Sonel AHV-3 Adapter Standard DC Voltage Measuremen User Manual
USER MANUAL
Version 1.00 21.02.2023
| · AHV-3 adapter is not a stand-alone measuring device. It is an accessory designed to coopera- te with a dedicated meter ( CMP-1015-PV and selected devices that measure DC voltage (with input impedance of 10 MΩ for V DC measure- ment ). The rules for using the adapter and per- forming measurements with it can be found in the meter’s manual. Please get acquainted with it before proceeding with measurements.
· The manufacturer reserves the right to intro- duce changes in the design, accessories and technical data of the adapter.
---|---
General information
Safety symbols
The following international symbols are used in the device and/or in this manual:
Measurement categories according to EN IEC 61010-2-030
in circuits directly connected to low volt- age installations,
-
CAT III – concerns measurements performed in buildings installations,
-
CAT IV – concerns measurements performed at the source of low voltage
-
Safety
AHV-3 adapter is used to measure high voltages in collaboration with a master meter. In order to provide con- ditions for correct operation, the following recommenda- tions must be observed:
- Before you proceed to operate the device, acquaint yourself thoroughly with this manual and observe the safety regulations and specifications defined by the producer
- Any application that differs from those specified in the present manual may result in a damage to the device and constitute a source of danger for the
- The adapter should be operated only by suitably qual- ified persons having the necessary permissions to carry out measurements on electrical systems. Op- erating the adapter by unauthorised personnel may result in damage to the device and constitute a source of danger for the
- The adapter must not be used for networks and de- vices in areas with special conditions, e.g. fire-risk and explosive-risk
- It is unacceptable to operate the device when:
it is damaged and completely or partially out of order,
its cable insulation is damaged,
it was stored for an excessive period of time in dis- advantageous conditions
(e.g. excessive humidity). After moving the device from a cool to a warm
place with a high level of relative humidity,
do not start measurements until the device is warmed up to the ambient
temperature (ap- proximately 30 minutes)
Description of outputs
INPUT-H – cable to connect the adapter to DC+ high voltage
INPUT-L – cable to connect the adapter to DC- high voltage
OUTPUT-H – connection socket for the cable connecting
OUTPUT-L – connection socket for the cable connecting the adapter with
COM socket of the master meter.
ATTENTION!
- The adapter may be damaged when connect- ed to the voltage exceeding 3000 V
- First connect the adapter to the meter, and then the adapter to the tested object.
Cleaning and maintenance
The adapter may be cleaned with a soft, damp cloth using all-purpose
detergents. Do not use any solvents or cleaning agents which may damage the
casing (powders, pastes, etc.).
The electronic system of the adapter does not require maintenance
Storage
Before storage, make sure that the adapter is dry.
Dismantling and utilization
Worn-out electric and electronic equipment should be gathered selectively,
i.e. it must not be placed with waste of another kind.
Worn-out electronic equipment should be sent to a collection point in
accordance with the regulations valid in a given region.
Before the equipment is sent to a collection point, do not dismantle any
elements.
Observe local regulations concerning disposal of packages, waste batteries and
accumulators
Technical data
Parameters of compatible master meter
-
a) Input impedance………………………………………… 10 MΩ (V DC)
Basic technical data -
measuring range……………………………………….. 0…3000 V DC
-
ratio (INPUT / OUTPUT)…………………………………………… V / 1 V
-
przelicznik wyniku
-
CMP-1015-PV meter (HV DC measurement)………………… 1 x v.
master meter (V DC measurement) 10 x m.v
Other technical data
- insulation type acc. to EN 61010-1………………………………. double
- measurement acc. to EN IEC 61010-2-030…………. III 3000 V DC
- protection class to EN 60529……………………………………. P54
- test leads length……………………………………………………. 2 m
- dimensions……………………………………….. 130 x 65 x 26 mm
- weight (with cables)………………………………………………… 275 g
- operating temperature…………………………………. .. -5°C…+40°C
- storage temperature………………………………….. .. -20°C…+60°C
- humidity……………………………………………………. 20%…90%
- reference humidity…………………………………………………… 30%
Standard accessories
The standard set of equipment supplied by the manufac- turer includes:
- 2x pin probe 3 kV – WASONBL3KV ,
- test leads set for CMM/CMP – WAPRZCMX1 ,
- M-13 carrying case – WAFUTM13 ,
- declaration of verification
The current list of accessories can be found on the manu- facturer’s website.
Manufacturer
The manufacturer of the device and provider of guar- antee and post-guarantee
service:SONEL S.A.
Wokulskiego 1158-100 Świdnica Polandtel. +48 74 858 38 60fax +48 74 858 38 09
E-mail: export@sonel.pl Web page:
www.sonel.pl
| ATTENTION!
Service repairs must be performed only by the manufacturer.
---|---