MEAN WELL RPS-120S Series 120W 3×2 Reliable Green Medical Power Supply
User,s Manual
Features
- 3″x2″ compact size
- 120W convention, 150W peak (1 0sec.)
- Medical safety approved (2 x MOPP) accroding to ANSI/AAMI ES60601 -1 and IEC/BS EN/EN60601 -1
- Suitable for BF application with appropriate system consideration
- EMI for both Class I & Class II configuration
- -30~+85°C wide range operating temperature
- No load power consumption<0.3W
- Extremely low leakage current
- Protections: Short circuit / Overload / Over voltage / Over temperature
- Operating altitude up to 4000 meters (Note.6~
- 3 years warranty
Applications
- Oral irrigator
- Hemodialysis machine
- Medical monitors
- Sleep apnea devices
- Pumps machine
GTIN CODE
MW Search: https://www.meanwell.com/serviceGTIN .aspx
Description
RPS-120S is a 120W highly reliable green PCB type medical power supply with a high power density on a 3″ by 2″ footprint. It accepts 80~264VAC input and offers various models with the output voltages between 12V and 48V. The working efficiency is up to 94% and the extremely low no load power consumption is down below 0.3W. RPS-120S is able to be used for both Class I (with FG) & Class TI (no FG) system design. The extremely low leakage current is less than 150μ A. In addition, it conforms to the international medical regulations (2*MOPP) and EMC BS EN/EN55011, perfectly fitting all kinds of BF rated “patient contact” medical system equipment.
Model Encoding
SPECIFICATION
MODEL| RPS-120S-12| RPS-120S-15| RPS-120S 24| RPS-
120S-27| RPS-120S-48
---|---|---|---|---|---
OUTPUT| DC VOLTAGE| 12V| 15V| 24V| 27V| 48V
CURRENT| Peak(10 sec.)| 11.8A| 9.5A| 6.25A| 5.55A| 3.125A
Convection| 9.5A| 7.6A| 5A| 4.44A| 2.5A
RATED POWER| Peak(10 sec.)| 141.6W| 142.5W| 150W| 149.8W| 150W
Convection| 114W| 114W| 120W| 119.9W| 120W
RIPPLE & NOISE (max.) **Note.2| 100mVp-p| 120mVp-p| 150mVp-p|
150mVp-p| 200mVp-p
VOLTAGE ADJ. RANGE| 11.4~12.6V| 14.3~15.8V| 22.8~25.2V| 25.6 ~ 28.4V|
45.6 ~50.4V
VOLTAGE TOLERANCE Note.3| ±2.0%| ±2.0%| ±1.0%| ±1.0%| ±1.0%
LINE REGULATION| ±0.5%| ±0.5%| ±0.5%| ±0.5%| ±0.5%
LOAD REGULATION| ±1.0%| ±1.0%| ±1.0%| ±1.0%| ±1.0%
SETUP, RISE TIME| 600ms, 30ms/230VAC 600ms, 30ms/115VAC at
full load
HOLD UP TIME (Typ.)| 15ms/230VAC 15ms/115VAC at full
load
INPUT| VOLTAGE RANGE Note.4| 80 ~ 264VAC 113 ~
370VDC
FREQUENCY RANGE| 47 ~ 63Hz
POWER FACTOR| PF>0.94/230VAC PF>0.98/115VAC at full load
EFFICIENCY (Typ.)| 91%| 92%| 93%| 94%| 93.5%
AC CURRENT (Typ.)| 2.3A/115VAC 1.1A/230VAC
INRUSH CURRENT (Typ.)| COLD START 30A/115VAC
60A/230VAC
LEAKAGE CURRENT(max.) Note.5| Earth leakage current < 150μA/264VAC ,
touch current < 80μA/264VAC
PROTECTION| OVERLOAD| 130~160% rated output power
Protection type : Hiccup mode, recovers automatically after fault condition is
removed
OVER VOLTAGE| 13.2 ~ 15.6V| 16.5 ~ 19.5V| 26.4 ~ 31.2V| 29.7 ~ 35V| 52.8
~ 62.4V
Protection type : Shut down o/p voltage, re-power on to recover
OVER TEMPERATURE| Protection type : Shut down o/p voltage, recovers
automatically after temperature goes down
ENVIRONMENT| WORKING TEMP.| -30 ~ +85℃ (Refer to “Derating Curve”)
WORKING HUMIDITY| 20 ~ 90% RH non-condensing
STORAGE TEMP.| -40 ~ +85℃
TEMP. COEFFICIENT| ±0.03%/℃ (0 ~ 50℃)
VIBRATION| 10 ~ 500Hz, 2G 10min./1cycle, 60min. each along X, Y, Z axes
OPERATING ALTITUDE Note.6| 4000 meters
SAFETY & EMC**
(Note 7)
| SAFETY STANDARDS| IEC 60601-1:2005+A1+A2, TUV BS EN/ EN 60601-1:2006+A1+A12+A2, ANSI/AAMI ES60601-1:2005+A2
CAN/CSA C22.2 No. 60601-1:2014+A2, EAC TP TC 004 approved; Design refer to BS EN/EN60335-1(by request)
ISOLATION RESISTANCE| Primary-Secondary: 2xMOPP, Primary-Earth:1xMOPP,
Secondary-Earth:1xMOPP
WITHSTAND VOLTAGE| I/P-O/P:4KVAC I/P-FG:2KVAC O/P-FG:1.5KVAC
ISOLATION RESISTANCE| I/P-O/P, I/P-FG, O/P-FG: 100M Ohms / 500VDC / 25℃/
70% RH
EMC EMISSION| Parameter| Standard| Test Level / Note
Conducted emission| BS EN/EN55011 (CISPR11)| Class B
Radiated emission| BS EN/EN55011 (CISPR11)| ClassⅠ: Class B , Class Ⅱ: Class A
Harmonic current| BS EN/EN61000-3-2| Class A
Voltage flicker| BS EN/EN61000-3-3| —–
EMC IMMUNITY| BS EN/EN60601-1-2
Parameter| Standard| Test Level / Note
ESD| BS EN/EN61000-4-2| Level 4, 15KV air ; Level 4, 8KV contact
RF field susceptibility| BS EN/EN61000-4-3| Level 3, 10V/m( 80MHz~2.7GHz )
Table 9, 9~28V/m( 385MHz~5.78GHz )
EFT bursts| BS EN/EN61000-4-4| Level 3, 2KV
Surge susceptibility| BS EN/EN61000-4-5| Level 4, 4KV/Line-FG; 2KV/Line-Line
Conducted susceptibility| BS EN/EN61000-4-6| Level 3, 10V
Magnetic field immunity| BS EN/EN61000-4-8| Level 4, 30A/m
Voltage dip, interruption| BS EN/EN61000-4-11| 95% dip 1 periods, 30% dip 25
periods,
95% interruptions 250 periods
OTHERS| MTBF| 4050.3K hrs min. Telcordia SR-332 (Bellcore) ;
468.0K hrs min. MIL-HDBK-217F (25℃)
DIMENSION (LWH)| 76.250.828mm or 3″ 2″ 1.1″ inch
PACKING| 0.13Kg; 100pcs/14Kg/1.13CUFT
NOTE|
- All parameters NOT specially mentioned are measured at 230VAC input, rated load and 25 of ambient temperature.
- Ripple & noise are measured at 20MHz of bandwidth by using a 12″ twisted pair-wire terminated with a 0.1μf & 47μf parallel capacitor.
- Tolerance : includes set up tolerance, line regulation and load regulation.
- Derating may be needed under low input voltages. Please check the derating curve for more details.
- Touch current was measured from primary input to DC output.
- The ambient temperature derating of 3.5℃/1000m with fanless models and of 5℃/1000m with fan models for operating altitude higher than 2000m(6500ft).
- The power supply is considered a component which will be installed into a final equipment. All the EMC tests are been executed by mounting the unit on a 360mm*360mm metal plate with 1mm of thickness. The final equipment must be re-confirmed that it still meets EMC directives. For guidance on how to perform these EMC tests, please refer to “EMI testing of component power supplies.”
(as available on https://www.meanwell.com//Upload/PDF/EMI_statement_en.pdf )
※ Product Liability Disclaimer: For detailed information, please refer tohttps://www.meanwell.com/serviceDisclaimer.aspx
Block Diagram
Derating Curve
AMBIENT TEMPERATURE(c)
Output Derating VS Input Voltage
INPUT VOLTAGE (VAC) 60HZ
Mechanical Specification
AC Input Connector (CN1): JST 83P-VH or equivalent
Pin No.
|
Assignment
|
Mating Housing
|
Terminal
---|---|---|---
1
|
AC/L
|
JST VHR or equivalent
|
JST SVH-21T-P1.1 or equivalent
2
|
No Pin
3
|
AC/N
DC Output Connector (CN100): JST 84P-VH or equivalent
Pin No.
|
Assignment
| Mating Housing| Terminal
---|---|---|---
1,2
|
+V
|
JST VHR or equivalent
|
JST SVH-21T-P1.1 or equivalent
3,4
|
-V
- HS1 must have safety isolation distance with system case.
Note:
- RPS-120S model delivers EMI Class B for both conducted emission and radiated emission for the power supply, when configured into Class I (with FG) system.
- RPS-120S model delivers EMI Class 8 conducted emission and Class A radiated emission with King Core K58 RC (12157) in output cable for the power supply when configured into Class U (no FG) system.