DENTALEZ 263758 Hiflo Swivel Connectors
Before operating the swivel connector, carefully read and follow all of these instructions. Save these instructions for future reference and observe all warnings and cautions.
INDICATIONS FOR USE
Swivel connectors are used to adapt the handpiece to the dental unit delivery tubing. Drive air, coolant air, water, and as applicable fiber optic lighting is supplied through the swivel.
SPECIFICATIONS
NOTE: All swivel connectors listed above, except 263758, contain an anti- retraction valve to prevent the backflow of water into the handpiece tubing. See Cleaning & Sterilization for special guidance regarding cleaning and maintenance.
DESCRIPTION
The Fiber Optic (F.O.) 4-Line HiFlo Swivels are designed to connect to ISO Type B 4-Hole Fiber Optic Tubing. The Fiber Optic (F.O.) 6-Pin ISO Type C, HiFlo Swivels are designed to connect to StarBright® Fiber Optic Handpiece Tubing or standard (non-Star) ISO C tubing.
SWIVEL INSTALLATION
Note: Use an air pressure gauge to adjust the pressure to the appropriate psi as recommended in the specified handpiece manual (30-34 for HiFlo swivels). The handpiece must be connected prior to taking pressure reading.
-
Slide tubing coupler back until the end of the ISO connector is exposed.
-
Align and insert fittings of the swivel into the ISO connector holes of the tubing.
-
Slide tubing coupler forward and thread clockwise until hand tight against the swivel.
-
Use the disconnect wrench, StarDental Part #203048, to tighten the swivel to the tubing coupler. Be sure to turn the tubing coupler, while holding the swivel in place. DO NOT OVER-TIGHTEN. Note: The swivel tubing connection may loosen over time, resulting in water leaks and/or loss of power. Check for tightness periodically.
-
Snap the handpiece onto the swivel.
-
Adjust water volume at the delivery head to obtain a satisfactory spray mist.
-
Slide tubing coupler back until the end of the ISO connector is exposed. The lamp controller must be putting out 3.1-3.3 volts DC.
-
Use the digital multimeter to verify that the positive female contact is above the larger hole (exhaust port) at the tubing coupler. If the polarity is incorrect, then reverse the connection of tubing wires at Lamp Control electrical connection ports.
NOTE: Step 8 is not applicable to StarDental Tri-Lamp Controller. -
Align and insert fittings of the swivel into the ISO connector holes of the tubing. Slide tubing coupler forward and thread clockwise until hand tight against the swivel.
-
Verify that the LED swivel operates. If it does not light, immediately shut off and recheck connection and polarity per steps 7 and 8.
CLEANING & STERILIZATION
TROUBLESHOOTING
- Problem: WATER LEAKAGE (Refer to O-Rings & Rear Seal Replacement)
- Cause: Brittle swivel silicone rubber rear seal
- Correction: Replace black silicone rubber seal: for 4-line HiFlo FO swivel, replace StarDental Part #263734 and for 6-pin HiFlo swivels, replace StarDental Part #263770.
- Cause: Swivel O-ring worn
- Correction: Replace both O-rings at the front of any swivel with StarDental Part #259884
- Problem: SWIVEL WILL NOT DISENGAGE FROM HANDPIECE
- Cause: Buildup of debris
- Correction: Place a few drops of DentaLube II, StarDental Part #262539, into the threaded area of the bearing/spring assembly to disengage the handpiece. (Refer to Cleaning.) To prevent reoccurrence, clean the swivel handpiece connector.
- Problem: NO LIGHT OR INTERMITTENT LIGHT
- Cause: Corrosion on lamp contact pins
- Correction: Remove lamp; clean contact pins of the lamp. (Refer to Cleaning.)
- Cause: Oil on lamp contact spring
- Correction: Remove lamp from tubing; clean spring, front and rear sockets, and lamp contact with isopropyl alcohol.
- Problem: NO LIGHT
- Cause: Lamp burned out in 6-Pin swivel
- Correction: Replace the lamp. (Refer to Lamp Replacement.)
- Problem: YELLOWISH LIGHT
- Cause: Debris on optical surfaces
- Correction: Clean all exposed optical surfaces.(Refer to Cleaning.)
Note: If a problem persists, contact the dealer for further assistance; or, contact
LAMP REPLACEMENT
O-RINGS AND REAR SEAL REPLACEMENT
Periodically replace the swivel O-rings with StarDental Part #259884. Seals wear with repeated use. Once a year or as needed, replace the black silicone rubber seal:
- For the 4-line HiFlo FO swivel, replace it with StarDental Part #263734
- For 6-pin HiFlo swivels, replace with StarDental Part #263770
REPLACEMENT PART NUMBERS
ALWAYS INSIST ON STAR DENTAL PARTS FOR BEST PERFORMANCE.
IMPORTANT NOTICE ABOUT RETURNING PRODUCT TO DENTALEZ INC.
Anyone wishing to return any DENTALEZ Inc. product to a DENTALEZ facility MUST
obtain the proper RETURN AUTHORIZATION NUMBER from the Customer Service
Department. The RA Number must appear clearly marked on both the outside and
inside of the returned product carton. When calling for a return authorization
number you will be asked to provide the following information:
To Obtain A Return Authorization Number Call 1-866-DTE-INFO
- Product Name, Model Number or Part Number
- Serial Number of Product
- Dealer Return Purchase Order Number
- Reason For Return
- Copy of Invoice Showing Purchase
Items will be evaluated. Return Authorization Number is not a guarantee for
credit. Your cooperation with this procedure will help us expedite processing
of any returns.
FOR NEW PRODUCTS, MANUALS, AND TECHNICAL INFORMATION, CALL 1-866-DTE-INFO OR
VISIT DENTALEZ.COM
LIMITED WARRANTY
O-Rings, Seals, and Lamps are specifically excluded from this warranty. DENTALEZ warrants Swivel Connectors for 1 year from the date of purchase*. Please note the following additional terms of our warranty policy:
- The warranty covers manufacturing defects only and does not cover defects resulting from abuse, accidents, misuse, improper handling, cleaning, care or maintenance or the failure to follow or observe our operating/use, maintenance, and/or installation instructions. Damage resulting from the use of chemicals, cleaners, disinfectants or sterilization is not covered under this warranty nor is normal wear and tear. Failure to use authorized parts or an authorized repair facility voids this warranty.
- Liability is limited to repair or replacement of the defective part(s) or products at our sole discretion. All other liabilities, in particular liability for damages, including, without limitation, special, exemplary, consequential, or incidental damages are excluded
- THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. NO EMPLOYEE, REPRESENTATIVE, OR DEALER IS AUTHORIZED TO CHANGE THIS WARRANTY IN ANY WAY OR TO GRANT ANY OTHER WARRANTY.
- Provided conditions in the warranty are met.
WARRANTY REGISTRATION
Please visit the website below to register your StarDental product. This will
ensure proper registration and prompt and accurate service should any of your
purchases need evaluation or repair during the warranty period.
StarDental Products:
www.dentalez.com/warranty
Address:
DENTALEZ, Inc., StarDental ® Division 1816 Colonial Village Lane Lancaster,
PA 17601 USA 066234 Rev A Date of Issue: 2022-08-19