Alandental KD-202D-3B Wall Mounted Medical Examination Light User Manual

July 9, 2024
Alandental

Alandental KD-202D-3B Wall Mounted Medical Examination Light

Alandental-KD-202D-3B-Wall-Mounted-Medical-Examination-
Light

Brief introduction

  • It adopts 12pcs brand 3W LED bulbs as the light source, double switch control, and the brightness adjustable. The lamp head and arm rotate freely, multi-angle lighting. lean on the wall when not in use.
  • The product is mainly composed of lamp panel, power box, spring arm, hanging bracket.
  • The diameter of lamp panel is200mm, larger than the ordinary examination light, smaller than traditional LED shadowless lamp. Make up for small light spot, low illumination, short lighting depth of the examination lamp.
  • This product installed on the wall, easy to operate.
  • This product adopts wide voltage input, low voltage output power supply, stable and reliable performance. Suitable for clinic, dental, gynecology, medical beauty, and pet medical.

Performance Summary
The design and production of this examination light strictly executes the first part of the safety requirements of medical electrical equipment GB9706. 1-2007: general requirements which is guaranteed in safety.

Main technical parameters

  1. Input voltage: AC90 240V 50H, 60Hz
  2. Input power: < 50VA
  3. Power of bulb: 3W*12
  4. Color temperature: 5500 ± 500K
  5. Diameter of lamp holder: 200 mm
  6. Working distance: 700 mm

Security tips and symbol instructions

  1. When the examination light is in using, you should always abide by the basic safety precautions in order to avoid fire, electric shock or the risk of injury.
  2. Please read the instructions carefully, and operating according to the specifications and requirements
  3. Before clean the product, unplug the power line, and do not use liquid or spray cleaners.
  4. Please don’t use the cold light source near the water. Don’t put anything on the power line of the examination light.
  5. When replace the bulbs, you must turn off the power switch of the light source and unplug the power cord.
  6. Maintenance should be performed by professiona·1 s.
  7. The brightness of this examination light is very high, when you use, don’t contact with eyes directly, in case of damage.
  8. Symbol:
    O Disconnect (main power supply) I Connect (main power supply)

Working environment

  • Environment temperature: -5 °C  40°C
  • Relative humidity: 35% 75%
  • Atmospheric pressure: 700hPa !060hPa

Product Interface

Alandental-KD-202D-3B-Wall-Mounted-Medical-Examination-
Light-1

  1. Lamp holder
  2. Lamp switch
  3. Brightness adjusting knob
  4. Decoration cover
  5. Lift arm
  6. Horizontal arm
  7. Rotation axis
  8. Expansion screws
  9. Hanging board
  10. Indicator light
  11. Power box
  12. Power switch
  13. Power cord interface
  14. Handle

Usage

  1. Check before use
  2. Open the package, check whether it is damaged or not.
  3. Check whether the model, name and quantity are in keeping with the “configuration list”; Please contact with factory or distributor in time to get properly resolved if you find something incorrect.
  4. Read the instructions carefully and follow the installation steps.
  5. Installation
  6. First positioning the four fixed holes in the Hanging board on the wall, the holes are shown as Picture 2-A. The upper end is drilled with two embedded holes, depth of 110mm, the drill size must be $12mm ;The bottom end is drilled with two embedded holes, depth of 30mm, the drill size must be $6mm, then load the expansion pipe.
  7. Remove the four screws from the Rotation axist, connect with Hanging board9, fix the screws, according to Picture 2-B.
  8. Install the Hanging board9 on the wall according to Picture 2-C, use two expansion screws to fix the upper end, use two self- tapping screws to fix the bottom end , then install the power box according to step 2-D. E. F.
  9. Install the Decoration cover 4, connect the docking plug, to finish the whole steps according to 2-G. H. IAlandental-KD-202D-3B-Wall-Mounted-Medical-Examination-Light-2Alandental-KD-202D-3B-Wall-Mounted-Medical-Examination-Light-3
  10. Usage
  11. Connect the power cord with internet electric power supply and Power cord interface(3), turn on Power switch(2), the Indicator light bright up; open the lamp switch2, the lamp bright up, the examination light start to work.
  12. Adjust the brightness control3 could change the brightness level from zero to maximum.
  13. Hold the handle(4) could control the height of lamp holder, rotate the lamp arm left and right around the Rotation axisO.
  14. Hold the handle(4, take the lamp holder support as a fulcrum, rotate the lamp holder left and right to change the lighting angle. [Noticel Take the joint between Lift arm5 and Horizontal arm® as fulcrum, tuck the lamp against the wall when not in use.

Attention

  1. Please pull out the power plug before clean the lamp, do not use liquid spray to clean it.
  2. When install the lamp, please assemble and reinforce it according to the instruction of the technical requirements.
  3. Do not use the lamp near water.
  4. The repair, installation and commissioning of the operation lamp should be performed by professionals.
  5. Never put any item on the power line.
  6. The old light bulb should be put into the waste barrel. Don’t throw it carelessly.
  7. The light intensity of illumination is higher, you should be cautious in use with your eyes.

Failure judgment and eliminate methods

(This link requires professional operation)

Failure phenomenon Reason Solution
Open the power switch, the indicator is not bright up 1. Power lines have no
electricity Connect to the circuit where there is electricity supply

2. The power cord is not connected well

3. The system faulty

| Reconnect the power cord

Connect the after-sales department to get solved

Open the power switch, the indicator is bright, but the lamp is not bright up| 1. The lamp switch is not turned on (Turn on the switch 2)
2. The brightness at the minimum position (Adjust the brightness adjustment knob 3 to the maximum)
3. Poor contact with docking plug (Reconnect the 2-G)

Bulb replacement
The lamp bulb is LED light source, the normal life up to 10000 hours, no need to replace normally. In case of any failure, please inform the manufacturer in time for processing.

After-sales service

  • The product warranty period is 18 months.
  • During warranty period, Send or give us your warranty and purchase invoices, we will provide free maintenance service when you have quality problems.
  • Light bulbs are consumable accessories.
  • If you modified the function by yourself and then it does not work, is not covered under warranty.
  • If the marks on the product such like model, serial number, date of manufacture has been altered, deleted, or unidentifiable, will not provide free warranty service.
  • If moving falling, squeezing, soaking, moisture and other failures caused by human factors can not enjoy our free warranty service.
  • The production date shown on label, service life is 5 years, we provide lifetime maintenance.

XII Transportation and storage condition
Relatone temper 05 805100-550 Atmospheric pressure: 500hpa™ 1060hpa

Electrical schematic diagram and components list

Alandental-KD-202D-3B-Wall-Mounted-Medical-Examination-
Light-4

Alandental-KD-202D-3B-Wall-Mounted-Medical-Examination-
Light-6

29 Resistance 15 n /0. 25VA
30 Resistance 120K <l /0. 25VA
JI lfosi stance 12K fl /0. 25VA
32 Resistance !OK n /0. 25VA
33 Resistance lK n /0. 25VA
34 Resistance 22K0/0.25VA
35 Resistance 15K n /0. 25VA
36 Resistance 5. 6KO /0. 25VA
37 ffosi stance 560 n /0. 25VA
38 Resistance I. 2KQ /0. 25VA
39 Resistance 3. 9KO /0. 25VA
40 lfosi stance 680KQ /0. 25VA
41 Resistance I. 5KQ /0. 25VA
42 Resistance 100K<l /0. 25VA
43 Resistance 4. 7KQ /0. 25VA
44 Resistance 220 Q /0. 25VA
45 Resistance 150K <l /0. 25VA
46 Resistance 2 <l /0. 25VA
17 lfosistance 22K/O. 25VA
48 Resistance 24 n / 0. 25VA
49 Resistance 3. 9KIO. 25VA
50 Resistance 100 n /3VA
51 Resistance 680 n /3VA
52 Diode Fl07
53 Diode I N4752
54 Diode TN4148
55 Audion 1815
56 l’hermi stor 51J-]5
57 Pie1.ores i stor I0D47 IK
58 Integrated circuit LM494 I
--- --- --- ---
59 Rectifier bridge KBUSIO I
60 Hui b I.Ell 54
61 Powe1 socket I
62 Powe1 line I
63 Fuse holder 2
64 Power switch I

Warranty card and maintenance card

Alandental-KD-202D-3B-Wall-Mounted-Medical-Examination-
Light-7

Configuration list

Alandental-KD-202D-3B-Wall-Mounted-Medical-Examination-
Light-8

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals