SYMPTOM: | PROBABLE CAUSE: | SOLUTION: |
---|
Alarm:
Short beeps, long pause Concentrator not operating, power switch On.
Beep……Beep……
| Main Power Loss:
1. Power cord not plugged in.| 1. Insert plug into outlet.
2. No power at outlet.| 2. Inspect house circuit breakers or fuses. If
problem recurs, use a different outlet.
3. Tripped circuit breaker.| 3. Call Service Provider.
Alarm:
Continuous Concentrator not operating, Power Switch On.
Beeeeeeeep….
| System Failure:
1. Unit overheating due to blocked air intake. | 1a. Remove and clean cabinet
filters.
1b. Move oxygen concentrator at least three inches away from walls, draperies
or furniture.
2. Insufficient power at outlet.| 2. DO NOT use extension cords. Move to
another electrical outlet or circuit.
3. Internal repairs required.| 3. Call Service Provider.
Alarm:
RapidBeep….Beep…
Beep…Beep| Low Flow Alarm:
1. Kinked or blocked tubing, cannula or humidifier.| 1. Inspect for kinks or
blockages. Correct, clean or replace item. Once corrected, turn power Off for
60 seconds and then turn power back ON.
2. Flowmeter set at 0.5 L/min or less.| 2. Reset flowmeter to prescribed
flowrate. If prescribed flowrate is below 0.5 L/min, use the Pediatric
Flowmeter IRCPF16 (Only with Perfecto2 5)
YELLOW or RED light Illuminated.
Alarm:
Continuous On RED light only.
- Only applies to IRC5PO2.
| 1. Low oxygen purity.| 1. Clean or Replace filters.
2. Kinked or blocked tubing, cannula or humidifier. | 2. Inspect for kinks
or blockages. Correct, clean or replace item. Once corrected, turn power Off
for 60 seconds and then turn power back On.
3. Flowmeter set at 0.5 L/min or less.*| 3. Check flowmeter is set to 1.0
L/min or more.
4. Unit overheating due to blocked air intake. | 4a. Remove and clean cabinet
filters.
4b. Move concentrator at least three inches from walls, draperies, and
furniture.
5. Insufficient power at outlet.| 5. DO NOT use extension cords. Move to
another electrical outlet or circuit.
6. Internal repairs required.| 6. Call Service Provider.
GREEN light with YELLOW Light Flashing.| 1. Internal repairs required.| 1.
Call Service Provider.