Sonel MRU-10 Earth Resistance Meter User Manual

June 4, 2024
Sonel

Sonel MRU-10 Earth Resistance Meter

Over view

Safety

The following international symbols are used in the Analyzer and in this manual:

| Warning;
See explanation in the manual| | Dual insulation (Protection class II)
---|---|---|---
| Do not dispose of with
other household waste| | Declaration of Conformity with EU directives

MRU-10 meter is designed for measuring parameters important for safety of electrical installations. Therefore, in order to provide conditions for correct operation and accuracy of obtained results, the following recommendations must be observed:

  • Before you proceed to operate the meter, acquaint yourself thoroughly with the present manual and observe the safety regulations and specifications provided by the producer.
  • MRU-10 meter is designed to measure earth resistance values. Any application that differs from those specified in the present manual may result in a damage to the device and constitute a source of danger for the user.
  • The meter must be operated solely by appropriately qualified personnel members holding required certificates for carrying measurements in electric installations. Unauthorized use of the meter may result in its damage and may be a source of serious hazard to the user.
  • Using this manual does not exclude the need to comply with occupational health and safety regulations and with other relevant fire regulations required during the performance of a particular type of work. Before starting the work with the device in special environments, e.g. potentially fire-risk/explosive environment, it is necessary to consult it with the person responsible for health and safety.
  • It is unacceptable to operate the device when:
  • a damaged meter which is completely or partially out of order,
  • a meter with damaged insulation,
  • a meter stored for an excessive period of time in disadvantageous conditions (e.g. excessive humidity). If the meter has been transferred from a cool to a warm environment with a high level of relative humidity, do not start measurements until the meter is warmed up to the ambient temperature (approximately 30 minutes).
  • Before measurement make sure that test leads are connected to appropriate measuring terminals.
  • Do not power the meter from sources other than those listed in this manual.
  • The inputs of the meter are protected electronically against overload e.g. due to having been connected to a live circuit, for all combinations of inputs – up to 276 V for 30 seconds.
  • Factory calibration does not include the resistance of the test leads. The result displayed by the meter is a sum of the resistance of the measured object and the resistance of leads.
  • The device meets the requirements of standards EN 61010-1 and EN 61557-1, -5.

Note
Due to continuous product development, the manufacturer reserves the right to introduce changes to the functionality, appearance, accessories and technical data of the meter. Due to continuous development of the meter’s software, the actual appearance of the display, in case of some of the functions, may slightly differ from the display presented in this operating manual.

Turning the meter ON and activating screen backlight

To turn ON the meter, press ON/OFF button. The screen of the device will activate its all segments (self-test), and then it will briefly show the software version.

When the meter is ON, a short press on ON/OFF button will turn ON and then OFF the screen backlight;

To turn OFF the device, press and hold ON/OFF button for approx. 2 seconds. A screen indicating shut-ting off the device is shown briefly

Meter Configuration

Turn on the meter by pressing and holding DOWN button. After switching the meter ON, a screen is shown with network frequency settings Fn.

When the Fn screen is displayed, use UP and DOWN buttons to set the network frequency of 50 Hz or 60 Hz (default: 50 Hz default).

Short press START but-ton to accept the selected value. You will enter the screen with settings of audio messages bEEP.

Use UP and DOWN but-tons to turn the audio messages ON ( ) or

Short press START but-ton to accept the selected option. The screen of Auto- OFF time settings will be shown: oFF.

Use UP and DOWN but-tons to set the auto-off time at 300 s, 600 s, 900 s or “- – – -” (Auto-OFF inactive). Auto-OFF function is used to turn-off in-active meter after a pre-selected time, which is signalled by a beep

Short press START but-ton to accept the selected option. You will enter the screen with measuring voltage selection Un.

Use UP and DOWN but-tons to set the measuring voltage at 25 V or 50 V. The measuring voltage set relates to all measurement functions in the meter.

Short press START but-ton to accept the selected option and return to the network frequency selection screen Fn.

Long press (> 2 s) START button to the introduced changes and proceed to the measurement screen RE3P

Long press (> 2 s) ON/OFF button to turn off the meter without accepting the changes introduced at the current set-ting position

Short press (> 2 s) ON/OFF button to enter the measurement screen RE3P without accepting the changes introduced at the current setting position.

Measurements

Earth resistance measurements significantly different from other measurements performed to assess the protection against electric shock. They require a thorough knowledge of the structure of the earthing system, the phenomena occurring during the measurements and the skills of coping with adverse outdoor conditions. Earthing system tests/ measurements require adequate knowledge and measuring equipment, which will be able to maximally facilitate these examinations.

Measurement of interference voltages DC + AC

In measuring functions, before START button is pressed, the meter monitors the measuring volt-age at the terminals (be-tween E socket and S / H sockets) and the interference voltage value is displayed on the screen.

Additional information displayed by the meter

U N > 100V!, >100V and
a continuous beep U ,
NOISE!” and| Voltage at test terminals is higher than 100 V, the measurement is blocked.
---|---
U N xxV!, >40V and a continuous beep U ,
NOISE!” and| Where xx is the value of interference voltage. Voltage at test terminals is higher than 40 V but lower than 100 V, the measurement is blocked.
U N xxV!, >24V,
NOISE!” and| Where xx is the value of interference voltage. Voltage at test terminals is higher than 24 V but lower than 40 V, the measurement is blocked.
NOISE!”| The interference voltage is lower than 24 V, but with a high value – the measurement result may be affected by additional uncertainty.

Measurement of earth resistances with 3-pole method (RE3P)

For earth resistance measurements, the most commonly used method is the ‘technical method’, often called the potential drop method, or 3-pole method. During the measurement, the voltage drop at the earthing is measured with current flowing through it, then the Ohm’s law is used to calculate the resistance.

The scheme of measuring the earthing resistance with the ‘technical method’ is shown above. The figure shows the measurement of earthing resistance RE. To perform the measurement, use two additional earth contact probes (hereinafter probes):

  • H probe (current probe) to allow excitation of current flow in the circuit: tested earth electrode RE→ meter → H current electrode → earth → tested earth electrode;
  •  S probe (voltage probe) for measuring the voltage drop across the resistance of the measured earthing as a result of current flow.

Disconnect the tested earth electrode from the installation of the facility.

Turn on the meter using ON/OFF button. The meter enters the measurement function screen RE3P. The meter is in the mode of measuring the interference voltage between the test terminals. The measuring voltage is compatible with the volt-age selected when set-ting up the device.

  • Test leads should be connected to the measurement terminals in the device, as shown above.
  • The current probe H (driven into earth) should be connected to H socket of the meter.
  • The voltage probe S (driven into earth) should be connected to S socket of the meter.
  • The earth electrode being tested should be connected to E socket of the meter with the lead.
  • The earth electrode being tested and the current probe and the voltage electrode should be located in one line.

Press START.
The progress of the measurement is indicated by horizontal lines on the screen.

After completing the measurement, results are displayed for all the measurements carried-out: at the bottom of the screen, the main result is shown for RE, whereas the upper part of the screen shows additional results for RH. The result is displayed for 20 seconds. The result can be recalled by pressing UP button.

Use UP button to view the component results in the following order:

  • RH RS ER UN, where:
  • RH – resistance of probe H RS – resistance of probe S ER – additional uncertain-ty resulting from the probes
  • UN – interference voltage

Repeat the measurements (steps 4, 5, 6) moving the voltage probe a few meters – placing it farther and closer to the measured earth electrode.
If RE measurement results differ from one an-other by more than 3%, the distance of the cur-rent probe from the earth electrode being tested should be considerably increased and the measurements should be repeated.

Note:
Measurement of resistance-to-earth may be carried out if voltage of interferences does not exceed 24 V. Voltage of interferences is measured up to the level of 100 V, but above 40 V it is signalled as dangerous. The meter must not be connected to voltages exceeding 100 V.

  •  Particular attention should be paid to quality of connection between the object being tested and the test lead – the contact area must be free from paint, rust, etc.
    If resistance of probes is too high, RE earth electrode measurement will include an additional uncertainty. Particularly high uncertainty of measurement occurs when a small value of resistance-to-earth is measured with probes that have a weak contact with earth (such a situation occurs frequently when the earth electrode is well made and the upper soil layer is dry and slightly conductive). In such a case, the ratio of resistance of the probes to resistance of the tested earth electrode is very high and consequently, uncertainty of measurement that depends on this ratio is also very high. Then, you can make a calculations according to the formulas given in sec. 9.2 to estimate the influence of measurement conditions. You can improve the contact between the probe and soil, for example, by dampening with water the place where the probe is driven into earth, driving the probe into earth in a different place or using a 80 cm-long probe. Check also the test leads for possible insulation damage and for corroded or loosened connection between the banana plug and the test lead. In majority of cases the achieved measurement accuracy is satisfactory. However, you should always be aware of the un- certainty included in the measurement.

Additional information displayed by the meter

R E > 9999 Ω Measuring range is exceeded.

U N > 100 V, >100 V
and a continuous beep U ,NOISE!” and| Voltage at test terminals is higher than 100 V, measurement is blocked.
U N xxV, >40 V and a
continuous beep
U ,NOISE!” and ****| Where xx is the value of interference voltage. Voltage at test terminals is higher than 40 V, measurement is blocked.
U N xxV, >24 V,

NOISE!” and

| Where xx is the value of interference voltage. Voltage at test terminals is higher than 24 V but lower than 40 V, the measurement is blocked.
NOISE!”| The interfering signal (noise signal) is too high – the measurement result may be affected by additional uncertainty.
and ER together with the value ex- pressed in %| Measurement due to the resistance of the probes > 30%. (Measured values are used in calculation of uncertainty.)
and R H or R S with the value of Ω| Resistance of probes H and S, or one of them exceeds 19.9k Ω, correct measurement is not possible.
Flashing edges


| Flashing edges of symbols: E or H or S, two or all three at the same time: disconnected one, two or three leads to the terminals, or the resistance of the auxiliary probe/s is outside the measuring range.

Measurement of earth resistances using 2-pole method (RE2P)

RE2P method may be also used for measuring earth resistance. When the earthing system is known and earthing is available with a known resistance value, the measurement result will be the sum of the earth resistances: measured and known.

Turn on the meter. After turning the meter ON, the screen with measurement method RE3P is shown.

To enter the screen of RE2P measurement method, press DOWN button once.

The meter is in the mode of measuring the interference voltage between the test terminals. The measuring voltage is compatible with the volt-age selected when set-ting up the device.

Test leads should be connected to the measurement terminals in the device, as shown above.

In order to start the measurement, press START button.

After the measurement, its result is shown: at the bottom of the screen, the main result is shown for RE2P, whereas the upper part of the screen shows the measured interference voltage UN. The result is displayed for 20 seconds. The result can be recalled by pressing UP button.

Additional information displayed by the meter

R > 9999 Ω Measuring range is exceeded.

U N > 100 V, > 100 V and a continuous beep U ,

NOISE!” and

| Voltage at test terminals is higher than 100 V, measurement is blocked.
U N xxV, > 40 V and a continuous beep U ,
NOISE!” and| Where xx is the value of interference voltage. Voltage at test terminals is higher than 40 V, measurement is blocked.
U N xxV, > 24 V,
NOISE!” and| Where xx is the value of interference voltage. Voltage at test terminals is higher than 24 V but lower than 40 V, the measurement is blocked.
NOISE!”| The interfering signal (noise signal) is below 24 V, but with a high value – the measurement result may be affected by additional uncertainty.

Power supply

Before the measurements, make sure that the status of the batteries or rechargeable batteries in the meter is sufficient for performing tasks related to the operation of the device.

Monitoring the power supply voltage

The charge level of the batteries or rechargeable batteries is indicated by the symbol in the right upper corner of the display on a current basis:

When all segments of the battery symbol are lit, it means that the batter- ies/rechargeable batter-ies are fully charged.

When all segments of the battery symbol are dark, it means that the batter- ies/rechargeable batter-ies are discharged and must be replaced.

Message bAtt indicates that batteries/rechargeable batter-ies are totally discharged, all measurements are blocked. The meter switches off automatically after 5 sec

Replacing (rechargeable) batteries

MRU-10 is powered by four AA alkaline LR6 batteries or rechargeable batteries of NiMH type. The (rechargeable) batteries are placed in the compartment at the bottom of the enclosure. The de-vice is not equipped with an internal battery charger. Rechargeable batteries must be recharged in an external charger.

Note
Do not power the meter from sources other than those listed in this manual. Before replacing the (rechargeable) batteries, disconnect the test leads from the meter.

Disconnect the unit from the object!

Turn off the device using ON/OFF button.

Remove the screws that secure the battery cover at the bottom of the compartment(4 pcs), Remove all batteries (re-chargeable batteries). Observe the correct polarity when inserting new batteries/rechargeable batteries. Place and tighten the battery compartment cover.

Note
Reverse polarity will not damage the meter or the batteries, but the me-ter will not work. Have the meter serviced in case of battery leakage inside the compartment.

General principles regarding using Ni-MH rechargeable batteries
  • Store the he rechargeable batteries (the meter) in a dry, cool and well ventilated place and protect them from direct sunlight. The temperature of the environment in the case of prolonged storage should not exceed 30°C. If the rechargeable batteries are stored for a long time in a high temperature, then the occurring chemical processes may reduce their lifetime.
  • Rechargeable batteries NiMH usually lasts for 500-1000 charging cycles. The rechargeable batter-ies reach their maximum capacity after being formatted (2-3 charge and discharge cycles). The most important factor which influences the lifetime of rechargeable batteries is the level of their discharge. The deeper the discharge level of the batteries, the shorter their lifetime.
  • The memory effect is limited in the case of NiMH batteries. These batteries may be charged at any point with no serious consequences. However, it is recommended to discharge them completely every few cycles.
  • During storage of Ni-MH rechargeable batteries they are discharged at the rate of approximately 20% per month. Keeping rechargeable batteries at high temperatures may accelerate this process even 100%. In order to prevent excessive discharge of rechargeable batteries, after which it would be necessary to format them, it is recommended to charge them from time to time (even if they are not used).
  • Modern fast chargers detect both too low and too high temperature of the battery pack and react to the situation adequately. Too low temperature should prevent starting the process of charging, which might irreparably damage rechargeable batteries. An increase of the temperature of the rechargeable batteries is a signal to stop charging and is a typical phenomenon. However charging at a high ambient temperature apart from reducing batteries’ lifetime causes an accelerated increase of their temperature and the result is that the batteries are not charged to their full capacity.
  • Please note that when the batteries are charged with a fast-charger they are charged only to approx. 80% of their capacity – better results can be achieved by continuing charging: the charger enters trick-le-charging mode and during the next few hours batteries are charged to their full capacity.
  • Do not charge or use the batteries in extreme temperatures. Extreme temperatures reduce the life-time of batteries and rechargeable batteries . Avoid placing devices powered by rechargeable batter-ies in very hot environments. The nominal working temperature must be absolutely observed.

Cleaning and maintenance

Note
Use only the maintenance methods specified by the manufacturer in this manual.

  • The casing of the meter may be cleaned with a soft, damp cloth using all-purpose detergents. Do not use any solvents or cleaning agents which might scratch the casing (powders, pastes, etc.).
  • Clean the probe with water and dry it. Before the probe is stored for a prolonged period of time it is recommended to grease it with any machine lubricant.
  • The reels and test leads should be cleaned with water and detergents, and then dried. The electronic system of the meter does not require maintenance.

Storage

In the case of storage of the device, the following recommendations must be observed:

  • Disconnect all the test leads from the meter.
  • Clean the meter and all its accessories thoroughly.
  • Wind the long test leads onto the reels.
  • If meter is to be stored for a prolonged period of time, the batteries must be removed from the device.
  • In order to prevent a total discharge of the battery pack in the case of a prolonged storage, charge it from time to time.

Dismantling and Disposal

  • Worn-out electric and electronic equipment should be gathered selectively, i.e. it must not be placed with waste of another kind.
  • Worn-out electronic equipment should be sent to a collection point in accordance with the law of waste electrical and electronic equipment.
  • Before the equipment is sent to a collection point, do not dismantle any elements.
  • Observe local regulations concerning disposal of packages, waste batteries and accumulators.

Technical data

  • The specified accuracy relates to the meter terminals.
  •  The abbreviation „m.v.” in the accuracy definition means the measured value.
Basic data

Measurement of interference voltage UN (RMS)

Range Resolution Accuracy
0…100 V 1 V ±(10% m.v. + 1 digit)
  • Measurement for fN 45…65 Hz.
  •  The frequency of measurement – at least 2 measurements / sec.

Measurement of earth resistances RE2P (2-pole method)

Range Resolution Accuracy
0.01 Ω     19.99 Ω 0.01 Ω ±(3% m.v. + 3 digits)
20.0 Ω…199.9 Ω 0.1 Ω
200 Ω…1999 Ω 1 Ω ±5%
2000 Ω…9999 Ω 1 Ω ±8%
  • Measuring current at the short circuit of > 20 mA.
  • Measuring frequency of 125 Hz or 150 Hz.
  • Selected test voltage: 25V or 50 V.
  • Maximum interference voltage for the RE measurement is 24 V.

Measurement of earth resistances RE3P (3-pole method)

  • Measurement method: technical, conforming to EN 61557-5.
  • Measurement range according to EN 61557-5: 0.53 Ω…9999 Ω for Un = 50 V.
Range Resolution Accuracy
0.00 Ω   19.99 Ω 0.01 Ω ±(3% m.v. + 3 digits)
20.0 Ω…199.9 Ω 0.1 Ω
200 Ω…1999 Ω 1 Ω ±5%
2000 Ω…9999 Ω 1 Ω ±8%
  • Measuring current at the short circuit of > 20 mA.
  • Measuring frequency of 125 Hz or 150 Hz.
  • Selected test voltage: 25V or 50 V.
  • Maximum interference voltage for the RE measurement is 24 V.

Measurement of resistance of earth contact probes RH and RS

Range Resolution Accuracy
0…999 Ω 1 Ω ±(5% + 8 digits)
1.00   9.99 kΩ 0.01 kΩ
10.0…19.9 kΩ 0.1 kΩ

Other technical data

  • type of insulation ……………………………………….. double, according to EN 61010-1 and IEC 61557

  • measurement category (for 2000 m a.s.l.) ………………………………….. III 300 V acc. to EN 61010-1

  • degree of housing protection acc. to EN 60529: …………………………………………………………… IP67

  • maximum interference of AC + DC voltages, for the measurement …………………………………. 24 V

  • maximum measured voltage of interferences …………………………………………………………….. 100 V

  • measuring current frequency …………………………………………………….. 125 Hz for 50 Hz networks ……………………………………………………………………………………………… 150 Hz for 60 Hz networks

  •  measuring voltage for RE2P, RE3P ………………………………………………………………….. 25 V or 50 V

  • measuring current (short circuit) for RE2P, RE3P……………………………………………………….>20 mA

  • maximum resistance of earth contact probes …………………………………………………………….. 20 k

  • meter power supply …………….. alkaline batteries or NiMH rechargeable batteries size AA – 4 pcs.

  • number of measurements for RE3P…………………………………………………………………………..>3000
    ……………………………………………. (RE=10 , RH=RS=100 , 25V 50 Hz, 2 measurements/minute)

  • time of performing the resistance measurement with 2-pole method ……………………………….. <4 s

  • time of performing the resistance measurement with 3-pole method ……………………………….. <8 s

  • dimensions ……………………………………………………………… 221 x 102 x 62 mm (without test leads)

  • weight of the meter with batteries ……………………………………………………………………………… 660 g

  • operating temperature ………………………………………………………………………………….. -10C..+50C

  • reference temperature …………………………………………………………………………………………. 23 ±2C

  • storage temperature ………………………………………………………………………………………… -20..+60C

  • relative humidity …………………………………………………………………………………………………. 20..90%

  • nominal relative humidity …………………………………………………………………………………….. 40..60%

  • altitude a.s.l. ……………………………………………………………………………………………………..≤2000 m*

  • quality standard, design and manufacturing are ……………………………………… ISO 9001 compliant

  • the device meets the EMC requirements according to ……………… EN 61326-1 and EN 61326-2-2

NOTE

  • Information about the use of meter at altitude from 2000 m to 5000 m As for voltage inputs E, S, H the instrument is to be considered downgraded to measurement category CAT III 150 V to ground (max 150 V between inputs) or CAT IV 100 V to ground (max 100 V be-tween inputs). Markings and symbols indicated on the instrument are to be considered valid when using it at altitude lower than 2000 m.
Additional data

Data on additional uncertainties are useful mainly when the meter is used in non-standard conditions and for metrological laboratories for the purpose of calibration.

Influence of serial interference voltage on the resistance measurements for function RE3P

R E U N Additional uncertainty [Ω]
0.00…. 10.00 Ω 25 V ±(0,001 R E +0,01) U z +0,007 U z 2
50 V ±(0,001 R E +0,01) U z +0,004 U z 2
10.01..2000 Ω 25 V, 50 V ±(0,001 R E +0,01) U z +0,001 U z 2
2001..9999 Ω 25 V, 50 V ±(0.003 R E + 0.4) U z

Influence of the earth contact probes on earth resistance measurements for function RE3P

RE[Ω], RS[Ω] and RH[Ω] are values displayed by the device.

Additional uncertainties according to IEC 61557-5 (RE3P)

Significant parameter Designation Additional uncertainty
Position E1 0%
Supply voltage E2 0% ( is not lit)
Temperature E3 ±0.2 digit/°C for R<1 kΩ

±0.07%/°C ±0.2 digit/°C for R≥1 kΩ
Serial interference voltage| E4| According to the formulas shown in par. 9.2.1 (UN=3V 50/60Hz)
Resistance of earth contact probe| E5| According to the formula in par. 9.2.2

Accessories

The current list of accessories can be found on the manufacturer’s website.

Standard accessories
NAME INDEX QUANTITY
Earth contact probe, 25 cm WASONG25 2 pcs
Cable 2.2 m, black, 1 kV (banana plugs) WAPRZ2X2BLBB 1 pc.
Cable 15 m, red, for measuring earthing values at a winder
WAPRZ015REBBN 1 pc.
Cable 30 m, yellow, for measuring earthing values at a winder
WAPRZ030YEBBN 1 pc.
Black crocodile clip 1 kV 20 A WAKROBL20K01 1 pc.
Case M-6 WAFUTM6 1 pc.
Harness for the device (type M-1) WAPOZSZE4 1 pc.
Holder – a hanger for housing M-1 WAPOZUCH1 1 pc.
AA batteries LR6 4 pcs.
User manual 1 pc.
Calibration certificate issued by an accredited laboratory 1 pc.
Optional accessories

Additionally, the following items that are not included in the scope of standard equipment can be purchased from the manufacturer or the distributors:

NAME INDEX PICTURE
Earth contact probe (30 cm) WASONG30
Earth contact probe (80 cm) WASONG80
Vise clamp (banana plug) WAZACIMA1
Case L-3 (for probes 80cm) WAFUTL3
Cable on the reel, 25 m, red, for measuring earthing values (banana plugs)
WAPRZ025REBBSZ
Cable on the reel, 50 m, yellow, for measuring earthing values (banana plugs)
WAPRZ050YEBBSZ
Cable on the reel, 100 m, red, for measuring earthing values
WAPRZ100REBBSZ
Cable on the reel, 200 m, yellow, for measur- ing earthing values
WAPRZ200YEBBSZ
Red crocodile clip 1 kV 20 A WAKRORE20K02
Yellow crocodile clip 1 kV 20 A WAKROYE20K02

Manufacturer

The manufacturer of the device and provider of guarantee and post-guarantee service:

SONEL SA

Note:
Service repairs must be performed only by the manufacturer.

Laboratory services

SONEL Testing and Calibration Laboratory has been accredited by the Polish Center for Accreditation (PCA) – certificate no. AP 173.
Laboratory offers calibration for the following instruments that are used for measuring electrical and non-electrical parameters.

  • METERS FOR MEASUREMENTS OF ELECTRICAL PARAMETERS
    • voltage meters,
    • current meters (including clamp meters),
    • resistance meters,
    • insulation resistance meters,
    • earth resistance and resistivity meters,
    • RCD meters,
    • short-circuit loop impedance meters,
    • power quality analyzers,
    • portable appliance testers (PAT),
    • power meters,
    • multimeters,
    • multifunction meters covering the functions of the above-mentioned instruments,
  • ELECTRICAL STANDARDS
    • calibrators,
    •  resistance standards,
  • METERS FOR MEASUREMENTS OF NON-ELECTRICAL PARAMETERS
    • pyrometers,
    •  thermal imagers,
    • luxmeters.

The Calibration Certificate is a document that presents a relation between the calibration standard of known accuracy and meter indications with associated measurement uncertainties. The calibration standards are normally traceable to the national standard held by the National Metrological Institute.
According to ILAC-G24 „Guidelines for determination of calibration intervals of measuring instruments”, SONEL S.A. recommends periodical metrological inspection of the instruments it manufactures no less frequently than once every 12 months.
For new instruments provided with the Calibration Certificate or Validation Certificate at the factory, re-calibration should be performed within 12 months from the date of purchase, however, no later than 24 months from the date of purchase.

ATTENTION !
The person performing the measurements should be absolutely sure about the efficiency of the device being used. Measurements made with an inefficient meter can contribute to an in-correct assessment of the effectiveness of health protection and even human life.

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

Related Manuals