MOBI 70117 Air Non-Contact Forehead Thermometer Instruction Manual

June 6, 2024
MOBI

**

mobi logo**

MOBI 70117 Air Non-Contact Forehead Thermometer

MOBI 70117 Air Non-Contact Forehead Thermometer

Item No: 70117
Please read this instruction manual completely before operating

Package Contents

Parts

  • Air Non-Contact Forehead Thermometer  * 1pc
  • 1.5V (AAA) Batteries  *  2pc
  • User Manual * 1pc

Product Specification

MOBI 70117 Air Non-Contact Forehead Thermometer fig1

  1. Distance Sensor
  2. Infrared Sensor
  3. LCD Display
  4. Power/Scan Button
  5. MEM – Memory Button
  6. Temperature Mode Selection
  7. Battery Cover

MOBI 70117 Air Non-Contact Forehead Thermometer fig2

  1. Forehead Temperature Symbol
  2. Ambient Temperature Symbol
  3. Object Temperature Symbol
  4. Low Battery Symbol
  5. Temperature Display
  6. C°/F° Symbol
  7. Memory Mode Symbol

Installing / Replacing Battery

  • Located at the back lower base of the thermometer is the battery cover.
  • Insert two AAA batteries to match the “+” and “- “as illustrated in the battery compartment.
  • Make sure no plastic tabs are in between the batteries and the springs.
  • Close the battery cover and ensure it is locked in place before operation.
  • Replace battery when low battery indicator appears.
  • When replacing batteries, always replace with new AAA batteries ensuring to again match polarities indicated in the battery compartment.
    • If the thermometer does not power on after replacing the batteries, please remove the batteries and allow 60 seconds to pass before reinstalling the batteries once more.MOBI 70117 Air Non-Contact Forehead Thermometer fig3Note: Please follow local and state guidelines for proper battery disposal.

Turning the Unit On/Off

  • To turn on the unit, simply press and release the ‘Power/Scan’ button.
  • For conservation of battery, this unit will automatically power off, after inactivity. There is no manual off button.

Celsius and Fahrenheit Settings

  • Press the power button to turn the thermometer on.
  • In the bottom right corner, you will see either ‘F’ or ‘C’. This is the unit the thermometer is currently recording measurements.
  • To change the unit of measurement please press and hold the “MEM/SET” button to switch to the alternative unit. The newly selected unit will be displayed on the screen.

Adjusting Sound: On/Off

  • To turn the sound on/off long press the ‘MODE’ button.
  • If sound is being turned ‘ON’ device will ‘beep’ twice and display the word “ON”.
  • If sound is being turned ‘OFF’ device will ‘beep’ once and display the word “OFF”.

Memory Function

  • Each temperature taken by this unit is stored in the memory.
  • This unit has the capability to hold 20 readings in the memory.
  • To access the memory data please press and release the ‘MEM’ button.
  • The latest reading will be displayed. The Memory number will be displayed on the screen.
  • Continue to press (and release) the ‘MEM’ button to filter through the stored information.
  • Once the memory is full the oldest data will automatically be deleted and overwritten with the new information.

Temperature Indicators

  • This thermometer has a fever indication light.
  • A Green LED will light up around the Power/Scan button if the temperature is between 93.2°F(34.0°C) ~ 100.3°F(37.9°C).
  • A Red LED will light up around the Power/Scan button if the temperature is between 100.4°F(38.0°C) ~ 109.4°F(43.0°C)

Note: If thermometer is placed too far away to take temperature the LED will light up BLUE and beep until in the proper position.

Measurement Prepar ation for Best Results

  • For more accurate measurements always store the device in same environment it is being used, for at least 30 minutes prior. An unexpected fluctuation of ambient temperature is likely to lower measuring accuracy.
  • Do not talk or move body during measurement.
  • Make sure to dry forehead & clear any excessive hair which may cause inaccurate reading.
  • The infrared sensor must be kept clean, dry, and undamaged to ensure accurate results.
  • External factors may influence forehead temperatures, particularly when the individual has been:
    • Exposed to very hot or cold temperatures.
    • Recently exercised, bathed, or other physical exertion as this increase’s temperature.
  • Ensure the thermometer is correctly placed before taking the reading. Reference the sections below for how to best position the thermometers for each reading method.
  • Use of this device is not intended as a substitution for consultation with your physician. Measurement results are for reference only. Contact your physician if you have or suspect and health concerns.

Forehead Temperature

Forehead results will display from 89.6 °F to 109.2 °F (32.0 °C to 42.9 °C). ‘HI’ is displayed for all results above 109.2 °F (42.9°C), and ‘LO’ is displayed for result below 89.6 °F (32.0°C).

  • Ensure the unit is in ‘forehead mode’. In the upper left-hand corner, there will be an icon of a head. If you do not see the head icon in the upper left-hand corner, press and release the ‘MODE’ button to switch mode until the correct icon displays on screen.
  • For optimal results point the unit towards the forehead holding the probe 0.4 – 2 inches away from the forehead.
  • Ensure you point the unit first, then press and release the ‘ON/Scan’ button. Scanning before correctly positioning the thermometer will result in inaccurate readings.
  • Allow 1-2 seconds after pressing the ‘ON/Scan’ button for temperature to display (if the sound is on the thermometer will ‘beep’).
  • Please remove any obstructions like hats, beanies, headbands, and hair for accurate reading.
  • Please remember to take temperature in a controlled environment where direct sunlight or cold air will not affect the measurement.
  • Please wait a few minutes before taking the temperature of an individual who has been exposed to extreme heat or sever cold weather.

Object / Food Temperature

  • Ensure the unit is in ‘Object mode’. In the upper right-hand corner, there will be an icon of a bottle. If you do not see the bottle icon in the upper right hand corner press and release the ‘MODE’ button to switch modes until the correct icon displays on screen.
  • For optimal results point the unit towards the ‘object’ holding the probe 0.4 – 2 inches away from the object.
  • Ensure you point the unit first, and then press and release the ‘ON/Scan’ button. Scanning before correctly positioning the thermometer will result in inaccurate readings.
  • Allow 1-2 seconds after pressing the ‘ON/Scan’ button for temperature to display (if the sound is on thermometer will ‘beep’ once)

Room Temperature

  • When the thermometer is turned on, it will display the ambient room temperature. In the upper middle section, there will be an icon of a house.

Troubleshooting

  • Resetting the Thermometer
    To reset your thermometer, take out the batteries for 1 hour while storing the unit in a warm place with sunlight. Make sure the sensor of the thermometer isn’t in direct contact with the sunlight, only the body of the thermometer. After the 1 hour, insert the batteries and turn on the device.

  • ErrE
    If you receive and ‘ErrE’ message on the screen, place the thermometer in sunlight or a warm place and let sit for 30 minutes.

  • HI
    ‘HI’ is displayed for all results above 109.2 °F (42.9°C). Please ensure that you are in correct reading method.

  • LO
    ‘LO’ is displayed for result below 89.6 °F (32.0°C). Please ensure that you are in correct reading method.

Care and Maintenance

  • It is recommended that the thermometer is stored in a dry place free from dust.
  • Do not expose sensor to direct sunlight or air conditioning/fan that will tamper with the thermometer sensor.
  • Please do NOT expose thermometer to any humidity or water (non-waterproof).
  • To clean the unit, please use an alcohol swab or wipe. Do not use water directly. Ensure all residual alcohol has evaporated before use.

Thermometer Specification

  • Measurement Unit:
    °C/°F

  • Body Temperature Measuring Range:
    32.0°C ~ 42.9°C (89.6°F ~ 109.2°F)

  • Laboratory Accuracy:
    ±0.2°C(±0.4°F) for 35.0°C~42.0°C (95.0°F~107.6°F)
    ±0.3°C(±0.5°F) for other range

  • Object Temperature Measure Reading:
    20°C~60°C (68.0°F~140.0°F)

  • Object Temperature Measuring Accuracy:
    ±2.0°C(±4.0°F)

  • Dimension:
    138.5(L)x40.0(W)x32.6(H)mm( ±2mm)

  • Weight:
    About 72g (without battery)

  • Forehead Operating Condition:
    15.0°C~40.0°C (59.0°F~104.0°F) with a relative humidity of 15%~85%, Atmosphere pressure: 700hPa-1060hPa

  • Object Operating Condition:
    10.0°C~40.0°C (50.0°F~104.0°F) with a relative humidity of 15%~85%, Atmosphere pressure: 700hPa-1060hPa

  • Storage and Transport Condition:
    20°C~55°C (-4°F~131°F) with a relative humidity of 15%~90%, Atmosphere pressure: 700hPa-1060hPa

  • Service Life of the Battery:
    With a new battery (Carbon) approx. 10 months based on the use frequency at 5 times/day.

  • IP Classification system:
    IP22

  • Classification of ME Equipment:
    Internally powered/Continuous Operation

  1. The reference site is core, the mode is adjusted mode.
  2. This thermometer is intended for home use by customer only. The measure site is forehead.
  3. This infrared thermometer meets requirements established in ASTM Standard E 1965-98 for the thermometer system. Full responsibility for the conformance of the product to the standard is assumed by MOBI Technologies Inc. Los Angeles, CA, 90015.
  4. ASTM laboratory accuracy requirements for the thermometer only in the display range of 98 °F to 102.2 °F (37 °C to 39 °C) for infrared thermometers is ± 0.4 °F (± 0.2 °C), whereas for mercury-in-glass and electronic thermometers, the requirement per ASTM Standards E 667-86 and E 1112-86 is ± 0.2 °F ( ± 0.1 °C).

Equipment Symbols

MOBI 70117 Air Non-Contact Forehead Thermometer fig4 MOBI 70117 Air Non-Contact Forehead Thermometer
fig5

Product Safety & Warnings

WARNING:

  • This product was designed to measure human body temperature on the forehead, do not use this for any other purpose.
  • Do not use this thermometer if it has been malfunctioning or broken or damaged.
  • This product is not waterproof. Do not use this product under any wet conditions and do not immerse in any liquid or water.
  • This product should not be left alone with children or pets when not in use by an adult.
  • Do not store the thermometer in extreme temperatures below (-4°F / -20°C) or over (131°F/ 55°C).

Limited Product Warranty

MOBI® warrants its products to be free from defects in material and workmanship for period of (1 Year) from the date of purchase. Should the product prove to be defective at any time during the warranty period, we would either replace or repair the product at our option without charge. After the warranty period, a service charge will apply for replacement of parts and labor costs. To obtain warranty service, you must return the product complete with its original parts & accessories along with a copy of the original purchase, gift, or registry receipt to us. This warranty does not cover damage caused by any tampering, accident, misuse, abuse, improper maintenance, unauthorized modification, or connection to an improper power supply. There will be a charge for repair caused for such damage. This warranty becomes void upon purchase from an unauthorized, third-party seller. This warranty excludes all incidental or consequential damages and any liability other than stated above.
Questions?
Please visit: www.getmobi.com
Support Portal: https://mobisupport.zendesk.com
Support Ticket: https://www.getmobi.com/gethelp

Read User Manual Online (PDF format)