Invia® Motion™ Quick Card

The Invia Motion Quick Card does not replace the Invia Motion
instructions for use. For information concerning intended use, indications, contraindications, warnings and cautions please refer to the
appropriate Invia Motion Clinician instructions for use and Invia Motion
Patient instructions for use included with the Invia Motion pump or at
www.medelahealthcare.com.

Charging Instructions Fully charge battery for the first time before starting the therapy.
Charging time can be up to 4 hours.
Recharge battery throughout the length of the therapy. The expected
battery time generally exceeds 10 hours.

Overview

Positioning and wearing Invia Motion Negative Pressure Wound Therapy
(NPWT) pump

  • Upright
    Positioning and Wearing

  • Laying with display side up

  • Upside down
    Positioning and Wearing

  • Laying with display side down
    Positioning and Wearing

  • Shoulder strap
    Positioning and Wearing

  • Belt strap
    Positioning and Wearing

If placed in an inappropriate position or when canister is full, “Canister full” notification will appear
and Invia Motion canister/tubing set must be changed to continue the therapy.

Information on the display

Symbol Description
Pump is running/On-state in intermittent mode (Symbol is blinking)
Constant mode

Negative Pressure Setting

| Negative pressure setting in mmHg
| Battery charge level (symbol steady) Battery charging (bars moving)
| Off-state in intermittent mode (Symbol is blinking)
| Intermittent mode
| Battery fully charged
|  – shows runtime already used in days () and hours () (Example for pump that has operated for 6 days and 23 hours).

Not applicable for Invia Motion Endure.

Notification

Symbol Description Troubleshooting

blinking + acoustic signal

|

Leak in system

|

Check the system for leaks – refer to instructions for use, chapter Notifications and troubleshooting

blinking + acoustic signal

|

Blockage

|

Check the system for blockage – refer to instructions for use, chapter Notifications and troubleshooting

blinking + acoustic signal

|

Canister full

|

Change the Invia Motion canister/tubing set – refer to description in chapter “Change Invia Motion canister/tubing set”

blinking + acoustic signal

|

Battery low

| Connect the Invia Motion pump to the power supply and fully charge battery. Charging time can be up to 4 hours – refer to the instructions for use, chapter Notifications and troubleshooting

blinking + acoustic signal

|

Battery fully discharged

|

Connect the Invia Motion pump to the power supply – refer to the instructions for use, chapter Notifications and troubleshooting

blinking + acoustic signal

|

Battery missing

| Ensure that battery compartment cover is mounted and closed properly – refer to the instructions for use, chapter Notifications and troubleshooting / Overview notifications

blinking (example for 15 days

version)

|

Pump time expired

|

Dispose of the Invia Motion NPWT pump. Attach another device if necessary – refer to the instructions for use, chapter Notifications and troubleshooting

acoustic signal

|

Error (Internal fault)

| Switch the Invia Motion NPWT pump off by pressing the  button and holding for 3 seconds.

Patients: contact your healthcare professional immediately for further instructions. Refer to the instructions for use, chapter Notifications and troubleshooting / Overview notifications

Change Invia Motion canister/tubing set

Acoustic signals are muted by pressing the Mute Button button.

  1. Wash hands and prepare clean paper towel or tissue.
  2. Remove the Invia Motion NPWT pump from the carrying case. Switch the Invia Motion NPWT pump off by pressing and holding the button for 3 seconds. Press only once.
  3. Position the clamp next to the quick-connector and close the clamp on the pump tubing and also close the clamp on the dressing tubing.
    Clamp open

Clamp closed

  1. Disconnect the Invia Motion NPWT pump tubing from the dressing tubing by pressing on the sides of the Quick-connector.
    Use a paper towel or tissue to cover the tip of the used tubing to protect against wound fluid leakage.
    Dressing TubePump
Tube

  2. Detach the full canister/tubing set by rotating the canister to the left.

  3. Wash hands.

  4. Obtain and open a new Invia Motion canister/tubing set package.
    CAUTION
    Do not use an Invia Motion canister/tubing set if the sterile packaging is damaged or has expired. Make sure the correct canister is used: Ref 087.4016. Contact your healthcare professional if the correct canister is not available.

  5. Remove the canister/tubing set from the packaging and attach it to Invia Motion NPWT pump.
    Canister tubing
set

  6. Connect the Invia Motion NPWT pump tubing to the dressing tubing. Push together until you hear a click.
    Pump and Dressing tubeCanister tubing
set

  7. Open the clamp on the dressing tubing. Also make sure the clamp on the pump is open.

  8. Switch the Invia Motion NPWT pump on by pressing the button. Slide the Invia Motion NPWT pump into the carrying case.

  9. Ask your clinician how to properly dispose of a used canister/tubing set.