ALANDENTAL WOODPECKER Dental LED.E Curing Light
Introduction
Guilin Woodpecker Medical Instrument Co., Ltd. is a high-tech enterprise in researching, developing, and producing dental equipment, and ha a perfect quality assurance system, main products including ultrasonic scaler, curing light, apex locator and ultrasurgery etc.
Principle And applicatuon
LED.E adopts the principle of ray radiation to solidify the light-sensitive
resin by shooting at it in a short time.
This product is used for dentistry. It has the function of accelerating dental
restoration and solidifying the material of dental whitening.
Product Performance Structure and Components
LED.E curing light (dentistry) is mainly composed by high power LED, optical fiber, main unit and charge device.
Basic Technical Specifications
-
Adapter:
-
Rechargeable Lithium battery:
Battery voltage and capacity: 3.7V, 1400mAh
Battery model: ICR18490 -
Adapter input: AC100~240V 50Hz/60Hz Output: DC5V 1 A
-
-
Applied part: Optical fiber
-
Light source:
- 5W high power blue light LED
- Wave length: 420nm-480nm
- Light intensity: 1000mW/cm2~1200mW/cm2
-
Work condition:
Environment temperature: +5″C to +40″C
Relative humidity: :S80%
Atmosphere pressure: 70kPa tol06kPa -
Size: 40x50x260mm
-
Net weight: 14 7g
-
Power consume: :S8W
-
Protection type against electrical shock: Class II
-
Protection degree against electrical shock: Type B
-
Protection against harmful ingress of water or particular matter: ordinary equipment (IPX0)
-
Safety in the presence of flammable anesthetic mixture with air, oxygen or nitrous oxide: not suitable under this condition.
In~allation and Demounting
- Take off the red cap from the optical fiber, then insert the metal part into the front of main unit (make sure to screw when installing, oblique insert should be prevented).
- Fix the light hood on the bottom of the optical fiber.
- Un install the LED, just reverse the procedure above.
- Battery replacement method: open the battery cover of the main unit, take the battery out, than disconnect the plug slightly. Connect theplug of the new battery correctly, put the new battery in, and then fix the battery cover.
- After using the machine or it is in need of charging, make the main unit headon inserted in the pedestal and compress it to make sure the main unit and pedestal chucking. When there is no need of charging, please pull out the pedestal Adapter or put main unit into the pedestal inversely.
Operation
-
Press the mode button to set the working mode, the corresponding indicator will on when a mode set.
-
Full power: the blue light radiates in full power.
-
Ramping: the blue light power increases stronger continually, after 5 seconds reaches to the highest power.
-
Pulse: the blue light works on the pulse condition.
-
Press the time button to set the solidifying time, 4 working time is available: 5, 10, 15, 20seconds.
-
When operating, aim the optical fiber at the correct position, press the power button to start or stop to emit of the blue light.
-
During operation, the blue light can be stopped by press the power button at any time.
-
A battery detect circuit is fixed inside the main unit, when low power is detected, the indicator of the main unit will wink, please charge in time. Connect the Adapter and the charge device correctly, put the main unit into the charge device, the yellow indicator will enlightened, when charge finish, the green indicator will enlightened and the yellow one will off. The green light is the power indicator of the charging finish. If a battery error is detected, the indicator of the charge device will wink.
-
When operating finish, please clean the optical fiber with calico to avoid infecting the light intensity.
-
This equipment will turn off automatically if don’t any action within 2 minutes, tum it on by press any button.
-
The effective light intensity of this equipment is much more higher than Halogen Lamp, The solidified depth of the curing light composites resin for 10 seconds will not less than 4mm.
-
This equipment has the function of light intensity measuring. Connect the Adapter of the light intensity measure device correctly, aim the top of optical fiber at the sensor, turn on the main unit to emit blue light, the green indicator of the light intensity measure device will
indicate the value of the light intensity. -
The optical fiber can be autoclaved under the high temperature of 135°C and pressure 0.22 Mpa. (Refer to EN13060 standard)
-
The curing light is equipped with an over-heat protection system. It can continuously work 200s, For example, continuously operate the curing light for 10 times under 20s working mode (even the curing light works less than 20s, it is counted as a full operation), then it will come into over-heat protection status. And only after 2-minute sleep, it can restart working 200s continuously.
light intensify measurement
- Connect the output plug of power adapter to the plug of DC5 .OV in the pedestal.
- Choose general mode and aim the optical fiber at the measurement point, press on / off button, the present light intensity is displayed on the indicator of pedestal.
Precaution
- Please recharge the battery at least 4 hours before first time usage.
- During operation, the light should be aimed straightly at the resin to ensure the effect of solidification.
- Avoid aiming the blue light at eyes directly.
- Only the original pedestal charger, adapter and Lithium battery could be used, because other brand pedestal charger, adapter and Lithium battery are likely to damage the circuit.
- It is forbidden to touch the charging connector with metal or other conductor, to avoiding damage the circuit of charge or the battery.
- Charging the battery in the condition of cool and ventilated. Please make sure of pressing out the buckle between the main unit and the pedestal, otherwise the battery charging might be failed because of the poor contact.
- Do not disassemble the Lithium battery, it will lead to the circuit short or the electrolyte leakage.
- Do not squeeze, shake and short the battery, do not store the batterywith metal material.
- WARNING: If the curing light works for 40s continously, the temperature of the top of optical fiber may reach 56’C.”
- WARNING: Do not modify this equipment without authorization of the manufacturer.
Contraindication
The heart disease patients, pregnant women and children should be cautious to use this equipment.
Maintenance
- Only the optical fiber of this equipment can be autoclaved under high temperature and high pressure, other parts should be cleaned by clean water or neutral sterilized liquid, but do not soak the equipment in the water. Do not clean by volatile or soluble liquid, otherwise the marks of the control panel will fade or some spare part will be corrosion. Especially for the charging base, please use medical alcohol (75%) to clean it.
- Please clean the optical fiber to avoid the remaining resin on the surface and infect the life-span and the effectiveness of solidification.
Troubleshooting
If all the above solutions have been completed, the machine still can not work normally. Please contact our special repair shop or us.
Packing list
The components of the machine are listed in the packing list.
Transportation
- Excessive impact and shake should be prevented in transportation. Lay it carefully and lightly and don’t invert it.
- Don’t put it together with dangerous goods during transportation.
- Avoid solarization and getting wet in rain or snow during transportation.
Storage and transportation
- The equipment should be handled carefully and lightly, kept away from the shaking source, and installed or stored at shadowy, dry, cool and ventilated places.
- Don’t store the equipment together with articles that are combustible, poisonous, caustic, and explosive.
- This equipment should be stored in an environment where the humidity is :S80%, the atmosphere pressure is 70kPa~106kPa and the temperature is -10°C~55°C.
- Excess impact or shake should be prevented during transportation. Handle with care. Do not place upside down.
- Don’t put it together with dangerous articles during transportation.
- Keep it away form the sun, rain or snow during transportation.
After service
From the date this equipment has been sold, base on the warranty card, we will repair this equipment free of charge if it has quality problems, please refer to the warranty card for the warranty period.
For technical data, lease contact
Wellkang Ltd (www.CE-Marking.eu) 29 Harley St.,LONDON,W1G 9QR,UK
Symbol instruction
Authorised Representative in the EUROPEAN
COMMUNITY
Environmental protection
There are no harmful factors in our product. You can deal with it based on the local law.
Product conforms to the following standards
- EN 60601-1 :2006 EN 60601-1-2:2007
- EN 60601-1-6:2010 EN 62366:2008
- EN 60825-1 :2007 EN 980:2008
- ISO 9687:1993 EN 1041:2008
- ISO 15223-1-2012 EN ISO 14971:2012
- EN ISO 17664:2004 EN ISO 17665-1:2006
- EN ISO 7405:2008 + Al:2003 EN ISO 10993-1:2009
EN ISO 10993-5:2009 EN ISO 10993-10:2010
EMC – Declaration of conformity
The device has been tested and homologated in accordance with EN 60601-1-2 for EMC. This does not guarantee in any way that this device will not be effected by electromagnetic interference Avoid using the device in high electromagnetic environment.
Statement
All rights to modify the product are reserved to the manufacturer without further notice. The pictures are only for reference. The final interpretation rights belong to GUILIN WOODPECKER MEDICAL INSTRUMENT CO., LTD. The industrial design, inner structure, etc, have claimed for several patents by WOODPECKER, any copy or fak product must take legal responsibility.
Scan and Login website for more intonation
Guilin Woodpecker Medical Instrument Co., Ltd. Information lndusb’ial Park, National High-Tech Zone, Guilin, Guangxi, 541004 P. R. China
Tel:
- Europe Sales Dept: +86-~73196, +86-TT3-2125222
- North America, South America &
- Oceania Sales Depl.:+86-773-5873198, +86-TT3-2125123
- Asia & Africa Sales Dept:+86-TT3-5855350, +86-773-2125896
Fax: +86-TT3-5822450
E-mail: woodpedtel@glwoodpecker.oom
sales@glwoadpecker.com
Website: http://www.glwoodpecker.com
Wellkang ltd (www.CE-Marking.eu) 29HarleySt.,LONDON,W1G9QR,UK