Nidec TNC Series Torque Cap Tester User Manual
Warning
Operators should wear protection such as a mask and gloves in case pieces or components break away from the unit under test. Whether the unit is ON or OFF, DO NOT exceed the capacity of the sensor. NEVER exceed 120% of the rated capacity, or the torque sensor will be damaged. At 110% of the rated capacity the display will flash a warning. Measure in line torque only. DO NOT attempt to measure forces at an angle to the sensor – damage to sensor may result. Do not attempt to repair or alter this instrument. Warranty will be voided and damage to the unit may result. Use and store within the stated temperature and humidity ranges, or damage and failure may result. Ensure during testing unit is properly mounted & secured to stable sur- face. Mount display with the two integral 8.5 mm mounting holes only. If not using this instrument for extended periods of time, remove the batteries to prevent potential battery leakage from causing product damage.
The TNC Digital Torque Cap Tester is a simple operation test in- strument for quickly measuring Open and Close torque values up to 14.8 ft-lbs (20 N-m). The TNC meter is used in medical, pharmaceutical, industrial quality control, R&D and countless oth- er torque applications.
The TNC’s provide long operation life and power flexibility with the ability to work from the internal rechargeable battery or included AC adapter.
The TNC’s have six modes of operation: Track for live readings, Peak which records the highest level over a test, First Peak which records a reading after a programmed decrease, Auto Peak which resets after a period of time, Auto First Peak which resets the first peak after a period of time, Double Peak which provides two separate peak readings from programmed decreas- es. Preset operation with programmable tolerance thresholds for quick pass fail test is available in all modes.
The analog bar graph on the backlit, reversible LCD aids users by providing the resultant torque’s current position compared to the full scale range along with the direction of the force. Pass/fail icons produce live test feedback allowing instant determination of the results.
Every tester comes standard with USB and RS-232 outputs. The TNC’s software enables the uploading of data and the additional statistical analysis with its auto-calculation of the selected values.
These excellent features make the TNC Torque Cap Testers a valuable and versatile addition to the production, R & D and qual- ity control departments.
SPECIFICATIONS
- Measuring Range: See Capacity & Resolution chart on last page
- Accuracy: ± 0.3% of reading over 5 to 100% capacity
- Test Sample Diameter Range: 0.77” – 7.48” (19.5 – 190 mm)
- Units of Measure: N-cm, N-m, in-lb, ft-lb, kg-cm (depending on range)
- Measure Modes: T rack, Preset, Peak, First Peak, Auto Peak, Auto First Peak, Double Peak
- Overload Protection: 120% of Full Scale
- Sampling Rate: 2000 Hz
- Display: 160* 128 Dot matrix backlit LCD
- Display Update Rate: 10 times/sec
- Resolution: See Capacity & Resolution chart on last page
- Memory: 1000 data points
- Set Point: Programmable High and Low Limits in Preset mode
- Output: USB, serial port RS-232, High and low limit NPN Battery
- Indication: Battery icon flashes when low
- Battery Life: Approx. 25 Hours continuous on full charge
- Charger/Adapter: Universal USB charger, input 110 – 240 V ac 50/60 Hz
- Power: 3.6 V dc 1600 mAH Ni-MH battery pack Operating
- Temperature: 14 to 104ºF (-10 to 40ºC) Humidity Limit: 20 – 80% RH
- Dimensions: 7.75 x 7.87 x 5.59” (197 x 200 x 142 mm)
- Product Weight: 5.5 lb (2.5 kg)
- Package Weight: 7.7 lb (3.5 kg)
- Warranty: 1 year
- Certification: CE
- Included Accessories: 4 torque pins, USB cable, charger adapt- er, 2 mounting bolts, table-top rubber pads, cal. cert.
- Optional Accessories: RS-232 cable, printer
LCD SCREEN STANDARD VIEW
Test Mode Icons:
| Track: Real Time, live measuring mode
---|---
| Peak: Reading will not change until a higher value is mea- sured
| First Peak: Captures First Peak after a decrease has been detected.
Drop Ratio set in menus.
| Auto Peak: Resets Peak after a programmed time period
| Auto First Peak: Resets the First Peak after a programmed time period.
First Peak drop Ratio set in menus
| Double Peak: Captures two First Peak values. Both drop ratios set in
drop ratio menu
- Battery Icon: Battery level or charging status. Flashes when gauge needs to be recharged
| Preset: Set the upper & lower limit for GO/NG testing Activated only in the Measurement menu.
---|--- - OK/OV Preset Indicator:
****| Under Lower Limit
---|---
| Between Low Limit & Upper Limit
| Over Upper Limit - Torque Icons: Indicates force
| Clockwise (CW) Dotted line area indicates CW overload
---|---
| Counterclockwise (CCW) Dotted line area indicates CCW overload - Current measured value
- Analog Bar: Indicates current position within full scale. When the bar enters the area enclosed by the dotted line, this signifies the full scale capacity is exceeded by an overload
- Storage Icon: Indicates data is being
- System time
- Units Indicator: Selected engineering
- Statistics X : Average; R: Variance
- Data list: Five Point Measurement in Peak & First Peak Modes
- Key Setting Mode: Printer Icon Key press will Disc icon key will store reading.
- Double Peak Mode Percentages
MODE DIAGRAM DETAILS
Digital View
Graphic View
Double Peak Mode View
KEY FUNCTIONS
POWER: Push for 2 seconds to power On or Off SAVE/PRINT: Print the current
force value or
storedata, depending on the key setting. (See 7 for key setting) Changing
Values: Select next digit location.
MENU/ENTER: Enter the menus. In Menus: Select or Enter
ZERO/EXIT: Zeros Values In Menus: Back or Exit
MODE/UP: Changes the measurement mode In Menus: Moves selection up or
increases the value.
DOWN: Moves selection down or decreases the value.
MENU STRUCTURE
Menu | Measurement | Unit |
---|
Group
Test Mode
Preset
Memory| Browse
Print All
Print Select
Upload
Delete Select
Delete All
System| Display Mode
Auto Power Off
Backlight
Key Tone
Language
Date/Time
Diagram Scale
Key Setting
Display Direction
Default
Information
Calibration| Torque
Sensitivity
Weight
Force
ADVANCED MENU OPTIONS
Measurement
The Measurement menu contains four selectable items: Unit, Group, Test Mode, Preset.
From the home screen, touch “MENU/ENTER” to enter the main menu.
Unit
The measuring unit can be selected under this menu. Different range models may have different unit selection capabilities. Touch “MOVE/UP” or “DOWN” keys to shift to the next selection. Press “ZERO/EXIT” to cancel or touch “MENU/EXIT” to confirm and exit.
Group
When several test samples need to be measured, the samples can be coded into
groups. The range is 01-99. When set to “00”, changes to “01 automatically as
“00” is not an available choice. Press “MODE/UP” to adjust the value,
touch “PRINT/SAVE” to shift to the next position. Touch “ZERO/EXIT” to
cancel; press “MENU” to confirm and exit.
Test Mode
Change the mode of operation between the six modes. Press “MODE/UP” or
“DOWN” keys to select. Press “ZERO/EXIT” to cancel or “MENU/ENTER”
to confirm and exit. This adjustment can also be changed at the home screen
display by simply press- ing “MODE”. First Peak Mode will display a drop
ratio setting menu. This drop ratio actives the first peak recording. Auto
Peak Mode will display a Peak Time setting menu. This resets the peak value
after programmed time has elapsed. Double Peak Mode will display a dual drop
ratio settings menu. This will enable two separate peak values to be recorded.
Preset
In the Preset menu, program high and low limit values to enable ok/ov testing.
The lower limit value cannot be greater than the upper limit value, and
neither limit value can be greater than 110% of the rated capacity. Select
proper mode to have presets active. Press “MODE/UP” to adjust the value,
touch “SAVE/PRINT” to shift to the next position. Press “ZERO/EXIT” to
cancel; touch “MENU/ENTER” to confirm and exit.
MEMORY
In the Memory menu, the user can browse, delete, or print the data.
Browse All
The data will be displayed. Touch “MODE/UP” or “DOWN” keys to go to
the next position. Touch “MENU/ENTER” to see additional details of the
data. Touch “ZERO/EXIT” to go back.
Print All
In the Print All Menu the data saved in memory can be output to a printer
through the serial port RS232 connection. (See 10.3 RS232 in Communications
section for more informa- tion)
Print Selected
In this menu, select the data range to print. Touch “MODE/UP” to adjust
the value, touch “SAVE/PRINT” to shift to the next position. Press
“ZERO/EXIT” to cancel; touch “MENU/ENTER” to confirm.
Upload
- Choose in EDMS software the ‘Upload Memory Data From Gauge’
- On the TNC screen, select The screen should display ‘Upload…’ as shown below.
- While ‘Upload…’ is visible on the screen, click ‘Start Acquisition’ on the EDMS
- Data should populate the table on the right hand side of the software screen
Delete Selected
Select the range of data to be deleted. Touch “MODE/UP” to ad- just the
value. Press “SAVE/PRINT” to shift to the next position. Touch
“ZERO/EXIT” to cancel; touch “MENU/ENTER” to confirm.
Delete All
In this menu, a prompt will appear. All data will be deleted by selecting
“YES” and canceled by selecting “NO” or pressing “ZERO/EXIT”.
SYSTEM
Under the System menu, several parameters may be set.
Display Mode
Two display modes may be selected: Digital and Graphic.
Auto Power Off
To maximize battery life, the power can be set to shutdown after non-use. The
time can be set in this menu. The range is 01-99 minutes. When set to “99” the
gauge will never turn off. Touch “MODE/UP” to adjust the value, touch
“SAVE/PRINT” to shift to the next position. Press “ZERO/EXIT” to
cancel; Push “MENU/EN- TER” to confirm and exit.
Backlight
Under this menu, the backlight can be set to ON, OFF or have an auto
shutdown time. Touch “MODE/UP” or “DOWN” keys to shift to the next
position. Press “ZERO/EXIT” to cancel. Press “MENU/ ENTER” to confirm
and exit.
Key Tone
Turn the key sound ON or OFF. Touch “MODE/UP” or “DOWN” keys to shift
to the next position. Touch “LOG” to cancel; Press “MENU/ENTER” to
confirm and exit.
Language
Select between English, German and Chinese.
Date/Time
The system time may be set under this menu. Touch “MODE/ UP” to adjust the
value. Press “SAVE/PRINT” to shift to the next position. Touch
“ZERO/EXIT” to cancel. Press “MENU/ENTER” to confirm and exit.
Diagram Scale
Select the visible are of the graph when in diagram mode in this menu. Select
between 3 levels.
Key Setting
Set the default function of the “SAVE/PRINT” key from the home screen. The
function can be set to print or store the current displayed value. Press
“MODE/UP” or “DOWN” to select the proper setting. Press “LOG” to
cancel; touch “MENU/ENTER” to confirm and exit.
Display Direction
Select the mode of the LCD display: Automatic, Obverse or Re- verse. Touch
“MODE/UP” or “DOWN” keys to shift to the next po- sition. Press
“ZERO/EXIT” to cancel; Push “MENU/ENTER” to confirm and exit.
Default Setting
If you make a selection that you feel has caused the gauge to operate
improperly, you can restore the instrument to the facto- ry default settings.
Carefully use this function! The password to accomplish this factory default
is 1, 2, 3. A confirm screen will appear to complete.
Information
Information includes the model, version of the software and the serial number.
CALIBRATION
To Obtain access to the calibration sub menus, input the pass- word 1, 2, 3.
Four methods of calibration may be selected.
Torque Calibration
If applying a known standard torque such as with a calibrated torque tool or torque test machine apparatus, use this mode and apply the torque to the TNC.
Input in the standard values by pressing “MODE/UP” to adjust the value,
touch “SAVE/PRINT” to shift to the next position. Press “MENU/ENTER”
to view the next available standard value input screen.
After the last standard value input screen, Press “MENU/ENTER” to view the
confirmation selection. Here you can select YES or NO with the “MODE/UP”
or “DOWN” keys. Press “MENU/ENTER” to confirm the desired selection.
Sensitivity Calibration
Sensitivity Calibration is specific to higher capacity models where the sensor
has a mV/V value equal to a torque value. Input in the values by pressing
“MODE/UP” to adjust the value, touch “SAVE/ PRINT” to shift to the next
position. Press “MENU/ENTER” to view the confirmation selection. Here you
can select YES or NO with the “MODE/UP” or “DOWN” keys. Press
“MENU/ENTER” to con- firm the desired selection.
Weight Calibration
In this mode, “Weights” means mass weight. When a lever and a calibrated mass weight are used creating a standard torque, this mode calculates the value properly to calibrate the instrument. The torque generated by weights on a lever (L) is associated with the local gravity acceleration (g) that the user inputs.
Input in the standard lever length (L), gravity (g) and weight val- ues by
pressing “MODE/UP” to adjust the value, touch “SAVE/ PRINT” to shift
to the next position. Press “MENU/ENTER” to view the next available
standard value input screen.
After the last standard value input screen, Press “MENU/ENTER” to view the
confirmation selection. Here you can select YES or NO with the “MODE/UP”
or “DOWN” keys. Press “MENU/ENTER” to confirm the desired selection.
Force Calibration
When using a standard force (or standard calibrated force via a weight) and a
lever (L), select Force Calibration. The standard force applied to the lever
may be produced by force from weights, a test machine or force gauge that
applies a standard force.
Input in the standard lever length (L) and force values by pressing “MODE/UP”
to adjust the value, touch “SAVE/PRINT” to shift to the next position. Press
“MENU/ENTER” to view the next available standard value input screen.
After the last standard value input screen, Press “MENU/ENTER” to view the
confirmation selection. Here you can select YES or NO with the “MODE/UP” or
“DOWN” keys. Press “MENU/ENTER” to confirm the desired selection.
CHARGING
Metal Hydride AAA rechargeable batteries, which are supplied fully charged to allow immediate use. Users need to recharge bat- teries when a low battery icon flashes. Users should connect the gauge and the charger using the USB cable. Then connect the charger to an AC socket to start charging. Laptops and other USB devices can also charge the gauge. A fully charged battery pack will provide approximately 25 hours of constant use.
Rechargeable battery pack:
- Type: Ni-MH 6VDC 1600 mAH rechargeable batteries
- Charging time: 3~4 hours
- Battery life: approx. charge and discharge 500 times
COMMUNICATIONS
USB
The TNC Series Digital Torque Cap Tester is designed in accor- dance with
USB2.0 standard protocol. (Figure 10-1) The USB Port can be connected to a
charger with the USB cable for charging the internal Ni-MH battery and can be
connected to a computer for uploading the measured values. Connect the gauge
and the computer with the USB cable, then open the computer software. Upload
the values. Please refer to the software manual for addi- tional information.
Port Pin Assignments
PIN# | Definition |
---|---|
1 | RS232-Transmit (Tx) |
2 | RS232-Receive (Rx) |
3 | RS232-Ground |
4 | Output A+ |
5 | Reserved |
6 | Output Com- |
7 | Output B+ |
8 | Reserved |
RS232
The RS232 serial port is used to connect a printer to print the memory data.
RS-232 specifications are as follows:
- Data transmission: serial interface
- Synchronization: asynchronous
- Signal Level: RS-232 level, logic 1:-5v, logic 0: +5v
- Hardware Flow Control: None
- Data word length: 8 bits
- Stop bit: 1bit
- Parity: None
- Baud rate: Autodetect 38400, 19200 or 9600
Comparison Output
Comparison Output internal circuit shown as below.
When the measured value is less than the lower limit tolerance value, the “pc2” operates, 7pin and 6pin line conduction. When the measured value is more than the upper limit tolerance value or 110% of the rated capacity, the “pc1” operates, 4pin and 6pin line conduction. Maximum permissible voltage: 7pin to 6pin, 4pin to 6pin 35V; 6pin to 7pin, 6pin to 4pin 6V.
MISC
Dimensions
Parts List
Items not shown: Mounting bolts, rubber sticker foot pads
CAPACITY & RESOLUTION
Range by Unit (Resolution)
Model| N-m| N-cm| kgf-cm| in-lbf| ft-lbf
TNC-2| 0.1- 2 (0.0002)| 10-200 (0.02)| 1-20 (0.002)| 0.89-17.7 (0.002)|
0.07-1.5 (0.0002)
TNC-5| 0.25 – 5 (0.0005)| 25-500 (0.05)| 2.5-50 (0.005)| 2.2-44.3
(0.005)| 0.18-3.7 (0.0005)
TNC-10| 0.5-10 (0.001)| 50-1000 (0.1)| 5-100 (0.01)| 4.4-88.5 (0.01)|
0.37-7.4 (0.001)
TNC-20| 1-20 (0.002)| 100-2000 (0.2)| 10-200 (0.02)| 8.9-177 (0.02)|
0.7-14.8 (0.002)