlined in the manual. Use the equipment only for its intended purpose as stated in the manual to ensure safety. Before installation and use, familiarize yourself with all  safety procedures, installation standards, and the contents of this document.

Symbols and Markings

The equipment carries CE and UKCA marks indicating compliance with European safety directives and UK Statutory Instruments. Pay attention to warning symbols on the equipment and refer to the user manual when necessary. Ensure proper disposal of the equipment to minimize environmental impact.

Intrinsically Safe Devices

The DPI104 is intrinsically safe and comes in blue color, while the DPI104-IS is yellow. Both models feature static dissipative properties.

Product Markings

Refer to Figure E1 in the manual for product markings including certificate holder’s name and address, CE Mark, UKCA Mark, pressure range, date of manufacture, and country of assembly.

FAQ

Can I repair the equipment myself?

No, do not attempt repairs on the equipment. Return it to the manufacturer or an approved service agent for repair.

What should I do if I need more information on recycling?

If you need information on recycling systems or disposal of the equipment, contact your local or regional waste administration. You can also visit the link provided in the manual for further instructions.

Introduction

The Druck DPI104 / DPI104-IS pressure indicator uses a silicon transducer to produce a pressure reading in pressure units. These user instructions include the operations for all DPI104 / DPI104-IS Pressure Indicators, safety instructions and the requirements for intrinsically safe instruments.
The DPI104 / DPI104-IS is equipped with ISO 228 G1/4 male, or 1/4 NPT male, pressure connector up to 700 bar (10,000 psi). Above 700 bar (10,000 psi), autoclave male pressure connectors are supplied.
For the full specification and user manual, refer to Druck website:

https://druck.com/essential

WARNING

  • Do not use with media that has an oxygen concentration > 21 % or other strong oxidizing agents.
  • This product contains materials or fluids that may degrade or combust in the presence of strong oxidizing agents.
  • Do not use the non-intrinsically safe DPI104 in locations where explosive gas, vapor or dust are present. There is a risk of an explosion.
  • Some liquid and gas mixtures are dangerous. This includes mixtures that occur because of contamination. Make sure that the DPI104 / DPI104-IS is safe to use with the necessary media.
  • It is dangerous to ignore the specified limits (refer to data sheet) for the DPI104 / DPI104-IS or to use the DPI104 / DPI104-IS when it is not in its normal condition. Use the applicable protection and obey all safety precautions.
  • To prevent a dangerous release of pressure, isolate and bleed the system before disconnecting a pressure connection. A dangerous release of pressure can cause injury.
  • To prevent an explosion or fire, use only the Druck specified battery and external supply.

Safety
The DPI104 / DPI104-IS has been designed to be safe when operated using the procedures detailed in this manual. Do not use this equipment for any other purpose than that stated, the protection provided by the equipment may be impaired.
Before installing and using the DPI104 / DPI104-IS, read and understand all the related data. This includes: all local safety procedures and installation standards (for example: EN 60079-14), and this document.

Symbols

  • This equipment meets the requirements of all relevant European safety directives. The equipment carries the CE mark.
  • This equipment meets the requirements of all relevant UK Statutory Instruments. The equipment carries the UKCA mark.
  • This symbol, on the equipment, indicates a warning and that the user should refer to the user manual.
  • Druck is an active participant in the UK and EU Waste Electrical and Electronic Equipment (WEEE) take-back initiative (UK SI 2013/3113, EU directive 2012/19/EU).
    • The equipment that you bought has required the extraction and use of natural resources for its production. It may contain hazardous substances that could impact health and the environment.
    • In order to avoid the dissemination of those substances in our environment and to diminish the pressure on the natural resources, we encourage you to use the appropriate take-back systems. Those systems will reuse or recycle most of the materials of your end life equipment in a sound way. The crossed-out wheeled bin symbol invites you to use those systems.
    • If you need more information on the collection, reuse, and recycling systems, please contact your local or regional waste administration.
    • Please visit the link below for take-back instructions and more information about this initiative.

https://druck.com/weee

Repair
Do not do repairs to this equipment. Return the equipment to the manufacturer or an approved service agent.

Intrinsically Safe Devices

Note: In this document, a “safe area” includes non-hazardous location, unclassified area.

Product Markings

Refer to Figure E1 and key below:

  1. Certificate holder’s name and address.
  2. CE Mark and Notified Body Number (####).
  3. UKCA Mark and Approved Body Number (####).
  4. Pressure range. Example: 20 bar g (g: gauge; a: absolute; sg: sealed gauge).
  5. Date of manufacture (month/year).
  6. Country of assembly: Made in UK/China.
  7. Serial number.
  8. ATEX certificate number.
  9. UKEX certificate number.
  10. European ATEX directive markings.
  11. Hazardous area markings.
  12. IECEx certificate number.
  13. Warning text: “WARNING: Replace battery in safe area only”.

Installation

WARNING

  • Do not open the DPI104-IS when an explosive atmosphere is present – this can cause an explosion.
  • Do not use tools on the DPI104-IS that might cause incendive sparks – this can cause an explosion.
  • Do not connect an energized electrical circuit in a hazardous area while explosive atmospheres are present – this can cause an explosion. Isolate the power supply to the equipment first.
  • Batteries can cause incendive sparks. Install the battery in DPI104-IS in a safe area only.
  • Use only Panasonic Industrial 6LR61AD.

CAUTION
Do not use force to turn the pressure connector or the bezel farther than the end stops. Using force can damage the DPI104 / DPI104-IS.
Do not use the body of the DPI104 / DPI104-IS to tighten the pressure connection, this can cause damage. Use the flat faces on the pressure connector to hold the body and tighten the pressure union.

Batteries

CAUTION Remove the battery from the unit when placed in storage for an extended period of time.
If the battery has leaked return the unit to an approved Druck Service Centre.

For DPI104-IS use only 1 x 9V, IEC 6LR61, ANSI/NEDA 1604 manganese alkaline primary cell from the following list:

Panasonic Industrial 6LR61AD.

Replacing the Batteries

This symbol indicates that the battery is low.
The non intrinsically safe (Blue, DPI104) unit can operate from a 9V, IEC 6LR61, ANSI/NEDA 1604 alkaline battery.

The intrinsically safe (Yellow, DPI104-IS) unit shall only use the batteries specified in section “Batteries” on page 2.

Fit as shown in Figure A1 using the following procedure:

  1. If applicable, set the power to off and isolate the external power supply.
  2. Remove the display bezel (steps 1 and 2).
  3. Make sure that the o-ring (item 8) and the related surfaces are serviceable. Use only original parts supplied by the manufacturer.
  4. Remove the battery clamp/cover (step 3, item 7).
  5. Disconnect the battery connector (step 4, item 9) and discard the used battery (step , item 10).
  6. Use an applicable recycling facility.
  7. Attach the battery connector (item 9) to the new battery.
  8. Install the new battery (item 10) and re-attach the battery clamp/cover (item 7).
  9. Push the display bezel back into position until it is fully engaged.

Specific Conditions of Use
When the power supply is through the 8-pin connector, use only a Type A or a Type B cable as specified in
IEC 60079-14.

Declaration RequirementsEU Directive 2014/34/EU
This equipment is designed and manufactured to meet the essential health and safety requirements not covered by EU Type Examination Certificate ITS07ATEX25517X when installed as detailed in this guide.

Declaration Requirements – UK SI 2016/1107 (as amended by SI 2019/696)
This equipment is designed and manufactured to meet the essential health and safety requirements not covered by UK-Type Examination Certificate ITS21UKEX0076X when installed as detailed above.

Pressure Safety
This equipment meets the requirements of the European Pressure Equipment Directive 2014/68/EU and UL 61010-1 for pressure safety.

Return Goods/Material Procedure
If the unit requires calibration or is unserviceable, return it to the nearest Druck Service Centre listed at: https://druck.com/service.
Contact the Service Department to obtain a Return Goods/Material Authorization (RGA or RMA). Provide the following information for a RGA or RMA:

  • Product (e.g. DPI104 / DPI104-IS)
  • Serial number.
  • Details of defect/work to be undertaken.
  • Calibration traceability requirements.
  • Operating conditions.

Quick Start

Before Using the Instrument

  • Read and understand the “Safety” section.
  • Check the DPI104 / DPI104-IS for signs of damage. Do not use a damaged DPI104 / DPI104-IS.

Power On Sequence
Refer to Figure B1 and the procedure below.

  1. Press the button.
  2. The display initially shows the full-scale rating of the instrument.
  3. After a short delay, the pressure measurement is displayed.

Power Off Sequence
Refer to FigureB2 and the procedure below.

  1. Display showing pressure measurement.
  2. Press and hold the button.
  3. The display shows (off).
  4. Press the button to turn off.

Change the Pressure Units
Refer to Figure C1 and the procedure below.

  1. Display showing pressure measurement in mbar pressure units.
  2. Press and hold the button.
  3. The display shows (off).
  4. Press the button again.
  5. The display shows (units).
  6. Press the button to cycle through the available pressure units.
  7. Example shows psi pressure units flashing.
  8. Press the button to accept psi pressure units.
  9. Display showing pressure measurement in psi pressure units.

Set Tare (or Set Zero)
Refer to Figure D1 and the procedure below.

  1. Display showing pressure measurement of 70mbar.
  2. Press and hold the button to tare the instrument to a zero reading.
  3. The display shows (done).
  4. Display showing pressure measurement with a tare value of 70mbar.

Office Locations

https://druck.com/contact

Services and Support Locations

https://druck.com/service

Copyright 2018 Baker Hughes Company. This material contains one or more registered trademarks of Baker Hughes Company and its subsidiaries in one or more countries. All third-party product and company names are trademarks of their respective holders.

bakerhughes.com

Copyright 2018 Baker Hughes Company.