Aseptico AEU-1070 Dental Implant Motor Systems Instruction Manual

May 15, 2024
Aseptico

Aseptico AEU-1070 Dental Implant Motor Systems

Aseptico-AEU-1070-Dental-Implant-Motor-Systems-product

SAFETY PRECAUTIONS

Aseptico accepts no liability for direct or consequential injury or damage resulting from improper use, arising in particular through the non-obser-vance of the operating instructions, or improper preparation and maintenance of this product. To prevent injury to people and property damage, please heed relevant warnings and remarks. They are marked as follows:

  • WARNING: Serious injury or death may result if ignored.
  • CAUTION: Damage to property or the environment may re-sult if ignored.
  • NOTE: Important additional information and hints.
  • WARNING: The system is supplied non-sterile! Before first use, and before each patient use thereafter, sterilize specified components as recommended in the Cleaning, Maintenance, and Sterilization section.
  • WARNING: Use for intended purposes only. Failure to observe the operating instructions may result in the patient or user suffering serious injury or the product being damaged, possibly beyond repair. Before using this product, make sure that you have studied and understood the operating instructions.
  • CAUTION: Federal law restricts this device to sale by or on the order of a dentist.
  • CAUTION: Use of other dental accessories or sub-assemblies from third-party manufacturers is the sole responsibility of the user.
  • CAUTION: All repairs are to be performed by authorized service personnel only.
  • CAUTION: Always follow these guidelines when operating the unit:
  • Never touch drills, burs, files, or other handpiece tips when they are still rotating.
  • The handpiece should only be attached when the motor has stopped running.
  • WARNING: Do not install where there is a risk of an explosion. The system is not intended for operation in the presence of flammable anesthetics or gases.
  • WARNING: All handpieces have inherent inefficiencies that can lead to torque variations. Routine calibration is recommended, even if using the same handpiece or whenever a handpiece is changed. If further verification of torque accuracy is desired, then it is suggested that a torque wrench be used.
  • WARNING: Always comply with the handpiece and implant/file manufacturers’ instructions regarding maximum speeds, torques, forward and reverse directions, and use of all instrumentation, drills, burs, etc., used in endodontics, implantology, and other oral surgery applications.
  • CAUTION: The irrigation supply system is designed for use with a saline solution or sterile water. For implants, use only suitable irrigants as recommended by the manufacturer’s instructions.
  • CAUTION: Connect the mains power cable to a properly grounded outlet only.
  • CAUTION: The motor is sensitive to shock and may be damaged if dropped or impacted against a hard surface.
  • WARNING: Do not disassemble or alter the system motor, console, or foot switch. No modification of the unit is allowed.
  • CAUTION: Use only appliance cord Type C13, 10A per IEC / EN 60320-1. Note: North America, Denmark, Australia, and New Zealand may require hospital-grade plugs. Consult local codes.
  • WARNING: Do not use this device in conjunction with an electric scalpel or on patients with pacemakers.
  • CAUTION: Never connect or disconnect the bag spike to the irrigation bag over the console. Water spilled onto the console can damage the unit.
  • CAUTION: It is recommended to always have the patient wear a rubber dam during implant procedures.
  • WARNING: Using a different handpiece model from those recommended can cause significant under- or over-torque.
  • WARNING: Using the incorrect ratio setting can cause significant under- or over-torque and under- or over-speeds.
  • WARNING: Do not position the unit so that the power cord is inaccessible for disconnection during use.
  • WARNING: Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result in improper operation. If such use is necessary, this equipment and the other equipment should be observed to verify that they are operating normally.
  • WARNING: Use of accessories, transducers, and cables other than those specified or provided by the manufacturer of this equipment could result in increased electromagnetic emissions or decreased electromagnetic immunity of this equipment and result in improper operation.
  • WARNING: Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be used no closer than 30 cm (12 inches) to any part of the Unit, including cables specified by the manufacturer. Otherwise, degradation of the performance of this equipment could result.
  • NOTE: Any malfunction or deterioration in the characteristics and/ or performance of this device, as well as any inadequacy in its labeling or the instructions for use which might lead to or might have led to the death of a patient or user or a serious deterioration in his state of health (e.g. Serious Adverse Event); this information shall be reported immediately to Aseptico Inc. and/or the Authorized Representative (within the European Union).
  • DANGER RISK OF EXPLOSION IF USED IN THE PRESENCE OF FLAMMABLE ANESTHETICS

This equipment meets all requirements for safety and performance, related to Electromagnetic Compatibility Standard IEC 60601-1-2:2014.

NOTE: The EMISSIONS characteristics of this equipment make it suitable for use in industrial areas and hospitals (CISPR 11 class A). If it is used in a residential environment (for which CISPR 11 class B is normally required) this equipment might not offer adequate protection to radio-frequency communication services. The user might need to take mitigation measures, such as relocating or re-orienting the equipment. Information concerning the accuracy and precision of this product may be obtained upon request by contacting Aseptico at the address shown on this page.

INTENDED USE

The AEU-1070-70V Dental Implant Motor System 70Ncm w/ Variable-Speed Footswitch and AEU-1070 Dental Implant Motor System 70 Ncm are drive systems for instruments and tools used in dentistry for implant procedures. The systems include an irrigation supply and a wide range of user controls designed to provide precision drilling during osteotomy preparation and implant placement.

RX: FEDERAL LAW RESTRICTS THIS DEVICE TO SALE BY OR ON THE ORDER OF A DENTIST

Congratulations!
Your new Aseptico AEU-1070-70V / AEU-1070 system is one of the finest and most affordable dental implant motor systems available to the dental profession.

Features include:

  • Powerful brushless 40,000 RPM motor for use with AHP-85MB-X or AHP-85MB-CX 20:1 reduction handpiece, or AHP-64 1:1 handpiece.
  • Large, easy-to-read touchscreen display.
  • 8 programmable presets.
  • Multi-function foot controller (AEU-1070-70V) or basic on-off foot controller (AEU-1070)
  • Water-resistant cabinet design helps prevent liquid spills from entering the internal electronics and allows easy clean-up.
  • An isolated peristaltic pump chamber prevents leaks from entering the internal electronics.
  • Vibration vibration-isolated peristaltic pump motor provides extremely quiet operation.
  • Large 1000ml saline bag capacity.
  • Made in the USA.

This system is engineered to provide many years of reliable service. Please read the instructions provided in this manual to ensure the optimum service from your Aseptico equipment. Separate manuals may be provided to cover the operation and maintenance of handpieces or other accessories for your unit.

PACKAGE CONTENTS

  • Electronic Control Console (P/N 120407)
  • AE-230A-40 Autoclavable 40K Brushless Micromotor
  • Autoclavable Motor Holder (P/N 461561)
  • AE-70V2 Multi-function Foot Controller (AEU-1070-70V only)
  • AE-7 PM On-Off Foot Controller (AEU-1070 only)
  • AE-24 Autoclavable Irrigation Tubing Set
  • AE-24-PUMP Peristaltic Pump Tubing Set (10 pieces)
  • Irrigation Bag Hanger Rod (P/N 461541)
  • Hospital-grade Power Cord (P/N 840079)

PURCHASED SEPARATELY

  • AHP-85MB-X or AHP-85MB-CX 20:1 Mont Blanc® Reduction Contra-Angle Handpiece
  • AHP-64 1:1 Straight Handpiece
  • Replacement AE-24 Autoclavable Irrigation Tubing Set
  • Replacement AE-24-PUMP Peristaltic Pump Tubing (10 Pieces)
  • Replacement AE-24-BOT Autoclavable Irrigation Tubing Set for Bottles

PACKAGE CONTENTS

Aseptico-AEU-1070-Dental-Implant-Motor-Systems-fig-2

SYSTEM SETUP

This section provides step-by-step instructions for setting up the unit.

  • The handpiece, motor/cord assembly, and irrigation tubing set must be sterilized before each procedure as instructed in the Maintenance and Sterilization section (page 10).
  1. Unpack the unit and identify each of the components shown to the left (page 4).
  2. Insert the bag hanger A into the hole on top of the console. Make sure that tab B on the hanger fully enters into slot C on the console, locking the hanger in position.
  3. Plug the power cord A into the power inlet on the back of the console, then plug the other end of the power cord into a hospital-grade, grounded electrical outlet.
  4. Connect the foot controller plug A (for either the AE-7 PM or AE-70V2 foot controller) to the socket on the back of the console.
    • The foot controller plug is “keyed” to the socket and can only be inserted in one position, with the notch B pointing to the left. Hold the plug against the socket and rotate it until you feel the plug enter the socket, then turn the locking sleeve clockwise until it clicks to secure the connection.
  5. Connect the motor plug A to the socket on the front of the console.
    • The motor plug is “keyed” to the socket and can only be inserted in one position, with the larger tab B pointing up. Hold the plug against the socket and gently rotate it until you feel the plug enter the socket, then turn the locking sleeve clockwise to secure the connection.
  6. Install an AHP-85MB-X or AHP-85MB-CX 20:1 handpiece A, or an AHP-64 1:1 handpiece (handpieces purchased separately) onto motor B.
  7. Place the motor handpiece onto cradle A.
  8. Pull lever A on top of the console to release the pump door, then pull pump door B fully open.
  9. While installing the irrigation tubing it will be helpful to be able to identify the bayonet A, large tubing connector B, peristaltic tube segment C, Luer connector D, and the small tubing connector E.
  10. To install the irrigation tubing, start by threading the peristaltic tube segment through the top slot A in the pump door. The large tubing connector B should be above the slot, as shown.
  11. Pull the tube down through the top slot until the large tubing connector A seats securely into the slot.
  12. Thread the peristaltic tube segment through the bottom slot A on the pump door.
  13. With one hand, pull the irrigation tube firmly from the bottom of the pump door A . Use the other hand to guide the irrigation tube into the notch B on the console while closing the pump door.
  14. Attach the small end of the irrigation tube A to the handpiece irrigation fitting, following the instructions that came with your handpiece. Attach the irrigation tube to the motor cable at evenly spaced intervals using the AHP-03 clips B provided in the AE-24 tubing set bag.
  15. Hang a saline bag (1000ml max.) from the hanger with the saline port pointing towards the rear of the console. Swing the bag away from the console (to prevent dripping on the console) then spike the saline bag with the bayonet A . Wipe up any spills on the console immediately.
  16. Press the power switch A on the rear of the console to the on (I) position to turn the unit on.Aseptico-AEU-1070-Dental-Implant-Motor-Systems-fig-17
  17. Calibrating the handpiece takes about 20 seconds and should be done every time you turn the unit on or change handpieces. See Calibrating the Handpiece on page 9, after becoming familiar with the layout of the control panel.

CONTROL PANEL

This section describes the functions of the touch-screen.

Home Screen
The home screen displays the name of the active preset A and the current settings for each of the adjustable parameters B, C, D, D, E

Aseptico-AEU-1070-Dental-Implant-Motor-Systems-fig-19

PRESET — This segment of the screen displays the active preset name. Press here A to select presets sequentially. Below is a table of the factory settings for each preset.

PRESET DIRECTION SPEED RATIO TORQUE WATER
PRST-1 FORWARD 1200 20:1 50 70
PRST-2 FORWARD 800 20:1 50 70
PRST-3 FORWARD 15 20:1 35 0
PRST-4 FORWARD 15 20:1 50 0
PRST-5 REVERSE 25 20:1 50 0
PRST-6 FORWARD 20 20:1 30 0
PRST-7 FORWARD 21 20:1 40 0
PRST-8 FORWARD 22 20:1 50 0

MOTOR DIRECTION
This segment of the screen displays the current motor rotation direction. Press here to alternate between forward (F) and reverse (R).

SPEED / RATIO
This segment of the screen displays both the current speed and ratio settings. The ratio setting (20:1 or 1:1) is indicated by the format (whole number or scientific notation, respectively) in which the speed setting is displayed:

RATIO INDICATOR : The speed setting is displayed as a whole number (e.g. 2000) to indicate when the 20:1 ratio is currently selected. The speed setting is displayed in scientific notation (e.g. 40E3) to indicate when the 1:1 ratio is currently selected. Replace “E3” with “000” to calculate the speed of a 1:1 handpiece (e.g. 40E3 = 40,000 RPM).

Press here C to open the speed/ratio adjustment screen:

  • SPEED SETTING : Press the buttons to adjust the speed setting in the following ranges, according to the current ratio setting:
RATIO SETTING SPEED RANGE SPEED DISPLAYED
20:1 15–2,000 15–2000
1:1 1,000–40,000 1E3–40E3

RATIO SETTING : Press the number (1 or 20) that appears between the UP and DOWN buttons to select the ratio setting:

RATIO SETTING NUMBER DISPLAYED
20:1 20
1:1 1

Press anywhere on the right side of the screen to return to the
home screen.

TORQUE — This segment of the screen displays the current torque setting. Press here to open the torque adjustment screen, then press the UP and DOWN buttons to adjust the torque setting in the following ranges, according to the current ratio setting:

RATIO SETTING TORQUE RANGE
20:1 5–70
1:1 1–4

Press anywhere on the right side of the screen to return to the home screen.

  • The speed and ratio settings determine whether the system is in “Torque-Limit” mode or “Torque Auto-Stop” mode. See the table below for details.

TORQUE-LIMIT MODE

When speed setting is ≥500 RPM @ 20:1 ratio
When speed setting is ≥10,000 RPM @ 1:1 ratio
A torque warning tone will beep when the handpiece torque reaches 75% of the torque setting. The handpiece will slow down when a load greater than the torque setting is reached. The handpiece will resume speed once the load is removed.
TORQUE AUTO-STOP MODE
When speed setting is < 500 RPM @ 20:1 ratio
When speed setting is < 10,000 RPM @ 1:1 ratio
A torque warning tone will beep when the handpiece torque reaches 75% of the torque setting. The handpiece will stop one second after the handpiece reaches the torque setting. The handpiece will resume operation once the foot pedal is released and reapplied.

  • Pressing the Torque screen segment also activates the Factory Restore and Calibration screens, as described in the sections below.

WATER — This segment of the screen displays the current irrigation flow setting. Press here to open the flow adjustment screen, then press the UP and DOWN buttons to adjust the flow setting (0–100%). A setting of 0% turns irrigation off. Press anywhere on the right side of the screen to return to the home screen.

Customizing Presets
The Preset segment A of the screen will flash if you have made any adjustments to the settings of the current preset: To clear any adjustments made to the current preset, simply press and immediately release the flashing Preset segment A of the screen. To save the adjustments that you have made (to customize the current preset), press and hold the flashing Preset segment A of the screen until you hear a beep. You may write down the settings for your custom presets in the blank tables on page 15.

Factory Restore
To restore all presets to their factory settings:

  1. Press the Torque segment D of the home screen.
  2. Press and hold the “CAL” button that appears between the UP and DOWN buttons until you hear a beep.
  3. Press and hold the flashing “RECALL” button that appears in the Preset segment A of the screen until you hear another beep. The factory settings have been restored for all presets (all user-customized presets have been erased).
  4. To cancel the factory restore procedure, press anywhere on the right side of the screen before completing step 3 above.

Calibrating Handpieces

Because the mechanical condition of motors and handpieces changes over time, it is recommended that you calibrate hand-pieces every time you power up the unit especially when you change handpiece models:

  1. Confirm that the correct handpiece ratio is selected for the current handpiece in use. See the Speed / Ratio section C (page 8) to change the ratio setting, if necessary.
  2. Insert a drill, bur, or file into the handpiece.
  3. Position the handpiece in the cradle so that the drill/bur/file can rotate freely without touching anything.
  4. Press the Torque segment of the home screen.
  5. Press the “CAL” button that appears between the UP and DOWN buttons to start the calibration process. The CAL button will flash and the motor will run through several speed changes, then stop.
  6. If calibration is successful, the CAL button will stop flashing when the motor stops.
    • If the CAL button continues to flash after the motor stops, calibration has failed. Try calibrating again. Repeated failures can indicate a dirty, damaged, or defective handpiece and/or a motor in need of repair or replacement.
  7. Press anywhere on the right side of the screen to return to the home screen.

OPERATION

This section explains how to operate the unit using either the AE-70V2 or AE- 7PM foot controller.

AE-70V2 Foot Controller
The AE-70V2 foot controller provides multiple-function control of the system.

  • VARIABLE SPEED PEDAL — This pedal activates the motor and irrigation pump simultaneously. The motor speed varies (up to the current Speed setting) in proportion to how far down you press the pedal. The irrigation pump activates at the current water setting, regardless of how far down the pedal is pressed.
  • PRESET — Press repeatedly to step through each of the presets sequentially. This provides the same function as pressing the Preset segment A (page page 8) on the touch screen.
  • MOTOR DIRECTION — Press to alternate between forward and reverse. This provides the same function as pressing the Motor Direction icon B (page page 8) on the touch screen.
  • HANDLE — Use to manually position the foot controller. Install the handle by inserting the handle prongs into the holes on the rear of the foot controller.
  • TORQUE STEP-INCREASE — For 20:1 handpieces, press to increase the torque setting in 2 Ncm steps for a maximum of 5 consecutive steps (or until the maximum torque setting is reached). On the sixth press (or the press after the maximum torque setting is reached), the torque setting will return to the original, lowest setting.
  • WATER ON-OFF / STEP-INCREASE — Press to switch the irrigation pump on and off. When the pump is switched off, the Water segment E (page 8) of the screen will display “— —” instead of 0%. Press and hold this button to step through the water settings in +10% increments; the steps occur at 1-second intervals with a beep sounding to signal each step.

AE-7PM Foot Controller
The AE-7PM foot controller provides basic on-off control of the system.

  • ON/OFF — Depress this footswitch to activate the motor and irrigation pump at the settings displayed on the screen.Aseptico-AEU-1070-Dental-Implant-Motor-Systems-FIG-18

MAINTENANCE AND STERILIZATION

Failure to adhere to the following instructions or blatant misuse of the device and/or related components will be considered abuse of the device and will void the warranty.

  • Call Aseptico at 1-800-426-5913 with any questions on sterilization procedures.

Handpiece
Follow the instructions provided with the handpiece for complete cleaning, maintenance and sterilization instructions.

Motor & Cord Assembly

  • CAUTION: The motor is sensitive to shock. Do not drop or impact the motor against a hard surface.
  • WARNING: Do not oil or lubricate the motor.
  • CAUTION: Do not allow oil from the handpiece to drain into the motor. After lubricating and before sterilizing the handpiece, stand it on a paper towel to allow excess oil to drain out of the handpiece.
  • WARNING: Do not attach a handpiece to the motor while the motor is running.
  • WARNING: Do not bend the motor cord sharply.
  • WARNING: Do not attempt to disassemble the motor or motor connector.
  • WARNING: Use of a sterilization method or temperatures other than what are prescribed below may damage components or present a risk of cross-contamination between patients.
  • CAUTION: Do not soak or submerge the motor in any liquid.
  • CAUTION: Do not use ultrasonic cleaners.
  • CAUTION: Do not use cold sterilization or dry heat.
  1. Brush off debris from the motor and cord.
  2. Thoroughly clean the motor and cord with a moist cloth or towel to remove any remaining signs of debris. Avoid pulling the cord away from the motor. Always wipe towards the motor.
  3. 3. Install the provided autoclaving plug A onto the end of the motor cord connector.
  4. Sterilize the motor and cord assembly using one of the STERILIZATION METHODS listed below (A, B or C).
  5. Loosely coil the motor cord when autoclaving. Avoid sharply bending the cord.

STERILIZATION METHODS:
Wrapped Sterilization — Place in an appropriately sized sterilization pouch and seal.

  • A. Standard autoclaving (Gravity displacement method)
    • Time : 15 min
    • Temperature : 132° C (270° F)
    • Dry time : 30 minutes
  • B. Pre-vacuum (Dynamic air removal method)
    • Time : 4 minutes
    • Temperature : 132° C (270° F)
    • Dry time : 40 minutes
    • Flash Sterilization — For immediate use only.
  • C. Unwrapped standard autoclaving (Gravity displacement method)
    • Time: 10 minutes
    • Temperature: 132° C (270° F)
    • Dry time : No dry time is required for flash sterilization.

Silicone Water Lines

  1. Flush the tubing with clean water for 30 seconds after each use. Do not use any disinfectants on the tubing.
  2. Sterilize the tubing:
    • Wrapped Sterilization — Place in an appropriately sized sterilization pouch and seal.
    • Standard autoclaving (Gravity displacement method)
    • Time: 30 min
    • Temperature: 132° C (270° F)
    • Dry time : 30 minutes

Console and Foot Controller

  1. Clean the exterior of the console and foot controller by wiping with a soft cloth moistened with mild detergent or 1:10 bleach solution (1 part bleach to 9 parts water). Use of solvents or other solutions may cause damage and void the warranty.

MESSAGES

System Messages

  1. When the system is powered on, the software part number and version will be displayed briefly on the console screen.

Error Messages
NONE

Fault Messages
NONE

WASTE DISPOSAL
Dispose of all medical waste products, residues, etc. per your local, state, and country laws. Dispose of the AEU‑1070‑70V/AEU‑1070 and accessories at the end-of-life under your local, state and country laws.

CONTRAINDICATIONS
None known

ADVERSE REACTIONS
None known

TROUBLESHOOTING

The console   does not light up when turned on:

•     Check the power cord connection.
•     Check fuse. If blown, see Fuse Replacement (page 13)
The console lights up when turned on, but the handpiece does not operate:
•     Check motor plug connection.
•     Check foot controller connection.
•     Depress the main foot controller switch.
•     Increase the speed setting.
•     Increase the torque setting.
•     Check that the connector pins in the plug socket are not bent.
•     Check that the rotary tool is properly seated in the handpiece and that the latch/collet is closed.
No water flow from pump to handpiece:
•     Check that the water setting is 10% or above.
•     Check that the water container seal is completely punctured.
•     Make sure the irrigation tubing is properly installed in pump door and flow is in the correct direction.
Motor slowing down or sluggish:
•     Check for dirty, under-lubricated handpieces.
•     Verify that the ratio setting matches the handpiece.
•     Check if handpiece lubricant is draining into the motor.
Display error:
•     Switch off power, wait 5 seconds, then switch back on to reset.
Pump leaking:
•     Replace the peristaltic pump segment of the irrigation line with a new one from the extra tube set provided.

FUSE REPLACEMENT

WARNING
Switch off the power and unplug the unit before following the steps below.

  • The system features auto-sensing global voltage compatibility. The fuse indicated is correct for 100V-240V 50/60 Hz line voltage.
    1. Remove the fuse holder from the power inlet socket by prying it out using two small blade screwdrivers, one on the top A  and the other on the bottom B of the fuse holder.
    2. Replace both fuses in the fuse holder with the same type and rating: 5 x 20mm, 1.6A, 250V slow-blow.
    3. Reinstall the fuse holder.

SYMBOL DESCRIPTIONS

Aseptico-AEU-1070-Dental-Implant-Motor-Systems-FIG-33 Aseptico-AEU-1070-Dental-Implant-Motor-Systems-
FIG-34

SPECIFICATIONS

  • CONSOLE DIMENSIONS: 19.6 cm W x 18.0 cm D x 16.0 cm H (7.72” W x 7.09” D x 6.30” H)
  • CONSOLE WEIGHT : 1.41 kg (3.11 lbs)
  • POWER SOURCE: 100-240V, 50-60Hz
  • CURRENT RATING : 100V/2A, 240V/2A
  • DUTY CYCLE : 16.7%

ENVIRONMENTAL CONDITIONS

  • Operating Temperature : 10°C to 28°C (50°F to 82.4°F)
  • Transport/Storage Temperature : –20°C to 60°C (–4°F to 140°F)
  • Relative Humidity : 10 to 90% non-condensing
  • Altitude: 0 to 3,000 meters (0 to 9,842 feet)

Classifications:

  • Class 1 Equipment
  • Type B Equipment
  • Ordinary Equipment – degree of protection against ingress of water
  • Not suitable for use in the presence of a flammable anesthetic mixture with air or with oxygen or nitrous oxide.
  • IPX1 – Footswitch

Expected Service Life:
7 years

Essential Performance:
DENTAL EQUIPMENT does not have ESSENTIAL PERFORMANCE.

MEDICAL ELECTRICAL EQUIPMENT concerning ELECTRIC SHOCK, FIRE AND MECHANICAL HAZARDS ONLY by UL 60601-1: 1988 +A1+A2, ANSI/AAMI EN 60601-1: 2005/(R)2012 & A1:2012, C1:2009/(R)2012, A2:2010/(R)2012- Pt 1, CAN/CSA C22.2 No 60601-1:14-Pt 1

WARRANTY

Aseptico Inc. warrants its new products against defects in material and workmanship under normal and proper use, care, and maintenance for two (2) years from the date of the original invoice. This two (2) year warranty does NOT apply nor is it extended to products that are not manufactured by Aseptico. These products may be covered by a separate limited warranty provided by the particular manufacturer, and all claims and questions regarding the same are to be directed to the particular manufacturer.

Expendable components, such as batteries, fuses, light bulbs, and tubing sets installed on Aseptico products are specifically excluded and have no warranty. Consumable goods are warranted for the stated expiration date of such goods.

Repair or replacement of any product(s) or part(s) under this warranty does not extend the term of this warranty, and such product (s) or part(s) shall remain covered by the unexpired portion of the warranty period, or for ninety (90) days from the date of return to Aseptico, whichever is later. This limited warranty applies only to the initial or first installation of the product or part.

During the specific warranty periods set forth above, Aseptico will, at its option, repair or replace the product(s) or particular part(s) that are found to be defective in either material or workmanship in part or whole. Aseptico shall be the sole arbiter of such action. In the event of an alleged defect under warranty, the purchaser is to notify Aseptico’s Customer Service department promptly. Customer Service will provide Return Material Authorization (RMA) instructions, usually directing that the product be returned for service, shipping prepaid by the buyer or end user, to Aseptico or its designated and authorized warranty service center.

This warranty shall not apply to products (1) that have been subjected to neglect, abuse, misuse, improper installation, inadequate maintenance, or damage due to improper use of cleaning materials or chemicals, or non- compliance with Aseptico’s storage, installation, operation, maintenance or environmental requirements; (2) that have undergone any modification or repair not previously authorized by Aseptico in writing, or service, repair or modification by or from any facility other than an authorized Aseptico service center or technician, or that use non-authorized software or spare or replacement parts; or (3) that fail due to reasonable and normal use or wear and tear, or materials made, furnished or specified by the buyer or end user.

Aseptico does not assume under this warranty any risks or liabilities arising from the clinical use of its products, whether or not such use involves coincidental utilization of products manufactured by others. Under no circumstances will Aseptico be liable or responsible for special, compensatory, incidental, consequential or punitive damages, lost profits, lost sales, or loss of use or loss of business opportunity by or through the use of the product. Aseptico’s sole and maximum liability for the product, other than its obligations set forth above, shall be the total purchase price paid for the product.

CUSTOM PRESETS

Use the blank tables below to write down your custom preset settings:

PRESET DIRECTION SPEED RATIO TORQUE WATER
PRST-1        
PRST-2        
PRST-3        
PRST-4        
PRST-5        
PRST-6        
PRST-7        
PRST-8        
  • P.O. Box 1548
  • 8333 216th Street S.E.
  • Woodinville, WA 98072
  • Phone (425) 487-3157
  • Web: www.aseptico.com
  • P/N 421116 • Rev. E • ECO 15776 • 08/2023 • Printed in the USA