NOVA LONGMANN PAP-REMS-2 Inline Standby DC UPS 12V User Guide
- June 3, 2024
- nova
Table of Contents
NOVA LONGMANN PAP-REMS-2 Inline Standby DC UPS 12V
Description
No. | Description |
---|---|
1 | 4 Bar LED Power Level Indicator |
2 | Blue LED ‘charging’ Indicator |
Green ‘Full charge’ Indicator
3| Press ‘HERE’ for Battery Level
4| Front of Device
5| Active Air vent – do not obstruct
6| Output: Cable to PAP Machine (Cable Included)
7| LED ‘backup is active’
8| Input: Use AC/DC Adaptor-Referenced below under
“Use”
9| Power Switch
10| Important Device details
Use
This Device is designed for Philips sleep machines.
Supports:
- Philips Respironics Dorma
- Philips Respironics System One REMStar (60 series)
- Philips Respironics System One REMStar SE™
All sleep machines use the following wall adapters only:
- REF: 1091398
- REF: 1091399
The details will be under your machine and on the wall adapter/charger
Backup power time
TIME: 12 + HOURS DEVICE WITH HUMIDIFIER SETTING 0 OR DISCONNECTED**
TIME: 3 TO 5 HOURS DEVICE + HUMIDIFIER SETTING 1* @25 DEGREES CELSIUS
DEVICE CHARGE TIME: 6 TO 11+ HOURS @25 DEGREES CELSIUS
Operation Instructions
- Insert your PAP machine wall adapter into the input port of the Longmann device and the Longmann device cable into your PAP machine.
- If your pap machine has an option under settings for ‘auto on’, make sure it is active under menu-settings.
- Do not cover the device. Make sure the Longmann device is placed at least 10cm away from your PAP machine.
Humidifier settings
This device has been set to work with the humidifier setting 1, and can also power settings 2, 3, 4, and 5 with significantly reduced backup time.
Parameters for humidifier setting
· Room temperature is above 28 degrees – the backup time increases to more than 5 hours, setting 1
· Room temperature is below 24 degrees – the backup time decreases
substantially
· Mask not on face – the Longmann Device will shut down within 1 hour –
the device will reset and restart when mains power resumes
· Water level is above the waterline mark – the backup time decreases
· Water Humidification container is not properly secured with Phillips
PAP machine – the backup time decreases substantially
· Air filters are not changed regularly – the backup time decreases
substantially
· Initial water temperature below 18 degrees – the backup time decreases
substantially
Humidifier Troubleshooting
On UPS Backup Mode: If your sleep machine was initially set to higher than humidifier setting 1- in the event of a power failure after 5 minutes of runtime your sleep machine will beep “Low power alert”, In this case, change the humidifier setting to setting 1.
If your sleep machine does not beep the Longmann device will automatically switch the humidifier off but still keep the sleep machine on. When mains power resumes, the sleep machine will resume its humidifier settings.
Warranty
- 3 Years from the date of purchase
Safety instructions
This product is not waterproof and is suitable for indoor use only. Only for use with the original Philips AC/DC wall adaptor, this device is maintenance- free; do not try to disassemble this device.
Care and Handling
Do not clean or wipe the device with any abrasive material or chemical. Use only dry cloth to dust the device. No liquid cleaning materials are allowed to be used. Do not drop or throw this device. This product should only be operated by an adult.
Warning
Use with caution. This product is not a toy and should be kept away from
children. Lithium Ion Batteries are explosive if damaged or used incorrectly.
Check your local laws for correct battery disposal. All care should be taken
in using this device the same with all electrical rechargeable battery
devices.
Technical Specifications & Product Identifiers
If you have any questions, please contact your local representative
- Device Model NOVA PAP-REMS-2
- Reference P870311-1
- Catalogue Number P173095
- Barcode Number 6946322429205
- Batch Number Bottom of device
- Manufactured Date Bottom of device
- Serial Number Bottom of device
- Battery Use Type Rechargeable
- Battery Output Voltage 12V Constant
- Battery Type Lithium-Ion
- Battery Capacity 126 Watt-hours
- Charging Philips AC/DC WALL ADAPTER REF: 1091398, REF: 1091399, (Optional Longmann Solar panel, and Longmann 12/24v Car Charger for version3)
- Reset Button Automatic reset
- Display Press Button 4 LED Power Indicator, (Solar Charging Indicator If Applicable version 3)
- Visible LED’s 6 (Solar option add 2)
- Input Philips WALL ADAPTER REF: 1091398, REF: 1091399, (Longmann Solar panel, Longmann 12/24v Car Charger for version 3)
- Output Philips cable for Philips REF: 1091398, REF: 1091399
- Aviation Regulated Allowed under FAA Regulations, Please call your airline to confirm.
- Dimensions 127mm x 99mm x 69mm (Height)
- Weight 920 grams e
Extended Specifications
Contents
- Rechargeable lithium-ion battery device
Compatibility
- Supports Dorma range, System One (60 series)
- Utilizes Philips AC/DC WALL ADAPTER
- REF: 1091398, REF: 1091399. Only
Use with Non-Philips Healthcare Equipment
- No
Product Category
- Sleep Therapy
Product Type
- Sleep Therapy Accessories
Prescription
- Not required
Run Time
- 12 hours +
Recharge
- 6- 11+ hours at room temperature
Warranty
- Three years
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>