SECTION 7–TROUBLESHOOTING GUIDE


SECTION 7–TROUBLESHOOTING GUIDE
SYMPTOM: PROBABLE CAUSE: SOLUTION:

Alarm:
Short beeps, long pause Concentrator not operating, power switch On.

Beep……Beep……

| Main Power Loss:
1. Power cord not plugged in.| 1. Insert plug into outlet.
2. No power at outlet.| 2. Inspect house circuit breakers or fuses. If problem recurs, use a different outlet.
3. Tripped circuit breaker.| 3. Call Service Provider.
Alarm:
Continuous Concentrator not operating, Power Switch On.

Beeeeeeeep….

| System Failure:
1. Unit overheating due to blocked air intake. | 1a. Remove and clean cabinet

filters.

1b. Move oxygen concentrator at least three inches away from walls, draperies or furniture.
2. Insufficient power at outlet.| 2. DO NOT use extension cords. Move to another electrical outlet or circuit.
3. Internal repairs required.| 3. Call Service Provider.
Alarm:
RapidBeep….Beep…
Beep…Beep| Low Flow Alarm:
1. Kinked or blocked tubing, cannula or humidifier.| 1. Inspect for kinks or blockages. Correct, clean or replace item. Once corrected, turn power Off for 60 seconds and then turn power back ON.
2. Flowmeter set at 0.5 L/min or less.| 2. Reset flowmeter to prescribed flowrate. If prescribed flowrate is below 0.5 L/min, use the Pediatric Flowmeter IRCPF16 (Only with Perfecto2 5)
YELLOW or RED light Illuminated.

Alarm:
Continuous On RED light only.

  • Only applies to IRC5PO2.

| 1. Low oxygen purity.| 1. Clean or Replace filters.
2. Kinked or blocked tubing, cannula or humidifier.
 | 2. Inspect for kinks or blockages. Correct, clean or replace item. Once corrected, turn power Off for 60 seconds and then turn power back On.
3. Flowmeter set at 0.5 L/min or less.*| 3. Check flowmeter is set to 1.0 L/min or more.
4. Unit overheating due to blocked air intake. | 4a. Remove and clean cabinet filters.
4b. Move concentrator at least three inches from walls, draperies, and furniture.
5. Insufficient power at outlet.| 5. DO NOT use extension cords. Move to another electrical outlet or circuit.
6. Internal repairs required.| 6. Call Service Provider.
GREEN light with YELLOW Light Flashing.| 1. Internal repairs required.| 1. Call Service Provider.


Part No 1143482 Perfecto 2 ™ Series