DENTALEZ 7717-006A Ramvac Aeras Vacuum Tank Installation Guide
- June 5, 2024
- DENTALEZ
Table of Contents
INSTALLATION MANUAL
VACUUM TANK INSTALLATION MANUAL
NOTES:
SECTION I – SPECIFICATIONS
TANKS
| 1 15-GALLON TANK| 30-GALLON TANK
---|---|---
Typical No. of Treatment Rooms| I 1-6| 7-12
Nominal Width (in)| 1 29″| 29″
Nominal Depth (in)| I 19″| 19″
Nominal Height (in)| 1 24″| 36″
Floor to Drain Centerline (in)| I 8″| 8″
this assumes one chair per person. For bays, let”Rooms”equal chairs.|
- Connects to waste piping (see SECTION IV – PLUMBING – DRAIN AND WASTE DETAILS)
- Connects to operatories (see SECTION IV – PLUMBING – FACILITY VACUUM LINE FROM OPERATORIES)
- Connects to vacuum unit (see SECTION IV – PLUMBING – TANK TO VACUUM POWER UNIT LINE)
- Connects to Aeras control TANK FLOAT (opt.) connector (see SECTION III – AERAS CONTROL – CONNECTIONS)
VACUUM TANK INSTALLATION MANUAL
I | r INFIWITANK | 50-GALLON CYLINDRICAL TANK |
---|---|---|
Typical No. of Treatment Roomst | I 1-20 | 13-20 |
Nominal Width (in) | I 29″ | 26″ |
Nominal Depth (in) | I 21″ | 26″ |
Nominal Height (in) | I 38″ | 62″ |
Floor to Drain Centerline (in) | I 8″ | 5″ |
IThis assumes one chair per person. For bays, let “Rooms” equal chairs. |
- Connects to waste piping (see SECTION IV – PLUMBING – DRAIN AND WASTE DETAILS)
- Connects to operatories* (see SECTION IV – PLUMBING – FACILITY VACUUM LINE FROM OPERATORIES)
- Connects to vacuum unit** (see SECTION IV – PLUMBING – TANK TO VACUUM POWER UNIT LINE)
- Connects to Aeras control INFINITANK SOLENOID (opt.) connector (see SECTION III – AERAS CONTROL – CONNECTIONS)
- Connects to Aeras control TANK FLOAT (opt.) connector (see SECTION III – AERAS CONTROL – CONNECTIONS)
*Tank inlet for 50-gallon tank is 3″.
**Tank outlet for 50-gallon tank is 3″.
TANKS/VACUUMS STACKED
TANKS/VACUUMS STACKED| |
---|---|---
| 15-GALLON TANK WITH RV4-12| 15-GALLON TANK WITH RV5-2, RV7-2 OR RV10-2
Typical No. of Treatment Rooms†| 1-6| 1-6
Nominal Width (in)| 29″| 33″
Nominal Depth (in)| 19″| 21″/22″ (RV10-2)
Nominal Height (in)| 48″| 59″
Floor to Drain Centerline (in)| 8″| 8″
†This assumes one chair per person. For bays, let “Rooms” equal chairs.
| 30-GALLON TANK
WITH RV5-2, RV7-2, OR RV10-2| INFINITANK
WITH RV5-2, RV7-2, OR RV10-2
---|---|---
Typical No. of Treatment Rooms†| 7-12| 1-20
Nominal Width (in)| 33″| 33″
Nominal Depth (in)| 21″/ 22″ (RV10-2)| 21″/22″ (RV10-2)
Nominal Height (in)| 71″| 71″
Floor to Drain Centerline (in)| 8″| 8″
†This assumes one chair per person. For bays, let “Rooms” equal chairs.
- Connects to waste piping (see SECTION IV – PLUMBING – DRAIN AND WASTE DETAILS)
- Connects to operatories (see SECTION IV – PLUMBING – FACILITY VACUUM LINE FROM OPERATORIES)
- Connects to vacuum unit (see SECTION IV – PLUMBING – TANK TO VACUUM POWER UNIT LINE)
- Platform (optional)
- Connects to Aeras control TANK FLOAT (opt.) connector (see SECTION III – AERAS CONTROL – CONNECTIONS)
- Connects to Aeras control INFINITANK SOLENOID (opt.) connector (see SECTION III – AERAS CONTROL – CONNECTIONS)
SECTION II – INSTALLATION
AERAS 15-GALLON/30-GALLON TANK SETUP
- Connects to waste piping (see SECTION IV – PLUMBING – DRAIN AND WASTE DETAILS)
- Connects to operatories (see SECTION IV – PLUMBING – FACILITY VACUUM LINE FROM OPERATORIES)
- Connects to vacuum unit mounted tank (see SECTION IV – PLUMBING – TANK TO VACUUM POWER UNIT LINE)
- Connects to vacuum unit located remotely (see SECTION IV – PLUMBING – TANK TO VACUUM POWER UNIT LINE)
- Vacuum check valve (see SECTION IV – PLUMBING – VACUUM CHECK VALVE INSTALLATION)
- Suspension Hooks – Remove tank legs. Use leg bolts to attach and suspend tank from ceiling with chain or cable (customer supplied).
- Connects to Aeras control TANK FLOAT (opt.) connector (see SECTION III – AERAS CONTROL – CONNECTIONS)
VACUUM TANK INSTALLATION MANUAL
AERAS INFINITANK SETUP
- Connects to waste piping (see SECTION IV – PLUMBING – DRAIN AND WASTE DETAILS)
- Connects to operatories (see SECTION IV – PLUMBING – FACILITY VACUUM LINE FROM OPERATORIES)
- Connects to vacuum unit (see SECTION IV – PLUMBING – TANK TO VACUUM POWER UNIT LINE)
- Connects to Aeras control TANK FLOAT (opt.) connector (see SECTION III – AERAS CONTROL – CONNECTIONS)
- Connects to Aeras control INFINITANK SOLENOID (opt.) connector (see SECTION III – AERAS CONTROL – CONNECTIONS)
AERAS INFINITANK HIGH-FLOW SETUP
NOTE: Shipped with a parts bag labeled “Parts for Optional High-flow InfiniTank” containing a coupler and a plug. See below for proper assembly.
- Connects to waste piping (see SECTION IV – PLUMBING – DRAIN AND WASTE DETAILS)
- Connects to operatories* (see SECTION IV – PLUMBING – FACILITY VACUUM LINE FROM OPERATORIES)
- Connects to vacuum unit** (see SECTION IV – PLUMBING – TANK TO VACUUM POWER UNIT LINE)
- Connects to Aeras control INFINITANK SOLENOID (opt.) connector (see SECTION III – AERAS CONTROL – CONNECTIONS)
- Connects to Aeras control TANK FLOAT (opt.) connector (see SECTION III – AERAS CONTROL – CONNECTIONS)
*Tank inlet for 50-gallon tank is 3″.
**Tank outlet for 50-gallon tank is 3″. Tank outlet for InfiniTank is 2″.
a. Connect the two tank outlets together as shown.
b. See SECTION II – INSTALLATION – CYLINDRICAL TANK SETUP – MULTIPLE VACUUM for connection.
CYLINDRICAL TANK SETUP
-
Connects to waste piping (see SECTION IV – PLUMBING – DRAIN AND WASTE DETAILS)
-
Connects to operatories* (see SECTION IV – PLUMBING – FACILITY VACUUM LINE FROM OPERATORIES)
-
Connects to vacuum unit** (see SECTION IV – PLUMBING – TANK TO VACUUM POWER UNIT LINE)
-
Vacuum check valve (see SECTION IV – PLUMBING – VACUUM CHECK VALVE INSTALLATION)
*Tank inlet for 50-gallon tank is 3″.
**Tank outlet for 50-gallon tank is 3″. -
Connects to waste piping (see SECTION IV – PLUMBING – DRAIN AND WASTE DETAILS)
-
Connects to operatories* (see SECTION IV – PLUMBING – FACILITY VACUUM LINE FROM OPERATORIES)
-
Connects to vacuum unit** (see SECTION IV – PLUMBING – TANK TO VACUUM POWER UNIT LINE)
Tank locations (2 and 3) are reversible.
Sling kit for horizontal 50-gallon tank is Ramvac supplied.
Mounting hardware for the custom tank is customer-supplied. Tank support requirements: 1 gallon of liquid = 8.33 lbs. (e.g. 50-gallon tank × 8.33 lbs. = 416.5 lbs.)
SECTION III – AERAS CONTROL
VACUUM TANK INSTALLATION MANUAL
CONNECTIONS
- InfiniTank TM
- solenoid (optional)
- Autowash (optional)
- Vacuum (inHG/kPa)
- Motor temperature
- Pump temperature
- Oil level
- Tank float (optional)
- Water in oil
- Water in pump
- Ambient temperature/humidity
INFINITANK SWITCH
The INFINITANK switch should only be turned ON when using an InfiniTank
system. For all other vacuum tanks, the
INFINITANK switch should be turned OFF.
To access the INFINITANK switch, on the Technical screen, press ADVANCED >.
This will open a new screen showing the INFINITANK switch.
To enable the InfiniTank option, press the gray INFINITANK switch. The switch
will turn green, indicating that the
InfiniTank option is now ON. Pressing the green switch will turn the
InfiniTank option OFF.
Press < TECHNICAL to return to the previous screen.
SECTION IV – PLUMBING
VACUUM TANK INSTALLATION MANUAL
DRAIN AND WASTE DETAILS
MULTIPLE TANK SETUP
TANK DRAIN LINE
Waste pipe, slide valve and drain check valve:
-
1½” schedule 40 PVC with DWV fittings.
-
Slope: Minimum 1/4″ per 10 feet.
-
Use only Ramvac supplied 1½” drain check valve.
-
Drain into building sewer system.
Trap and drain branch: -
Trap and drain branch line two sizes larger than waste pipe.
-
Trap seal minimum 4″ deep.
Vents: -
Trap vent size no less than half the trap and drain branch size.
Vents no smaller than drain check valve.
Vents must extend a minimum of 6″ above tank top before turning horizontally.
NOTICE
Follow current NFPA 99c Edition, Ramvac instructions, and all local codes.
DRAIN CHECK VALVE INSTALLATION
Install drain check valve assembly with: “HORIZONTAL USE UP” facing UP
- Pipes butted inside connecting couplers (no gaps)
- Flow toward sewer
- Fits 1½” schedule 40 PVC
- Slide valve open
NOTICE
The slide valve is for service use only. Leave in the open position.
VACUUM CHECK VALVE INSTALLATION
Install vacuum check valve assembly with:
- “THIS SIDE UP” facing up
- Pipes butted inside connecting couplers (no gaps)
- Flow toward vacuum unit
- Fits 2″ schedule 40 PVC
- Isolates non-operating vacuum units
VACUUM TANK INSTALLATION MANUAL
TANK INLET MANIFOLD FOR MULTIPLE TANKS
FACILITY VACUUM LINE FROM OPERATORIES
NOTICE
Follow current NFPA 99c Edition, Ramvac instructions, and all local codes.
Vacuum Model RV4-12:
-
Up to 5 rooms: 1½” (or larger) schedule 40 PVC with DMW fittings or equivalent.
-
5 to 15 rooms: 2″ (or larger) schedule 40 PVC with DMW fittings or equivalent.
Vacuum Models RV5-2, RV7-2 and RV10-2: -
2″ (or larger) schedule 40 PVC with DMW fittings or equivalent.
All Vacuum Models: -
Slope: Minimum 1/4″ per 10 feet.
-
If an in-line low spot is unavoidable, place it in a known location and incorporate a clean-out.
-
For overhead plumbing, contact Ramvac.
-
For nitrous oxide scavenging, contact Ramvac.
TANK TO VACUUM POWER UNIT LINE
Vacuum Model RV4-12:
-
1½” (or larger) schedule 40 PVC with DMW fittings or equivalent.
Vacuum Models RV5-2, RV7-2 and RV10-2: -
2″ (or larger) schedule 40 PVC with DMW fittings or equivalent.
All Vacuum Models: -
Use new, clean pipe to connect tank to power unit; do not use old or existing pipe.
SECTION V – AUTOWASH
HOW THE SYSTEM WORKS
The auto wash is an automatic device that controls washdown of the vacuum
tank. The system is built to connect to an electrically operated water valve
connected to the facility water source. The outlet of the water valve is
connected to a spray nozzle mounted in the Ramvac vacuum tank. When the vacuum
pumps are turned off, a “delay” timer is started and begins to count down. The
“delay” timer is set for
2 hours and when the timer reaches zero, power is applied to the autowash
solenoid for 2 minutes to supply water to the spray nozzle. If the vacuum
system is turned on during that period, the auto wash process will immediately
cease.
SYSTEM SCHEMATIC
- Connects to waste piping (see SECTION IV – PLUMBING – DRAIN AND WASTE DETAILS)
- Connects to operatories* (see SECTION IV – PLUMBING – FACILITY VACUUM LINE FROM OPERATORIES)
- Connects to vacuum unit** (see SECTION IV – PLUMBING – TANK TO VACUUM POWER UNIT LINE)
- Facility water supply piping
- Connects to Aeras control INFINITANK (opt.) connector (see SECTION III – AERAS CONTROL – CONNECTIONS)
- Connects to Aeras control TANK FLOAT (opt.) connector (see SECTION III – AERAS CONTROL – CONNECTIONS)
- Connects to Aeras control AUTO WASH (opt.) connector (see SECTION III – AERAS CONTROL – CONNECTIONS)
*Tank inlet for 50-gallon tank is 3″.
**Tank outlet for 50-gallon tank is 3″. Tank outlet for InfiniTank is 2″.
SECTION VI – ACCESSORIES
VACUUM TANK INSTALLATION MANUAL
TANK VACUUM PLATFORM
TANK 12″ HEIGHT EXTENDER
TANK SUSPENSION HOOKS
VACUUM TANK INSTALLATION MANUAL
NOTES:
LIMITED WARRANTY
AERAS VACUUM TANK FROM RAMVAC
DENTALEZ and its employees are proud of the products we provide to the dental
community. We stand behind these products with a warranty against defects in
material and workmanship as provided below and have our own in-house repair
facility to service our products.
In the event that you experience difficulty with the application or operation
of any of our products, please contact our customer service department at our
expense at (866) DTE-INFO.
If we cannot resolve the issue by telephone, we will arrange for a
representative to contact you or suggest that the product be returned to our
factory for inspection.
If product return or repair is required, we will provide you with a Return
Authorization number and shipping instructions to return the product to the
proper facility. If the product is under warranty, we will ask you to provide
proof of purchase such as a copy of your invoice. Please be sure to include
the Return Authorization number on the package you are returning. Products
returned without a return authorization number cannot be repaired.
Freight costs for product returns are the responsibility of the customer.
Products under warranty will be repaired or replaced, at our sole discretion,
and returned at our expense. Products outside the warranty limits will be
repaired and returned with costs invoiced to the customer. We are not
responsible for shipping damages. We will, however, help you file a claim with
the freight carrier. Written repair estimates are available.
SCOPE OF WARRANTY
DENTALEZ warrants all Aeras Dry Vacuums to be free of defects in material and
workmanship, under normal usage and service.
DENTALEZ’s obligations under this warranty is limited to the repair or
replacement, at our option, of the defective parts which are reported to
DENTALEZ during the applicable warranty period.
APPLICABLE WARRANTY PERIOD
Aeras Dry Vacuums|
6 Years/4,200 hours from the date of installation
---|---
Aeras Dry Vacuums with Aeras Smart Monitoring Service| 8 Years /5,600 hours
from the date of installation provided the following additional terms
applicable to this 8 years extended warranty are met:
–As Smart Monitoring Service must be continuously maintained throughout the
eight (8) year warranty period
–Aeras Vacuum must be serviced and maintained by DENTAL-EZ or by an authorized
DENTAL-EZ dealer in accordance with DENTALEZ’s maintenance schedule and as
provided in the manual
–Only genuine Ramvac parts are used for all repairs and maintenance
Rotary Vane Pumps| 10 Years from date of installation
OWL Touch| 2 Years from the date of installation
EXCLUSIONS
This warranty covers manufacturing defects only and does not cover defects resulting from abuse, negligence, alterations, accidents, misuse, improper handling, cleaning, care or maintenance. Damage resulting from the use of chemicals, cleaners, disinfectants [or sterilization] is not covered under this warranty nor is normal wear and tear. Failure to follow or observe our operating/use, maintenance, and/or installation instructions voids this warranty as does [installation or other work performed by anyone other than DENTAL-EZ or its authorized dealers] the use of non-authorized parts and repairs made by an unauthorized repair facility.
EXCLUSIVE REMEDY
Liability is limited to repair or replacement of the defective part(s) at our
sole discretion. All other liabilities, in particular liability for damages,
including, without limitation, special, exemplary, consequential or incidental
damages are excluded and DENTAL-EZ shall not be liable for loss of profits or
loss of use.
AUTHORIZATION
THIS WARRANTY IS DENTALEZ’S ONLY WARRANTY AND IS IN LIEU OF ALL OTHER
WARRANTIES, EXPRESS OR IMPLIED, .INCLUDING ANY IMPLIED WARRANTIES OF
MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. NO EMPLOYEE,
REPRESENTATIVE OR DEALER IS AUTHORIZED TO CHANGE THIS WARRANTY IN ANY WAY OR
TO GRANT ANY OTHER WARRANTY.
WARRANTY REPAIRS
Parts repaired or replaced on a product that is in warranty will be warranted
for the duration of that product’s original warranty.
© 2021 DENTALEZ Alabama, Inc. DENTAL-EZ is a registered trademark and Aeras is
a trademark of DENTALEZ, Inc.
Ramvac is a registered trademark and OWL Touch is a trademark of Ramvac Dental
Products, Inc.
DENTALEZ Alabama, Inc.
2500 Highway 31 South
Bay Minette, AL 36507
T: 866-DTE-INFO
DENTALEZ.COM
© 2021 DENTALEZ, Inc. DENTAL-EZ is a registered trademark & Aeras is a
trademark of DENTALEZ, Inc.
Ramvac is a registered trademark of Ramvac Dental Products, Inc.
PN: 7717-006A
January 2021
Printed in USA