Nidec FG-3000R Digital Force Gauge User Manual
- July 4, 2024
- Nidec
Table of Contents
FG-3000R Digital Force Gauge
Operation Manual
FG-3000R Digital Force Gauge
WARNING
Operators should wear protection such as a mask and gloves in case pieces or
components break away from the unit under test.
Whether the unit is ON or OFF, DO NOT exceed the capacity of the gauge. NEVER
exceed 150% of the rated capacity, or the load cell will be damaged. At 110%
of the rated capacity, the display will flash a warning.
When mounting FG-3000 Series Digital Force Gauges, use M4 mounting screws with
a maximum insertion depth of 7 mm into the gauge.
Hand tighten mounting screws, DO NOT use tools. Do not use damaged clamp.
Measure in line tension and compression forces only. DO NOT attempt to measure
forces at an angle to the measuring shaft – damage to load cell and/or shaft
may result.
Do not attempt to repair or alter this instrument. Warranty will be voided and
damage to the unit may result.
Use and store within the stated temperature and humidity ranges, or damage and
failure may result.
When using adapter measuring heads, do not use tools. Hand tighten only.
The new FG-3000R Series digital force gauges are the choice for simple, cost-
effective compression testing. Combining one of the most compact housings, yet
maintaining a large back-lit LCD, these units were designed to fit perfectly
in the hand for ease of use. The multi-language FG-3000’s provide menu
programming for intuitive set-up of the instrument to your desired
requirements.
Three modes of operation are selectable: Track mode displays live readings,
Peak mode records the maximum reading sensed during the test, and Pre-set mode
which activates user defined high and low limit set points. The programmable
limits provide a quick visual and audible indication if a test passes or
fails. In addition, a comparator output enables integration of the instrument
into your quality system for repetitive testing such as on production lines.
The integral button style load-cell attached via the flexible, metal braided
cable enables reading to be taken in machinery or fixtures where other force
gauge adapters are unable to reach or fit.
The display graphics facilitate user comprehension and operation. An analog
bar graph provides perspective of current reading in comparison to the full
scale range. Pass/Fail icons provide an instant response of the testing
outcome while a storage symbol acknowledges when a reading is logged.
A menu-selectable display orientation streamlines switching from push to pull
testing for portable or test stand applications.
SPECIFICATIONS
Accuracy: ± 0.5% F.S.
Selectable Units: kN, kgf, N, klbf and lbf. (Depending on Range)
Overload Capacity: 150% of F.S. (LCD flashes beyond 110% of F.S.)
Measurement Method: Peak, Track, Preset
Data Sampling Rate: 1000 Hz
Display: 160*128 dot matrix LCD with Backlight
Display Update Rate: 10 times/second
Resolution: (See chart)
Memory: 500 data
Set Point: Programmable high and low limits in Preset Mode
Battery Indicator: Display flashes battery icon when battery is low
Power: 3.6 VDC 400 mAH Ni-MH rechargeable batteries
Battery Life: Approximately 16 hours continuous use per full charge
Charger / Adaptor: Universal USB/BM charger, Input: 110 ~ 240VAC
Temperature Effects: <0.054% per °F (0.03% FS per °C)
Outputs: USB, RS-232; High & Low Limit NPN’s
Operating Temperature: 14 to 104°F (-10 to 40°C)
Storage Relative Humidity: 20 to 80%
Housing: Aluminum
Storage Temperature: -4 to 122°F (-20 to 50°C)
Oper. Relative Humidity: 5 to 95%
Dimensions: 5.5 x 2.8 x 1.4” (140 x 71 x 35.5 mm)
Product Weight: 1.1 lb (0.5 kg)
Package Weight: 2.8 lb (1.3 kg)
Warranty: 1 year
Certification: CE
Included Accessories: AC Adaptor/Charger, USB cable, calibration cert.,
carrying case
LCD Screen
- Battery icon: Battery level or charging status. Flashes when gauge needs to be recharged.
- OK/OV Indicator: Measured value between low limit and upper limit; Value over upper limit Value between lower limit and 75% of lower limit
- Force icon: Indicates force direction. Tension Compression
- Test mode icon: Three measurement modes: Track, Peak and Preset
- Current meaured value
- Analog bar: Indicates current position within full scale. When the bar enters the area enclosed by the dotted line, it means full scale capacity is exceeded and overload.
- Storage icon: Indicates data is being saved.
- System time
- Units Indicator: Selected engineering unit.
- Data Transmission icon
OPERATION
1.1 Key Functions
All keys are capacitive touch.
| ON/OFF: Push for 1 second to power On or Off
---|---
| During Measurement: Store data.
In Menus: Back or quit.
| During Measurement: Enter the menus.
In Menus: Select or Enter
| During Measurement: Track mode, tares weight of attachment. In Peak modes,
resets the peak value.
In Menus: Moves selection up or increases the value.
| During Measurement: Changes the measure mode from Track, Peak or Preset
In Menus: Moves selection down or decreases the value.
1.2 Modes
Track: Real time, live measuring mode.
Peak: Peak readings will not change until a higher value is measured.
Preset:
User-defined set points GO/NG testing with available and visual indicators.
1.3 Menu Structure
The FG-3000R Series Force Gauge has multi-level menu interface (Table 1-3)
that enables simple navigation and programming.
MENUS | SUBMENUS | SELECTIONS |
---|---|---|
Measurement | Unit | kN, N, kgf, lbf, klbf |
Test Mode | Track, Peak, Preset | |
Memory | Browse | |
Selected, All | ||
Delete All | Yes, No | |
System | Display | Obverse, Reverse |
Auto Power | On, Off | |
Backlight | On, Off | |
Key Sound | On, Off | |
Date/Time | ||
Calibration | Yes, No | |
Default | Yes, No | |
Language | English, Chinese, German | |
Information | Range SN, Version |
PREPARATION
2.1 Confirm the model
This series force gauge has 5 ranges available, each model corresponding to
the capacity and resolution shown on the last page of this manual. Select the
appropriate model you need before use.
SETUP
3.1 Measurement
The Measurement menu contains the Unit of measure and Measurement Mode sub-
menus, as shown in Fig. 3-1.
3.2 Select Units
The measuring units can be selected under this menu. Different range models
may have different unit selection capabilities. Touch “ZERO” or “MODE” keys to
shift to the next selection. Press “LOG” to cancel or touch “MENU” to confirm
and exit. (Fig. 3-2)
3.3 Select Test Mode
The FG-3000R has 3 types of Test Modes.
Track: The real time measuring mode. Under this mode, press the ZERO key to
tare any initial reading being displayed.
Peak: In Peak mode, the maximum force will be recorded and displayed. Press
the ZERO key to reset the peak value.
Preset: Enables the setting of an upper and lower limit to compare to the
measured force value.
A simple GO/NG analysis is displayed on screen via icon indicators for quick
pass/fail testing.
To guarantee an accurate test, make sure to zero the display and tare any
small force being displayed before beginning the test.
There are two means to select your appropriate Test Mode. At the home screen
simply press the MODE key to scroll through the three measuring modes.
You can also select the mode under the Measurement menu in the Test Mode sub-
menu. See Fig. 3-3(a)
If the Preset is selected, a new screen will pop up where you can set the
Upper and Lower limits. See Fig 3-3(b)
Press ZERO to adjust the number and press MODE to move to the next digit.
Note:
- The upper limit can not exceed 110% capacity of the force gauge.
- The lower limit must not be less than 10% of capacity.
- The upper limit must exceed the lower limit
SAVING THE MEASURED VALUE
Measured results can be stored in the force gauge’s memory. You can review or
print the stored data at a later time.
At the home screen press the LOG key to store a value. The storage icon will
be displayed.
The data stored is the current displayed force value in Track and Preset
modes. In Peak mode it is the peak value shown on the display.
4.1 Memory
Memory menu contains three submenus: Browse, Print, Delete all, as shown in
Fig. 4-1.
You can browse stored data or print all the data via the FG-PRINTER mini-
printer (sold separately). You may also delete all the records in the Delete
all sub-menu.
4.2 Browse
In the Browse sub-menu. The data in memory can be reviewed in the order saved
which is oldest to newest. See Fig. 4-2(a) Press ZERO/UP or MODE/DOWN to
scroll.
Press MENU. A small window will pop out. Here you can select Delete or Print.
See Fig. 4-2(b).
If you select Delete, a confirm window will appear asking you to confirm.
Press MENU to confirm or LOG to exit.
4.3 Print
You can print the data in memory. Enter Print. (Fig. 4-3) Choose Selected or
All.
If Selected is chosen, the total Range of available data points will be
indicated. Adjust the value points to be printed to the right of Select. Fig.
4-3(a)
If All is selected, a confirm window will appear asking you to confirm. See
Fig. 4-3(b).
4.4 Delete All
All data points can be cleared from memory under the Delete all sub-menu (Fig.
4-4). A confirm window will appear asking you to confirm. See Browse for
details on deleting individual points one at a time.
SYSTEM
Under the System menu, the Display, Auto Power, Backlight, Key Sound, Date/Time, Calibration and Default sub menus are present.
5.1 Display
There are two display modes: Obverse and Reverse (Fig. 5-1(a)).
Obverse will allow the display to be up-right with the keypad underneath,
while Reverse will allow the display to be up-right with the keypad above.
Fig. 5-1(b)
5.2 Auto Power
The FG-3000R has an automatic power off function. With Auto Power on, if there
is no operation performed within five minutes it will power off automatically.
(Fig. 5-2)
5.3 Backlight
The backlight can be set to turn on or off. See Fig.5-3. Choosing the
backlight to be off will reduce the consumption of the battery.
5.4 Key Sound
The Key Sound can be turned on or off as shown in Fig. 5-4.
5.5 Date/time
Date and time can be adjusted under this menu. Press ZERO to adjust the number
and press MODE to move to the next digit. Fig. 5-5
5.6 Calibration
Because of the sensor material performance or the influence of external
factors, there may be errors of a certain level after a period of usage.
It is recommended to send the force gauge to a specialized testing
organization for calibration.
If you have standard force weights or other standard load and a test stand,
you may utilize this function and procedure to calbrate the sensor.
-
Mount the force gauge.
-
Use the tare by use of the ZERO key.
-
Enter Calibration sub-menu as in Fig. 5-6(a).
The calibration interface is shown in Fig. 5-6(b). -
Load a standard force. Now the value in the standard input area is just equal to the current measured value. Wait a moment for the force to stabilize.
-
Press ZERO and MODE to input the standard force value.
-
Press MENU to enter the next calibration. Press LOG to interrupt the calibration.
When 3 calibration points have been finished, a confirm window will pop up
asking to “Save and Exit”(YES)/(NO). Fig. 5-6(d)
Press ZERO or MODE to select, then press MENU.
If “YES” is selected, Calibration is complete. Fig. 5-6(e)
5.7 Default
With this function, the force gauge can be restored back to the original
factory settings. Only perform this function when all other troubleshooting
tactics have first been attempted.
LANGUAGE
The force gauge can display in various languages. Set the language as desired. See Fig. 6.
INFO
Information about the force gauge such as range, version and serial number is provided in this menu. Fig. 7
COMMUNICATION PORT
The force gauge has a USB for recharging and communicating with a PC, plus an 8 pin connection for printer connection and set point output. Fig. 8-1
The RS232 serial port is used to connect the mini-printer to print the memory
data stored on the gauge.
RS-232 Specifications:
-Hardware Flow Control: None
-Data word length: 8 bits
-Stop bit: 1bit
-Parity: None
-Baud rate: 38400
8.2 Setpoint Output
Two NPN open collector setpoint outputs are available.
The internal circuit of the setpoint output is shown as Fig 8-2.
Pin7 with Pin6 will be connected when an overload alarm occurs.
In Preset Mode, Pin7 to Pin6 is connected when the measured value exceeds the
upper limit. Pin4 to Pin6 is connected when the measured value passes below
the lower limit.
CAUTION: Maximum permissible voltage: pin 7 to 6, pin 4 to 6 must be
lower than 35V ; pin 6 to 7, pin 6 to 4 must be lower than 6V .
Remember to remove the load after measurement. Applying a load for a long time
period may affect the accuracy of the instrument.
MAINTENANCE
9.1 Maintenance
After use, please keep the instrument body clean. Do not let oil and other
substances persist on the body and screen so as not to damage the instrument.
Remember to remove the load after measurement. Applying a load for a long time
period may affect the accuracy of the instrument.
9.2 Charging
When the battery is low, the icon “ ” will be displayed. The batteries should
be charged immediately.
Connect the gauge and the charger with the USB cable. Then connect the charger
with AC socket to start charging.
TROUBLESHOOTING
According to the following table, review possilbe solutions for problems encountered. Do not disassemble the gauge by yourself or attempt to repair. If you cannot resolve the fault yourself, please contact Nidec-Shimpo.
Failure | Possible Causes | Potential Solutions |
---|---|---|
Unit will not turn on | Low battery | Recharge and then reboot. If after 3-4 |
hours of charging time the battery does not properly hold a charge, the
battery needs to be replaced. Contact Nidec-Shimpo.
No key sound| Key sound is turned off| Turn on the key sound in menu
No backlight| Backlight is turned off| Turn on the backlight in menu
Error is too large| The gauge is not calibrated| Calibration of force gauge is
required. After calibration if the error remains outside of the
specifications, sensor may be damaged. Contact NidecShimpo to get RMA for
return.
CAPACITY AND RESOLUTION
Model | N | kN | kgf | lbf | klbf |
---|---|---|---|---|---|
FG-3000R-1 | Capacity | 1000 | 1 | 100 | 220 |
Resolution | 1 | 0.001 | 0.01 | 0.01 | |
FG-3000R-5 | Capacity | 5000 | 5 | 500 | 1100 |
Resolution | 1 | 0.001 | 0.1 | 1 | |
FG-3000R-10 | Capacity | 10 | 1000 | 2250 | |
Resolution | 0.01 | 1 | 1 | ||
FG-3000R-20 | Capacity | 20 | 2000 | 4500 | |
Resolution | 0.01 | 1 | 1 | ||
FG-3000R-50 | Capacity | 50.0 | 5000 | ||
Resolution | 0.01 | 1 | 0.01 |
Model | D | H1 | H2 | d | L |
---|---|---|---|---|---|
FG-3000R-1 | 32 | 23 | 21. | 8 | 222 |
FG-3000R-5
FG-3000R-10
FG-3000R-20| 38| 24| 22| 11| 225
FG-3000R-50