Alandental TR-DOM01 Dental Surgical Microscope User Guide

June 26, 2024
Alandental

Microscope Specification
Dental stereo microscope instructions


(This manual describes in detail the installation method of the dental microscope. Instructions for use and maintenance and care, please read the instructions carefully before use.)


Cautions

Operation

  1. DO NOT exposed to sunlight. Should be placed in a dry, clean environment, avoid high temperatures and severe vibration.
  2. The microscope is a precision instrument and should be handled carefully to avoid shocks and collisions during transportation.
  3. DO NOT leave dirt or fingerprints on the surface of the lens, so as not to affect the clarity of the image.

Maintenance

  1. All lenses should be kept clean. If there is any fine dust, blow it off with a blower or gently wipe off the cotton yarn. If there are traces of oil and fingers, use a small amount of cotton yarn with a 3:7 ratio of ethyl alcohol and ether to gently wipe off.
  2. Do not use organic solvents to wipe the surface of the microscope, especially the surface of plastics, and use neutral detergents for cleaning.
  3. Do not disassemble the microscope by yourself so as not to affect the performance of the microscope
  4. Do not touch dust, cover it with a dust cover, store it in a place away from moisture, so as not to rust or mold, when the microscope is not in use.
  5. To maintain the performance of the microscope, regular inspections are recommended.

Part Names and Functions

Parts Function Diagram

Eyepieces, objectives, light source, fine focus,
power supply box, 250° rotation, 360° rotation,
360° adjustment, power cord, 80cm, 33cm,
32cm, 38.5x33cm, 34x22cm, 42x36cm

Installation Instructions

Step 1

Screw on the picture and tighten.

Step 2

The elastic arm is inserted into the metal fixing ring.
Remember that there is a side with +/- adjustment of elasticity!
Then use the Allen screws to tighten as shown below.

Step 3

Insert one end of the large port of the power box with robotic arm The elastic arm is then tightened with the Allen screw.

Step 4

Insert short metal arm and tighten it.

Step 5

Put in the sleeve, set it to the end, tighten the screws, and remember to tighten.

Step 6

Plug the power cord into the flashlight port.

Step 7

Insert the flashlight and screw down the small screws below.

Installation Method

  • The normal use of the instrument room temperature +5 ° – + 40 ° , the power cord is plugged into the power grid, power switch to “NO” position.
  • Home, irradiation light is on. Place the object under observation at the center of the objective lens, hold the robotic arm and move up and down.
  • Micromirror, so that the observed object is at the working distance (about 230mm), then slowly turn the focusing handwheel,
  • Makes the right eyepiece observe a clear object image; then observe the left eyepiece, if the object is not clear, adjust the viewing tube.
  • Circle, so that it can get a clear image, and then turn the left and right eyepiece tube, so that the two objectives of the exit distance and operation of the pupil.
  • The viewer’s eyesight is appropriate, so that you can see a clear, three-dimensional image.

Configuration List

  • 1,Microscope section: eyepiece 10X (optional 20X), objective lens lx.
  • 2, the microscope arm bending: three angles can rotate arm angle.
  • 3, Worktable: total height 80CM, upper disk 38.5CM33CM, middle disk 34CM22CM,Lower disk 42CM*36CM.

Applicable Place

  • Stomatological hospital, dental clinic, beauty salon, pet hospital, ENT clinic, etc.

Technical Parameters

Comprehensive Rate 10X 20X
Model TR-DOM01 TR-DOM01
Eyepieces 10X2Omm 20X2Omm
Objectives lens 1X 1X
Working Distance 230mm 230mm
Light Source 5W LED 5W LED

Troubleshooting

Question Reason Method
1.Double images do not coincide Incorrect pupil distance adjustment amend

Pupil distance
Incorrect vision adjustment| Re- adjust visibility
Lett and light eyepieces with different
magnification| Install the same eyepiece
2.Diny things in the field of view| Din on Specimen| Clean specimens
Din on eyepiece surface| Clean eyepiece
3.Unclear image| Objective lens surface with dirt| Clean objective lens
4.Unclear zoom| Incorrect diopter adjustment| Re – adjust visibility
Incorrect diorites adjustment| Re – adjust visibility
 | Incorrect Focusing| Refocus
5.The microscope
Tess body descends to
make the image clear| Arm bending elasticity is too weak| Increased flexibility
6.Eyes arc easily fatigued| Incorrect diopter adjustment| Adjust diopters correctly
7.Light is off when switch is on| No power| Check the connection of the power cord
Bulb mot inserted| insert correctly
The light bulb doesn’t work| Replace bulb
8. The bulb suddenly
burnt out| Unspecified bulbs are used| Use the specified bulb
Overtension| Control voltage

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

Alandental User Manuals

Related Manuals