Banggood FA-2 Frequency Counter Kit Meter User Manual
- June 6, 2024
- banggood
Table of Contents
Banggood FA-2 Frequency Counter Kit Meter
This manual describes how to install and used FA-2 frequency counter.
For assistance ,contact:
WUTONG electronic
http://bg7tbl.taobao.com
Key Feature
- High speed,high precision,up to 11 avalid digit/second@10MHz
- CH2 input up to 6GHz
- High precison OCXO inside,can connect extern 10M frequency
- With data statis ,display average,max,min,p-p frequency.
- With USB port,connect to PC,data compact HP53131 mode
- CH1 with power ,dymnic up to 60dB,
overview
FA-2 frequency counter is high precision, high speed, with status function, auto status input frequency, It is convenient for long-term monitoring of external frequency changes. Channel 2 can measure frequencies up to 6 GHz. Built-in OCXO, or external 10M standard.
specification
-
CH1 frequency range:1Hz-200MHz
-
CH1 input impedance:50 OHM/1M OHM
-
CH1 input sensitive:
50 OHM:25mVrms-2Vrms
1MR:25mVrms-2Vrms -
CH1:
0.1: 0.001Hz@10M
1:0. 0001Hz@10M
10: 0.00001Hz@10M -
CH1 frequency accuary
0.1s:0.001Hz@10M
1s:0.0001Hz@10M
10s:0.00001Hz@10M -
CH1 test speed:12digits/second@100M-200M
11digits/second@10M-99.999999M,
10digits/second@1M-9.999999M
9digits/second@0.1M-0.999999M
8digits/second@90K-99K
8digits/second@1Hz-89.999K -
CH1 power test range: 50dBm-+20dBm,1M-550M@50 OHM
-
CH1 power accuary:0.5dB@10MHz
-
CH1 plug type:BNC
-
CH2 frequency range:30MHz-6GHz
-
CH2 input impedance:50 OHM
-
CH2
0.1: 1Hz@1GHz
1: 0.01Hz@1GHz
10: 0.001Hz@1GHz -
CH2 frequency accuary:
0.1s:1Hz@1GHz
1s:0.01Hz@1GHz
10s:0.001Hz@1GHz -
CH2 input power range:-20dBm- +13dBm@0.5-5G,
-
CH1 plug type:SMA
-
gate time:0.1S/1S/10S
-
inside frequency stadnard:10MHz
-
inside frequency stadnard type:OCXO
-
inside frequency stadnard ageing:0.5Hz/year
-
10M frequency output power:4dBm
-
extern frequency input range:0dBm to +20dBm
-
run mode:
CH1 frequency
CH1 statUs
CH2 frequency
CH2 statis -
front panel:
MODE, change mode
GATE: change gate time.
RST: restart test
CH1: channel 1 input
CH2: channel 2 input -
rear panel
10M REF INPUT: extern 10M input
10M REF OUTPU: 10M output
USB: USB port
STD ADJ:inside OCXO frequency adjust
power switch
dc plug -
status mode data type: average, max, min, p-p
-
power : DC11.7-12.5V, star: less than 0.4A, stabilize: less than 0.2A
-
size:LHD=10655105mm
-
weight:350g
-
accessory: DC12V adapter 1pcs
install used
rear panel
- 10MHz frequency standard input and output
- USB port, inside is ft232rl chip, usb driver can download from ftdi website
- Inside OCXO frequency adjust
- Power on/off
- DC12V input
Front panel
- 12864 LCD
- MODE key, turn run mode
- GATE key, change gate time
- RST key, restart test, RST+GATE, open/closed ch1 lpf, RST+MODE, CH1 select 50R/1MR
- Channel 2 input
- Channel 1 input
LCD display data
- CH1 input power
- Frequency display area
- Units
- Gate time,0.1S,1S,10S
- Inside/outside frequency standard
- CH1 input impedance,1MR/50R
- Gate state,display * means gate is open,else gate closed
- Display P means high precison mode
- 150KHz LPF
channel status mode
- 1-4000000000, statis points.
- now test frequency
- average value
- max value
- min value
- peak-peak value
- 1MR,input impendence,50OHM OR 1MOHM
- std is inside or extern
- gate times
- LPF select, channel 2 fequency test mode
- now frequency
- std select
- gate times
P reference for CH1 FREQ mode
channel 2 status mode
- reference channel 1 status mode
Connect PC
Usb print port connect to PC,usb chip is ft232rl,can search in web.
USB, config command
- Default bauard is 9600bps,used ascii code
- $R*press RST key,return KEY RST ROK
- $M*press MODE key,return KEY MODE MOK
- $G* Press GATE key,return KEY GATE GOK
- $CXX* XX:0-63, setting LCD contrast,save in EEPROM,return SET LCD CONTCOK
- $BXXXXX* setting baudrate ,XXXXX 00480,00960,01920,03840,05760,11520,Power off without save,retun BOK
- $S* read statis data
- The following instructions are powered off and saved
- $E2020* power up select CH1 frequency mode
- $E2121*power up select CH1 frequency statis mode
- $E2222* power up select CH2 frequency mode
- $E2323*power up select CH2 frequency statis mode
- $E3030* CH1 select 50R
- $E3131* CH1 select 1MR
- $E3232* BEEP ON.
- $E3333* BEEP off
- $E3434*opened high precision mode
- $E3535* closed high precison mode,all input freqney is 8 digitS/second
- $E4040select CH1=50R,input 10M 0dBm,calibration power meter, used with $E4141 command
- $E4141* select CH1=50R,input 10M -20dBm,calibration power meter
use notes
Away from high temperate,away from jam,power input must be less than DC12.5V
Block diagram
FAQ
Q : how can I know the machine is good
A: connect front panel 10M to rear panel CH1, LCD will display 10M means
counter is good.
Q: test low frequency, frequency is not accuary, how to do
A: open LPF
Q: can used oscillator pen to test
A: not suggest, because X1 frequency is limit , X10 is HIGH impedance, it is
easy to signal attenuator, can not test
Q: what is the application of the usb port.
A: communication, control counter with PC.
Read User Manual Online (PDF format)