INVACARE Matrx Contour Visco NG User Manual
- June 6, 2024
- INVACARE
Table of Contents
Yes, you can°.
This manual MUST be given to the user of the product.
BEFORE using this product, this manual MUST be read and saved for future
reference.
General
1.1 Introduction
This user manual contains important information about the handling of the
product. To ensure safety when using the product, read the user manual
carefully and follow the safety instructions.
Invacare reserves the right to alter product specifications without further
notice.
Before reading this document, make sure you have the latest version. You find
the latest version as a PDF on the Invacare website.
If you find that the font size in the printed document is difficult to read,
you can download the PDF version from the website. The PDF can then be scaled
on-screen to a font size that is more comfortable for you.
In case of a serious incident with the product, you should inform the
manufacturer and the competent authority in your country.
1.2 Compliance
Quality is fundamental to the company’s operation, working within the
disciplines of ISO 13485.
This product features the CE mark, in compliance with the Medical Device
Regulation 2017/745 Class I. The launch date of this product is stated in the
CE declaration of conformity.
We are continuously working towards ensuring that the company’s impact on the
environment, locally and globally, is reduced to a minimum.
We only use REACH-compliant materials and components.
1.3 Service Life
The expected service life of this product is three years when used daily and
in accordance with the safety instructions, maintenance intervals, and correct
use, stated in this manual.
The effective service life can vary according to frequency and intensity of
use.
1.4 Warranty Information
We provide a manufacturer’s warranty for the product in accordance with our
General Terms and Conditions of Business in the respective countries.
Warranty claims can only be made through the provider from whom the product
was obtained.
1.5 Limitation of Liability
Invacare accepts no liability for damage arising from:
- Non-compliance with the user manual
- Incorrect use
- Natural wear and tear
- Incorrect assembly or set-up by the purchaser or a third party
- Technical modifications
- Unauthorized modifications and/or use of unsuitable spare parts
Safety
2.1 General Safety Information
Education, clinical judgment, and action-based planning based on vulnerability
are fundamental factors in the prevention of pressure injuries.
A range of assessment scales can be used as a formal method of assessing risk
from pressure injury development and should be used in conjunction with an
informal assessment (informed nursing judgment). Informal assessment is
considered to be of greater importance and clinical value.
WARNING!
Risk of Serious Injury or Damage
Improper use of this product may cause injury or damage.
– If you are unable to understand the warnings, cautions, or instructions,
contact a health care professional or provider before attempting to use this
equipment.
– Do not use this product or any available optional equipment without first
completely reading and understanding these instructions and any additional
instructional material such as the user manual, service manual, or instruction
sheet supplied with this product or optional equipment.
WARNING!
Risk of Developing Pressure Injuries
Any object between the user and the pressure-reducing surface may result in
the development of pressure injuries.
– Make sure that the support surface in contact with the user is kept free
from crumbs and other food debris.
– Drip cables and other foreign objects must not become trapped between the
user and the pressure-reducing surface.
CAUTION!
Risk of Injury or Damage to Property
– Before use, check the product for any defects.
– Do not use a defective product.
2.2 Safety Information for Transport
- Take care when handling the product to avoid any damage.
- Avoid contact with jewelry, nails, abrasive surfaces, etc.
- Avoid contact with walls, door frames, door catches or locks, etc.
- Do not transport in roll cages unless completely protected from the sharp edges of the cage.
Product Overview
3.1 Product Description
The Invacare Matrx Contour Visco NG is a light anatomically contoured 100%
viscoelastic (memory) foam cushion The cover is multi-stretch, water-
resistant, and vapor-permeable.
It features an anti-slip base and carries a handle.
Certain sizes are available with an HR foam base layer.
3.2 Intended Use
This cushion is intended to be used in an armchair or wheelchair as part of an
overall pressure injury prevention program of care.
There are no contraindications known.
3.3 Symbols on the Product
**** | European Conformity | **** | Medical Device |
---|---|---|---|
**** | Manufacturer | **** | Date of manufacture |
**** | European representative | **** | Read user manual |
**** | Lot number | **** | Maximum user weight |
**** | Machine wash (maximum temperature see label) | **** | Tumble dry low |
heat
---|---|---|---
| Do not iron| | Do not dry clean
| Do not bleach| | Avoid burning cigarette
| Do not put near flame| | Do not pierce or cut
****| Orientation label (depending on the product)|
Usage
4.1 Safety information
WARNING!
Invacare advises that the patient is assessed by a trained healthcare
professional to ensure the correct size and seating requirements. This is to
achieve the best pressure and postural care and that position for the
individual’s needs are met. – Always consult a healthcare professional before
using the product.
WARNING!
It is very important for the patient to reposition themselves, or to be
repositioned, on a regular basis. This relieves pressure which helps prevent
both tissue compression and potential pressure injury formation. Skin redness
is one clinical indicator of skin breakdown.
– Regularly inspect/monitor if your skin develops any redness.
– Immediately contact a healthcare professional, if your skin develops
redness.
IMPORTANT!
Accidental cover damage allows fluid ingress and staining.
– Do not place sharp objects on the cover.
– When using patient transfer aids, ensure not to damage the cover. Check for
sharp edges or burrs before use.
– Ensure that the cushion is not jammed or damaged by sharp edges when used on
wheelchairs.
– Avoid burning cigarettes or open flames near the cushion.
– If pets are around, ensure that no claws puncture the cover.
IMPORTANT!
The water-resistant cover allows fluids to roll off the surface, however with
sustained periods of fluid contact the cover may absorb moisture. – Wipe away
any fluid as soon as it is identified.
4.2 Using Cushion
- Remove all packaging before use.
- Place cushion on wheelchair or chair with the grey non-slip base facing down.
- Ensure that the cushion is placed in coordination with the orientation label, located on the left-hand side of the cushion.
Maintenance
5.1 Inspection
Check cushion (foam and cover) for strike-through (this may include fluid
ingress, stains, rips, or damage) after the use of each patient and on a
minimum monthly basis depending on which occurs first) by a suitably
qualified and competent person.
Check cushions
- Unzip the cover completely.
- Check for any staining on the interior foam.
- Replace any damaged cover or stained foam and dispose of it as per local authority procedure.
5.2 Cleaning and Disinfection
General Safety Information
CAUTION!
Risk of Contamination
– Take precautions for yourself and use appropriate protective equipment.
IMPORTANT!
Wrong fluids or methods can harm or damage the product.
– All cleaning agents and disinfectants used must be effective, compatible
with one another, and must protect the materials they are used to clean.
– Never use corrosive fluids (alkalines, acid, etc.), abrasive cleaning
agents, or solvents (cellulose thinner, acetone, etc.). We recommend a mild
detergent.
– Always make sure that the product is completely dried before taking it into
use again.
For cleaning and disinfection in clinical or long-term care environments, follow your in-house procedures.
Cleaning Intervals
IMPORTANT!
Regular cleaning and disinfection enhance smooth operation, increase the
service life and prevent contamination.
Clean and disinfect the product
– regularly while it is in use,
– before and after any service procedure,
– when it has been in contact with anybody’s fluids,
– before using it for a new user.
Cleaning Instructions
Cleaning Cover
(Removal of contaminants such as dust and organic matter)
- Remove cover for laundering.
- Launder the cover using a diluted detergent solution (Instructions and maximum temperature on label).
Drying Cover
- Hang cover from a line or bar and drip dry in a clean indoor environment. or Tumble dry on a low heat setting.
IMPORTANT!
– Tumble dry setting must not exceed 40 °C.
– Do not tumble dry for longer than 10 minutes.
– Dry thoroughly before use.
Disinfection Instructions
IMPORTANT!
– Only use disinfectants and methods approved by your local infection control
institution and follow your local infection control policy.
Disinfecting cover
(Reducing the number of microorganisms)
IMPORTANT!
Failure in the disinfection process may result in the accumulation of reagents
that could damage the polyurethane coating, react with the frame, or negate
the biocompatibility results.
– Ensure that all cleaning agents, and disinfectants, are thoroughly rinsed
off.
– Dry thoroughly before use.
- Wipe down the cover with a suitable detergent.
- Rinse the cover thoroughly with clean water using a single-use nonabrasive cloth.
- Dry the cover thoroughly.
Heavy Soilage
When badly soiled, we recommend cleaning with a diluted detergent solution at a maximum temperature in the washing machine.
– Clean up all spillages of bodily fluids i.e. blood, urine, feces, sputum, wound exudate, and all other bodily secretions as soon as possible using a suitable detergent.
– Large spillages of blood should be absorbed and removed with paper towels
first, followed by as above.
When badly soiled, we recommend cleaning with a diluted detergent solution at
a maximum temperature in the washing machine.
– Clean up all spillages of bodily fluids i.e. blood, urine, feces, sputum,
wound exudate, and all other bodily secretions as soon as possible using a
suitable detergent.
– Large spillages of blood should be absorbed and removed with paper towels
first, followed by as above.
Replacing Cover
-
Unzip the cover and remove it carefully from the foam core.
-
Place the new cover onto the foam core.
Ensure that the corners of the foam core are positioned correctly into the corners of the cover. -
Close the zipper. Where ever possible ensure the slider /sliders are positioned at the rear of the cushion.
WARNING!
– In event of contamination, contact your hygiene specialist.
– Remove contaminated foams from use.
After Use
6.1 Storage
- Store cushions in a dry environment.
- Store cushions within a protective cover.
- Store items on a clean, dry surface free from sharp edges to avoid any possible damage.
- Never store other items on top of a cushion.
- Do not store cushions next to radiators or other heating devices.
- Protect cushions from direct sunlight.
6.2 Disposal
Be environmentally responsible and recycle this product through your recycling
facility at its end of life.
Disassemble the product and its components, so the different materials can be
separated and recycled individually.
The disposal and recycling of used products and packaging must comply with the
laws and regulations for waste handling in each country. Contact your local
waste management
company for information.
Technical Data
7.1 Dimensions and Weights
Total height [mm] | 100 |
---|---|
Product weight [kg] | 1.5 – 2.5 |
User weight per size
Width [mm] | Depth [mm] | Firmness | User Weight [kg] |
---|---|---|---|
380 | 405 | Firm | 40 – 80 |
380 | 430 | ||
380 | 455 | ||
405 | 405 | ||
405 | 430 | ||
405 | 455 | ||
Width [mm] | Depth [mm] | Firmness | User Weight [kg] |
--- | --- | --- | --- |
430 | 405 | Very Firm | 60 – 100 |
430 | 430 | ||
430 | 455 | ||
455 | 405 | ||
455 | 430 | ||
455 | 455 | 60 – 110 | |
480 | 405 | ||
480 | 430 | ||
480 | 455 | ||
505 | 405 | 60 – 120 | |
505 | 430 | ||
505 | 455 | ||
355 | 530 | 60 – 100 | |
405 | 530 | 60 – 120 | |
455 | 530 | 60 – 130 | |
505 | 530 | 60 – 140 |
7.2 Materials
Viscoelastic foam HR foam1 | Polyurethane foam components |
---|---|
Cover | PU coated and Polyester fabric |
1 Certain size only
All product components do not contain natural rubber Latex.
Notes
Belgium & Luxemburg:
Invacare nv
Autobaan 22
B-8210 Loppem
Tel: (32) (0)50 83 10 10
Fax: (32) (0)50 83 10 11
belgium@invacare.com
www.invacare.be
United Kingdom:
Invacare Limited
Pencoed Technology Park, Pencoed
Bridgend CF35 5AQ
Tel: (44) (0) 1656 776 222
Fax: (44) (0) 1656 776 220
uk@invacare.com
www.invacare.co.uk
Invacare UK Operations Limited
Pencoed Technology Park, Pencoed
Bridgend CF35 5AQ
UK
Invacare Portugal, Lda
Rua Estrada Velha 949
4465-784 Leça do Balio
Portugal Making Life’s Experiences Possible®
Yes, you can°.
Documents / Resources
|
INVACARE Matrx Contour Visco
NG
[pdf] User Manual
Matrx Contour Visco NG
---|---
References
- | Invacare AT
- | Invacare BE
- | Invacare CH
- | Invacare GB
- | Invacare DE
- | Invacare DK
- | Invacare ES
- | Invacare IE
- | Invacare IT
- | Invacare NL
- | Invacare PT
- | Invacare SE