INVACARE 2478444 Premier Dress Toileting High Sling for Patient Lifts User Manual
- June 6, 2024
- INVACARE
Table of Contents
INVACARE 2478444 Premier Dress Toileting High Sling for Patient Lifts
This manual MUST be given to the user of the product.
BEFORE using this product, this manual MUST be read and saved for future
reference.
General
Introduction
This user manual contains important information about the handling of the product. To ensure safety when using the product, read the user manual carefully and follow the safety instructions.
Invacare reserves the right to alter product specifications without further notice.
Before reading this document, make sure you have the latest version. You find the latest version as a PDF on the Invacare website.
If you find that the font size in the printed document is difficult to read, you can download the PDF version from the website. The PDF can then be scaled on screen to a font size that is more comfortable for you.
In case of a serious incident with the product, you should inform the manufacturer and the competent authority in your country.
Symbols in this Document
Symbols and signal words are used in this document and apply to hazards or unsafe practices which could result in personal injury or property damage. See the information below for definitions of the signal words.
| WARNING
Indicates a hazardous situation that could result in serious injury or death
if it is not avoided.
---|---
| CAUTION
Indicates a hazardous situation that could result in minor or slight injury if
it is not avoided.
| IMPORTANT
Indicates a hazardous situation that could result in damage to property if it
is not avoided.
| Tips and Recommendations Gives useful tips, recommendations and information
for efficient, trouble-free use.
Service Life
Expected service life of the product is 1-5 years. Service life time varies depending on the fabric, frequency of use, laundering practices, and weight carried.
Warranty Information
We provide a manufacturer’s warranty for the product in accordance with our General Terms and Conditions of Business in the respective countries. Warranty claims can only be made through the provider from whom the product was obtained.
Limitation of Liability
Invacare accepts no liability for damage arising from:
- Non-compliance with the user manual
- Incorrect use
- Natural wear and tear
- Incorrect assembly or set-up by the purchaser or a third party
- Technical modifications
- Unauthorised modifications and/or use of unsuitable spare parts
Compliance
Quality is fundamental to the company’s operation, working within the disciplines of ISO 13485.
This product features the CE mark, in compliance with the Medical Device Regulation 2017/745 Class I. The launch date of this product is stated in the CE declaration of conformity.
We are continuously working towards ensuring that the company’s impact on the environment, locally and globally, is reduced to a minimum.
We only use REACH compliant materials and components.
Product-specific standards
The product has been tested and conforms to ISO 10535 (Hoists for the transfer of disabled persons) and all related standards.
For further information about local standards and regulations, contact your local Invacare representative. See addresses at the end of this document.
Safety
General Safety Information
| WARNING!
Risk of Serious Injury or Damage
Improper use of this product may cause injury or damage.
- If you are unable to understand the warnings, cautions or instructions, contact a health care professional or provider before attempting to use this equipment.
- Do not use this product or any available optional equipment without first completely reading and understanding these instructions and any additional instructional material such as user manual, service manual or instruction sheet supplied with this product or optional equipment.
---|---
| WARNING!
Risk of Injury or Damage
Improper use of this product may cause injury or damage. – Do not make any
unauthorized alterations or modifications to the product. – Do not use the
sling for any other purpose than to transfer an individual from one resting
surface to another.
---|---
| WARNING!
The maximum safe working load must not be exceeded
- Do not exceed the maximum safe working load of this product or other components of the lifting system. See documentation or the labeling for the stated maximum safe working load.
- The component with the lowest load limit determines the maximum safe working load of the entire system.
---|---
| WARNING!
Risk of Injury or Damage
Ignition sources can cause burns or fire.
- Patient transfer must be performed with a safety space between the lift and possible ignition sources (heater, stove, fireplace, etc.).
- The patient and assistants must not smoke during transfer.
- The sling must not be placed over heat sources (heater, stove, fireplace, etc.).
---|---
| CAUTION!
Compatibility of Slings with Attachment System
Invacare uses a common attachment system based on hooks and loops. Loops on
the sling are attached to hooks on the lift.
- A risk assessment is always to be carried out by a professional prior to use of Invacare slings with lifts of other manufacturers.
- If you are in doubt about the use of Invacare slings with equipment of other manufacturers, contact your Invacare provider for advice.
---|---
Labels and Symbols on the Product
The labels are placed on the upper backside of the sling.
Symbols — Fig. 1
Ⓐ | Read user manual |
---|---|
Ⓑ | European Conformity |
Ⓒ | Manufacturer |
Ⓓ | Medical device |
Ⓔ | Reference number |
Ⓕ | Serial number |
Ⓖ | Date of manufacture |
Ⓗ | Lot number |
Ⓘ | Max. safe working load |
Ⓙ | Do not bleach |
Ⓚ | Tumble dry, low temperature |
Ⓛ | Do not iron |
Ⓜ | Washing max. 90 °C |
Ⓝ | Field for patient name |
Ⓞ
| QR code with link to instruction video
Product Overview
Intended Use
The Invacare Dress Toileting slings are non-rigid body-support units to transfer an individual from one surface to another. The slings are intended to be combined with a passive lift using a loop to hook attachment with 2 or 4 connection points.
The slings are not intended to be left underneath the patient after the transfer.
A healthcare professional or private person who has received proper training is the intended operator of this product.
Indications
The slings are indicated for the transfer of partially immobile patients with the following minimum degrees of body control:
Contraindications
The Dress Toileting slings are contraindicated both for immobile individuals, and those with high and bilateral amputee legs.
Accessories
- Extension loops 20 cm or 30 cm length Intended to be attached to the outer loops and extend the suspension straps of the sling for more reach or positioning options.
Usage
General Safety Information
| WARNING!
Risk of injury or damage
Improper use of this product may cause injury or damage.
- Do not attempt any transfer without approval of the patient’s healthcare professional.
- Ensure that the patient is not prone to spasm or a risk assessment has been completed to address the possibility of forward pitching.
---|---
| WARNING!
Risk of injury
Improperly attached or adjusted slings can cause the patient to fall or cause
injury to assistants.
- See the user manual of the lift in addition and follow the instructions for safety and use.
- Check for proper sling attachment before starting the transfer.
- Check again for proper sling attachment when the straps are tensioned before lifting the patient off the surface.
- Do not use any kind of plastic back incontinence pad or seating cushion between patient and sling material that may cause the patient to slide out of the sling during transfer.
- Make sure that there is sufficient head support when lifting a patient
- Position the patient in the sling as directed by the instructions provided with the sling.
- Adjustments for patient safety and comfort should be made before moving the patient.
---|---
**| WARNING!
Risk of injury**
Usage of damaged slings can cause the patient to fall or cause injury to
assistants.
- Bleached, torn, cut, frayed, or damaged slings are unsafe and could result in injury. Discard immediately.
---|---
To select the appropriate sling, a risk assessment must be performed by a healthcare professional. The risk assessment must consider:
- The patient’s weight, size, physical ability and medical condition.
- The type of transfer and the environment.
- The compatibility to the other lifting equipment used.
| For more information see the Invacare Sling Selection Guide
---|---
Applying the Sling
Sling Application in Sitting Position
- Lean the patient forward ensuring that he is well supported to reduce the risk of falling.
- Slide the sling down behind the back of the patient in a central position. Fig. 2a
- Lean the patient back against the sling.
- Pull the leg support forward along the outside of the thigh on each side. Fig. 2b
- Pass each leg support under the corresponding thigh.
- Close the belt around the patients chest and make sure that it is well adjusted to the patient.
Sling Application in Lying Position
- Turn the patient to the side.
- Fold up half of the sling and place it centrally along the spine. Fig. 2c
- Turn the patient to the other side and unfold the sling.
- Turn the patient on his back onto the sling. Ensure the sling is symmetrically placed.
- Lift the legs by bending the knees and apply the sling under the legs.
- Close the belt around the patients chest and make sure that it is well adjusted to the patient.
| If the patient has sufficient trunk control and is able to sit up from the
bed or floor, this will ease the application and steps 1–4 can be performed
similar to applying the sling in sitting position.
---|---
Positioning of Patient Arms
The arms of the patient must be placed outside the sling
Positioning of the Leg Supports
The leg supports can be positioned in different ways depending on patient conditions or preferences and tasks to be performed:
- with crossed leg straps – Fig. 3a: The leg straps are crossed one through the other and are attached to the opposite hook.
| Recommended for maximum security as it reduces the risk of forward pitching and also maintains a better alignment of the hips and lower limbs.
---|--- - with uncrossed leg straps – Fig. 3b: The leg straps are not crossed and attached to the hook on the corresponding side.
| Recommended when the patient has abdominal sensitivity or requires intimate hygiene.
---|---
Sling Attachment to the Lift
The suspension straps of the sling are equipped with color coded loops providing different lengths to place the patient into different positions. Shorter straps at the shoulders and longer straps at the legs will produce a more vertical lift, which will assist a transfer into a sitting position. By lengthening the straps at the shoulders and shortening the straps at the legs, a more reclined position can be achieved which is more suitable for a transfer into a lying position. Match the loop colors of the corresponding straps on each side of the sling for an even lift of the patient. Fig. 4
| IMPORTANT!
For a transfer into a sitting position, place the patient as vertical as
possible in the sling
---|---
- Attach the desired loop of each strap to the corresponding hook on the lift:
a. With 2 connection points attach the shoulder strap before the leg strap on each of the hooks. Fig. 5a
b. With 4 connection points attach each of the shoulder and leg straps to an individual hook. Fig. 5b
| For models with head support, attach the head straps to the same hooks as the shoulder straps.
---|---
Removing the Sling
- After completion of the transfer, detach the sling from the lift.
- Reverse the procedure described in 4.2 Applying the Sling, page 5 .
Maintenance
Inspection
Daily Inspection Before
Use Inspection to be performed daily, before use and after each laundering:
- ·Visually inspect the sling and check all parts for signs of damage, wear or potential failure.
Periodic Inspection
| IMPORTANT!
Inspections must be performed by a qualified person who is well acquainted
with the design, use and care of slings.
---|---
A periodic safety inspection of the sling must be performed at least every 6 months unless otherwise stated in local requirements.
| A detailed safety inspection checklist is available from Invacare as a
separate document.
---|---
LOLER Statement
The UK Health and Safety Executive’s Lifting Operations and Lifting Equipment
Regulations 1998, require any equipment that is used in the workplace to lift
a load be subject to safety inspection on a six monthly basis. Please refer to
the HSE web site for guidance www.hse.gov.uk.
The person responsible for the equipment must ensure adherence to LOLER regulations.
Cleaning and Disinfection
**| WARNING!
Risk of injury
** Usage of damaged slings can cause the patient to fall or cause injury to
assistants.
- After each laundering, inspect the sling for wear, tear, and loose stitching.
---|---
| IMPORTANT!
Regular cleaning and disinfection prevents contamination.
Clean and disinfect the product
– regularly while it is in use,
– when it has been in contact with any body fluids,
– before using it for a new user
---|---
| IMPORTANT!
Always make sure that the product is completely dried before taking it into
use again.
---|---
Cleaning Instructions
Launder and dry the sling according to the washing instructions on the label.
Frequent use of high temperatures and tumble drying can reduce the service
life of the product.
We recommend the use of an ordinary household laundry detergent. Do not use
softener.
For disinfection, select a disinfectant that is compatible with the sling
materials and follow the instructions of the disinfectant manufacturer.
After Use
Storage Conditions
The sling must be stored in dry conditions with a maximum relative humidity of 50% and without direct sunlight.
Reconditioning
This product is suitable for reuse. To recondition the product for a new user, carry out the following actions:
- Inspection
- Cleaning and disinfection
For detailed information, see 5 Maintenance, page 5 .
Make sure that the user manual is handed over with the product.
If any damage or malfunction is detected, do not reuse the product.
Disposal
Be environmentally responsible and recycle this product through your recycling
facility at its end of life.
Disassemble the product and its components, so the different materials can be
separated and recycled individually.
The disposal and recycling of used products and packaging must comply with the
laws and regulations for waste handling in each country. Contact your local
waste management company for information.
Technical Data
Maximum Safe Working Load
**** | XS , S , M & L | XL |
---|---|---|
Dress Toileting Low | 200 kg | 250 kg |
Dress Toileting High
Materials
Material | Part |
---|---|
Polyester | Main fabric, straps, edge band (solid), main label (woven) |
Polyamide | Edge band (net & spacer), secondary label |
Polyurethane | Padding |
The materials used are not flame retardant.
Invacare Distributors
Eastern Europe, Middle East & CIS:
Invacare EU Export
Am Achener Hof 8
D-88316 Isny
Tel: (49) (0)7562 700 397
eu-export@invacare.com
www.invacare-eu-export.com
Documents / Resources
|
INVACARE 2478444 Premier Dress Toileting High Sling for Patient
Lifts
[pdf] User Manual
2478444, Premier Dress Toileting High Sling for Patient Lifts
---|---
References
- HSE: Information about health and safety at work
- Home | Invacare Europe
- | Invacare AT
- | Invacare BE
- | Invacare CH
- | Invacare GB
- | Invacare DE
- | Invacare DK
- | Invacare ES
- Invacare France | Large gamme de matériel médical
- | Invacare IE
- | Invacare IT
- | Invacare NL
- | Invacare PT
- | Invacare SE
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>