RADIOLUCENT
Will not shadow X-rays.
FUNCTIONAL DESIGN
Will not warp or become brittle. Unaffected by chloroform, germicides or irrigating solutions.
WRAPPED STERILIZATION
A standard sterilization packaging material must be used. Ensure that the pack is large enough to contain the Visi-Frame without stressing the seals
GRAVITY TYPE AUTOCLAVE
- Minimum of 15 minutes at 132° C: Dry time: minimum 30 minutes
- Minimum of 30 minutes at 121° C: Dry time: minimum 40 minutes
PRE-VACUUM AUTOCLAVE
- Minimum of 4 minutes at 132° C Dry time: minimum 30 minutes
NOTE: Do not exceed 135° C
Dry using the sterilization dry cycle and allow to cool to room temperature in
bag prior to use. If moisture
is present, increase your sterilizer dry cycle appropriately