TSI 8540 and 8543 Pump Replacement Instructions
- July 10, 2024
- tsi
Table of Contents
TSI 8540 and 8543 Pump Replacement
Specifications :
- Product: Replacement Pump (854043) for Environmental DustTrakTM Aerosol Monitor
- Compatible Models: 8540 and 8543
- Manufacturer: TSI
Product Usage Instructions
Unpacking:
- Carefully unpack the Replacement Pump from the shipping container.
- Verify that all items listed in the table are present.
- Contact TSI immediately if any items are missing or broken.
Tools Needed for Installation:
- Phillips Screw Driver (not included)
Replace Pump:
- Remove photometer service panel.
- Remove the four (4) Phillips-head screws from the left side vpanel, ensuring to remove only the raised-head screws.
- Pull the pump from the instrument case.
- Disconnect the vacuum tube, muffler, and power cable from the pump module.
- Obtain replacement pump. Connect muffler to (P), vacuum tube to (V), and power cable.
- Place the new pump into the photometer case with the muffler oriented at the bottom.
- Secure pump module in the instrument case with four (4) Phillips-head screws on the left side.
- Replace photometer service panel and reinstall photometer inside the environmental enclosure.
FAQ :
-
Q: What should I do if I encounter a hot pump during installation?
A: Take caution as the pump can be hot to the touch. Allow it to cool down before proceeding with the installation. -
Q: How do I contact TSI for assistance?
A: You can reach out to TSI using the contact information provided in the manual based on your location.
Replacement Pump (854043) for Environmental DustTrak™ Aerosol Monitor
Models 8540 and 8543
Quick Start Guide
This quick start will guide you through the steps to install the
replacement pump in the Model 8540/8543 Environmental DustTrak™ Aerosol
Monitor.
| C A U T I O N
---|---
Pump can be hot to the touch.
Unpacking
- Carefully unpack the Replacement Pump from the shipping container and verify that all the items listed in the following table are present.
- Contact TSI immediately if items are missing or broken.
- Additional items may be included if you ordered accessories or spare parts.
Qty.
| Item Description|
Reference Picture
---|---|---
1| Replacement Pump and muffler (854043)|
Tools Needed for Installation
Phillips Screw Driver (not included)
Remove Photometer Unit from Environmental Enclosure
-
Disconnect all cabling from photometer and heated inlet if present.
-
Remove inlet column. If the inlet column includes a heated inlet remove the ferrite from the heated inlet cable and carefully pull the heated inlet cable through the key hole on the mounting ring when removing the inlet column.
-
Loosen retainer ring on top of enclosure while holding photometer.
-
Pull photometer down and out of enclosure.
Replace Pump
-
Remove photometer service panel.
-
Remove the four (4) Phillips-head screws from the left side panel.
NOTICE
Remove only the raised-head screws. DO NOT remove any flush-head screws. -
Pull the pump from the instrument case
-
Disconnect the vacuum tube, muffler, and power cable from pump module.
-
Obtain replacement pump. Connect muffler to (P), vacuum tube to (V), and power cable.
-
Place new pump into photometer case with muffler orientated at the bottom.
-
Secure pump module in instrument case with four (4) Phillips-head screws on left side.
-
Replace photometer service panel and reinstall photometer inside environmental enclosure.
TSI Incorporated – Visit our website www.tsi.com for more information.
- USA Tel: +1 800 874 2811
- UK Tel: +44 149 4 459200
- France Tel: +33 1 41 19 21 99
- Germany Tel: +49 241 523030
- India Tel: +91 80 67877200
- China Tel: +86 10 8219 7688
- Singapore Tel: +65 6595 6388
- P/N 6008823 Rev. B ©2024 TSI Incorporated Printed in U.S.A.
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>