hygiena KIT230215 SL GMO A2704-12 Soya Detection Kit Instructions
- June 1, 2024
- Hygiena
Table of Contents
- Introduction
- Intended Use
- Principle of PCR detection
- Contents
- Additional Materials, Reagents and Devices Required
- General precautions
- Sampling and handling
- Protocol
- Data analysis
- Troubleshooting
- Stability and Storage
- Specifications
- Quality control
- Ordering information
- Supplementary Information
- Change Index
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
Product Instructions
foodprooffi SL GMO A2704-12 Soya Detection Kit
Revision A, March 2024
PCR kit for the qualitative detection of A2704-12 DNA using real-time PCR
instruments.
Product No. KIT230215
Kit for 50 reactions for a maximum of 48 samples
Store at -15 to -25 °C
Introduction
Many countries worldwide have implemented legislation for the use, cultivation
and labeling of foodstuffs containing genetically modified organisms (GMOs).
These regulations allow the usage of GMOs under certain conditions, often
including a defined threshold for labeling or where the import and use of GMOs
is prohibited.
Thus, reliable methods for the detection and identification of GMOs in food
and feed are required.
With the foodprooffi SL GMO product line, Hygienafi Diagnostics offers a wide
range of easy and reliable assays for the detection of GMOs. The foodproof SL
GMO A2704-12 Soya Detection Kit allows fast, safe and easy detection in food
and feed samples.
Intended Use
The foodproof SL GMO A2704-12 Soya Detection Kit is designed to detect GM-Soy
A2704-12 event-specific gene in various processed foods, raw materials, feed,
seeds, etc. The kit detects A2704-12 DNA directly and is, therefore, highly
sensitive and specific.
This kit provides a real-time PCR Master Mix with enzyme components and the
specific primer/probe set for rapid testing by real-time PCR assay, as well as
the internal positive control (IC) system for reliable results.
Principle of PCR detection
The foodproof SL GMO A2704-12 Soya Detection assay is a qualitative Duplex
real-time PCR test, which contains the amplification of GM Soy A2704-12
specific gene and the internal positive control (IC) using specific primers
and probes labeled with different fluorescent dyes. The target sequences are
detected through the FAM and HEX (VIC) channels, respectively.
The primer and probe mixture provided is based on the so-called TaqManfi
principle. During PCR amplification, forward and reverse primers hybridize to
the target DNA. A fluorogenic probe is included in the same reaction mixture,
which consists of an oligonucleotide labeled with a 5‘-reporter dye and a
downstream 3‘-quencher.
During PCR amplification, the probe is cleaved and the reporter dye and
quencher are separated. The resulting increase in fluorescence can be detected
on a range of real-time PCR platforms.
The monitoring of the fluorescence intensities during the real-time PCR allows
the detection of accumulating product without re-opening the reaction tubes
after the PCR run.
The kit minimizes contamination risk and contains all reagents needed for
detection (except for PCR-grade H2O).
3.1 Internal Amplification Control
This kit contains the Internal Positive Control (IC) as a PCR inhibition
Control. The IC allows the user to detect and control possible PCR inhibition.
The IC reagents are included in the primer/probe mixture and the IC is co-
amplified with target DNA from the sample. The results can be visualized in
the HEX (VIC) channel.
3.2 Carry-over prevention using UNG system
The foodproof SL GMO A2704-12 Soya Detection Kit utilizes the UNG system.
Carry-over contamination between PCR reactions can be prevented by including
uracil-N-glycosylase (UNG) in the reaction mix. UNG can only prevent carry-
over from PCR reactions that include deoxyuridine triphosphate (dUTP) in the
original PCR reaction.
Contents
This kit is intended for 50 reactions, including controls.
Table 1: Kit Contents
Reagent | Cap Label | Volume | Description |
---|---|---|---|
2x Real−time PCR Master Mix | 2xM | 625 μL | Buffer containing dNTPs, MgCl2, UNG |
and Taq DNA polymerase
Primer / Probe Mixture| A2704−12| 200 μL| Primer/ probe mixture:
· GM Soya A2704−12−specific primer and probe
· IC−specific primer and probe
· IC DNA
Control DNA| C| 50 μL| Positive control DNA
Additional Materials, Reagents and Devices Required
- Disposable powder-free gloves and laboratory coat
- Pipettors (0.5 to 10 µL, 2 to 20 µL, 20 to 200 µL, 200 to 1,000 µL)
- Sterile aerosol-barrier pipette tips
- Ice or benchtop cooler
- Vortex mixer
- Clean bench area or PCR box
- Tabletop centrifuge with rotor for 2 mL reaction tubes
- Real-time thermal cycler with FAM and HEX (VIC) detection channels
- Disposable polypropylene microcentrifuge tubes for PCR
- PCR-grade H2O
- For DNA Extraction: foodproof Sample Preparation Kit
General precautions
-
Store extracted positive material (samples, controls and other amplicons) away from all other reagents and add it to the reaction mix in a separate area.
-
Thaw all components thoroughly on ice before starting the experiment.
-
When thawed, mix the components and centrifuge briefly.
-
Do not pipette by mouth.
-
Do not eat, drink, smoke, apply cosmetics or handle contact lenses in laboratory work areas.
-
Do not use a kit beyond its expiration date.
-
Safety Data Sheets (SDS) can be found at www.hygiena.com/documents.
-
Use disposable gloves, laboratory coats and eye protection while handling samples and reagents.
Thoroughly wash hands afterward. -
Dispose of all samples and unused reagents in compliance with local regulations.
-
Specimens should be considered potentially infectious and handled in a biological cabinet in accordance with Biosafety Level 2 or other appropriate biosafety practices.
-
Clean and disinfect all sample or reagent spills using a disinfectant such as 0.5% sodium hypochlorite or other suitable disinfectant.
-
Avoid contact of specimens and reagents with the skin, eyes and mucosa. If contact occurs with skin, eyes or mucosa, immediately flush with water and seek medical attention.
-
Use of this product should be limited to personnel trained in laboratory DNA amplification techniques.
-
To avoid carry-over contamination with PCR product or control DNA, please note the following points:
1. Be careful not to contaminate the Primer/Probe Mixture and 2x real-time PCR Master Mix with other PCR products or Control DNA through pipetting. To prevent contamination, the use of aerosol-barrier tips is recommended.
2. Open and close all sample tubes carefully. Avoid splashing or spraying PCR samples.
3. It is important to have designated lab areas where PCR reactions are set up, preferentially separated in space from the areas where PCR reactions are analyzed.
4. The laboratory process must be one-directional; it should begin in the Extraction Area and move to the Amplification and Detection Area. Do not transport samples, equipment and reagents to the areas where you performed previous steps.
Sampling and handling
7.1 Sample Collection
Various processed food, raw material, feed and seed samples are routinely
examined.
7.2 Sample Storage
The assay sensitivity can be reduced if you routinely freeze the samples
before testing or store them for an extended period of time. Avoid repeated
freezing and thawing of samples, which may lead to DNA degradation and
decreased sensitivity.
7.3 Nucleic Acid Extraction
Carry out DNA isolation according to the extraction kit’s product
instructions. For more information, please see
www.hygiena.com.
Protocol
8.1 DNA Isolation
Hygiena Diagnostics provides sample preparation kits suitable for all kinds of
foods and raw materials.
(See 5. “Additional Required Materials, Reagents and Devices”)
8.2 Preparing the PCR
To prevent the risk of contamination with foreign DNA, we recommend that all
experiment steps be performed in a PCR cleanroom or separated environment
area. Aerosol-barrier pipette tips are recommended for each step.
8.2.1 Thawing the Kit Components
The use of ice or a benchtop cooler is recommended during experiments to
maintain enzyme activity.
8.2.2 Prepare Reaction Master Mix
Each reaction has a total volume of 25 µL; the volume of the DNA sample is 5
µL.
- Prepare the reaction mixture according to Table 2 below.
Table 2: PCR reaction mixture Composition| Volume
---|---
Primer / Probe Mixture| 4 μL
2x Real−time PCR Master Mix| 12.5 μL
PCR−grade H2O| 3.5 μL
Total| 20 μL - Add 5 µL of extracted DNA sample into the tube.
8.2.3 Prepare Control Amplification Reactions
• Positive control amplification: Add 5 µL of Control DNA instead of sample
DNA.
• Negative control amplification: Add 5 µL of PCR-grade H2O instead of sample
DNA
8.2.4 Mixing
Mix the reagents in the PCR reaction tubes by tapping a minimum of 5 times.
Briefly centrifuge the tubes to remove any air bubbles or drops inside the
cap.
8.3 Amplification
- Program your real-time PCR instrument according to the manufacturer’s manual.
- Create a temperature-time profile on your instrument as follows in Table 3.
Table 3: Temperature Time Profile
Temperature | Time | Cycle(s) |
---|---|---|
50 ˚C* | 2 min | 1 |
95 ˚C | 10 min | 1 |
95 ˚C | 15 seconds | 45 |
60 ˚C** | 1 min |
*The UDG activation step inhibits contamination by PCR product.
**Detect the fluorescence at this step.
Data analysis
The fluorescence curves are analyzed in FAM and HEX (VIC) fluorescence
detection channels (see Table 4).
You can predict the presence or absence of the target gene in your samples by
analyzing the real-time PCR results.
Table 4: Specific Detection on Fluorescence Channel
Target Gene | Fluorophore |
---|---|
A2704−12 | FAM |
IC | HEX (VIC) |
9.1 Interpretation of Results
- The signal is considered to be positive if the corresponding fluorescence accumulation curve crosses the threshold line.
- Results are accepted as relevant if both positive and negative amplification controls pass.
- IC: When amplifying a target sample with a high copy number, the IC may not produce an amplification plot. This does not invalidate the test and should be interpreted as a positive experimental result.
Table 5: Interpretation of Results
| Positive Control| Negative Control| A2704-12|
IC| Interpretation
---|---|---|---|---|---
FAM| HEX (VIC)
Case 1| +| −| +| +| A2704−12 gene is detected.
Case 2| +| −| +| −*
Case 3| +| −| −| +| A2704−12 gene is not detected.
Case 4| +| −| −| −| Invalid result; retest
Case 5| +| +| +/−| +/−
Case 6| −| +/−| +/−| +/−
*Detection of the Internal Amplification Control in the respective channel is not required for a positive result.
A high copy number of the target gene can lead to reduced or absent Internal
Amplification Control signal.
Troubleshooting
Situation | Possible cause | Recommendation |
---|---|---|
**** Negative control samples are positive. | Carry−over contamination. | · |
Exchange all critical solutions.
· Repeat the analysis of all tests with fresh aliquots of all reagents.
· Take measures to detect and eliminate the source of contamination.
No signal is detected for amplification positive controls.| Incorrect
programming of the real−time PCR instrument.| · The PCR should be repeated
after checking the programming of instruments, storage conditions and the
expiration date.
The kit reagents have expired.
Kit components have not been stored according to the manufacturer’s
instructions.
· Incorrect PCR reaction.
· Pipetting errors.
· Omitted reagents.| · The PCR should be repeated after checking for correct
pipetting scheme and reaction setup.
No signal is detected for IC in HEX (VIC) channel and A2704−12−specific gene
in FAM channel.| PCR inhibitors are present at a high concentration.| · DNA
extraction should be repeated.
Stability and Storage
Store the kit at -15 to -25 °C through the expiration date printed on the label.
Specifications
-
Sensitivity
Limit of detection (LOD) at 0.05% -
Specificity
100% exclusivity for non-target genes
Quality control
In compliance with the Federal State Institution of Science “Central Research
Institute of Epidemiology” ISO 13485
– certified Quality Management System, each lot of foodproof SL GMO A2704-12
Soya Detection Kit has been tested against predetermined specifications to
ensure consistent product quality.
Ordering information
Product | Order No. | # Tests |
---|---|---|
foodproof SL GMO A2704−12 Soya Detection Kit | KIT230215 | 50 reactions |
foodproof Sample Preparation Kit III | KIT230174 | 50 reactions |
Supplementary Information
15.1 Ordering Information
Hygiena Diagnostics offers a broad range of reagents and services. For a
complete overview and for more information, please visit our website at
www.hygiena.com.
15.3 Trademarks
foodprooffi, microprooffi, vetprooffi, ShortPrepfi, StarPrepfi, RoboPrepfi and
LyoKitfi are registered trademarks of Hygiena Diagnostics GmbH. Hygienafi is a
registered trademark of Hygiena. Other brand or product names are trademarks
of their respective holders.
Other brand or product names are trademarks of their respective holders.
15.4 Contact and Support
If you have questions or experience problems with this or any other product of
Hygiena Diagnostics GmbH, please contact our Technical Support staff
(www.hygiena.com/support). Our scientists
commit themselves to providing rapid and effective help. We also want you to
contact us if you have suggestions for enhancing our product performance or
using our products in new or specialized ways. Such customer information has
repeatedly proven invaluable to us and the worldwide research community.
15.5 Reference Number
The reference number and original Hygiena Diagnostics GmbH article number: Z
722 01
Change Index
Version 1, July 2015
First version of the package insert.
Revision A, March 2024
Rebranding and new layout.
Z 722 01 20 -> INS-KIT230215-RevA
Hygienafi®
Camarillo, CA 93012
USA
diagnostics.support@hygiena.com
Manufactured by
Hygiena Diagnostics GmbH
Hermannswerder 17
14473 Potsdam
Germany
www.hygiena.com
References