medela Invia Motion Quick Card
The Invia Motion Quick Card does not replace the Invia Motion instructions for use. For information concerning intended use, indications, contraindications, warnings and cautions please refer to the appropriate Invia Motion Clinician instructions for use and Invia Motion Patient instructions for use included with the Invia Motion pump or at www.medela.com/healthcare
**Overview
**
Positioning and wearing Invia Motion Negative Pressure Wound Therapy (NPWT) pump.
If placed in an inappropriate position or when canister is full, “Canister full” notification will appear and Invia Motion canister/tubing set must be changed to continue the therapy.
Charge battery for the first time before starting the therapy (approximate charging time is 4 hours). Recharge battery throughout the length of the therapy. The expected battery time generally exceeds 10 hours.
Information on the display
Description
- Pump is running/On-state in intermittent mode (Symbol is blinking)
- Constant mode
- Negative pressure setting in mmHg
- Battery charge level (symbol steady)
- Battery charging (bars moving)
- Off-state in intermittent mode (Symbol is blinking)
- Intermittent mode
- Battery fully charged
- Info used time – shows runtime already used in days ( ) and hours ( ) (Example for pump that has operated for 6 days and 23 hours). Not applicable for Invia Motion Endure.
Change Invia Motion canister / tubing set
Acoustic signals are muted by pressing the button.
-
Wash hands and prepare clean paper towel or tissue.
-
Remove the Invia Motion NPWT pump from the carrying case. Switch the Invia Motion NPWT pump off by pressing and holding the button for 3 seconds. Press only once.
-
Position the clamp next to the quick-connector and close the clamp on the pump tubing and also close the clamp on the dressing tubin.
-
Disconnect the Invia Motion NPWT pump tubing from the dressing tubing by pressing on the sides of theQuick-connector. Use a paper towel or tissue to cover the tip of the used tubing to protect against wound fluid leakage.
-
Detach the full canister/tubing set by rotating the canister to the left.
-
Wash hands.
-
Obtain and open a new Invia Motion canister/tubing set package.
CAUTION- Do not use an Invia Motion canister/tubing set if the sterile packaging is damaged or has expired.
- Make sure the correct canister is used: Ref 087.4016.
- Contact your healthcare professional if the correct canister is not available.
-
Remove the canister/tubing set from the packaging and attach it to Invia Motion NPWT pump.
-
Connect the Invia Motion NPWT pump tubing to the dressing tubing. Push together until you hear a click.
-
Open the clamp on the dressing tubing. Also make sure the clamp on the pump is open.
-
Switch the Invia Motion NPWT pump on by pressing the button. Slide the Invia Motion NPWT pump into the carrying case.
-
Ask your clinician how to properly dispose of a used canister/tubing set.
Notification
Symbol | Description | Troubleshooting |
---|
blinking + acoustic signal
|
Leak in system
|
Check the system for leaks – refer to instructions for use, chapter Notifications and troubleshooting
blinking + acoustic signal
|
Blockage
|
Check the system for blockage – refer to instructions for use, chapter Notifications and troubleshooting
blinking + acoustic signal
|
Canister full
|
Change the Invia Motion canister/tubing set – refer to description in chapter “Change Invia Motion canister/tubing set”
blinking + acoustic signal
|
Battery low
|
Connect the Invia Motion pump to the power supply – refer to the instructions for use, chapter Notifications and troubleshooting
blinking + acoustic signal
|
Battery fully discharged
|
Connect the Invia Motion pump to the power supply – refer to the instructions for use, chapter Notifications and troubleshooting
blinking + acoustic signal
|
Battery missing
| Ensure that battery compartment cover is mounted and closed properly – refer to the instructions for use, chapter Notifications and troubleshooting / Overview notifications
blinking (example for 15 days
version)
|
Pump time expired
|
Dispose of the Invia Motion NPWT pump. Attach another device if necessary – refer to the instructions for use, chapter Notifications and troubleshooting
acoustic signal
|
Error (Internal fault)
| Switch the Invia Motion NPWT pump off by pressing the button and holding for 3 seconds.
Patients: contact your healthcare professional immediately for further instructions. Refer to the instructions for use, chapter Notifications and troubleshooting / Overview notifications
CAUTION : U.S. Federal law restricts this device to sale by or on the
order of a physician. Medela, Invia and Invia Motion are registered trademarks
of Medela Holding AG, Switzerland.
Medela AG
Lättichstrasse 4b
6341 Baar, Switzerland
www.medela.com
References
- Medela Healthcare US | Medela US icon_question
- Homepage Medela | Medela
- Homepage Medela | Medela
- Breastfeeding & Pumping
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>