Specifications:

  • Product: METRACABLE TDR PRO
  • Manufacturer: Gossen Metrawatt GmbH
  • Type: Time Domain Reflectometer
  • Model Number: 3-447-089-03 1/12.20

Product Information:

The METRACABLE TDR PRO is a time domain reflectometer designed
for localizing faults in electrical cables such as two-core,
coaxial, and power cables, and for measuring their lengths. It
transmits a pulse to the voltage-free cable and displays the
pulse’s reflection (TDR trace) on its panel, indicating cable
length, components, wiring type, and faults. Suitable for indoor,
laboratory, industrial, and construction site use, it is portable
and can be handheld, worn with a strap, or placed on a surface with
a tilt stand. It comes with a pouch for storage and transport. The
instrument can manage data and transfer measurements to a PC using
METRACABLE MANAGER software.

Product Usage Instructions:

1. Safety Instructions:

Observe all safety information in the documentation to prevent
injury and damage to the instrument.

2. Applications:

2.1 Intended Use / Use for Intended Purpose:

The METRACABLE TDR PRO is specifically designed for localizing
faults in electrical cables and measuring their lengths. Ensure it
is used only for this purpose to guarantee operator and instrument
safety.

2.2 Use for Other than Intended Purpose:

Avoid using the instrument for purposes other than localizing
faults in electrical cables as it may compromise safety and
accuracy.

2.3 Liability and Guarantee:

Follow the complete operating instructions to maintain liability
and guarantee coverage.

2.4 Opening the Instrument / Repairs:

Refer to the complete operating instructions for guidance on
opening the instrument or repairs.

3. Initial Startup:

3.1 Power Supply:

Ensure proper power supply as per the instructions provided in
the manual.

3.2 Switching the Instrument On/Off and Displays:

Follow the instructions to power on/off the instrument and
interpret the displays.

3.3 Basic Settings:

Set up basic configurations according to the user manual before
starting measurements.

4. Measurements:

Perform cable fault localization and length measurements
following the guidelines in the operating instructions.

5. Product Support:

Contact product support for assistance with any operational
issues.

6. Repair and Replacement Parts Service / Calibration Center

and Rental Instrument Service:

Refer to the manual for information on repair, replacement
parts, calibration, and rental services.

7. CE Declaration:

Review the CE declaration section for compliance
information.

8. Return and Disposal:

Follow the guidelines for returning the instrument and proper
disposal methods.

FAQ:

Q: Where can I find the complete operating instructions?

A: The complete operating instructions are available in PDF
format at www.gossenmetrawatt.com.

“`

Condensed Operating Instructions

METRACABLE TDR PRO

Time Domain Reflectometer

3-447-089-03 1/12.20

Read the complete operating instructions which are available in PDF format at www.gossenmetrawatt.com.
The condensed operating instructions do not replace the complete operating instructions!

METRACABLE TDR PRO

Gossen Metrawatt GmbH

Table of Contents
1 Safety Instructions………………………………………………… 1
2 Applications …………………………………………………………. 4 2.1 Intended Use / Use for Intended Purpose……………….. 4 2.2 Use for Other than Intended Purpose …………………….. 5 2.3 Liability and Guarantee ………………………………………… 5 2.4 Opening the Instrument / Repairs………………………….. 5
3 The Instrument……………………………………………………… 6 3.1 Scope of Delivery ……………………………………………….. 6 3.2 Symbols on the Instrument…………………………………… 6 3.3 Instrument Overview……………………………………………. 7 3.4 Technical Data …………………………………………………… 9 3.5 Supported Cable Types …………………………………….. 11 3.6 Menu ……………………………………………………………… 12
4 Initial Startup………………………………………………………. 16 4.1 Power Supply…………………………………………………… 16 4.2 Switching the Instrument On/Off and Displays……….. 17 4.3 Basic Settings ………………………………………………….. 18
5 Measurements ……………………………………………………. 19
6 Product Support …………………………………………………. 26
7 Repair and Replacement Parts Service / Calibration Center and Rental Instrument Service26
8 CE Declaration ……………………………………………………. 27
9 Return and Disposal ……………………………………………. 28

Gossen Metrawatt GmbH

METRACABLE TDR PRO

1 Safety Instructions

Observe this documentation, in particular all included safety information, in order to protect yourself and others from injury, and to prevent damage to the instrument.
· Carefully and completely read and adhere to these condensed operating instructions, as well as the instrument’s operating instructions. The respective documents can be found at http:// www.gossenmetrawatt.com. Retain these documents for future reference.
· Use only the recommended accessories with the instrument!
· Tests may only be performed by a qualified electrician, or under the supervision and direction of a qualified electrician. The user must be instructed by a qualified electrician concerning performance and evaluation of the tests.
· Wear suitable and appropriate personal protective equipment (PPE) whenever working with the instrument.
· Always comply with all safety rules and other regulations which are applicable at the place of use.
· Comply with applicable safety regulations, for example in accordance with DIN VDE 0100, DIN VDE 0800 and DIN VDE 0805.
· If the instrument doesn’t function flawlessly, remove it from operation and secure it against inadvertent use.
· The instrument may only be used as long as it’s in good working order. Inspect the housing before use. Pay particular attention to any possible cracks and the insulation around the sockets.

1

METRACABLE TDR PRO

Gossen Metrawatt GmbH

· Accessories and cables may only be used as long as they’re fully intact. Inspect all cables and accessories before use. Pay particular attention to damaged housings, interrupted insulation or kinked cables.
· Do not use the instrument after long periods of storage under unfavorable conditions (e.g. humidity, dust or extreme temperature).
· Do not use the instrument after extraordinary stressing due to transport.
· Use the instrument only within the specified ambient conditions.
· Use the instrument only in accordance with the specified protection class (IP code).
· The instrument must not be exposed to direct sunlight.
· The instrument and the included accessories may only be used for the measurements described here and in the operating instructions for the instrument.
· Do not apply any external voltage to the instrument.
· Do not use the instrument if the battery compartment cover has been removed. Touch contact with dangerous voltage is otherwise possible.
· Hands must be kept behind the probe/terminal guards during testing.
· The operator must check the safety of the circuit before starting the test, and appropriate precautions must be implemented.
· Circuits must be de-energized and isolated before making any test connections.
· The instrument may only be used with voltage-free cables.

2

Gossen Metrawatt GmbH

METRACABLE TDR PRO

· The instrument is equipped with a Bluetooth® module. Determine whether or not use of the implemented frequency range of 2.402 to 2.480 GHz is permissible in your country.
· Always create a backup copy of your measurement data.

3

METRACABLE TDR PRO

Gossen Metrawatt GmbH

2 Applications

2.1 Intended Use / Use for Intended Purpose
The METRACABLE TDR PRO is a time domain reflectometer for localizing faults in electrical cables (e.g. two-core, coaxial and power cables), and for measuring their lengths. The instrument is connected to the voltage-free cable to this end, to which it transmits a pulse. The pulse’s reflection, i.e. its TDR trace, appears at the instrument’s display panel. The shape of the trace indicates the length of the cable, any included components, type of wiring and even any cable faults.
The METRACABLE TDR PRO can be used in indoor environments, in laboratories, in industrial settings and on construction sites. It’s a portable instrument which can be held in the hand during measurement, or hung around the neck using the carrying strap. Alternatively, the METRACABLE TDR PRO can be set up on a suitable surface with the help of the tilt stand. It can be stored and transported in the included pouch.
The instrument and the cable database can be managed and measurements can be transferred to a PC and viewed with the help of included METRACABLE MANAGER software.
Safety of the operator, as well as that of the instrument, is only assured when it’s used for its intended purpose.

! Attention!
Read and comply with the complete operating instructions (available at http://wwww.gossenmetrawatt.com).

4

Gossen Metrawatt GmbH

METRACABLE TDR PRO

2.2 Use for Other than Intended Purpose
Using the instrument for any purposes other than those described in these condensed operating instructions, or in the instrument’s operating instructions, is contrary to use for intended purpose.

2.3 Liability and Guarantee
Gossen Metrawatt GmbH assumes no liability for property damage, personal injury or consequential damage resulting from improper or incorrect use of the product, in particular due to failure to observe the product documentation. Furthermore, all guarantee claims are rendered null and void in such cases.
Nor does Gossen Metrawatt GmbH accept any liability for data loss.

2.4 Opening the Instrument / Repairs
The instrument may only be opened by authorized, trained personnel in order to ensure flawless, safe operation and to assure that the guarantee isn’t rendered null and void.
Unauthorized modifications to the instrument are prohibited.
If it can be ascertained that the instrument has been opened by unauthorized personnel, no guarantee claims can be honored by the manufacturer with regard to personal safety, measuring accuracy, compliance with applicable safety measures or any consequential damages.

5

METRACABLE TDR PRO

Gossen Metrawatt GmbH

3 The Instrument

3.1 Scope of Delivery Please check for completeness.

1

METRACABLE TDR PRO (with carrying strap)

(M281A)

4

LR6 batteries, 1.5 V, type AA

1

Set of test probes

(1.3 m, banana plugs; 300 V, Cat II)

2

Alligator clips (plug-on)

1

BNC adapter (banana plug to coaxial cable)

1

Pouch (with carrying strap)

1

Operating instructions (this document)

Included “METRACABLE MANAGER” software is available from our website at http://www.gossenmetrawatt.com. For further information please refer to the following section: “METRACABLE MANAGER Software 25”.

3.2 Symbols on the Instrument

Warning concerning a point of danger

!

(attention, observe documentation!)

European conformity marking

The instrument may not be disposed of with household trash.

6

Gossen Metrawatt GmbH
3.3 Instrument Overview Front

METRACABLE TDR PRO
Eyelet for Carrying Strap Fall Protector Battery Display

LCD Panel

ESC Key
SEL Key
Scroll Keys (for control)
On/Off Key Fall Protector and Battery Compartment Lid

7

METRACABLE TDR PRO Top
Back

Gossen Metrawatt GmbH
4 mm Banana Plug Sockets
Eyelet for Carrying Strap

Tilt Stand

Battery Compartment Lid and Fall Protector

8

Gossen Metrawatt GmbH

METRACABLE TDR PRO

3.4 Technical Data

Power supply Operating time Connections Place of use Ambient conditions
Electrical safety
Electromagnetic compatibility (EMC)

4 ea. LR6 battery, 1.5 V, type AA or
4 ea. rechargeable NiMH battery,1.2 V, type AA

Up to 30 hours (depending on battery type and quality), automatic shutdown can be selected

2 ea. 4mm banana plug safety sockets

Indoor environments, laboratories, industrial settings, construction sites

Operating

-10 … +50 °C

temperatures:

Storage

-25 … +75 °C

temperatures:

Relative humidity:

No condensation allowed

Elevation:

Max. 2000 m

Pollution

1

degree:

Protection category:

II per DIN EN 61140/ VDE 0140-1

Overvoltage protection:

DC: 100 V AC: 230 V / 50 Hz

Interference emission:

EN 55011: 2015

Interference immunity:

EN 61000-4-2: 2009 EN 61000-4-3: 2006

9

METRACABLE TDR PRO

Gossen Metrawatt GmbH

Mechanical design
Internal memory Languages Interface

Housing (W × H × D):

Approx. 19.5 × 10.0 × 4.5 cm

Mechanical protection:

Impact-resistant ABS housing with fall protector and display protection (2 mm Plexiglas with hardened safety glass)

Protection:

IP 52 per DIN VDE 0470, part 1/ EN 60 529 (protection against ingress of solid foreign objects: 1.0 mm diameter, protection against ingress of water: protection against falling dripping water, when the housing is inclined up to 15°)

Weight:

Approx. 390 g (without batteries)

Display:

LCD, monochrome, luminous, 240 x 128 pixels, adjustable on-time and contrast for background illumination

Up to 32 entries in the cable database Up to 510 measurements in the instrument

German, English, French Bluetooth®

10

Gossen Metrawatt GmbH

METRACABLE TDR PRO

Measurements

Signal type: Range: Accuracy: Resolution:
Output pulse: Pulse lengths:
Velocity factor:
Impedance: Signal type: Zoom (magnification factor):

Symmetrical search signal 14 km +1% ± pixels at 0.66 VF 3.125 ns or 0.3 m depending on cable Max. 20 V pp
12, 25, 50, 100, 200, 500, 1000 and 2500 ns Variable from 0.2 to 0.99 in steps of 0.01 50, 75, 100, 125 Symmetrical In steps of 6 dB

3.5 Supported Cable Types
Symmetrical cables with maximum attenuation of 80 dB can be measured using the TDR method. Examples:

Coaxial Cable

Shielded Cable

Twisted Pair

Multi-wire Line

11

METRACABLE TDR PRO

Gossen Metrawatt GmbH

3.6 Menu
The instrument has two menus: the main menu with basic settings and the TDR menu with specific settings for measurements.

Main Menu Overview
Press and hold the ESC key in order to open the main menu. Use the scroll keys to navigate within the menu, and to make selections and entries. Press the SEL key in order to acknowledge a selection. Select the TDR Measurement item in order to return to the TDR trace.

TDR measurement Velocity factor
Cable type
Data transmission

Exit menu / return to measurement
VF value The material-dependent velocity of propagation of the signal within the cable is taken into consideration by means of the VF value. It must be selected individually for each respective cable so that distance can be calculated correctly.
Access to the cable database and selection of the cable type (including VF value) (max. 32 entries)
Activate Bluetooth for data transmission to the PC.

12

Gossen Metrawatt GmbH

METRACABLE TDR PRO

Settings

Language
Unit of measure
Battery type
Illumination
Shutdown
Contrast Program
update Time Date

German, English, French
Meters (m) or feet (ft) in combination with velocity factor VF or half of velocity factor VF/2
Battery, rechargeable battery (depending on which is used)
Display illumination time after last entry 0 … x sec. (0 = continuously on)
Auto-shutdown of the instrument after last entry 0 … x min. (0 = continuously on)
-20 … +50 (default: 20)
Start the update process
Time of day in 24-hour format
Date in format DD/MM/YY

13

METRACABLE TDR PRO

Gossen Metrawatt GmbH

TDR Menu Overview
The TDR menu is opened by pressing and holding the SEL key. Use the scroll keys to navigate within the menu, and to make selections and entries. Press the SEL key in order to acknowledge a selection. Press the ESC key in order to return to the TDR trace.

Reference

Sets the momentary measurement as a reference curve in the background for comparing two measurements.
Note! The reference curve is not changed during Y zooming.

Save
Freeze Input

Saves the momentary measurement with timestamp. (Data can be transferred subsequently to the PC via Bluetooth.)
Freezes the momentary trace at the display.
Selection of either AC or DC.
· AC = capacitive coupling to the cable. Increased protection against interference voltage.
· DC = direct connection to the cable. Less protection against interference voltage.
Use DC for long cables (long pulse times influence the graphic representation of the TDR trace when using AC).

14

Gossen Metrawatt GmbH

METRACABLE TDR PRO

Pulse length
Z

Selection of the desired pulse length. Pulse transmission duration changes when the TDR range is adjusted. Pulse length must be increased in order for the instrument to overcome signal attenuation and still be able to determine cable length. However, shorter pulse lengths result in better resolution.
Impedance of the cable to be measured. These values are available from the respective cable manufacturers (determined automatically when the AUTO Test function is used).

15

METRACABLE TDR PRO

Gossen Metrawatt GmbH

4 Initial Startup

Proceed as follows:
1. Supply the instrument with electrical power 16. 2. Switch the instrument on and familiarize yourself with the
display 17. 3. Enter the basic settings 18.

4.1 Power Supply
The instrument is battery-operated. Charged batteries are shipped loose. These must be inserted before initial use.

! Attention!
Risk of injury due to contact with dangerous voltages. Only use the instrument when the battery compartment lid is inserted and firmly screwed into place.
Required tools: Philips screwdriver
The instrument has been switched off. All measurement cables and accessories have been re-
moved from the instrument.
1. Place the instrument face down on a stable surface. 2. Loosen and remove the 2 screws from the battery com-
partment lid. 3. Remove the battery compartment lid by pulling it down. 4. Insert the 4 included batteries into the battery compart-
ment, making sure that the plus and minus poles match up with the provided polarity symbols.

16

Gossen Metrawatt GmbH

METRACABLE TDR PRO

! Attention!
Insert only the included batteries during initial startup. Information concerning use of other batteries or rechargeable batteries can be found in the operating instructions (available at http://wwww.gossenmetrawatt.com).

5. Slide the battery compartment lid onto the battery compartment.
6. Secure the battery compartment lid with the screws. The batteries have now been inserted. The instrument can
now be used.

4.2 Switching the Instrument On/Off and Displays
Press the ON/OFF key in order to switch the instrument on and off. The TDR trace is displayed immediately after pressing the ON/OFF key.

C1:38,7m 0.67

Cursor Selection Cursor VF Display Battery Display

Display Range Lower Limit

0m

Cursor 150 ~ 75m

Function Selected with SEL Key

Display Range Upper Limit Input Display AC (~) or DC (­)
Impedance Display

The cursor can be moved to the right or the left with the help of the scroll keys. Pressing and holding the scroll key causes the cursor to move more quickly.

17

METRACABLE TDR PRO

Gossen Metrawatt GmbH

By pressing the SEL key, a function is selected which can then be controlled with the help of the two scroll keys, namely Cursor, Length, Y-Zoom or X-Zoom. For example if length is selected, length can be reduced or increased with the help of the scroll keys.

4.3 Basic Settings
1. Press the ON/OFF key at the instrument. The instrument is switched on and the TDR trace is displayed.
2. Open the main menu by pressing and holding the ESC key. The main menu appears.
3. Navigate to the Settings submenu with the help of the and keys.
4. Acknowledge by pressing the SEL key. The Settings menu is displayed.
5. Set time and date, select a language, select a unit of measure in combination with velocity factor and enter settings for illumination, shutdown and contrast in accordance with your preferences “Menu” 12.
The instrument is ready to perform measurements “Measurements” 19.

18

Gossen Metrawatt GmbH

METRACABLE TDR PRO

5 Measurements

This section describes basic performance of measurements with the METRACABLE TDR PRO. Detailed information, as well as information concerning additional functions and other measurements (e.g. for coaxial cables), can be found in the operating instructions (available at http://www.gossenmetrawatt.com).
The METRACABLE TDR PRO is connected to the voltage-free cable to be measured. It performs a TDR measurement (time domain reflectometry) whose results appear at the display as a TDR trace.
The instrument transmits an electrical pulse during measurement. Cable length can be calculated based on the amount of time it takes for the pulse signal to be returned to the instrument (reflection). The shape of the reflection provides information concerning:
· Any included components (e.g. a splitter)
· The wiring itself (e.g. splices, branches etc.)
· Cable faults (short-circuits, broken cable, pinching etc.)
The locations of these items are also calculated ­ accurate to approximately 0.3 m ­ based on the time it takes for the signal to return to the instrument.
Cables with lengths of up to 14 km can be examined with the METRACABLE TDR PRO.
Proceed as follows:
1. Perform measurement 20. 2. Evaluate the reflection “Typical Reflection Curves and
Their Meanings” 24. 3. View the information at a PC for detailed observation and
evaluation “METRACABLE MANAGER Software” 25.

19

METRACABLE TDR PRO

Gossen Metrawatt GmbH

Performing the Measurement
The momentary TDR trace is always shown at the display. If you change a setting, the TDR trace is adjusted automatically.

The electrical cable to be measured is not in use.
The velocity factor (VF value) of the cable to be measured is known, or there’s a corresponding entry in the cable database. (Detailed information concerning the VF value can be found in the operating instructions which are available at http:// wwww.gossenmetrawatt.com.)
1. Connect the test probes to the METRACABLE TDR PRO.
2. Optional (recommended): Plug the alligator clips on to the test probe.
3. Hold the probes against the cable to be measured or connect the alligator clips to the cable to be measured. One test probe or alligator clip must directly contact each of the two conductors (without insulation).
4. Press the ON/OFF key at the instrument. The instrument is switched on and the TDR trace is displayed.
5. Adjust the velocity factor (VF value) to match the cable under test.
­ The instrument includes a cable database with entries for various cable types. Select an appropriate cable type: Open the main menu by pressing and holding the ESC key. Navigate to the Cable Type submenu with the help of the scroll keys and acknowledge your selection by pressing the SEL key. Select the appropriate entry from the database which then appears with the help of the scroll keys, and acknowledge your selection by pressing the SEL key. In order to return to the TDR trace, navigate to the TDR Measurement submenu and acknowl-

20

Gossen Metrawatt GmbH

METRACABLE TDR PRO

edge your selection by pressing the SEL key. (Enter the value manually if you don’t find an appropriate entry. See instructions below.)
­ Enter the value manually. Open the main menu by pressing and holding the ESC key. Navigate to the Velocity Factor submenu and acknowledge your selection by pressing the SEL key. Enter the value with the help of the scroll keys and acknowledge by pressing the SEL key. In order to return to the TDR trace, navigate to the TDR Measurement submenu and acknowledge your selection by pressing the SEL key.
6. Enter the basic settings for the momentary measurement: input (AC or DC) and pulse length (14). Press and hold the SEL key to this end, in order to open the TDR menu. Use the scroll keys to navigate through the options, and to make selections and entries. Press the SEL to acknowledge your selection and entries. Then return to the TDR trace by pressing the ESC key.
7. Activate the AUTO Test function by briefly pressing the ESC key. Impedance (Z) is determined automatically. The TDR trace is generated automatically.
8. Adjust length (display range) in the TDR trace: Briefly press the SEL key repeatedly until the Length function is selected. Adjust length with the help of the and keys. (Maximum length depends on the cable.)

Note! Use the AUTO Test function! Instead of adjusting impedance (Z) manually, it can be determined automatically by the instrument. Briefly press the ESC key in order to activate the AUTO Test function.

21

METRACABLE TDR PRO

Gossen Metrawatt GmbH

9. Adjust the display range of the TDR trace as required: Briefly press the SEL key repeatedly until the X Zoom or Y Zoom function is selected. Zoom in with the scroll key and zoom out with the scroll key. ­ Y Zoom: Magnifies/reduces the measuring range display in steps of 6 dB.
­ X Zoom: Magnifies/reduces the display of the measurement segment. A cursor which has been positioned within the display can be used as a zoom reference.
10. Navigate to the desired position within the TDR trace with the help of the cursor: Briefly press the SEL key repeatedly until the Cursor function is selected. Move the cursor with the help of the and scroll keys (press and hold a scroll key for rapid motion). You can display a second cursor to measure two reflection points (e.g. measurement of distance between distributor and fault or distributor and distributor). Use the scroll key to switch back and forth between the two cursors and absolute position and difference: ­ C1: absolute position of cursor 1, cursor 1 is active
­ C2: absolute position of cursor 2, cursor 2 is active
­ C1: difference between cursors 1 and 2, cursor 1 is active
­ C2: difference between cursors 1 and 2, cursor 2 is active
C1:40.2m 0.67

0m

Cursor 150 ~ 300m

22

Gossen Metrawatt GmbH

METRACABLE TDR PRO

Note! Distance is calculated on the basis of the velocity factor (VF value 12). If this has not been entered correctly for the respective cable, distance is displayed incorrectly.
11. Optional: determine cable length. Move the cursor to the end of the trace where either the reflection curve for “open end” or “short-circuit” is displayed (24). The displayed cursor position corresponds to the length of the cable.
12. Optional: ascertain and analyze any detected reflection curve(s) 24.
13. Optional: save your measurement. Open the TDR menu to this end by pressing and holding the SEL key. Navigate with the and scroll keys until the Save function is selected. Acknowledge by pressing the SEL key. Enter a name with the help of the and scroll keys. Acknowledge by pressing the SEL key. The measurement is saved to memory.
14. Switch the METRACABLE TDR PRO off by pressing the ON/OFF key. The instrument is switched off.
15. Remove all measurement cables from the cable under test and from the instrument.
Measurement has now been completed.

23

METRACABLE TDR PRO

Gossen Metrawatt GmbH

Typical Reflection Curves and Their Meanings
Typical reflection curves and associated components, wiring and cable faults:

Open End

Short-Circuit

Poor Contact

Tap

Split/Resplit

Water in Cable

Tap, Short

Splice

Water Ingress

Splitter
Comprehensive, helpful instructions for troubleshooting and fault identification can be found in the operating instructions (available at http://www.gossenmetrawatt.com).

24

Gossen Metrawatt GmbH

METRACABLE TDR PRO

METRACABLE MANAGER Software
With the help of included METRACABLE MANAGER software, measurement results saved to the instrument can be transferred to a PC for more precise observation and evaluation. The software can also be used to manage the cable database and update the instrument.
Complete information can be found in the operating instructions (available at http://www.gossenmetrawatt.com).

25

METRACABLE TDR PRO

Gossen Metrawatt GmbH

6 Product Support

If required please contact:

Gossen Metrawatt GmbH

Product Support Hotline

Phone: +49-911-8602-0

Fax:

+49-911-8602-709

e-mail: [email protected]

7 Repair and Replacement Parts Service / Calibration Center and Rental Instrument Service

If required please contact:

GMC-I Service GmbH

Service Center

Beuthener Str. 41 90471 Nürnberg, Germany

Phone: +49-911-817718-0

Fax:

+49-911-817718-253

e-mail: [email protected]

Web: www.gmci-service.com

This address is only valid in Germany. Please contact our representatives or subsidiaries for service in other countries.

26

Gossen Metrawatt GmbH

METRACABLE TDR PRO

8 CE Declaration

The instrument fulfills all requirements of applicable EU directives and national regulations. We confirm this with the CE mark.

27

METRACABLE TDR PRO

Gossen Metrawatt GmbH

9 Return and Disposal

This instrument is subject to directive 2012/19/EC on Waste Electrical and Electronic Equipment (WEEE) and its German national equivalent implemented as the Waste Electrical and Electronic Equipment Act (ElektroG) on the marketing, return and environmentally sound disposal of electrical and electronic equipment. The instrument is a category 9 product (monitoring and control instrument) in accordance with ElektroG (German Waste Electrical and Electronic Equipment Act).
The symbol at the left indicates that this instrument and its electronic accessories must be disposed of in accordance with applicable legal regulations, and not together with household trash. In order to dispose of the instrument, bring it to a designated collection point or contact our product support department (26).
This instrument is also subject to directive 2006/66/EC on batteries and accumulators and waste batteries and accumulators and its German national equivalent implemented as the Battery Act (BattG) on the marketing, return and environmentally sound disposal of batteries and accumulators.
The symbol at the left indicates that batteries and rechargeable batteries must be disposed of in accordance with applicable legal regulations. Batteries and rechargeable batteries may not be disposed of with household trash. In order to dispose of the batteries or rechargeable batteries, remove them from the instrument and bring them to a designated collection point.
Segregated disposal and recycling conserves resources and protects our health and the environment.
Current and further information is available on our website at http://www.gossenmetrawatt.com under the search terms “WEEE” and “environmental protection”.

28

Gossen Metrawatt GmbH

METRACABLE TDR PRO

29

METRACABLE TDR PRO

Gossen Metrawatt GmbH

30

Gossen Metrawatt GmbH

METRACABLE TDR PRO

31

Gossen Metrawatt GmbH

METRACABLE TDR PRO

© Gossen Metrawatt GmbH Edited in Germany · Subject to change, errors excepted · PDF version available on the Internet
All trademarks, registered trademarks, logos, product names and company names are the property of their respective owners.

Gossen Metrawatt GmbH Südwestpark 15 90449 Nürnberg · Germany

Phone: Fax: e-mail: Web

+49 911 8602-111 +49 911 8602-777 [email protected] www.gossenmetrawatt.com

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

GOSSEN User Manuals

Related Manuals